2D Perfusion, release R1.0.x, R1.1.x, R1.2, and R1.2.1 recallZ-1417-2023 · Philips Medical Systems Nederland B.V. · initiated Feb 20, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips Medical Systems Nederland B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There are technical issues related to signal generation and processing, which can lead to inaccurate presentations.

FDA root cause category: Software design

What the company is doing

A recall notification letter was issued February 20, 2023. The Urgent Medical Device Correction letter includes the following instructions for the Customer: " Stop using SmartPerfusion and 2D Perfusion. NOTE: This message does not impact (other parts of) the Philips Allura or Azurion systems or other tools within the Interventional Workspot. The Philips Allura and Azurion systems as well as other Interventional Workspot tools may continue to be used. " Place the Urgent Medical Device Correction Letter with the documentation of the system. " Circulate the notice to all users so that they are aware of the product issue. " Return the reply form attached to the Urgent Medical Device Correction letter to Philips to confirm that the users of the system have reviewed and understood it.

Recall details

Recall number
Z-1417-2023
Status
Open, Classified
Initiated by the company
Feb 20, 2023
Posted by the FDA
Apr 20, 2023
Recalling firm
Philips Medical Systems Nederland B.V.
Quantity
549 units (75 US, 474 OUS)
Product code
OWB
FDA submissions
K132147
Distribution
Worldwide - US Nationwide distribution.

Product description

2D Perfusion, release R1.0.x, R1.1.x, R1.2, and R1.2.1

Code information (models, lots, serial numbers)

UDI: (01)00884838059276 2D Perfusion with Allura Xper or Azurion model number: 722003, 722006, 722008, 722010, 722012, 722013, 722020, 722023, 722026, 722027 722028, 722029, 722035, 722038, 722039, 722058, 722059, 722064, 722067 722068, 722078, 722079, 722223, 722224, 722225, 722226, 722227 and 722228. Software License Keys: 459801071511 sw key 2D Perfusion (IW) 459800240601 sw key 2D Perfusion (XV)

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 91806).