Medical device recalls: April 2023
The FDA posted 147 medical device recall records in April 2023, from 73 recall events by 65 companies. 9 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Philips North America Llc · 3 events, 5 records
- Baxter Healthcare Corporation · 3 events, 3 records
- Medtronic Sofamor Danek USA Inc · 2 events, 17 records
- Stryker Corporation · 2 events, 7 records
- Philips Respironics, Inc. · 2 events, 4 records, 4 Class I
- Covidien · 2 events, 2 records, 1 Class I
- Coloplast Manufacturing US, LLC · 1 event, 17 records
- ROi CPS LLC · 1 event, 8 records
FDA root cause
- Under Investigation by firm · 43
- Process control · 33
- Process change control · 22
- Nonconforming Material/Component · 10
- Device Design · 9
- Software design · 8
- Employee error · 5
- Process design · 2
Class I recalls (9)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-1357-2023 · Covidien · posted Apr 19, 2023 · Open, Classified
A manufacturing error, resulted in a less than specified diameter of adult flexible tracheostomy tube connectors, which results in unsecure connections with 15mm caps and other 15mm circuit components and accessories. An unsecure connection could lead to…
- Z-1313-2023 · Fresenius Kabi USA, LLC · posted Apr 18, 2023 · Open, Classified
Fluid ingress that can cause a loss of electrical funtion and failure of the Set ID Sensory, resulting in the prevention of infusion or halt of infusion.
- Z-1273-2023 · Abbott Diabetes Care, Inc. · posted Apr 6, 2023 · Open, Classified
Lithium-ion batteries in glucose monitoring system readers may swell, overheat, or pose fire hazard. Only use the charging cable/power adapter included with readers. Do not use reader if it is damaged, cracked, swelling, too hot to hold, it will no longer…
- Z-1274-2023 · Abbott Diabetes Care, Inc. · posted Apr 6, 2023 · Open, Classified
Lithium-ion batteries in glucose monitoring system readers may swell, overheat, or pose fire hazard. Only use the charging cable/power adapter included with readers. Do not use reader if it is damaged, cracked, swelling, too hot to hold, it will no longer…
- Z-1275-2023 · Abbott Diabetes Care, Inc. · posted Apr 6, 2023 · Open, Classified
Lithium-ion batteries in glucose monitoring system readers may swell, overheat, or pose fire hazard. Only use the charging cable/power adapter included with readers. Do not use reader if it is damaged, cracked, swelling, too hot to hold, it will no longer…
- Z-1322-2023 · Philips Respironics, Inc. · posted Apr 6, 2023 · Open, Classified
A limited number of remediated Philips DreamStation units may experience communication issues when connecting to the cloud-based care management application. Should this be the case, this may affect the ability to download prescription settings to the…
- Z-1287-2023 · Philips Respironics, Inc. · posted Apr 3, 2023 · Open, Classified
The accuracy of delivered oxygen may deviate below the required tolerance of 5% from setpoint when providing high concentration oxygen therapy. Additionally, if equipped, the internal FiO2 sensor may indicate a value higher than the device is actually…
- Z-1288-2023 · Philips Respironics, Inc. · posted Apr 3, 2023 · Open, Classified
The accuracy of delivered oxygen may deviate below the required tolerance of 5% from setpoint when providing high concentration oxygen therapy. Additionally, if equipped, the internal FiO2 sensor may indicate a value higher than the device is actually…
- Z-1289-2023 · Philips Respironics, Inc. · posted Apr 3, 2023 · Open, Classified
The accuracy of delivered oxygen may deviate below the required tolerance of 5% from setpoint when providing high concentration oxygen therapy. Additionally, if equipped, the internal FiO2 sensor may indicate a value higher than the device is actually…
Class II recalls (137)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-1474-2023 · Stryker Corporation · posted Apr 28, 2023 · Open, Classified
Fiber optic cables assembled with not enough epoxy on the proximal end, creating light output issues, which could cause a clinical delay or procedure being converted to open.
- Z-1475-2023 · Steris Corporation · posted Apr 28, 2023 · Open, Classified
Light handle covers may separate from the light handle (detach and fall off) during use, this could result in a procedure delay and/or impact the sterile field which could result in an injury to the patient
- Z-1476-2023 · Steris Corporation · posted Apr 28, 2023 · Open, Classified
Light handle covers may separate from the light handle (detach and fall off) during use, this could result in a procedure delay and/or impact the sterile field which could result in an injury to the patient
- Z-1477-2023 · Philips North America Llc · posted Apr 28, 2023 · Open, Classified
Potential table stop due to a broken tabletop cable: On the ProxiDiagnost N90 table there is a cable located under the tabletop that is covered by a plastic sleeve which can crack and damage the cable. If the tabletop cable breaks, the user may not be able…
- injeTAK Adjustable Tip NeedlesClass IIZ-1478-2023 · Labories Medical Technologies · posted Apr 28, 2023 · Open, Classified
The packaging may be damaged, compromising the sterile barrier.
- Z-1469-2023 · Siemens Medical Solutions USA, Inc · posted Apr 27, 2023 · Open, Classified
Examination room monitor connected with the hybrid cable may have a display issue and no image function is available
- Z-1470-2023 · FUJIFILM Healthcare Americas Corporation · posted Apr 27, 2023 · Open, Classified
Arm which holds the tube head in place is susceptible to breaking, if the device breaks it could potentially cause serious harm or injury to the patient and/or user.
- Z-1471-2023 · Draeger Medical Systems, Inc. · posted Apr 27, 2023 · Open, Classified
The software on the Infinity CentralStation drops peaks on narrow waveforms causing the Infinity M300 to fail to meet Frequency and Impulse Response requirements. In this condition, ECG waveforms with unusually narrow and/or high frequency QRS may be…
- Z-1472-2023 · Draeger Medical Systems, Inc. · posted Apr 27, 2023 · Open, Classified
The software on the Infinity CentralStation drops peaks on narrow waveforms causing the Infinity M300 to fail to meet Frequency and Impulse Response requirements. In this condition, ECG waveforms with unusually narrow and/or high frequency QRS may be…
- Z-1462-2023 · Datascope Corp. · posted Apr 26, 2023 · Open, Classified
The power cord supplied (Type J plug end) for the Cardiosave Hybrid IABP to Brazil is incompatible with the Brazilian Type N power receptacle.
- Z-1463-2023 · Covidien LP · posted Apr 26, 2023 · Open, Classified
There is a manufacturing non-conformance that may result in a metal shaving (burr) on the inside of the VersaOneTM Reusable Positioning Cannula, located where the Cannula and Body components are welded together.
- Z-1464-2023 · Covidien LP · posted Apr 26, 2023 · Open, Classified
There is a manufacturing non-conformance that may result in a metal shaving (burr) on the inside of the VersaOneTM Reusable Positioning Cannula, located where the Cannula and Body components are welded together.
- Z-1465-2023 · Covidien LP · posted Apr 26, 2023 · Open, Classified
There is a manufacturing non-conformance that may result in a metal shaving (burr) on the inside of the VersaOneTM Reusable Positioning Cannula, located where the Cannula and Body components are welded together.
- Z-1466-2023 · Wright Medical Technology, Inc. · posted Apr 26, 2023 · Open, Classified
Two lots of Ortholoc 3Di Fusion plates contain incorrect dorsiflexion descriptions laser marked on the parts.
- Z-1467-2023 · Wright Medical Technology, Inc. · posted Apr 26, 2023 · Open, Classified
Two lots of Ortholoc 3Di Fusion plates contain incorrect dorsiflexion descriptions laser marked on the parts.
- Z-1468-2023 · Mani, Inc. - Kiyohara Facility · posted Apr 26, 2023 · Terminated
A packaging defect may compromise the sterile barrier.
- HAMILTON-C6, REF: 160021Class IIZ-1429-2023 · Hamilton Medical AG · posted Apr 25, 2023 · Open, Classified
Software error causes, safety ventilation, in which ventilation continues in the "safety ventilation" mode with audible/visible alarm - patient inputs are not monitored, if the following coincide 1) A mode change to an adaptive mode (ASV, APVcmv, APVsimv,…
- Z-1430-2023 · Aesculap Implant Systems LLC · posted Apr 25, 2023 · Open, Classified
Mislabeled: Product is marked with GTIN # 04046955299592 however; the correct GTIN # is 04046955299607.
- Z-1431-2023 · ROi CPS LLC · posted Apr 25, 2023 · Open, Classified
The recalling firm was notified by their supplier that the light handle covers may separate from the light handle and fall off during use. The light handle covers are a kit component.
- Z-1432-2023 · ROi CPS LLC · posted Apr 25, 2023 · Open, Classified
The recalling firm was notified by their supplier that the light handle covers may separate from the light handle and fall off during use. The light handle covers are a kit component.
- Sterile surgical convenience kit: regard Item Number: 800718003, GY00923C - WPH Major Vaginal Pack.Class IIZ-1433-2023 · ROi CPS LLC · posted Apr 25, 2023 · Open, Classified
The recalling firm was notified by their supplier that the light handle covers may separate from the light handle and fall off during use. The light handle covers are a kit component.
- Z-1434-2023 · ROi CPS LLC · posted Apr 25, 2023 · Open, Classified
The recalling firm was notified by their supplier that the light handle covers may separate from the light handle and fall off during use. The light handle covers are a kit component.
- Z-1435-2023 · ROi CPS LLC · posted Apr 25, 2023 · Open, Classified
The recalling firm was notified by their supplier that the light handle covers may separate from the light handle and fall off during use. The light handle covers are a kit component.
- Sterile surgical convenience kit: regard Item Number 800706005, LD0091E - OH C-Section Pack.Class IIZ-1436-2023 · ROi CPS LLC · posted Apr 25, 2023 · Open, Classified
The recalling firm was notified by their supplier that the light handle covers may separate from the light handle and fall off during use. The light handle covers are a kit component.
- Z-1437-2023 · ROi CPS LLC · posted Apr 25, 2023 · Open, Classified
The recalling firm was notified by their supplier that the light handle covers may separate from the light handle and fall off during use. The light handle covers are a kit component.
- Z-1438-2023 · ROi CPS LLC · posted Apr 25, 2023 · Open, Classified
The recalling firm was notified by their supplier that the light handle covers may separate from the light handle and fall off during use. The light handle covers are a kit component.
- Z-1461-2023 · Baxter Healthcare Corporation · posted Apr 25, 2023 · Open, Classified
The product is brittle and potentially crumbling upon handling or when being removed from its package.
- Z-1422-2023 · Richard Wolf GmbH · posted Apr 24, 2023 · Open, Classified
Potentially unable to open forceps jaws when the endoscope shaft is bent by about 90 degrees. If, in addition, the proximal forceps shaft is bent, the forceps can no longer be opened or restrictions begin even with a slight bending of the endoscope.
- Z-1423-2023 · Richard Wolf GmbH · posted Apr 24, 2023 · Open, Classified
Potentially unable to open forceps jaws when the endoscope shaft is bent by about 90 degrees. If, in addition, the proximal forceps shaft is bent, the forceps can no longer be opened or restrictions begin even with a slight bending of the endoscope.
- Z-1424-2023 · Richard Wolf GmbH · posted Apr 24, 2023 · Open, Classified
Potentially unable to open forceps jaws when the endoscope shaft is bent by about 90 degrees. If, in addition, the proximal forceps shaft is bent, the forceps can no longer be opened or restrictions begin even with a slight bending of the endoscope.
- Z-1425-2023 · Richard Wolf GmbH · posted Apr 24, 2023 · Open, Classified
Potentially unable to open forceps jaws when the endoscope shaft is bent by about 90 degrees. If, in addition, the proximal forceps shaft is bent, the forceps can no longer be opened or restrictions begin even with a slight bending of the endoscope.
- Z-1426-2023 · Spineart SA · posted Apr 24, 2023 · Open, Classified
JULIET Lateral Lumbar Ti-Life Plate, a component of the JULIET Ti Lateral Lumbar Cage, may disassemble during the preparation stage before implantation, or during surgery when removing the implant holder, which could cause surgical delay.
- Z-1421-2023 · Carefusion 2200 Inc · posted Apr 21, 2023 · Open, Classified
A portion of the two products were incorrectly marked (etched) and packaged resulting in a product mismatch. The labels and laser etchings were switched on the two products.
- Z-1415-2023 · Becton Dickinson & Co. · posted Apr 20, 2023 · Open, Classified
Upon installation of BD Kiestra" InoqulA+" and BeA version 5.1 and 5.1.1 as part of the Icefall A platform, plate information was not visible in Synapsys" after processing has occurred. This could possibly lead to a delayed diagnosis and/or treatment or…
- Z-1416-2023 · Philips Medical Systems Nederland B.V. · posted Apr 20, 2023 · Open, Classified
There are technical issues related to signal generation and processing, which can lead to inaccurate presentations.
- Z-1417-2023 · Philips Medical Systems Nederland B.V. · posted Apr 20, 2023 · Open, Classified
There are technical issues related to signal generation and processing, which can lead to inaccurate presentations.
- Z-1418-2023 · Olympus Corporation of the Americas · posted Apr 20, 2023 · Open, Classified
There have been complaints that the clip did not come out of the tube sheath during the procedure.
- Z-1419-2023 · Beckman Coulter, Inc. · posted Apr 20, 2023 · Open, Classified
The reagent lot produces elevated results (approximately >30%) in some (15-20%)patient samples that do not correlate with the results of other thyroid markers nor Free T3 results generated on other platforms. This introduces a risk of the positive bias not…
- Z-1420-2023 · Baxter Healthcare Corporation · posted Apr 20, 2023 · Open, Classified
ExataMed 20 mL oral dispenser component of the Clear Pharmacy Pack may have particulate matter.
- Z-1400-2023 · Medacta Usa Inc · posted Apr 19, 2023 · Completed
There is a potential for breakage of the small screw used to affix the anti-backout plate.
- Z-1401-2023 · Medacta Usa Inc · posted Apr 19, 2023 · Completed
There is a potential for breakage of the small screw used to affix the anti-backout plate.
- Z-1402-2023 · Medacta Usa Inc · posted Apr 19, 2023 · Completed
There is a potential for breakage of the small screw used to affix the anti-backout plate.
- Z-1403-2023 · Medacta Usa Inc · posted Apr 19, 2023 · Completed
There is a potential for breakage of the small screw used to affix the anti-backout plate.
- Z-1404-2023 · Medacta Usa Inc · posted Apr 19, 2023 · Completed
There is a potential for breakage of the small screw used to affix the anti-backout plate.
- Z-1405-2023 · Medacta Usa Inc · posted Apr 19, 2023 · Completed
There is a potential for breakage of the small screw used to affix the anti-backout plate.
- Z-1406-2023 · Avanos Medical, Inc. · posted Apr 19, 2023 · Terminated
Incorrect expiration date.
- Z-1407-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Apr 19, 2023 · Open, Classified
The CirClamp subassembly found in the kit was the incorrect size.
- Z-1408-2023 · Medtronic Neuromodulation · posted Apr 19, 2023 · Open, Classified
Medtronic has received reports of SenSight Extension Tunneler kits that contain dual carriers with have been machined on only one side. As a result, the dual carrier cannot be used to pass two extensions simultaneously.
- INFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 24mm, REF 3604024Class IIZ-1409-2023 · Medtronic Sofamor Danek USA Inc · posted Apr 19, 2023 · Open, Classified
Incorrect type of titanium was used which reduces the gripping strength of the screw head of 5 batches of Infinity Multi-Axial Screws
- INFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 32mm, REF 3604532Class IIZ-1410-2023 · Medtronic Sofamor Danek USA Inc · posted Apr 19, 2023 · Open, Classified
Incorrect type of titanium was used which reduces the gripping strength of the screw head of 5 batches of Infinity Multi-Axial Screws
- Z-1411-2023 · Medtronic Sofamor Danek USA Inc · posted Apr 19, 2023 · Open, Classified
Incorrect type of titanium was used which reduces the gripping strength of the screw head of 5 batches of Infinity Multi-Axial Screws
- INFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 26mm, REF 3604026Class IIZ-1412-2023 · Medtronic Sofamor Danek USA Inc · posted Apr 19, 2023 · Open, Classified
Incorrect type of titanium was used which reduces the gripping strength of the screw head of 5 batches of Infinity Multi-Axial Screws
- INFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 34mm, REF 3604034Class IIZ-1413-2023 · Medtronic Sofamor Danek USA Inc · posted Apr 19, 2023 · Open, Classified
Incorrect type of titanium was used which reduces the gripping strength of the screw head of 5 batches of Infinity Multi-Axial Screws
- Z-1414-2023 · NeuroLogica Corporation · posted Apr 19, 2023 · Open, Classified
Issue related to the operation of the arm latch within the column that supports the arm, potential bodily harm due to a fall of the arm
- Z-1396-2023 · Ortho-Clinical Diagnostics, Inc. · posted Apr 18, 2023 · Open, Classified
Incorrect Number of Calibrator Levels (2) Set for the VITROS Immunodiagnostics Products Anti-SARS-CoV-2 Total N Antibody Calibrators in Assay Data Disks (ADD), resulting in delayed patients results
- Z-1397-2023 · Boston Scientific Corporation · posted Apr 18, 2023 · Open, Classified
Cease use of the Habib EndoHPB Bipolar RF Catheter with the ERBE VIO3 Generator immediately due to the potential for delivery of excessive energy leading to thermal injury or tissue damage. An inaccurate setting is currently provided in the HABIB EndoHPB…
- Z-1398-2023 · Compass Health Brands (Corporate Office) · posted Apr 18, 2023 · Open, Classified
Error in the Introduction section of the user manual that indicates the device may be used for the prevention of deep vein thrombosis (DVT). Marketed without a 510k for this indication
- Z-1399-2023 · Tobii Dynavox Llc · posted Apr 18, 2023 · Open, Classified
Problem maintaining the stable placement of the battery may come loose from the housing and become damaged.
- AQURE REF 933-599Class IIZ-1394-2023 · Radiometer Medical ApS · posted Apr 17, 2023 · Open, Classified
Due to potential software issue that may result in patient mix-up information.
- Z-1395-2023 · Gramercy Extremity Orthopedics · posted Apr 17, 2023 · Open, Classified
Instrument Kits for EPS plates may contain drill guides with incorrect drill depth range of 30mm - 60mm; not the correct drill depth range of 10mm - 40mm.
- Z-1391-2023 · Pro-Dex Inc · posted Apr 14, 2023 · Open, Classified
Incorrect Unique Device Identifier/GTIN code was used on product.
- FilmArray Gastrointestinal (GI) Panel for FILMARRAY Systems, REF: RFIT-ASY-0116, (30 pouches/kit)Class IIZ-1392-2023 · BioFire Diagnostics, LLC · posted Apr 14, 2023 · Open, Classified
Due to manufacturing issue, Gastrointestinal (GI) Panel may result in false negative results.
- Z-1393-2023 · Smith & Nephew, Inc. · posted Apr 14, 2023 · Open, Classified
A mispack occurred during the manufacturing process, resulting in the box incorrectly containing JOURNEY II UNI Tibia Insert Medical Size 1-2 12MM instead of the JOURNEY II UNI Tibia Insert Medial Size 7-8 8MM.
- Z-1359-2023 · Coloplast Manufacturing US, LLC · posted Apr 12, 2023 · Open, Classified
A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.
- Z-1360-2023 · Coloplast Manufacturing US, LLC · posted Apr 12, 2023 · Open, Classified
A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.
- Z-1361-2023 · Coloplast Manufacturing US, LLC · posted Apr 12, 2023 · Open, Classified
A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.
- Z-1362-2023 · Coloplast Manufacturing US, LLC · posted Apr 12, 2023 · Open, Classified
A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.
- Z-1363-2023 · Coloplast Manufacturing US, LLC · posted Apr 12, 2023 · Open, Classified
A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.
- Z-1364-2023 · Coloplast Manufacturing US, LLC · posted Apr 12, 2023 · Open, Classified
A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.
- Z-1365-2023 · Coloplast Manufacturing US, LLC · posted Apr 12, 2023 · Open, Classified
A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.
- Z-1366-2023 · Coloplast Manufacturing US, LLC · posted Apr 12, 2023 · Open, Classified
A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.
- Z-1367-2023 · Coloplast Manufacturing US, LLC · posted Apr 12, 2023 · Open, Classified
A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.
- Z-1368-2023 · Coloplast Manufacturing US, LLC · posted Apr 12, 2023 · Open, Classified
A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.
- Z-1369-2023 · Coloplast Manufacturing US, LLC · posted Apr 12, 2023 · Open, Classified
A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.
- Z-1370-2023 · Coloplast Manufacturing US, LLC · posted Apr 12, 2023 · Open, Classified
A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.
- Z-1371-2023 · Coloplast Manufacturing US, LLC · posted Apr 12, 2023 · Open, Classified
A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.
- Z-1372-2023 · Coloplast Manufacturing US, LLC · posted Apr 12, 2023 · Open, Classified
A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.
- Z-1373-2023 · Coloplast Manufacturing US, LLC · posted Apr 12, 2023 · Open, Classified
A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.
- Z-1374-2023 · Coloplast Manufacturing US, LLC · posted Apr 12, 2023 · Open, Classified
XXX
- Z-1375-2023 · Coloplast Manufacturing US, LLC · posted Apr 12, 2023 · Open, Classified
A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.
- Digital VIPR ↗Class IIZ-1376-2023 · Western/Scott Fetzer Company · posted Apr 12, 2023 · Open, Classified
Gauge malfunction: Digital gauge displays an incorrect tank pressure - displays pressure higher than actual tank pressure. This may lead to tank running out of oxygen during use without audible alarms, potentially leading to an out of oxygen event and no flow.
- Z-1377-2023 · Bio-Rad Laboratories, Inc. · posted Apr 12, 2023 · Open, Classified
Bio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has…
- Z-1378-2023 · Bio-Rad Laboratories, Inc. · posted Apr 12, 2023 · Open, Classified
Bio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has…
- Z-1379-2023 · Edwards Lifesciences, LLC · posted Apr 12, 2023 · Open, Classified
There is a potential that the distal end of the catheter may be bent beyond its intended curvature and when connected to patient monitoring device, the monitor may display an error message.
- Z-1380-2023 · Edwards Lifesciences, LLC · posted Apr 12, 2023 · Open, Classified
There is a potential that the distal end of the catheter may be bent beyond its intended curvature and when connected to patient monitoring device, the monitor may display an error message.
- Z-1381-2023 · Angiodynamics, Inc. · posted Apr 12, 2023 · Open, Classified
Not programmed in accordance with specification. The programming affects the RFID function and does not allow the NanoKnife probes to be recognized by the NanoKnife generator resulting in a delay in procedure
- Philips DigitalDiagnost C50 -intended for use in generating radiographic images of human anatomyClass IIZ-1382-2023 · Philips North America Llc · posted Apr 12, 2023 · Open, Classified
If the PM for the telescopic carriage is not performed annually as defined in the Instructions for Use (IFU), the tension force of the spring balancer inside of the telescopic carriage will unavoidably decay through normal wear and tear; as a result there…
- Z-1383-2023 · Philips North America Llc · posted Apr 12, 2023 · Open, Classified
If the PM for the telescopic carriage is not performed annually as defined in the Instructions for Use (IFU), the tension force of the spring balancer inside of the telescopic carriage will unavoidably decay through normal wear and tear; as a result there…
- Z-1384-2023 · Encore Medical, LP · posted Apr 12, 2023 · Open, Classified
Patella drills are marked with incorrect diameter size of 5.6mm, when they should be 5.4mm. Porous patella overdrills are marked with incorrect diameter size of 5.4 mm, when they should be 5.6 mm. If the incorrect drill is selected during surgery, there is…
- Z-1385-2023 · Howmedica Osteonics Corp. · posted Apr 12, 2023 · Open, Classified
Ccertain lots of the Restoris MultiCompartmental Knee (MCK) Tibial Baseplate Trials contain product with incorrect size laser marking
- Z-1386-2023 · Howmedica Osteonics Corp. · posted Apr 12, 2023 · Open, Classified
Certain lots of the Restoris MultiCompartmental Knee (MCK) Tibial Baseplate Trials contain product with incorrect size laser marking
- Z-1387-2023 · Covidien, LP · posted Apr 12, 2023 · Open, Classified
Medtronic was made aware of a supplier calibration issue during the contract sterilization process that may lead to unsterile product. Use of affected product may lead to an unspecified infection or, in the case of sutures used in the eye, pose a loss of…
- Z-1388-2023 · Covidien, LP · posted Apr 12, 2023 · Open, Classified
Medtronic was made aware of a supplier calibration issue during the contract sterilization process that may lead to unsterile product. Use of affected product may lead to an unspecified infection or, in the case of sutures used in the eye, pose a loss of…
- Z-1389-2023 · Covidien, LP · posted Apr 12, 2023 · Open, Classified
Medtronic was made aware of a supplier calibration issue during the contract sterilization process that may lead to unsterile product. Use of affected product may lead to an unspecified infection or, in the case of sutures used in the eye, pose a loss of…
- Z-1390-2023 · Covidien, LP · posted Apr 12, 2023 · Open, Classified
Medtronic was made aware of a supplier calibration issue during the contract sterilization process that may lead to unsterile product. Use of affected product may lead to an unspecified infection or, in the case of sutures used in the eye, pose a loss of…
- CombiDiagnost R90 1.0 and 1.1 multi-functional general R/F system, Model Numbers 709030 and 709031Class IIZ-1355-2023 · Philips North America Llc · posted Apr 10, 2023 · Open, Classified
If the system experiences an overvoltage / power surge it can cause a short circuit condition within the Main- cabinet. If this issue occurs, the user may experience an electrical fire within the cabinet, see or smell smoke emitting from the cabinet, or…
- ProxiDiagnost R90 1.0 and 1.1 multi-functional general R/F system, Model Numbers 706100 and 706110Class IIZ-1356-2023 · Philips North America Llc · posted Apr 10, 2023 · Open, Classified
If the system experiences an overvoltage / power surge it can cause a short circuit condition within the Main- cabinet. If this issue occurs, the user may experience an electrical fire within the cabinet, see or smell smoke emitting from the cabinet, or…
- Z-1350-2023 · Church & Dwight Inc · posted Apr 7, 2023 · Open, Classified
Church & Dwight First Response Pregnancy Kits was marketed with the Easy Read App without a new 510k. There is also a potential for Easy Read App to be unable to reads the test results (i.e., no results reported on the APP screen) or to misread the analog…
- Outset TabloClass IIZ-1351-2023 · Outset Medical, Inc. · posted Apr 7, 2023 · Open, Classified
An observed trend of high conductivity dialysate alarms in a hemodialysis system as a result of introducing a software tool into the manufacturing process.
- Z-1352-2023 · Merit Medical Systems, Inc. · posted Apr 7, 2023 · Open, Classified
Product that was built for design verification testing was inadvertently distributed to customers.
- Z-1353-2023 · Shanghai United Imaging Healthcare Co., Ltd. · posted Apr 7, 2023 · Open, Classified
Due to a software issue where the process of patient scanning, the scatter correction may occasionally fail with will potentially cause a failure of the PET image reconstruction and generation of PET images.
- Z-1354-2023 · Solta Medical Inc · posted Apr 7, 2023 · Open, Classified
A service unit was potentially improperly tested and calibrated during manufacturing on a failed test station.
- Z-1348-2023 · CooperSurgical, Inc. · posted Apr 6, 2023 · Open, Classified
It has come to CooperSurgical's attention that the affected Product may contain a medium other than the Global Total LP single step medium which does not contain the protein source required for embryo culture and development.
- Z-1349-2023 · Abbott Molecular, Inc. · posted Apr 6, 2023 · Open, Classified
The instrument was found to be installed without plunger assemblies in the clamp bar for an Amplification Detection Unit. After correction of the issue, the instrument was returned for use without calibration. This would result in a potential for delay…
- Z-1345-2023 · Terumo Cardiovascular Systems Corporation · posted Apr 5, 2023 · Open, Classified
Some Capiox NX, RX and FX Oxygenators may contain an arterial thermistor component that is improperly bonded to the oxygenator port. This may result in a leak at the point where the arterial thermistor is improperly bonded to the oxygenator port.
- Z-1346-2023 · Terumo Cardiovascular Systems Corporation · posted Apr 5, 2023 · Open, Classified
Some Capiox NX, RX and FX Oxygenators may contain an arterial thermistor component that is improperly bonded to the oxygenator port. This may result in a leak at the point where the arterial thermistor is improperly bonded to the oxygenator port.
- Z-1347-2023 · Terumo Cardiovascular Systems Corporation · posted Apr 5, 2023 · Open, Classified
Some Capiox NX, RX and FX Oxygenators may contain an arterial thermistor component that is improperly bonded to the oxygenator port. This may result in a leak at the point where the arterial thermistor is improperly bonded to the oxygenator port.
- Medtronic Grafton Putty:Class IIZ-1323-2023 · Medtronic Sofamor Danek USA Inc · posted Apr 4, 2023 · Open, Classified
Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.
- Medtronic Grafton CrunchClass IIZ-1324-2023 · Medtronic Sofamor Danek USA Inc · posted Apr 4, 2023 · Open, Classified
Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.
- Medtronic Grafton Flex:Class IIZ-1325-2023 · Medtronic Sofamor Danek USA Inc · posted Apr 4, 2023 · Open, Classified
Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.
- Medtronic Grafton Orthoblend:Class IIZ-1326-2023 · Medtronic Sofamor Danek USA Inc · posted Apr 4, 2023 · Open, Classified
Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.
- Medtronic Grafton Matrix:Class IIZ-1327-2023 · Medtronic Sofamor Danek USA Inc · posted Apr 4, 2023 · Open, Classified
Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.
- Medtronic Grafton Matrix Strips:Class IIZ-1328-2023 · Medtronic Sofamor Danek USA Inc · posted Apr 4, 2023 · Open, Classified
Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.
- Medtronic Grafton Plus PasteClass IIZ-1329-2023 · Medtronic Sofamor Danek USA Inc · posted Apr 4, 2023 · Open, Classified
Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.
- Medtronic Grafton DBF: ↗Class IIZ-1330-2023 · Medtronic Sofamor Danek USA Inc · posted Apr 4, 2023 · Open, Classified
Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.
- Medtronic Grafton DBF Inject: ↗Class IIZ-1331-2023 · Medtronic Sofamor Danek USA Inc · posted Apr 4, 2023 · Open, Classified
Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.
- Z-1332-2023 · Medtronic Sofamor Danek USA Inc · posted Apr 4, 2023 · Open, Classified
Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.
- Medtronic Grafton Gel:Class IIZ-1333-2023 · Medtronic Sofamor Danek USA Inc · posted Apr 4, 2023 · Open, Classified
Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.
- Medtronic XPANSE Bone Insert: ↗Class IIZ-1334-2023 · Medtronic Sofamor Danek USA Inc · posted Apr 4, 2023 · Open, Classified
Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.
- Z-1335-2023 · Maquet Cardiovascular, LLC · posted Apr 4, 2023 · Open, Classified
Some batches of product were not sterilized to their minimum sterilization specification.
- Z-1336-2023 · Maquet Cardiovascular, LLC · posted Apr 4, 2023 · Open, Classified
Some batches of product were not sterilized to their minimum sterilization specification.
- Z-1337-2023 · Maquet Cardiovascular, LLC · posted Apr 4, 2023 · Open, Classified
Some batches of product were not sterilized to their minimum sterilization specification.
- Z-1338-2023 · Moberg Research, Inc. · posted Apr 4, 2023 · Open, Classified
Battery leakage can cause corrosion to the metal enclosure near the electrical interface. Prolonged contact to corrosive materials could cause injury to skin. DOC
- Z-1339-2023 · Stryker Corporation · posted Apr 4, 2023 · Open, Classified
Not able to achieve desired pressure and not holding pressure result from two different causes which have been reported to occur during initial set-up of the cuff onto the patient, may lead to operative site blood loss, hemorrhage or, in exceedingly rare…
- Z-1340-2023 · Stryker Corporation · posted Apr 4, 2023 · Open, Classified
Not able to achieve desired pressure and not holding pressure result from two different causes which have been reported to occur during initial set-up of the cuff onto the patient, may lead to operative site blood loss, hemorrhage or, in exceedingly rare…
- Z-1341-2023 · Stryker Corporation · posted Apr 4, 2023 · Open, Classified
Not able to achieve desired pressure and not holding pressure result from two different causes which have been reported to occur during initial set-up of the cuff onto the patient, may lead to operative site blood loss, hemorrhage or, in exceedingly rare…
- Z-1342-2023 · Stryker Corporation · posted Apr 4, 2023 · Open, Classified
Not able to achieve desired pressure and not holding pressure result from two different causes which have been reported to occur during initial set-up of the cuff onto the patient, may lead to operative site blood loss, hemorrhage or, in exceedingly rare…
- Z-1343-2023 · Stryker Corporation · posted Apr 4, 2023 · Open, Classified
Not able to achieve desired pressure and not holding pressure result from two different causes which have been reported to occur during initial set-up of the cuff onto the patient, may lead to operative site blood loss, hemorrhage or, in exceedingly rare…
- Z-1344-2023 · Stryker Corporation · posted Apr 4, 2023 · Open, Classified
Not able to achieve desired pressure and not holding pressure result from two different causes which have been reported to occur during initial set-up of the cuff onto the patient, may lead to operative site blood loss, hemorrhage or, in exceedingly rare…
- Z-1314-2023 · Fujifilm Irvine Scientific, Inc. · posted Apr 1, 2023 · Open, Classified
Due to the detection of oil toxicity in several lots of Oil for Embryo Culture products.
- Z-1315-2023 · Stryker Sustainability Solutions · posted Apr 1, 2023 · Open, Classified
Increased incidence of disposable tourniquet cuffs "not able to achieve desired pressure" and "not holding pressure" due to 1) Cuff fabric tearing during use, and 2) Pinched or unseated O-ring at the connection between the cuff, disposable fill line,…
- Automated Peritoneal Dialysis System. Used for automatic control of dialysis solution exchanges.Class IIZ-1316-2023 · Baxter Healthcare Corporation · posted Apr 1, 2023 · Open, Classified
The electrical safety testing was not properly performed on the impacted devices and additional testing is required
- Z-1317-2023 · Limacorporate S.p.A · posted Apr 1, 2023 · Open, Classified
Shoulder glenoid fossa prosthesis was manufactured without tantalum markers so in the absence of the markers, on already implanted devices, a correct radiographic assessment of the hybrid glenoid position can be performed by the evaluation of the joint…
- Z-1318-2023 · Covidien · posted Apr 1, 2023 · Open, Classified
Due to potential leaks associate with the Oral/Nasal Tracheal Tube Cuffed inflation system.
- Z-1319-2023 · Biosense Webster, Inc. · posted Apr 1, 2023 · Open, Classified
Incorrect labeling; specifically, the product name, lot number and product code on the carton label for Bi-Directional Guiding Sheaths does not match the inner label pouch
- Z-1320-2023 · Biosense Webster, Inc. · posted Apr 1, 2023 · Open, Classified
Incorrect labeling; specifically, the product name, lot number and product code on the carton label for Bi-Directional Guiding Sheaths does not match the inner label pouch
- Z-1321-2023 · Remote Diagnostic Technologies Ltd. · posted Apr 1, 2023 · Open, Classified
During internal testing, Regulatory Compliance issues were identified regarding the Monitor and Power Supplies for fluid ingress and basic safety issues.
Class III recalls (1)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-1358-2023 · Agilent Technologies Inc./US · posted Apr 11, 2023 · Open, Classified
Distributed sample collection kit with an unapproved instruction for use.
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.