CoCr Head recallZ-1294-2023 · Zimmer, Inc. · initiated Feb 27, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Update the compatibility matrix as referred to in the Instructions for Use (IFU) for the CoCr Femoral Head XS- removes the compatibility with the Epsilon Durasul Constrained Acetabular Liners from the matrix due to the range of motion in flexion/extension being less than 100 degrees as recommended per an internationally recognized standard (ISO 21535:2009
FDA root cause category: Under Investigation by firm
What the company is doing
Zimmer Biomet issued Urgent Medical Device Correction letter on 2/27/23 via Fed'X .Letter to Distributors, Doctors, Risk Managers and states reason for recall, health risk and action to take: Hospital Responsibilities: 1. Review this field action notice and ensure that affected personnel are aware of the contents. 2. Complete Attachment 1 Certificate of Acknowledgement Form and send to [email protected]. This form must be returned even if you do not have affected products at your facility. 3. Retain a copy of the Attachment 1 Certificate of Acknowledgement Form with your field action notice records in the event of a compliance audit of your facility s documentation. 4. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected]. Surgeon Responsibilities: 1. Review this field action notice for awareness of the contents. 2. There are no specific patient monitoring instructions related to this notice that are recommended beyond your existing follow-up schedule. 3. Complete Attachment 1 Certificate of Acknowledgement Form and send to [email protected] . 4. Retain a copy of the Attachment 1 Certificate of Acknowledgement Form with your field action notice records in the event of a compliance audit of your facility s documentation. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an
Recall details
- Recall number
- Z-1294-2023
- Status
- Open, Classified
- Initiated by the company
- Feb 27, 2023
- Posted by the FDA
- Mar 29, 2023
- Recalling firm
- Zimmer, Inc.
- Quantity
- 1,168 untits
- Product code
- LPH
- FDA submissions
- K993259
- Distribution
- AL AZ CO FL LA MI OH OK TX Foreign: Canada
Product description
CoCr Femoral Head, XS, 38/-8, Taper 12/14-Intended to be used as a modular head component for articulation in total hip arthroplasty. Item Number: 01.01012.384
Code information (models, lots, serial numbers)
GTIN Number: 00889024283268 All lots
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- CoCr Head 01.01012.384Zimmer GmbH · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.