Skip to content

Tempus Pro recallZ-1321-2023 · Remote Diagnostic Technologies Ltd. · initiated Feb 23, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Remote Diagnostic Technologies Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

During internal testing, Regulatory Compliance issues were identified regarding the Monitor and Power Supplies for fluid ingress and basic safety issues.

FDA root cause category: Under Investigation by firm

What the company is doing

On 02/17/2023, the firm sent an "URGENT Medical Device Correction" Letter via priority mail to customers informing them that during internal testing, Regulatory Compliance issues regarding Fluid Ingress and Basic Safety issues were identified with the Tempus Pro Monitor, AC Mains Power Supply, and the Vehicle Adaptor. Customers are instructed to: " As there is currently no indication that any of the issues have resulted in a degradation in product performance, Philips advises that the Tempus Pro devices should remain in service. " Continue to follow the User/Operator Manual, including: o Take care to ensure that water or liquids are not spilt over the device or into its ventilation holes in the side corners. o Check connector covers, particularly the capnometer door cover, to ensure they close and latch acceptably. " Post this Urgent Medical Device Correction letter on or near your Tempus Pro device. " Complete and return the Urgent Medical Device Correction response form included, no later than 30 days from receipt. Philips will be providing a new and compliant AC Mains Power Supply to all affected customers. A new and compliant Vehicle Adaptor will also be provided to affected customers with devices that include that option. Philips anticipates providing this solution to customers in the second half of 2023. For questions or further information, contact the local Philips Representative, or Philip at (800)263-3342.

Recall details

Recall number
Z-1321-2023
Status
Open, Classified
Initiated by the company
Feb 23, 2023
Posted by the FDA
Apr 1, 2023
Recalling firm
Remote Diagnostic Technologies Ltd.
Quantity
5,540 units
Product code
MHX
FDA submissions
K201746
Distribution
Worldwide Distribution: U.S.(nationwide): AL, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI, WV, WY and O.U.S.(foreign): United Arab Emirates, Afghanistan, Austria, Australia, Belgium, Brunei Darussalam, Bolivia, Canada, Switzerland, Czechia, Germany, Denmark, Spain, France, United Kingdom, Hong Kong, Ireland, Israel, Iceland, Italy, Cayman Islands, Luxembourg, Malaysia, Netherlands, Norway, Portugal, Romania, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Mayotte, and South Africa.

Product description

Tempus Pro Patient Monitor is a portable vital signs monitor intended to be used by clinicians and medically qualified personnel, for the attended or unattended monitoring of single or multiple vital signs in clinical and pre hospital care applications. The affected power supplies are related to the following Tempus Pro Patient Monitoring units: Model Numbers: 00-1004 00-1004-R 00-1007 00-1007-R 00-1024-R 00-1026-R Model Numbers: 00-1004 00-1004-R 00-1007 00-1007-R 00-1024-R 00-1026-R

Code information (models, lots, serial numbers)

Model Numbers: 00-1004 00-1004-R UDI-DI Codes: 05060472440020 05060472442901 00-1007 00-1007-R UDI-DI Codes: 05060472440013 05060472442901 00-1024-R UDI-DI Codes: 05060472441027 05060472442925 00-1026-R UDI-DI Codes: 05060472441058 05060472442932 Lot Codes - All products manufactured are affected.

Devices named in this recall

4 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page