Skip to content

Moberg Component Neuromonitoring System (CNS… recallZ-1338-2023 · Moberg Research, Inc. · initiated Feb 20, 2023

Class II Open, Classified All recalls posted in April 2023

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Moberg Research, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Battery leakage can cause corrosion to the metal enclosure near the electrical interface. Prolonged contact to corrosive materials could cause injury to skin. DOC

FDA root cause category: Under Investigation by firm

What the company is doing

Natus issued Letters Urgent Medical Device Recall Letter Component Neuromonitoring System (Moberg CNS Monitor Model CNS-350) via first class beginning Feb 20, 2023. Letter states reason for recall, health risk and action to take: Ensure that the CNS is either shut down or plugged into a power outlet. Perform the next step while using proper personal protective equipment (PPE), including gloves. On the top of the CNS-350 device, open each of the three (3) battery ports and remove any backup battery. 2) Inspect the battery and the battery port for any white, brown, or green/blue buildup, or for any corrosion of the metal contact locations. a. Dispose of the battery at your local electronic e-waste facility. Note: The CNS-350 device may continue to be used (while plugged into a power outlet) with the backup battery removed. b. If battery leakage or corrosion is suspected, cease use of the CNS-350 device. New replacement battery will be shipped to you. The Moberg Technical Service department will be in contact with you in relation to the provision of the replacement product. For new battery replacement, installation instructions will be provided with the battery.

Recall details

Recall number
Z-1338-2023
Recall class
Class II
Status
Open, Classified
Initiated by the company
Feb 20, 2023
Posted by the FDA
Apr 4, 2023
Recalling firm
Moberg Research, Inc.
Quantity
89 units
Product code
OMA
FDA submissions
K080217
Distribution
Worldwide distribution - US Nationwide and the countries of France, Germany, Italy, Sweden.

Product description

Moberg Component Neuromonitoring System (CNS Monitor)-Intended as a short-term neurological monitoring system that collects, displays, and stores multiple physiological measurements Model: CNS-350

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

UDI Number: B485CNS3501 Serial Numbers: A-CNS-22502 A-CNS-22506 A-CNS-22507 A-CNS-22508 A-CNS-22509 A-CNS-22510 A-CNS-22511 A-CNS-22512 A-CNS-22513 A-CNS-22525 A-CNS-22526 A-CNS-22527 A-CNS-22528 A-CNS-22529 A-CNS-22530 A-CNS-22534 A-CNS-22535 A-CNS-22536 A-CNS-22537 A-CNS-22541 A-CNS-22542 A-CNS-22543 A-CNS-22544 A-CNS-22545 A-CNS-22546 A-CNS-22547 A-CNS-22548 A-CNS-22549 A-CNS-22550 A-CNS-22551 A-CNS-22552 A-CNS-22553 A-CNS-22554 A-CNS-22555 A-CNS-22556 A-CNS-22557 A-CNS-22558 A-CNS-22559 A-CNS-22560 A-CNS-22561 A-CNS-22562 A-CNS-22564 A-CNS-22565 A-CNS-22566 A-CNS-22567 A-CNS-22568 A-CNS-22569 A-CNS-22570 A-CNS-22571 A-CNS-22572 A-CNS-22573 A-CNS-22574 A-CNS-22575 A-CNS-22577 A-CNS-22578 A-CNS-22584 A-CNS-22585 A-CNS-22586 A-CNS-22587 A-CNS-22588 A-CNS-22589 A-CNS-22590 A-CNS-22591

Other devices under the same FDA submission

1 device cleared under K080217 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page