Skip to content

Automated Peritoneal Dialysis System. Used for automatic… recallZ-1316-2023 · Baxter Healthcare Corporation · initiated Feb 22, 2023

Class II Open, Classified All recalls posted in April 2023

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The electrical safety testing was not properly performed on the impacted devices and additional testing is required

FDA root cause category: Process design

What the company is doing

Baxter notified consignees via phone call on 02/22/2023. The notification outlined the issue and that Baxter was retrieving the units and then replacing them within 24 hours.

Recall details

Recall number
Z-1316-2023
Recall class
Class II
Status
Open, Classified
Initiated by the company
Feb 22, 2023
Posted by the FDA
Apr 1, 2023
Recalling firm
Baxter Healthcare Corporation
Quantity
16 units
Product code
FKX
FDA submissions
K201867
Distribution
US distribution to AL, AR, AZ, FL, IL, ND, NY, OH, TX, WA.

Product description

Automated Peritoneal Dialysis System. Used for automatic control of dialysis solution exchanges.

Code information (models, lots, serial numbers)

Product Code: 5C6M40; UDI/DI: 0085412676463; Serial Numbers: 2010013403, 2102050403, 2103016303, 2103061703, 2104009103, 2105007803, 2105022503, 2110006603, 2110034803, 2111034103, 2202044303, 2203059903, 2203071803, 2203073003, 2204078503, 2203109603.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page