GRAFTONand GRAFTON PLUSDemineralized Bone Matrix (DBM) recallZ-1324-2023 · Medtronic Sofamor Danek USA Inc · initiated Feb 8, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Sofamor Danek USA Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.
FDA root cause category: Process control
What the company is doing
Medtronic Sofamor Danek notified customers of the recall on about 02/08/2023, via letter. Customers were instructed to identify and quarantine any affected products, return any affected products, complete and return the Customer Confirmation Form, and provide the letter to all personnel that should be aware or where affected products have been transferred/distributed.
Recall details
- Recall number
- Z-1324-2023
- Status
- Open, Classified
- Initiated by the company
- Feb 8, 2023
- Posted by the FDA
- Apr 4, 2023
- Recalling firm
- Medtronic Sofamor Danek USA Inc
- Quantity
- 5308 units
- Product code
- MBP
- FDA submissions
- K051195
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Austria, Brunei Darussalam, Canada, Chile, Colombia, Denmark, Dominican Republic, Ecuador, Finland, Ghana, Greece, Guatemala, Hong Kong, India, Ireland, Israel, Jamaica, Japan, Jordan, Korea, Republic Of, Kuwait, Lebanon, Liechtenstein, Luxembourg, Macao, Malaysia, Mexico, Mongolia, Namibia, Netherlands, New Zealand, Norway, Oman, Panama, Paraguay, Peru, Philippines, Portugal, Puerto Rico, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, Taiwan, Thailand, Trinidad And Tobago, United Arab Emirates, United Kingdom, Viet Nam.
Product description
Medtronic Grafton Crunch a) DBM S44105 GRAFTON 5CC CRUNCH, REF S44105; b) DBM S44115 GRAFTON 15CC CRUNCH, REF S44115; c) DBM T44105 5CC GRAFTON CRUNCH, REF T44105; d) DBM T44105INT 5CC GRAFTON CRUNCH, REF T44105INT; e) DBM T44115 15CC GRAFTON CRUNCH, REF T44115; f) DBM T44115INT 15CC GRAFTON CRUNCH, REF T44115INT; Demineralized Bone Matrix (DBM) product
Code information (models, lots, serial numbers)
Show the full code information (64k characters)
a) DBM S44105 GRAFTON 5CC CRUNCH, REF S44105, GTIN 00613994961099, Serial Numbers: A41055, A41991; b) DBM S44115 GRAFTON 15CC CRUNCH, REF S44115, GTIN 00613994961082, Serial Numbers: A46383, A43751, A43940, A44572; c) DBM T44105 5CC GRAFTON CRUNCH, REF T44105, GTIN 00643169123120, Serial Numbers: A38507-041, A38507-042, A38507-043, A38507-044, A38507-045, A38507-046, A38507-047, A38507-048, A38507-049, A38507-050, A40379-096, A40379-097, A40379-098, A40379-099, A40379-100, A40379-101, A40379-102, A40379-103, A40379-104, A40379-105, A40379-106, A40379-107, A40379-108, A40379-109, A40379-110, A40379-111, A40379-112, A40379-113, A40379-114, A40379-115, A40379-116, A41167-031, A41167-032, A41167-033, A41167-036, A41167-037, A41167-038, A41167-039, A41167-040, A41167-041, A41167-042, A41167-043, A41167-044, A41167-045, A41167-046, A41167-047, A41167-048, A41167-049, A41167-050, A41167-051, A41167-052, A41167-053, A41167-054, A41167-055, A41167-056, A41167-057, A41167-058, A41167-059, A41167-060, A41167-061, A41167-062, A41167-063, A41167-064, A41167-065, A41167-066, A41167-067, A41167-068, A41167-070, A41169-181, A41169-182, A41169-183, A41169-184, A41169-185, A41169-186, A41169-187, A41169-189, A41169-192, A41169-193, A41169-194, A41169-195, A41169-196, A41169-197, A41169-198, A41169-199, A41169-200, A41169-201, A41169-202, A41169-203, A41169-204, A41568-135, A41568-136, A41568-137, A41568-138, A41568-139, A41568-140, A41568-141, A41568-142, A41568-143, A41568-144, A41568-145, A41568-146, A41568-147, A41568-148, A41568-149, A41989-232, A41989-233, A41989-234, A41989-235, A41989-236, A41989-237, A41989-238, A41989-239, A41989-240, A41989-241, A41989-242, A41989-243, A41989-244, A41989-245, A41989-246, A41989-247, A41989-248, A41989-249, A41989-250, A41989-251, A41989-252, A41989-253, A41989-254, A41989-255, A42288-123, A42288-124, A42288-125, A42288-126, A42288-127, A42288-128, A42288-129, A42288-130, A42288-131, A42288-132, A42288-133, A42288-134, A42288-135, A42288-136, A42288-137, A42288-138, A42288-139, A42288-140, A42288-141, A42288-142, A42288-143, A42288-144, A42288-145, A42288-146, A42288-147, A42288-148, A42288-149, A42288-150, A42288-151, A42288-152, A42753-135, A42753-136, A42753-137, A42753-138, A42753-139, A42753-140, A42753-141, A42753-142, A42753-143, A42753-144, A42753-145, A42753-146, A42753-147, A42753-148, A42753-149, A42753-150, A42753-151, A42833-077, A42833-078, A42833-079, A42833-080, A42833-081, A42833-082, A42833-083, A42833-084, A42833-085, A42833-086, A42833-087, A42833-088, A42833-089, A42833-090, A42833-091, A42833-092, A42934-052, A42934-053, A42934-054, A42934-055, A42934-056, A42934-057, A42934-058, A42934-059, A42934-060, A42934-061, A42934-062, A42934-063, A42934-064, A42934-065, A42934-066, A42934-067, A42934-068, A42934-069, A42934-070, A42934-071, A42934-072, A42934-073, A42934-074, A42934-075, A42934-076, A42934-077, A42934-078, A42934-079, A42934-080, A42934-081, A42934-082, A42934-083, A42934-084, A42984-055, A42984-056, A42984-057, A42984-058, A42984-059, A42984-060, A42984-061, A43116-077, A43116-078, A43116-079, A43116-080, A43116-081, A43116-082, A43116-083, A43116-084, A43116-085, A43116-086, A43116-087, A43116-088, A43116-089, A43116-090, A43116-091, A43116-092, A43116-093, A43116-094, A43116-095, A43116-096, A43344-174, A43344-175, A43344-176, A43344-177, A43905-096, A43905-097, A43905-098, A43905-099, A43905-100, A43905-101, A43905-102, A43905-103, A43905-104, A43905-105, A43905-106, A43905-107, A43905-108, A43905-109, A43905-110, A43905-111, A43905-112, A43905-113, A43905-114, A43905-115, A43905-116, A43905-117, A44372-105, A44372-106, A44372-107, A44372-108, A44372-109, A44372-110, A44372-111, A44372-112, A44372-113, A44372-114, A44372-115, A44372-116, A44372-117, A44372-118, A44372-119, A44372-120, A44372-121, A44372-122, A44372-123, A44372-124, A44372-125, A44372-126, A44372-127, A44372-128, A44852-074, A44852-075, A44852-076, A44852-077, A44852-078, A44852-079, A44852-080, A44852-081, A44852-082, A44852-083, A44852-084, A44852-085, A44852-086, A44852-087, A44852-088, A44852-089, A44852-090, A45132-059, A45132-060, A45132-061, A45132-062, A45132-063, A45132-064, A45132-065, A45132-066, A45132-067, A45132-068, A45132-069, A45132-070, A45132-071, A45132-072, A45132-073, A45132-074, A45132-075, A45132-076, A45132-077, A45132-078, A45132-079, A45281-101, A45281-102, A45281-103, A45281-104, A45281-105, A45281-106, A45281-107, A45281-108, A45281-109, A45281-110, A45281-111, A45281-112, A45281-113, A45281-114, A45281-115, A45281-116, A45364-089, A45364-090, A45364-091, A45364-092, A45364-093, A45364-094, A45364-095, A45364-096, A45364-097, A45364-098, A45364-099, A45364-100, A45364-101, A45364-102, A45364-103, A45364-104, A45364-105, A45364-106, A45364-107, A45364-108, A45364-109, A45364-110, A45364-111, A45364-112, A45364-113, A45364-114, A45364-115, A45364-116, A45364-117, A45364-118, A45364-119, A45364-120, A45607-089, A45607-090, A45607-091, A45607-092, A45607-093, A45607-094, A45607-095, A45607-096, A45607-098, A45607-099, A45607-100, A45607-102, A45607-103, A45607-104, A45607-105, A45607-106, A45607-107, A45607-108, A45607-110, A45607-111, A45607-112, A45607-113, A45607-114, A45607-115, A45607-116, A45607-117, A45607-118, A45607-119, A46579-166, A46579-167, A46579-168, A46579-169, A46579-170, A46579-171, A46579-172, A46579-173, A46579-174, A46579-175, A46579-176, A46579-177, A46579-178, A46579-179, A46579-180, A46579-181, A46579-182, A46579-183, A46579-184, A46579-185, A46579-186, A46579-187, A46579-188, A46579-189, A47032-099, A47032-100, A47032-101, A47032-102, A47032-103, A47032-104, A47032-105, A47032-107, A47032-108, A47032-109, A47032-110, A47032-111, A47032-112, A47032-113, A47032-114, A47032-116, A47032-117, A47032-118, A47032-119, A47087-021, A47087-022, A47087-023, A47087-027, A47087-028, A47087-029, A47087-030, A47087-031, A47087-032, A47087-033, A47087-034, A47087-035, A47087-036, A47087-037, A47087-038, A47087-039, A47087-040, A47087-041, A47087-042, A47087-043, A47087-044, A47087-045, A47087-046, A47087-047, A47087-049, A47087-050, A47087-051, A47087-052, A47087-053, A47087-054, A47087-055, A47087-056, A47087-057, A47087-058, A47087-059, A47129-073, A47129-074, A47129-075, A47129-076, A47129-077, A47129-078, A47129-079, A47129-080, A47129-081, A47129-082, A47129-083, A47129-084, A47129-085, A47129-086, A47129-087, A47129-088, A47129-089, A47129-090, A47129-091, A47129-092, A47129-093, A47129-094, A47129-095, A47129-096, A47129-097, A47129-098, A47129-099, A47129-100, A47129-101, A47129-102, A47129-103, A47129-104, A48098-037, A48098-038, A48098-039, A48098-040, A48098-041, A48098-042, A48098-043, A48098-044, A48098-045, A48098-046, A48098-047, A48098-048, A48145-050, A48145-051, A48145-052, A48145-053, A48145-054, A48145-055, A48145-056, A48145-057, A48145-058, A48145-059, A48145-060, A48145-061, A48145-062, A48145-063, A48145-064, A48145-065, A48145-066, A48145-067, A48145-068, A48145-069, A48145-070, A48145-071, A48145-072, A48145-073, A48145-074, A48145-075, A48145-076, A48145-077, A48145-078, A48145-079, A48145-080, A48593-138, A48593-139, A48593-140, A48593-141, A48593-142, A48593-143, A48593-144, A48593-145, A48593-146, A48593-147, A48593-148, A48593-149, A48593-150, A48593-151, A48593-152, A48593-153, A48593-154, A48593-155, A48593-156, A48593-157, A48593-158, A48593-159, A48593-160, A48593-161, A48593-162, A48593-163, A48593-164, A48991-120, A48991-121, A48991-122, A48991-123, A48991-124, A48991-125, A48991-126, A48991-127, A48991-128, A48991-129, A48991-130, A48991-131, A48991-132, A48991-133, A48991-134, A48991-135, A48991-136, A48991-137, A48991-138, A48991-139, A48991-140, A48991-141, A48991-142, A48991-143, A48991-144, A48991-145, A48991-146, A48991-147, A48991-148, A48991-149, A48991-150, A48991-151, A48991-152, A48991-153, A48991-154, A48991-155, A48991-156, A48991-157, A48991-158, A48991-159, A48991-160, A48991-161, A48991-162, A49650-046, A49650-047, A49650-048, A49650-049, A49650-050, A49650-051, A49650-052, A49650-053, A49650-054, A49650-055, A49650-056, A49650-057, A49650-058, A49650-059, A49650-060, A49650-061, A49650-062, A49650-063, A49650-064, A49650-065, A49650-066, A49650-067, A49650-068, A49650-069, A49650-070, A49650-071, A49710-104, A49710-105, A49710-106, A49710-107, A49710-108, A49710-109, A49710-110, A49710-111, A49747-113, A49747-114, A49747-115, A49747-116, A49747-117, A49747-118, A49747-119, A49747-120, A49747-121, A49747-122, A49747-123, A49747-124, A49747-125, A49747-126, A49747-127, A49747-128, A49747-129, A49747-130, A49747-131, A49747-132, A49747-133, A49747-134, A49747-135, A49747-136, A49864-050, A49864-051, A49864-052, A49864-053, A49864-054, A49864-055, A49864-056, A49864-057, A49864-058, A49864-059, A49864-060, A49864-061, A49864-062, A50081-080, A50081-081, A50081-082, A50081-083, A50081-084, A50081-085, A50081-086, A50081-087, A50081-088, A50081-089, A50081-090, A50081-091, A50081-092, A50081-093, A50081-094, A50081-095, A50081-096, A50081-097, A50081-098, A50081-099, A50081-100, A50081-101, A50580-026, A50580-027, A50580-028, A50580-029, A50580-030, A50580-031, A50580-032, A50580-033, A50580-034, A50580-035, A50580-036, A50580-037, A50580-038, A50580-039, A50580-040, A50580-041, A50580-042, A50580-043, A50580-044, A50580-045, A50580-046, A50580-047, A50844-064, A50844-065, A50844-066, A50844-067, A50844-068, A50844-069, A50844-070, A50844-071, A50844-072, A50844-073, A50844-074, A51585-064, A51585-065, A51585-066, A51585-067, A51585-068, A51585-069, A51585-070, A51585-071, A51585-072, A51585-073, A51585-074, A51585-075, A51585-076, A51585-077, A51585-078, A51585-079, A51585-080, A51585-081, A51585-082, A51585-083, A51585-084, A51585-085, A51585-086, A51585-087, A51585-088, A51585-089, A51585-090, A51585-091, A51585-092, A51585-093, A51585-094, A51585-095, A51585-096, A52421-050, A52421-051, A52421-052, A52421-053, A52421-054, A52421-055, A52421-056, A52421-057, A52421-058, A52421-059, A52421-060, A52421-061, A52421-062, A52421-063, A52421-064, A52421-065, A52421-066, A52421-067, A52421-068, A52421-069, A52421-070, A52421-071, A52421-072, A52421-073, A52421-074, A52421-075, A52421-076, A52421-077, A52421-078, A52421-079, A52421-080, A52421-081, A52421-082, A52421-083, A52421-084, A52421-085, A52421-086, A52421-087, A52421-088, A52421-089, A52421-090, A52460-181, A52460-182, A52460-183, A52460-184, A52460-185, A52460-186, A52460-187, A52460-188, A52460-189, A52460-190, A52711-031, A52711-032, A52711-033, A52711-034, A52711-035, A52711-036, A52711-037, A52711-038, A52711-039, A52711-040, A52711-041, A52711-042, A52711-043, A52711-044, A52711-045, A52711-046, A52711-047, A52711-048, A52711-049, A52711-050, A52711-051, A52711-052, A52711-053, A52711-054, A52711-055, A52711-056, A52711-057, A52711-058, A52711-059, A52711-060, A52711-061, A52711-062, A52711-063, A52711-064, A52711-065, A52711-066, A52711-067, A53585-026, A53585-027, A53585-028, A53585-029, A53585-030, A53585-031, A53585-032, A53585-033, A53585-034, A53585-035, A53585-036, A53585-037, A53585-038, A53585-039, A53585-040, A53585-041, A53585-042, A53585-043, A53585-044, A53585-045, A53585-046, A53585-047, A53585-048, A53585-049, A53585-050, A53585-051, A53585-053, A53751-053, A53751-054, A53751-055, A53751-056, A53751-057, A53751-058, A53751-059, A53751-060, A53751-061, A53751-062, A53751-063, A53751-064, A53751-065, A53751-066, A53751-067, A53751-068, A53751-069, A53751-070, A53751-071, A53751-072, A53751-073, A53751-074, A53751-075, A53751-076, A53751-077, A53751-078, A53751-079, A53751-080, A53751-081, A53756-127, A53756-128, A53756-129, A53756-130, A53756-131, A53756-132, A53756-133, A53756-134, A53756-135, A53756-136, A53756-137, A53756-138, A53756-139, A53756-140, A53756-141, A53756-142, A53756-143, A53756-144, A53756-145, A53756-146, A53756-147, A53756-148, A53756-149, A53756-150, A53756-151, A53756-152, A53756-153, A53756-154, A54021-026, A54021-027, A54021-028, A54021-029, A54021-030, A54021-031, A54021-032, A54021-033, A54021-034, A54021-035, A54021-036, A54021-037, A54021-038, A54021-039, A54021-040, A54021-041, A54021-042, A54021-043, A54021-044, A54119-097, A54119-098, A54119-099, A54119-100, A54119-101, A54119-102, A54119-103, A54119-104, A54119-105, A54119-106, A54119-107, A54119-108, A54119-109, A54464-113, A54464-114, A54464-115, A54675-026, A54675-027, A54675-028, A54675-029, A54675-030, A54675-031, A54675-032, A54675-033, A54675-034, A54675-035, A54675-036, A54675-037, A54675-038, A54675-039, A54675-040, A54675-041, A54675-042, A54675-043, A54675-044, A54675-045, A54675-046, A54675-047, A54675-048, A54675-049, A54675-050, A54675-051, A54675-052, A54675-053, A54675-054, A54675-055, A54675-056, A54675-057, A54675-058, A54675-059, A54675-060, A54675-061, A54675-062, A54675-063, A54675-064, A55091-031, A55091-032, A55091-033, A55091-034, A55091-036, A55091-037, A55091-038, A55091-039, A55091-040, A55091-041, A55091-042, A55091-043, A55091-044, A55091-045, A55091-046, A55091-047, A55091-048, A55091-049, A55091-050, A55091-051, A55091-052, A55091-053, A55091-054, A55091-055, A55091-056, A55091-058, A55091-059, A55091-060, A55091-061, A55091-062, A55091-063, A55091-064; d) DBM T44105INT 5CC GRAFTON CRUNCH, REF T44105INT, GTIN 00643169965034, Serial Numbers: A38507-051, A38507-052, A38507-053, A38507-054, A38507-055, A38507-056, A38507-057, A38507-058, A38507-059, A38507-060, A38507-061, A38507-062, A38507-063, A38507-064, A38507-065, A38507-066, A38507-067, A39972-046, A39972-047, A39972-048, A39972-049, A39972-050, A39972-051, A39972-052, A39972-053, A39972-054, A39972-055, A39972-056, A39972-057, A39972-058, A39972-059, A39972-060, A39972-061, A39972-062, A39972-063, A39972-064, A39972-065, A39972-066, A39972-067, A39972-068, A39972-069, A39972-070, A39972-071, A39972-072, A39972-073, A39972-074, A39972-075, A39972-076, A39972-077, A39972-078, A39972-079, A39972-080, A41213-031, A41213-032, A41213-033, A41213-034, A41213-035, A41213-036, A41213-037, A41213-038, A41213-039, A41213-040, A41213-041, A41213-042, A41213-043, A41213-044, A41213-045, A41213-046, A41213-047, A41213-048, A41213-049, A41213-050, A41213-051, A41213-052, A41213-053, A41213-054, A41213-055, A41516-026, A41516-027, A41516-028, A41516-029, A41516-030, A41516-031, A41516-032, A41516-033, A41516-034, A41516-035, A41516-036, A41516-037, A41516-038, A41516-039, A41516-040, A41516-041, A41516-042, A41516-043, A41516-045, A41516-047, A41516-048, A41516-049, A41516-050, A41516-052, A41516-053, A41516-056, A42221-281, A42221-282, A42221-283, A42323-034, A42323-035, A42323-036, A42323-037, A42323-038, A42323-039, A42323-040, A42323-041, A42323-042, A42323-043, A42323-044, A42323-045, A42323-046, A42323-047, A42323-048, A42323-049, A42323-050, A42323-051, A42323-052, A42323-053, A42323-054, A42323-055, A42323-056, A42323-057, A42323-058, A42323-059, A42753-152, A42753-153, A42753-154, A42753-155, A42753-156, A42753-157, A42760-066, A42760-067, A42760-068, A42760-069, A42760-071, A42760-073, A42760-075, A42760-076, A42760-077, A42760-078, A42760-081, A42760-082, A42760-084, A42760-085, A42760-086, A42760-087, A42760-088, A42760-089, A42760-090, A42760-091, A42760-092, A42760-093, A42760-094, A42934-087, A42934-088, A42934-089, A42934-090, A42984-033, A42984-034, A42984-035, A42984-036, A42984-037, A42984-038, A42984-039, A42984-040, A42984-041, A42984-042, A42984-043, A42984-044, A42984-045, A42984-046, A42984-047, A42984-048, A42984-049, A42984-050, A42984-051, A42984-052, A42984-053, A42984-054, A43103-183, A43103-184, A43103-185, A43103-186, A43103-187, A43103-188, A43103-189, A43103-190, A43103-191, A43103-192, A43103-193, A43103-194, A43103-195, A43103-196, A43103-197, A43103-198, A43103-199, A43103-200, A43103-201, A43103-202, A43103-203, A43103-204, A43103-205, A43103-206, A43103-207, A43103-208, A43103-209, A43103-210, A43344-155, A43344-156, A43344-157, A43344-158, A43344-159, A43344-160, A43344-161, A43344-163, A43344-164, A43344-165, A43344-168, A43344-169, A43344-170, A43344-171, A43344-172, A43344-173, A43546-101, A43546-102, A43546-103, A43546-104, A43546-105, A43546-106, A43546-107, A43546-108, A43546-109, A43546-110, A43546-111, A43546-112, A43546-113, A43546-114, A43546-115, A43546-116, A43571-095, A43571-096, A43571-097, A43571-098, A43571-099, A43571-100, A43571-101, A43571-102, A43571-103, A43571-104, A43571-105, A43571-106, A43571-107, A43571-108, A43571-109, A43571-110, A43571-111, A43571-112, A43571-113, A43571-114, A44852-091, A44852-092, A44852-093, A44852-094, A44852-095, A44852-096, A45084-086, A45084-087, A45084-088, A45084-089, A45084-090, A45084-091, A45084-092, A45084-093, A45084-094, A45084-095, A45084-096, A45084-097, A45084-098, A45084-099, A45084-100, A45084-101, A45084-102, A45084-103, A45084-104, A45084-105, A45084-106, A45084-107, A45084-108, A45084-109, A45084-110, A45084-111, A45084-112, A45084-113, A45084-114, A45084-115, A46628-221, A46628-222, A46628-223, A46628-224, A46628-225, A46628-226, A46628-227, A46628-228, A46628-229, A46628-230, A46628-231, A46628-232, A46628-233, A46628-234, A46628-235, A46628-236, A46628-237, A46628-238, A46628-239, A46628-240, A46628-241, A46628-242, A46628-243, A46628-244, A46628-245, A46628-246, A46628-247, A46628-248, A46628-249, A46628-250, A46628-251, A47081-045, A47081-046, A47081-047, A47081-048, A47081-049, A47081-050, A47081-051, A47081-052, A47081-053, A47081-054, A47081-055, A47081-056, A47081-057, A47081-058, A47081-059, A47081-060, A47087-060, A47087-061, A47087-062, A47087-063, A47087-064, A47087-065, A47087-066, A47087-067, A47087-068, A47087-069, A47087-070, A47087-071, A47087-072, A47087-073, A47087-074, A48098-049, A48098-050, A48098-051, A48098-052, A48098-053, A48098-054, A48098-055, A48098-056, A48098-057, A48098-058, A48098-059, A48098-060, A48098-061, A48098-062, A49710-112, A49710-113, A49710-114, A49710-115, A49710-116, A49710-117, A49710-119, A49710-120, A49710-121, A49710-122, A49710-123, A49710-124, A49710-125, A49710-126, A49710-127, A49747-137, A49747-138, A49747-139, A49747-140, A49747-141, A49747-142, A49747-143, A49747-144, A49747-145, A50082-058, A50082-059, A50082-060, A50082-061, A50082-062, A50082-063, A50082-064, A50082-065, A50082-066, A50082-067, A50082-068, A50082-069, A50082-070, A50082-071, A50082-072, A50082-073, A50082-074, A50082-075, A50082-076, A50082-077, A50082-078, A50082-079, A50082-080, A50082-081, A50082-082, A50082-083, A50082-084, A50082-085, A50082-086, A50082-087, A50082-088, A50082-089, A50968-031, A50968-032, A50968-033, A50968-034, A50968-035, A50968-036, A50968-037, A50968-038, A50968-039, A50968-040, A50968-041, A50968-042, A50968-043, A50968-044, A50968-045, A50968-046, A50968-047, A50968-048, A50968-049, A50968-050, A50968-051, A50968-052, A50968-053, A50968-054, A50968-055, A51981-026, A51981-027, A51981-028, A51981-029, A51981-030, A51981-031, A51981-032, A51981-033, A51981-034, A51981-035, A51981-036, A51981-037, A51981-038, A51981-039, A51981-040, A51981-041, A51981-042, A51981-043, A51981-044, A51981-045, A51981-046, A51981-047, A51981-048, A52460-162, A52460-163, A52460-164, A52460-165, A52460-166, A52460-167, A52460-168, A52460-169, A52460-170, A52460-171, A52460-173, A52460-180, A54119-110, A54119-111, A54119-112, A54119-113, A54119-114, A54119-115, A54119-116, A54119-117, A54119-118, A54119-119, A54119-120, A54119-121, A54464-086, A54464-087, A54464-088, A54464-089, A54464-093, A54464-094, A54464-095, A54464-103, A54464-107; e) DBM T44115 15CC GRAFTON CRUNCH, REF T44115, GTIN 00643169123113, Serial Numbers: A26631-051, A26631-052, A26631-053, A26631-054, A26631-055, A26631-056, A26631-057, A38377-125, A38377-126, A38377-127, A38377-128, A38377-129, A38377-130, A38377-131, A38377-132, A38377-133, A38471-041, A38471-042, A38471-043, A38471-044, A38471-045, A38471-046, A38471-047, A38472-047, A38472-048, A38472-049, A38472-050, A38472-051, A38472-052, A38796-048, A38796-049, A38796-050, A38796-051, A38796-052, A38796-053, A38796-054, A38796-055, A38796-056, A38796-057, A38796-058, A38796-059, A38796-060, A38796-061, A38796-062, A38796-063, A38796-064, A38796-065, A38796-066, A38796-067, A39047-108, A39047-109, A39047-110, A39047-111, A39047-112, A39047-113, A39091-085, A39091-086, A39091-087, A39091-088, A39091-089, A39091-090, A39091-091, A39091-092, A39091-093, A39091-094, A39111-182, A39111-183, A39111-184, A39111-185, A39111-186, A39111-187, A39111-188, A39111-189, A39111-190, A39111-191, A39111-192, A39111-193, A39111-194, A39111-195, A39111-196, A39202-053, A39202-054, A39202-055, A39202-056, A39202-057, A39202-058, A39202-059, A39202-060, A39202-061, A39202-062, A39202-063, A39202-064, A39202-065, A39202-066, A39202-067, A39202-068, A39202-069, A39202-070, A39202-071, A39245-046, A39245-047, A39245-048, A39245-049, A39245-050, A39245-051, A39245-052, A39245-053, A39245-054, A39245-055, A39245-056, A39283-114, A39283-115, A39283-116, A39283-117, A39283-118, A39283-119, A39413-267, A39413-268, A39413-269, A39413-270, A39413-271, A39413-272, A39413-273, A39413-274, A39413-275, A39413-276, A39413-277, A39512-026, A39512-027, A39512-028, A39512-029, A39512-030, A39512-031, A39512-032, A39512-033, A39512-034, A39512-035, A39512-036, A39512-037, A39512-038, A39512-039, A39512-040, A39512-041, A39512-042, A39512-043, A39512-044, A39512-045, A39512-046, A39512-047, A39516-092, A39516-093, A39516-094, A39516-095, A39516-096, A39520-103, A39520-104, A39520-105, A39520-106, A39520-107, A39520-108, A39520-109, A39520-110, A39520-111, A39811-170, A39811-171, A39811-172, A39811-173, A39811-174, A39811-175, A39811-176, A39811-177, A39811-178, A39811-179, A39811-180, A39866-164, A39866-165, A39866-166, A39866-167, A39866-168, A39866-169, A39866-170, A39866-171, A39866-172, A39866-173, A39867-049, A39867-050, A39867-051, A39867-052, A39867-053, A39867-054, A39867-055, A39867-056, A39867-057, A39962-052, A39962-053, A39962-054, A39962-055, A39962-056, A39962-057, A39962-058, A40005-103, A40005-104, A40005-105, A40005-106, A40005-107, A40005-108, A40005-109, A40005-110, A40005-111, A40005-112, A40108-144, A40108-145, A40108-146, A40108-147, A40108-148, A40108-149, A40108-150, A40108-151, A40108-152, A40108-153, A40108-154, A40108-155, A40108-156, A40112-053, A40112-054, A40112-055, A40112-056, A40112-057, A40112-058, A40112-059, A40112-060, A40112-061, A40112-062, A40112-063, A40138-042, A40138-043, A40138-044, A40138-045, A40138-046, A40138-047, A40138-048, A40138-049, A40173-162, A40173-163, A40173-164, A40173-165, A40173-166, A40173-167, A40173-168, A40173-169, A40173-170, A40246-056, A40246-057, A40246-058, A40246-059, A40246-060, A40246-061, A40246-062, A40246-063, A40309-097, A40309-098, A40309-099, A40309-100, A40309-101, A40309-102, A40309-103, A40309-104, A40309-105, A40309-106, A40309-107, A40309-108, A40309-109, A40309-110, A40309-111, A40453-026, A40453-027, A40453-028, A40453-029, A40453-030, A40453-031, A40453-032, A40453-033, A40453-034, A40453-035, A40543-075, A40543-076, A40543-077, A40543-078, A40543-079, A40543-080, A40543-081, A40543-082, A40543-083, A40543-084, A40543-085, A40652-126, A40652-127, A40652-128, A40652-129, A40652-130, A40652-131, A40733-040, A40733-041, A40733-042, A40733-043, A40733-044, A40733-045, A40733-046, A40733-047, A40733-048, A40733-049, A40733-050, A40733-051, A40733-052, A40733-053, A40736-081, A40736-082, A40736-083, A40736-084, A40736-085, A40736-086, A40736-087, A40736-088, A40736-089, A40736-090, A40736-091, A40913-086, A40913-087, A40913-088, A40913-089, A40913-090, A40913-091, A40913-092, A40913-093, A40913-094, A40913-095, A40913-096, A40915-101, A40915-102, A40915-103, A40915-104, A40915-105, A40915-106, A40915-107, A40915-108, A40915-109, A40915-110, A40915-111, A40915-112, A40915-113, A40915-114, A40915-115, A40915-116, A40915-117, A40915-118, A40953-174, A40953-175, A40953-176, A40953-177, A40953-178, A40953-179, A40953-180, A40953-181, A40953-182, A40989-196, A40989-197, A40989-198, A40989-199, A40989-200, A40989-201, A40989-202, A40989-203, A40989-204, A40989-205, A40989-206, A40999-031, A40999-032, A40999-033, A40999-034, A40999-035, A40999-036, A40999-037, A40999-038, A41012-047, A41012-048, A41012-049, A41012-050, A41012-051, A41012-052, A41012-053, A41012-054, A41012-055, A41012-056, A41012-057, A41012-058, A41012-059, A41012-060, A41012-061, A41012-062, A41012-063, A41270-121, A41270-122, A41270-123, A41270-124, A41270-125, A41270-126, A41270-127, A41270-128, A41270-129, A41270-130, A41270-131, A41270-132, A41270-133, A41270-134, A41270-135, A41270-136, A41270-137, A41270-138, A41270-139, A41270-140, A41273-031, A41273-032, A41273-033, A41273-034, A41273-035, A41273-036, A41397-031, A41397-032, A41397-033, A41397-034, A41397-035, A41397-036, A41400-052, A41400-053, A41400-054, A41400-055, A41400-056, A41400-057, A41400-058, A41400-059, A41400-060, A41439-177, A41439-178, A41439-179, A41439-180, A41439-181, A41464-210, A41464-211, A41464-212, A41464-213, A41464-214, A41464-215, A41464-216, A41464-217, A41464-218, A41472-174, A41472-175, A41472-176, A41472-177, A41472-178, A41472-179, A41472-180, A41472-181, A41472-182, A41472-183, A41630-031, A41630-032, A41630-033, A41630-034, A41630-035, A41630-036, A41630-037, A41630-038, A41630-039, A41630-040, A41630-041, A41630-042, A41630-043, A41630-044, A41630-045, A41630-046, A41630-047, A41630-048, A41784-086, A41784-087, A41784-088, A41784-089, A41784-090, A41784-091, A41784-092, A41784-093, A41784-094, A41784-095, A41784-096, A41784-097, A41879-091, A41879-092, A41879-093, A41879-094, A41879-095, A41879-096, A41879-097, A41879-098, A41879-099, A41879-100, A41879-101, A41879-102, A41879-103, A41879-104, A41905-026, A41905-027, A41905-028, A41905-029, A41905-030, A41905-031, A41905-032, A41905-033, A41905-034, A41991-068, A41991-069, A41991-070, A41991-071, A41991-072, A41991-073, A41991-074, A41991-075, A41991-076, A41991-077, A41991-078, A41995-118, A41995-119, A41995-120, A41995-121, A41995-122, A41995-123, A41995-124, A41995-125, A41995-126, A41995-127, A41995-128, A41995-129, A41995-130, A41995-131, A41995-132, A42080-046, A42080-047, A42080-048, A42080-049, A42080-050, A42080-051, A42080-052, A42080-053, A42080-054, A42080-055, A42102-046, A42102-047, A42102-048, A42102-049, A42102-050, A42102-051, A42102-052, A42102-053, A42102-054, A42102-055, A42102-056, A42102-057, A42102-058, A42102-059, A42102-060, A42102-061, A42193-088, A42193-089, A42193-090, A42193-091, A42193-092, A42193-093, A42193-094, A42193-095, A42193-096, A42225-043, A42225-044, A42225-045, A42225-046, A42225-047, A42225-048, A42225-049, A42225-050, A42225-051, A42225-052, A42225-053, A42225-054, A42225-055, A42225-056, A42225-057, A42225-058, A42225-059, A42225-060, A42225-061, A42234-061, A42234-062, A42234-063, A42234-064, A42234-065, A42234-066, A42234-067, A42234-068, A42251-078, A42251-079, A42251-080, A42251-081, A42251-082, A42251-083, A42251-084, A42251-085, A42251-086, A42251-087, A42251-088, A42251-089, A42251-090, A42251-091, A42251-092, A42282-133, A42282-134, A42282-135, A42282-136, A42282-137, A42282-138, A42282-139, A42282-140, A42282-141, A42282-142, A42282-143, A42282-144, A42539-046, A42539-047, A42539-048, A42539-049, A42539-050, A42539-051, A42539-052, A42539-053, A42539-054, A42539-055, A42539-056, A42607-057, A42607-058, A42607-059, A42607-060, A42607-061, A42607-062, A42607-063, A42607-064, A42607-065, A42607-066, A42607-067, A42607-068, A42607-069, A42607-070, A42607-071, A42607-072, A42607-073, A42607-074, A42607-075, A42607-076, A42607-077, A42661-092, A42661-093, A42661-094, A42661-095, A42661-096, A42661-097, A42662-129, A42662-130, A42662-131, A42662-132, A42662-133, A42662-134, A42662-135, A42662-136, A42662-137, A42665-085, A42665-086, A42665-087, A42665-088, A42665-089, A42665-090, A42665-091, A42665-092, A42665-093, A42665-094, A42665-095, A42665-096, A42665-097, A42665-098, A42665-099, A42665-100, A42754-147, A42754-148, A42754-149, A42754-150, A42754-151, A42754-152.A42754-153, A42754-154, A42754-155, A42791-064, A42791-065, A42791-066, A42791-067, A42791-068, A42791-069, A42791-070, A42791-071, A42791-072, A42791-073, A42791-074, A42791-075, A42791-076, A42791-077, A42791-078, A42791-079, A42826-053, A42826-054, A42826-055, A42826-056, A42826-057, A42826-058, A42826-059, A42826-060, A42826-061, A42826-062, A42826-063, A42826-064, A42826-065, A42872-053, A42872-054, A42872-055, A42872-056, A42872-057, A42872-058, A42872-059, A42872-060, A42872-061, A42872-062, A42872-063, A42872-064, A42872-065, A42872-066, A42872-067, A42872-068, A42872-069, A42872-070, A42872-071, A42932-026, A42932-027, A42932-028, A42932-029, A42932-030, A42932-031, A42932-032, A42932-033, A42932-034, A42932-035, A42932-036, A42932-037, A42935-155, A42935-156, A42935-157, A42935-158, A42935-159, A42935-160, A42935-161, A42935-162, A42935-163, A42935-164, A42972-089, A42972-090, A42972-091, A42972-092, A42972-093, A42972-094, A42972-095, A42972-096, A42972-097, A42972-098, A42972-099, A42972-100, A42972-101, A42972-102, A42982-031, A42982-032, A42982-033, A42982-034, A42982-035, A42982-036, A42982-037, A42982-038, A43058-034, A43058-035, A43058-036, A43058-037, A43058-038, A43058-039, A43058-040, A43058-041, A43058-042, A43058-043, A43058-044, A43058-045, A43058-046, A43058-047, A43058-048, A43058-049, A43100-081, A43100-082, A43100-083, A43100-084, A43100-085, A43100-086, A43100-087, A43100-088, A43100-089, A43100-090, A43100-091, A43104-056, A43104-057, A43104-058, A43104-059, A43104-060, A43104-061, A43104-062, A43104-063, A43104-064, A43104-065, A43104-066, A43104-067, A43104-068, A43104-069, A43104-070, A43104-071, A43104-072, A43104-073, A43112-064, A43112-065, A43112-066, A43112-067, A43112-068, A43112-069, A43112-070, A43112-071, A43112-072, A43112-073, A43191-057, A43191-058, A43191-059, A43191-060, A43191-061, A43191-062, A43191-063, A43191-064, A43191-065, A43191-066, A43191-067, A43191-068, A43191-069, A43191-070, A43191-071, A43191-072, A43193-040, A43193-041, A43193-042, A43193-043, A43193-044, A43193-045, A43193-046, A43193-047, A43193-048, A43193-049, A43193-050, A43193-051, A43284-221, A43284-222, A43284-223, A43284-224, A43284-225, A43284-226, A43284-227, A43285-046, A43285-047, A43285-048, A43285-049, A43285-050, A43285-051, A43285-052, A43285-053, A43285-054, A43285-055, A43285-056, A43285-057, A43285-058, A43298-044, A43298-045, A43298-046, A43298-047, A43298-048, A43298-049, A43298-050, A43298-051, A43298-052, A43298-053, A43298-054, A43298-055, A43298-056, A43298-057, A43298-058, A43340-036, A43340-037, A43340-038, A43340-039, A43340-040, A43340-041, A43340-042, A43340-043, A43340-044, A43340-045, A43340-046, A43340-047, A43340-048, A43340-049, A43340-050, A43340-051, A43340-052, A43340-053, A43374-040, A43374-041, A43374-042, A43374-043, A43374-044, A43374-045, A43374-046, A43374-047, A43374-048, A43374-049, A43374-050, A43395-054, A43395-055, A43395-056, A43395-057, A43395-058, A43395-059, A43395-060, A43395-061, A43497-157, A43497-158, A43497-159, A43497-160, A43497-161, A43497-162, A43497-164, A43497-165, A43501-026, A43501-027, A43501-028, A43501-029, A43501-030, A43501-031, A43501-032, A43501-033, A43501-034, A43501-035, A43577-095, A43577-096, A43577-097, A43577-098, A43577-099, A43577-100, A43577-101, A43577-102, A43577-103, A43577-104, A43577-105, A43634-112, A43634-113, A43634-114, A43634-115, A43634-116, A43634-117, A43634-118, A43634-119, A43634-120, A43698-131, A43698-132, A43698-133, A43698-134, A43698-135, A43698-136, A43698-137, A43698-138, A43698-139, A43698-140, A43698-141, A43698-142, A43698-143, A43698-144, A43751-121, A43751-122, A43751-123, A43751-124, A43751-125, A43751-126, A43751-127, A43904-036, A43904-037, A43904-038, A43904-039, A43904-040, A43904-041, A43904-042, A43904-043, A43904-044, A43904-045, A43909-113, A43909-114, A43909-115, A43909-116, A43909-117, A43909-118, A43909-119, A43909-120, A43909-121, A43936-089, A43936-090, A43936-091, A43936-092, A43936-093, A43936-094, A43936-095, A43936-096, A43936-097, A43936-098, A43936-099, A43936-100, A43936-101, A43936-102, A43936-103, A43936-104, A43936-105, A43936-106, A43936-107, A43938-046, A43938-047, A43938-048, A43938-049, A43938-050, A43938-051, A43938-052, A43938-053, A43938-054, A43938-055, A43938-056, A43938-057, A43938-058, A43938-059, A43938-060, A43938-061, A43948-045, A43948-046, A43948-047, A43948-048, A43948-049, A43948-050, A43948-051, A43948-052, A43948-053, A43948-054, A43948-055, A43948-056, A43948-057, A43948-058, A43983-105, A43983-106, A43983-107, A43983-108, A43983-109, A43983-110, A43983-111, A43983-112, A44032-036, A44032-037, A44032-038, A44032-039, A44032-040, A44032-041, A44032-042, A44032-043, A44032-044, A44032-045, A44032-046, A44131-187, A44131-188, A44131-189, A44131-190, A44131-191, A44131-192, A44131-193, A44131-194, A44131-195, A44131-196, A44131-197, A44131-198, A44173-165, A44173-166, A44173-167, A44173-168, A44173-169, A44173-170, A44173-171, A44173-172, A44173-173, A44173-174, A44186-213, A44186-214, A44186-215, A44186-216, A44186-217, A44186-218, A44186-219, A44186-220, A44186-221, A44186-222, A44219-056, A44219-057, A44219-058, A44219-059, A44219-060, A44219-061, A44219-062, A44219-063, A44219-064, A44219-065, A44219-066, A44219-067, A44219-068, A44219-069, A44219-070, A44219-071, A44222-051, A44222-052, A44222-053, A44222-054, A44222-055, A44222-056, A44222-057, A44222-058, A44222-059, A44222-060, A44222-061, A44222-062, A44222-063, A44222-064, A44222-065, A44222-066, A44222-067, A44222-068, A44222-069, A44231-031, A44231-032, A44231-033, A44231-034, A44231-035, A44231-036, A44231-037, A44231-038, A44231-039, A44231-040, A44231-041, A44231-042, A44235-113, A44235-114, A44235-115, A44235-116, A44235-117, A44235-118, A44235-119, A44235-120, A44235-121, A44315-240, A44315-241, A44315-242, A44315-243, A44315-244, A44315-245, A44315-246, A44315-247, A44315-248, A44315-249, A44315-250, A44315-251, A44315-252, A44315-253, A44315-254, A44315-255, A44315-256, A44315-257, A44366-184, A44366-185, A44366-186, A44366-187, A44366-188, A44366-189, A44366-190, A44444-098, A44444-099, A44444-100, A44444-101, A44444-102, A44444-103, A44444-104, A44444-105, A44444-106, A44444-107, A44444-108, A44572-090, A44572-091, A44572-092, A44646-036, A44646-037, A44646-038, A44646-039, A44646-040, A44646-041, A44646-042, A44646-043, A44646-044, A44646-045, A44646-046, A44646-047, A44646-048, A44646-049, A44646-050, A44646-051, A44646-052, A44646-053, A44649-176, A44649-177, A44649-178, A44649-179, A44649-180, A44649-181, A44649-182, A44649-183, A44650-171, A44650-172, A44650-173, A44650-174, A44650-175, A44650-176, A44650-177, A44650-178, A44650-179, A44650-180, A44650-181, A44650-182, A44650-183, A44650-184, A44650-185, A44650-186, A44650-187, A44650-188, A44650-189, A44713-240, A44713-241, A44713-242, A44713-243, A44713-244, A44713-245, A44713-246, A44724-050, A44724-051, A44724-052, A44724-053, A44724-054, A44724-055, A44724-056, A44724-057, A44724-058, A44724-059, A44724-060, A44724-061, A44724-062, A44724-063, A44724-064, A44724-065, A44724-066, A44724-067, A44724-068, A44724-069, A44754-042, A44754-043, A44754-044, A44754-045, A44754-046, A44754-047, A44754-048, A44754-049, A44754-050, A45039-036, A45039-037, A45039-038, A45039-039, A45039-040, A45039-041, A45039-042, A45039-043, A45039-044, A45039-045, A45039-046, A45124-056, A45124-057, A45124-058, A45124-059, A45124-060, A45124-061, A45124-062, A45124-063, A45124-064, A45124-065, A45134-102, A45134-103, A45134-104, A45134-105, A45134-106, A45134-107, A45134-108, A45134-109, A45134-110, A45134-111, A45135-046, A45135-047, A45135-048, A45135-049, A45135-050, A45135-051, A45135-052, A45135-053, A45135-054, A45135-055, A45135-056, A45135-057, A45135-058, A45135-059, A45181-135, A45181-136, A45181-137, A45181-138, A45181-139, A45181-140, A45181-141, A45181-142, A45181-143, A45181-144, A45181-145, A45185-046, A45185-047, A45185-048, A45185-049, A45185-050, A45185-051, A45185-052, A45185-053, A45185-054, A45185-055, A45185-056, A45185-057, A45185-058, A45185-059, A45185-060, A45185-061, A45185-062, A45185-063, A45185-064, A45412-093, A45412-094, A45412-095, A45412-096, A45412-097, A45412-098, A45412-099, A45412-100, A45412-101, A45413-036, A45413-037, A45413-038, A45413-039, A45413-040, A45413-041, A45413-042, A45413-043, A45413-044, A45413-045, A45413-046, A45415-114, A45415-115, A45415-116, A45415-117, A45415-118, A45415-119, A45415-120, A45415-121, A45415-122, A45499-104, A45499-105, A45499-106, A45499-107, A45499-108, A45499-109, A45499-110, A45499-111, A45499-112, A45580-041, A45580-042, A45580-043, A45580-044, A45580-045, A45580-046, A45580-047, A45580-048, A45580-049, A45580-050, A45580-051, A45580-052, A45580-053, A45581-052, A45581-053, A45581-054, A45581-055, A45581-056, A45581-057, A45581-058, A45581-059, A45581-060, A45581-061, A45591-031, A45591-032, A45591-033, A45591-034, A45591-035, A45591-036, A45591-037, A45591-038, A45591-039, A45594-036, A45594-037, A45594-038, A45594-039, A45594-040, A45594-041, A45594-042, A45594-043, A45594-044, A45594-045, A45594-046, A45594-047, A45594-048, A45597-118, A45597-119, A45597-120, A45597-121, A45597-122, A45597-123, A45597-124, A45597-125, A45597-126, A45597-127, A45598-046, A45598-047, A45598-048, A45598-049, A45598-050, A45598-051, A45598-052, A45598-053, A45598-054, A45598-055, A45598-056, A45598-057, A45598-058, A45598-059, A45598-060, A45598-061, A45598-062, A45598-063, A45606-041, A45606-042, A45606-043, A45606-044, A45606-045, A45606-046, A45606-047, A45606-048, A45606-049, A45606-050, A45606-051, A45613-036, A45613-037, A45613-038, A45613-039, A45613-040, A45613-041, A45613-042, A45613-043, A45613-044, A45671-124, A45671-125, A45671-126, A45671-127, A45671-128, A45671-129, A45671-130, A45671-131, A45671-132, A45671-133, A45746-115, A45746-116, A45746-117, A45746-118, A45746-119, A45746-120, A45746-121, A45746-122, A45746-123, A45746-124, A45755-026, A45755-027, A45755-028, A45755-029, A45755-030, A45755-031, A45755-032, A45755-033, A45755-034, A45755-035, A45755-036, A45755-037, A45755-038, A45755-039, A45755-040, A45755-041, A45874-036, A45874-037, A45874-038, A45874-039, A45874-040, A45874-041, A45874-042, A45874-043, A45874-044, A45874-045, A45874-046, A45874-047, A45874-048, A45874-049, A45874-050, A45926-075, A45926-076, A45926-077, A45926-078, A45926-079, A45926-080, A45926-081, A45926-082, A45926-083, A45941-021, A45941-022, A45941-023, A45941-024, A45941-025, A45941-026, A45941-027, A45941-028, A45941-029, A45941-030, A45941-031, A45941-032, A45941-033, A45941-034, A45941-035, A45941-036, A45941-037, A45941-038, A45941-039, A45946-111, A45946-112, A45946-113, A45946-114, A45946-115, A45946-116, A45946-117, A45952-177, A45952-178, A45952-179, A45952-180, A45952-181, A45952-182, A45952-183, A45952-184, A45952-185, A45962-036, A45962-037, A45962-038, A45962-039, A45962-040, A45962-041, A45962-042, A45962-043, A45962-044, A45962-045, A45962-046, A45962-047, A45962-048, A45964-031, A45964-032, A45964-033, A45964-034, A45964-035, A45964-036, A45964-037, A45964-038, A45964-039, A45964-040, A45964-041, A45964-042, A45964-043, A46006-059, A46006-060, A46006-061, A46006-062, A46006-063, A46006-064, A46006-065, A46006-066, A46006-067, A46006-068, A46021-041, A46021-042, A46021-043, A46021-044, A46021-045, A46021-046, A46021-047, A46021-048, A46165-036, A46165-037, A46165-038, A46165-039, A46165-040, A46165-041, A46165-042, A46165-043, A46165-044, A46165-045, A46165-046, A46165-047, A46165-048, A46165-049, A46165-050, A46171-055, A46171-056, A46171-057, A46171-058, A46171-059, A46171-060, A46281-041, A46281-042, A46281-043, A46281-044, A46281-045, A46281-046, A46281-047, A46281-048, A46281-049, A46281-050, A46281-051, A46281-052, A46281-053, A46281-054, A46281-055, A46283-026, A46283-027, A46283-028, A46283-029, A46283-030, A46283-031, A46335-048, A46335-049, A46335-050, A46335-051, A46335-052, A46335-053, A46335-054, A46335-055, A46335-056.A46383-046, A46383-047, A46383-048, A46383-049, A46383-050, A46383-051, A46383-052, A46383-053, A46383-054, A46383-055, A46383-056, A46383-057, A46383-058, A46383-059, A46507-031, A46507-032, A46507-033, A46507-034, A46507-035, A46507-036, A46507-037, A46507-038, A46507-039, A46507-040, A46507-041, A46507-042, A46512-047, A46512-048, A46512-049, A46512-050, A46512-051, A46512-052, A46512-053, A46512-054, A46512-055, A46512-056, A46512-057, A46512-058, A46512-059, A46512-060, A46512-061, A46512-062, A46512-063, A46512-064, A46515-095, A46515-096, A46515-097, A46515-099, A46515-100, A46515-101, A46515-102, A46515-103, A46515-104, A46515-105, A46515-106, A46515-107, A46515-108, A46539-202, A46539-203, A46539-204, A46539-205, A46539-206, A46539-207, A46539-208, A46539-209, A46623-021, A46623-022, A46623-023, A46623-024, A46623-025, A46623-026, A46623-027, A46623-028, A46623-029, A46623-030, A46623-031, A46623-032, A46623-033, A46623-034, A46623-035, A46718-041, A46718-042, A46718-043, A46718-044, A46718-045, A46718-046, A46718-047, A46718-048, A46718-049, A46718-050, A46718-051, A46718-052, A46718-053, A46718-054, A46718-055, A46727-130, A46727-131, A46727-132, A46727-133, A46727-135, A46727-136, A46727-137, A46727-138, A46776-041, A46776-042, A46776-043, A46776-044, A46776-045, A46776-046, A46776-047, A46776-048, A46776-049, A46776-050, A46776-051, A46776-052, A46776-053, A46776-054, A46776-055, A46776-056, A46776-057, A46833-031, A46833-032, A46833-033, A46833-034, A46833-035, A46833-036, A46833-037, A46833-038, A46833-039, A46833-040, A46833-041, A46833-042, A46833-043, A46849-116, A46849-117, A46849-118, A46849-119, A46849-120, A46849-121, A46849-122, A46849-123, A46849-124, A46849-125, A46849-126, A46890-073, A46890-074, A46890-075, A46890-076, A46890-077, A46890-078, A46890-079, A46890-080, A46991-087, A46991-088, A46991-089, A46991-090, A46991-091, A46991-092, A46991-093, A46991-094, A46995-065, A46995-066, A46995-067, A46995-068, A46995-069, A46995-070, A46995-071, A46995-072, A46995-073, A46995-074, A46995-075, A46995-076, A46995-077, A46995-078, A46995-079, A46995-080, A46995-081, A46997-031, A46997-032, A46997-033, A46997-034, A46997-035, A46997-036, A46997-037, A46997-038, A46997-039, A47084-036, A47084-037, A47084-038, A47084-039, A47084-040, A47084-041, A47084-042, A47084-043, A47084-044, A47084-045, A47084-046, A47084-047, A47084-048, A47084-049, A47084-050, A47084-051, A47084-052, A47084-053, A47084-054, A47282-048, A47282-049, A47282-050, A47282-051, A47282-052, A47282-053, A47282-054, A47282-055, A47282-056, A47282-057, A47282-058, A47282-059, A47282-060, A47282-061, A47282-062, A47282-063, A47394-031, A47394-032, A47394-033, A47394-034, A47394-035, A47394-036, A47394-037, A47394-038, A47429-041, A47429-042, A47429-043, A47429-044, A47429-045, A47429-046, A47429-047, A47429-048, A47429-049, A47430-031, A47430-032, A47430-033, A47430-034, A47430-035, A47430-036, A47430-037, A47430-038, A47430-039, A47430-040, A47430-041, A47430-042, A47430-043, A47431-116, A47431-117, A47431-118, A47431-119, A47431-120, A47431-121, A47431-122, A47431-123, A47431-124, A47431-125, A47431-126, A47591-080, A47591-081, A47591-082, A47591-083, A47591-084, A47591-085, A47591-086, A47591-087, A47591-088, A47591-089, A47591-090, A47591-091, A47591-092, A47591-093, A47591-094, A47591-095, A47591-096, A47591-097, A47591-098, A47591-099, A47591-100, A47591-101, A47597-031, A47597-032, A47597-033, A47597-034, A47597-035, A47597-036, A47597-037, A47597-038, A47597-039, A47597-040, A47599-036, A47599-037, A47599-038, A47599-039, A47599-040, A47599-042, A47599-043, A47645-093, A47645-094, A47645-095, A47645-096, A47645-097, A47645-098, A47645-099, A47645-100, A47645-101, A47645-102, A47645-103, A47645-104, A47645-105, A47649-046, A47649-047, A47649-048, A47649-049, A47649-050, A47649-052, A47649-053, A47649-054, A47724-026, A47724-027, A47724-028, A47724-029, A47724-030, A47724-031, A47724-032, A47724-033, A47724-034, A47724-035, A47724-036, A47724-037, A47724-038, A47724-039, A47724-040, A47724-041, A47781-021, A47781-022, A47781-023, A47781-024, A47781-025, A47781-026, A47781-027, A47781-028, A47781-029, A47938-031, A47938-032, A47938-033, A47938-034, A47938-035, A47938-036, A47938-037, A47938-038, A47938-039, A47938-040, A47938-041, A47938-042, A47938-043, A47938-044, A47992-049, A47992-050, A47992-051, A47992-052, A47992-053, A47992-054, A47992-055, A47992-056, A47992-057, A47992-058, A47992-059, A47992-060, A47992-061, A47992-062, A47992-063, A47992-064, A47992-065, A48094-026, A48094-027, A48094-028, A48094-029, A48094-030, A48094-031, A48094-032, A48094-033, A48094-034, A48094-035, A48094-036, A48095-026, A48095-027, A48095-028, A48095-029, A48095-030, A48095-031, A48095-032, A48095-033, A48095-034, A48095-035, A48095-036, A48142-134, A48142-135, A48142-136, A48142-137, A48142-138, A48142-139, A48142-140, A48142-141, A48142-142, A48142-143, A48244-066, A48244-067, A48244-068, A48244-069, A48244-070, A48244-071, A48244-072, A48244-073, A48244-074, A48244-075, A48275-034, A48275-035, A48275-036, A48275-037, A48275-038, A48275-039, A48275-040, A48275-041, A48275-042, A48275-043, A48275-044, A48275-045, A48275-046, A48275-047, A48275-048, A48275-049, A48278-031, A48278-032, A48278-033, A48278-034, A48278-035, A48278-036, A48278-037, A48278-038, A48278-039, A48278-040, A48278-041, A48278-042, A48278-043, A48278-044, A48278-045, A48278-046, A48280-077, A48280-078, A48280-079, A48280-080, A48280-081, A48280-082, A48280-083, A48280-084, A48280-085, A48280-086, A48378-036, A48378-037, A48378-038, A48378-039, A48378-040, A48378-041, A48378-042, A48378-043, A48378-044, A48378-045, A48378-046, A48378-047, A48378-048, A48385-041, A48385-042, A48538-253, A48538-254, A48538-255, A48538-256, A48538-257, A48538-258, A48538-259, A48538-260, A48538-261, A48538-262, A48538-263, A48541-032, A48541-033, A48541-034, A48541-035, A48541-036, A48541-037, A48541-038, A48541-039, A48541-040, A48541-041, A48541-042, A48541-043, A48541-044, A48541-045, A48541-046, A48541-047, A48541-048, A48541-049, A48541-050, A48590-026, A48590-027, A48590-028, A48590-029, A48590-030, A48590-031, A48590-032, A48590-033, A48590-034, A48590-035, A48590-036, A48591-021, A48591-022, A48591-023, A48591-024, A48591-025, A48591-026, A48591-027, A48591-028, A48591-029, A48652-091, A48652-092, A48652-093, A48652-094, A48652-095, A48652-096, A48652-097, A48652-098, A48652-099, A48652-100, A48729-069, A48729-070, A48729-071, A48729-072, A48729-073, A48729-074, A48729-075, A48729-076, A48729-077, A48729-078, A48730-062, A48730-063, A48730-064, A48730-065, A48730-066, A48730-067, A48730-068, A48730-069, A48730-070, A48730-071, A48730-072, A48730-073, A48731-041, A48731-042, A48731-043, A48731-044, A48731-045, A48731-046, A48731-047, A48731-048, A48731-049, A48731-050, A48731-051, A48997-041, A48997-042, A48997-043, A48997-044, A48997-045, A48997-046, A48997-047, A48997-048, A48997-049, A49002-041, A49002-042, A49002-043, A49002-044, A49002-045, A49002-046, A49002-047, A49002-048, A49002-049, A49280-041, A49280-042, A49280-043, A49280-044, A49280-045, A49280-046, A49280-047, A49280-048, A49280-049, A49280-050, A49345-026, A49345-027, A49345-028, A49345-029, A49345-030, A49345-031, A49345-032, A49345-033, A49345-034, A49647-036, A49647-037, A49647-038, A49647-039, A49647-040, A49647-041, A49647-042, A49647-043, A49647-044, A49706-193, A49706-194, A49706-195, A49706-196, A49706-197, A49706-198, A49706-199, A49706-200, A49706-201, A49706-202, A49706-203, A49706-204, A49708-103, A49708-104, A49708-105, A49708-106, A49708-107, A49708-108, A49708-109, A49708-110, A49708-111, A49708-112, A49708-113, A49708-114, A49864-063, A49864-064, A49864-065, A49864-066, A49864-067, A49864-068, A49864-069, A49864-070, A49864-071, A49864-072, A49864-073, A49864-074, A49913-026, A49913-027, A49913-028, A49913-029, A49913-030, A49913-031, A49913-032, A49976-074, A49976-075, A49976-076, A49976-077, A49976-078, A49976-079, A49976-080, A49976-081, A49976-082, A49976-083, A49976-084, A49976-085, A49976-086, A49976-087, A49976-088, A49976-089, A50311-061, A50311-062, A50311-063, A50311-064, A50311-065, A50311-066, A50311-067, A50311-068, A50311-069, A50311-070, A50405-050, A50405-051, A50405-052, A50405-053, A50405-054, A50405-055, A50405-056, A50405-057, A50405-058, A50405-059, A50405-060, A50407-134, A50407-135, A50407-136, A50407-137, A50407-138, A50407-139, A50407-140, A50407-141, A50407-142, A50408-060, A50408-061, A50408-062, A50408-063, A50408-064, A50408-065, A50408-066, A50408-067, A50408-068, A50408-069, A50408-070, A50408-071, A50408-072, A50468-052, A50468-053, A50468-054, A50468-055, A50468-056, A50468-057, A50721-026, A50721-027, A50721-028, A50721-029, A50721-030, A50721-031, A50721-032, A50721-033, A50721-034, A50721-035, A50721-036, A50721-037, A50721-038, A50721-039, A50755-079, A50755-080, A50755-081, A50755-082, A50755-083, A50755-084, A50755-085, A50755-086, A50755-087, A50757-042, A50757-043, A50757-044, A50757-045, A50757-046, A50757-047, A50757-048, A50757-049, A50757-050, A50757-051, A50757-052, A50757-053, A50757-054, A50757-055, A50757-056, A50757-057, A50757-058, A50844-075, A50844-076, A50844-082, A50844-083, A50844-084, A50844-085, A50844-086, A50844-087, A51014-031, A51014-032, A51014-033, A51014-034, A51014-035, A51014-036, A51014-037, A51014-038, A51014-039, A51014-040, A51014-041, A51015-059, A51015-060, A51015-061, A51015-062, A51015-063, A51015-064, A51015-065, A51015-066, A51015-067, A51015-068, A51015-069, A51015-070, A51015-071, A51054-058, A51054-059, A51054-060, A51054-061, A51054-062, A51054-063, A51054-064, A51054-065, A51054-066, A51054-067, A51054-068, A51054-069, A51054-070, A51166-038, A51166-039, A51166-040, A51166-041, A51166-042, A51166-043, A51166-044, A51166-045, A51166-046, A51172-026, A51172-027, A51172-028, A51172-029, A51172-030, A51172-031, A51172-032, A51172-033, A51172-034, A51172-035, A51172-036, A51172-037, A51172-038, A51172-039, A51172-040, A51182-171, A51182-172, A51182-173, A51182-174, A51182-175, A51182-176, A51182-177, A51182-178, A51182-179, A51182-180, A51182-181, A51182-182, A51182-183, A51182-184, A51182-185, A51182-186, A51246-031, A51246-032, A51246-033, A51246-034, A51246-035, A51246-036, A51246-037, A51246-038, A51246-039, A51246-040, A51246-041, A51246-042, A51246-043, A51246-044, A51442-048, A51442-049, A51442-050, A51442-051, A51442-052, A51442-053, A51442-054, A51442-055, A51442-056, A51442-057, A51442-058, A51442-059, A51442-060, A51547-026, A51547-027, A51547-028, A51547-029, A51547-030, A51547-031, A51547-032, A51547-033, A51547-034, A51547-035, A51547-036, A51547-037, A51547-038, A51547-039, A51547-040, A51604-176, A51604-177, A51604-178, A51604-179, A51604-180, A51604-181, A51604-182, A51604-183, A51604-184, A51604-185, A51604-186, A51604-187, A51604-188, A51604-189, A51608-021, A51608-022, A51608-023, A51608-024, A51608-025, A51608-026, A51608-027, A51608-028, A51608-029, A51608-030, A51608-031, A51608-032, A51608-033, A51709-036, A51709-037, A51709-038, A51709-039, A51709-040, A51709-041, A51709-042, A51709-043, A51710-062, A51710-063, A51710-064, A51710-065, A51710-066, A51710-067, A51710-068, A51710-069, A51710-070, A51710-071, A51710-072, A51710-073, A51710-074, A51710-075, A51710-076, A51710-077, A51930-094, A51930-095, A51930-096, A51930-097, A51930-098, A51930-099, A51930-100, A51930-101, A51930-102, A51930-103, A51942-135, A51942-136, A51942-137, A51942-138, A51942-139, A51942-140, A51942-141, A51942-142, A51942-143, A51942-144, A51942-145, A52248-039, A52248-040, A52248-041, A52248-042, A52248-043, A52248-044, A52248-045, A52248-046, A52248-047, A52248-048, A52248-049, A52248-050, A52248-051, A52248-052, A52248-053, A52248-054, A52248-055, A52248-056, A52248-057, A52250-041, A52250-042, A52250-043, A52250-044, A52250-045, A52250-046, A52250-047, A52250-048, A52250-049, A52250-050, A52250-051, A52250-052, A52250-053, A52250-054, A52250-055, A52250-056, A52341-036, A52341-037.A52341-038, A52341-039, A52341-040, A52341-041, A52341-042, A52341-043, A52341-044, A52341-045, A52341-046, A52341-047, A52341-048, A52341-050, A52341-051, A52341-052, A52341-053, A52524-051, A52524-052, A52524-053, A52524-054, A52524-055, A52524-056, A52524-057, A52524-058, A52524-059, A52524-060, A52524-061, A52524-062, A52524-063, A52524-064, A52653-045, A52653-046, A52653-047, A52653-048, A52653-049, A52653-050, A52653-051, A52653-052, A52653-053, A52653-054, A52654-036, A52654-037, A52654-038, A52654-039, A52654-040, A52654-041, A52654-042, A52654-043, A52654-044, A52654-045, A52654-046, A52716-174, A52716-175, A52716-176, A52716-177, A52716-178, A52716-179, A52716-180, A52716-181, A52784-036, A52784-037, A52784-038, A52784-039, A52784-040, A52784-041, A52784-042, A52784-043, A52784-044, A52784-045, A52784-046, A52784-047, A52784-048, A52784-049, A52784-050, A52784-051, A52857-039, A52857-040, A52857-041, A52857-042, A52857-043, A52857-044, A52857-045, A52857-046, A52857-047, A52857-048, A52857-049, A52857-050, A52857-051, A52862-036, A52862-037, A52862-038, A52862-039, A52862-040, A52862-041, A52862-042, A52862-043, A52862-044, A52862-045, A52862-046, A52862-047, A52862-048, A53112-041, A53112-042, A53112-043, A53112-044, A53112-045, A53112-046, A53112-047, A53112-048, A53112-049, A53112-050, A53112-051, A53112-052, A53112-053, A53112-054, A53112-055, A53112-056, A53113-063, A53113-064, A53113-065, A53113-066, A53113-067, A53113-068, A53113-069, A53113-070, A53113-071, A53113-072, A53113-073, A53212-041, A53212-042, A53212-043, A53212-044, A53212-045, A53212-046, A53212-047, A53212-048, A53212-049, A53212-050, A53220-046, A53220-047, A53220-048, A53220-049, A53220-050, A53220-051, A53220-052, A53220-053, A53220-054, A53220-055, A53220-056, A53220-057, A53220-058, A53220-059, A53220-060, A53242-116, A53242-117, A53242-118, A53242-119, A53242-120, A53242-121, A53242-122, A53242-123, A53453-021, A53453-022, A53453-023, A53453-024, A53453-025, A53453-026, A53453-027, A53453-028, A53453-029, A53581-041, A53581-042, A53581-043, A53581-044, A53581-045, A53581-046, A53581-047, A53581-048, A53581-049, A53581-050, A53581-051, A53581-052, A53581-053, A53581-054, A53581-055, A53581-056, A53581-057, A53641-042, A53641-043, A53641-044, A53641-045, A53709-041, A53709-042, A53709-043, A53709-044, A53709-045, A53709-046, A53709-047, A53709-048, A53709-049, A53709-050, A53709-051, A53709-052, A53709-053, A53709-054, A53712-021, A53712-022, A53712-023, A53712-024, A53712-025, A53712-026, A53712-027, A53712-028, A53712-029, A53712-030, A53712-031, A53712-032, A53712-033, A53743-044, A53743-045, A53743-046, A53743-047, A53743-048, A53743-049, A53743-050, A53743-051, A53743-052, A53743-053, A53743-054, A53745-036, A53745-037, A53745-038, A53745-039, A53745-040, A53745-041, A53745-042, A53745-043, A53745-044, A53745-045, A53745-046, A53745-047, A53745-048, A53745-049, A53805-036, A53805-037, A53805-038, A53805-039, A53805-040, A53805-041, A53805-042, A53805-043, A53805-044, A53805-045, A53805-046, A53805-047, A53805-048, A53805-049, A53805-050, A53808-045, A53808-046, A53808-047, A53808-048, A53808-049, A53808-050, A53808-051, A53808-052, A53808-053, A53808-054, A53808-055, A53808-056, A53808-057, A53808-058, A53808-059, A53808-060, A53809-113, A53809-114, A53809-115, A53809-116, A53809-117, A53809-118, A53809-119, A53809-120, A53809-121, A53966-049, A53966-050, A53966-051, A53966-052, A53966-053, A53966-054, A53966-055, A53966-056, A53966-057, A53966-058, A53966-059, A53966-060, A53966-061, A53966-062, A53966-063, A53966-064, A53987-051, A53987-052, A53987-053, A53987-054, A53987-055, A53987-056, A53987-057, A53987-058, A53987-059, A53987-060, A53987-061, A53987-062, A53987-063, A53987-064, A53987-065, A54023-031, A54023-032, A54023-033, A54023-034, A54023-035, A54023-036, A54023-037, A54023-038, A54023-039, A54023-040, A54023-041, A54123-036, A54123-037, A54123-038, A54123-039, A54123-040, A54123-041, A54123-042, A54123-043, A54123-044, A54124-021, A54124-022, A54124-023, A54124-024, A54124-025, A54124-026, A54124-027, A54124-028, A54124-029, A54125-031, A54125-032, A54125-033, A54125-034, A54125-035, A54125-036, A54125-037, A54125-038, A54125-039, A54125-040, A54233-031, A54233-032, A54233-033, A54233-034, A54233-035, A54233-036, A54233-037, A54233-038, A54233-039, A54233-040, A54308-031, A54308-032, A54308-033, A54308-034, A54308-035, A54308-036, A54308-037, A54308-038, A54308-039, A54308-040, A54308-041, A54308-042, A54308-043, A54308-044, A54315-036, A54315-037, A54315-038, A54315-039, A54315-040, A54315-041, A54315-042, A54315-043, A54315-044, A54315-045, A54315-046, A54315-047, A54315-048, A54315-049, A54315-050, A54315-051, A54315-052, A54342-041, A54342-042, A54342-043, A54342-044, A54342-045, A54342-046, A54342-047, A54342-048, A54342-049, A54342-050, A54342-051, A54342-052, A54342-053, A54342-054, A54342-055, A54342-056, A54342-057, A54342-058, A54342-059, A54342-060, A54342-061, A54342-062, A54426-046, A54426-047, A54426-048, A54426-049, A54426-050, A54426-051, A54426-052, A54426-053, A54426-054, A54426-055, A54426-056, A54426-057, A54426-058, A54426-059, A54426-060, A54426-061, A54426-062, A54426-063, A54426-064, A54426-065, A54674-021, A54674-022, A54674-023, A54674-024, A54674-025, A54674-026, A54674-027, A54674-028, A54674-029, A54674-030, A54753-031, A54753-032, A54753-033, A54753-034, A54753-035, A54753-036, A54753-037, A54753-038, A54753-039, A54753-040, A54759-021, A54759-022, A54759-023, A54759-024, A54759-025, A54759-026, A54759-027, A54759-028, A54759-029, A54759-030, A54759-031, A54759-032, A54759-033, A54759-034, A54759-035, A54759-036, A54827-033, A54827-034, A54827-035, A54827-036, A54827-037, A54827-038, A54827-039, A54827-040, A54827-041, A54827-042, A54827-043, A54960-031, A54960-032, A54960-033, A54960-034, A54960-035, A54960-036, A54960-037, A54960-038, A54960-039, A54999-091, A54999-092, A54999-093, A54999-094, A54999-095, A54999-096, A54999-097, A54999-099, A54999-100, A54999-101, A54999-102, A54999-103, A54999-104, A55011-026, A55011-027, A55011-028, A55011-029, A55011-030, A55011-031, A55011-032, A55011-033, A55011-034, A55011-035, A55011-036, A55011-037, A55011-038, A55011-039, A55011-040, A55016-041, A55016-042, A55016-043, A55016-044, A55016-045, A55016-046, A55016-047, A55016-048, A55016-049, A55094-026, A55094-027, A55094-028, A55094-029, A55094-030, A55094-031, A55094-032, A55094-033, A55094-034, A55094-035, A55094-036, A55095-036, A55095-037, A55095-038, A55095-039, A55095-040, A55095-041, A55095-042, A55095-043, A55095-044, A55095-045, A55095-046, A55095-047, A55095-048, A55095-049, A55099-041, A55099-042, A55099-043, A55099-044, A55099-045, A55099-046, A55099-047, A55099-048, A55099-049, A55099-050, A55099-051, A55158-031, A55158-032, A55158-033, A55158-034, A55158-035, A55158-036, A55158-037, A55158-038, A55158-039, A55312-026, A55312-027, A55312-028, A55312-029, A55312-030, A55312-031, A55312-032, A55312-033, A55312-034, A55312-035, A55312-036, A55313-031, A55313-032, A55313-033, A55313-034, A55313-035, A55313-036, A55313-037, A55313-038, A55313-039, A55313-040, A55313-041, A55313-042, A55313-043, A55313-044, A55313-045, A55313-046, A55314-031, A55314-032, A55314-033, A55314-034, A55314-035, A55314-036, A55314-037, A55314-038, A55314-039, A55314-040, A55314-041, A55314-042, A55314-043, A55314-044, A55314-045, A55314-046, A55320-021, A55320-022, A55320-023, A55320-024, A55320-025, A55320-026, A55320-027, A55320-028, A55320-029, A55320-030, A55320-031, A55320-032, A55320-033, A55320-034, A55368-021, A55368-022, A55368-023, A55368-024, A55416-026, A55416-027, A55416-028, A55416-029, A55416-030, A55416-031, A55416-032, A55416-033, A55416-034, A55416-035, A55416-036, A55416-037, A55416-038, A55416-039, A55448-041, A55448-042, A55448-043, A55448-044, A55448-045, A55448-046, A55448-047, A55448-048, A55448-049, A55448-050, A55448-051, A55448-052, A55448-053, A55448-054, A55448-055, A55448-056; f) DBM T44115INT 15CC GRAFTON CRUNCH, REF T44115INT, GTIN 00643169965041, Serial Numbers: A38958-122, A38958-123, A38958-124, A38958-125, A38958-126, A38958-127, A38958-128, A39666-174, A39666-175, A39666-176, A39666-177, A39666-178, A39666-179, A39666-180, A39666-181, A39666-182, A39666-183, A39962-059, A39962-060, A39962-061, A39962-062, A39962-063, A41397-037, A41397-038, A41397-039, A41397-040, A41397-041, A41435-048, A41435-052, A41435-053, A41435-054, A41435-055, A41435-056, A41435-057, A41435-058, A41707-026, A41707-027, A41707-028, A41707-029, A41707-030, A41707-031, A41707-032, A41707-033, A41707-034, A41707-035, A41707-036, A41707-037, A41707-038, A41707-039, A41707-040, A41707-041, A41707-042, A41710-079, A41710-080, A41710-081, A41710-082, A41710-083, A41710-084, A41710-085, A41710-086, A41710-087, A41710-088, A41710-089, A41710-090, A41710-091, A42282-145, A42282-146, A42282-147, A42282-148, A42282-149, A42283-161, A42283-162, A42283-163, A42283-164, A42283-165, A42283-166, A42283-167, A42283-168, A42283-169, A42283-170, A42283-171, A42283-172, A42325-031, A42325-032, A42325-033, A42325-034, A42325-035, A42325-036, A42325-037, A42325-038, A42325-040, A42325-041, A42325-042, A42325-043, A42325-044, A42325-045, A42503-101, A42503-102, A42503-103, A42503-104, A42503-105, A42503-106, A42503-107, A42503-108, A42503-109, A42503-110, A42616-266, A42616-267, A42616-268, A42616-269, A42616-270, A42616-271, A42616-272, A42616-273, A42616-274, A42616-275, A42940-064, A42940-065, A42940-066, A42940-067, A42940-068, A42940-069, A42940-070, A42940-071, A43023-191, A43023-192, A43023-193, A43023-194, A43023-195, A43023-196, A43023-197, A43023-198, A43023-199, A43023-200, A43023-201, A43300-087, A43300-088, A43300-089, A43300-090, A43300-091, A43300-092, A43300-093, A43300-094, A43300-095, A43300-096, A43300-097, A43300-098, A43696-059, A43696-060, A43696-061, A43696-062, A43696-063, A43696-064, A43696-065, A43696-066, A43696-067, A43740-157, A43740-158, A43740-159, A43740-160, A43740-161, A43740-162, A43740-163, A43740-164, A43740-165, A44192-041, A44192-042, A44192-043, A44192-044, A44192-045, A44192-046, A44192-047, A44192-048, A44192-049, A44192-050, A44192-051, A44192-052, A44192-053, A46021-049, A46021-050, A46171-061, A46171-062, A46171-063, A46171-064, A46171-065, A46171-066, A48385-043, A48385-044, A48385-045, A48385-046, A48385-047, A48385-048, A48385-049, A49638-031, A49638-032, A49638-033, A49638-034, A49638-035, A49638-036, A49638-037, A49638-038, A49638-039, A49638-040, A49638-041, A49638-042, A49638-043, A49638-044, A49709-175, A49709-176, A49709-177, A49709-178, A49709-179, A49709-180, A49709-181, A49709-182, A49709-183, A49709-184, A49709-185, A50083-090, A50083-091, A50083-092, A50083-093, A50083-094, A50083-095, A50083-096, A50083-097, A51016-026, A51016-027, A51016-028, A51016-029, A51016-030, A51016-031, A51016-032, A51016-033, A51016-034, A51016-035, A51016-036, A51016-037, A51016-038, A51016-039, A51016-040, A51016-041, A51247-042, A51247-043, A51247-044, A51247-045, A51247-046, A51247-047, A51247-048, A51247-049, A51247-050, A51247-051, A51247-052, A51247-053, A51247-054, A52561-069, A52561-078, A52561-083, A52561-084, A52561-085, A52561-086, A52812-041, A52812-042, A52812-043, A52812-044, A52812-045, A52812-046, A52812-047, A52812-049, A52812-050, A52812-051, A53641-035, A53641-038, A53641-039, A53641-040, A53641-041, A54024-051, A54024-052, A54024-053, A54024-054, A54024-055, A54024-056, A54024-057, A54024-058, A54024-059, A54024-060, A54024-061, A54024-062, A54024-063, A54024-064, A54024-067, A54024-068, A54024-069, A55317-026, A55317-027, A55317-028, A55317-029, A55317-030, A55317-031, A55317-032, A55317-033, A55317-034, A55317-035, A55317-036.
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T44115Medtronic Sofamor Danek, Inc. · Filler, bone void, osteoinduction (w/o human growth factor)2 recalls1 report
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T44105Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND1 recall0 reports
Product lines
Other devices under the same FDA submission
25 devices cleared under K051195 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T44135Medtronic Sofamor Danek, Inc. · Filler, bone void, osteoinduction (w/o human growth factor)2 recalls1 report
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T44150Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND2 recalls1 report
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T42275Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND2 recalls1 report
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T44145Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND2 recalls0 reports
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T42100Medtronic Sofamor Danek, Inc. · Filler, bone void, osteoinduction (w/o human growth factor)2 recalls0 reports
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T42200Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND2 recalls0 reports
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T42210Medtronic Sofamor Danek, Inc. · Filler, bone void, osteoinduction (w/o human growth factor)2 recalls0 reports
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T42280Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND2 recalls0 reports
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T43105Medtronic Sofamor Danek, Inc. · Filler, bone void, osteoinduction (w/o human growth factor)1 recall4 reports
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T43102Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND1 recall2 reports
- GRAFTON® DBM T44150INTMedtronic Sofamor Danek, Inc. · Filler, bone void, osteoinduction (w/o human growth factor)1 recall1 report
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T44125Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 91697).
- Z-1323-2023 · Open, ClassifiedMedtronic Grafton Putty: a) DBM S43101 GRAFTON 0.5CC PUTTY, REF S43101; b) DBM S43102 GRAFTON 1CC PUTTY, REF S43102; c) DBM S43103 GRAFTON 2.5CC PUTTY, REF S43103; d) DBM S…
- Z-1325-2023 · Open, ClassifiedMedtronic Grafton Flex: a) DBM S42090 GRAFTON 1.5CM X 1.5CM FLEX, REF S42090; b) DBM T42090 1.5CMX1.5CM FLEX GRAFTON, REF T42090; c) DBM T42090INT 1.5CMX1.5CM FLEX GRAFTON, REF T…
- Z-1326-2023 · Open, ClassifiedMedtronic Grafton Orthoblend: a) DBM S44125 5CC ORTHOBLEND LARGE DEFECT, REF S44125; b) DBM S44135 15CC ORTHOBLEND LARGE DEFECT, REF S44135; c) DBM S44145 5CC ORTHOBLEND SMALL DE…
- Z-1327-2023 · Open, ClassifiedMedtronic Grafton Matrix: a) DBM S42200 GRAFTON 2.5CMX10CM 2EA MATRIX, REF S42200; b) DBM S42210 GRAFTON 2.5CMX5CM 2EA MATRIX, REF S42210; c) DBM T42200 2.5CMX10CM 2 EACH GRAFTON…
- Z-1328-2023 · Open, ClassifiedMedtronic Grafton Matrix Strips: a) DBM S42275 8MMX1CMX10CM 2EA MATRIX STRPS, REF S42275; b) DBM S42280 8MMX1CMX20CM 2EA MATRX STRPS, REF S42280; c) DBM T42275 8MMX1CMX10CM 2 E…
- Z-1329-2023 · Open, ClassifiedMedtronic Grafton Plus Paste a) DBM S45001 GRAFTON 1CC PASTE, REF S45001; b) DBM S45005 GRAFTON 5CC PASTE, REF S45005; c) DBM S45010 GRAFTON 10CC PASTE, REF S45010; d) DBM…
- Z-1333-2023 · Open, ClassifiedMedtronic Grafton Gel: a) DBM S41120 GRAFTON GEL 1CC, REF S41120; b) DBM S41150 GRAFTON GEL 10CC, REF S41150; c) DBM T41110 0.5CC GRAFTON GEL, REF T41110; d) DBM T41110INT …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page