Color Cuff recallZ-1344-2023 · Stryker Corporation · initiated Mar 1, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Stryker Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Not able to achieve desired pressure and not holding pressure result from two different causes which have been reported to occur during initial set-up of the cuff onto the patient, may lead to operative site blood loss, hemorrhage or, in exceedingly rare situations, other major complications
FDA root cause category: Under Investigation by firm
What the company is doing
Stryker issued Urgent Medical Device Recall letter dated 2/27/23 via 2-day Fedex delivery. The purpose of this letter is to inform you of an increase in reported low pressure events for specific lots of Stryker Color Cuff Non-Sterile Disposable Tourniquet Cuffs and to reiterate the steps in the Instructions For Use (IFU) to mitigate the risk to achieve or maintain pressure. Letter states reason for recall, health risk and action to take: Please review all action steps below and complete the acknowledgement section (pg.3) of this letter and return by email to [email protected]. 1. Immediately review this notification and distribute this letter internally to all users of the Color Cuff Non-Sterile Disposable Tourniquet Cuffs. 2. Follow the subsequent steps from the Instructions For Use (IFU) to mitigate the risk of the cuff not being able to achieve or maintain pressure. Following these steps, you may continue to use the product as-is. a. Prior to using the Non-Sterile Tourniquet Cuffs, please review the Safety Directives, Inspection, and Instructions sections of the IFU. To view this IFU electronically, go to https://ifu.stryker.com/XHR/129532. b. Prior to connecting the Non-Sterile Disposable Tourniquet Cuffs to the Stryker SmartPump, please review the bullet points below from pages 18-20 of the IFU for the SmartPump Tourniquet System. For troubleshooting, please refer to page 28. To view this IFU electronically, go to https://ifu.stryker.com/XHR/129013. " ALWAYS monitor pneumatic connections during use, including the cuff(s) and cuff connector(s) with O-ring, the quick-lock connectors, and the fill line(s) and connectors. If leakage or damage is apparent, see Troubleshooting. " NOTE: If a cuff does not reach or maintain the target pressure value during initial inflation, discontinue using the cuff. See To Switch (or Backup) a Single Cuff. " ALWAYS respond to an alert condition before continuing to use the tourniquet pump. An ALERT symbol, a flashing
Recall details
- Recall number
- Z-1344-2023
- Status
- Open, Classified
- Initiated by the company
- Mar 1, 2023
- Posted by the FDA
- Apr 4, 2023
- Recalling firm
- Stryker Corporation
- Quantity
- 1707 units (packs of 10)
- Product code
- KCY
- Distribution
- US Nationwide distribution.
Product description
Stryker Color Cuff Non-Sterile Disposable Tourniquet Cuff -NS DISP 18X4,1BLA,2PRT QUICK Catalog number: 5921-218-235NS Indicated for use in surgical procedures that require the temporary occlusion of blood flow in a patient s extremities
Code information (models, lots, serial numbers)
GTIN: 07613327606164 Lot Numbers: 2022111802 2022112904 2022121301 2022121302 2022121905 2022121906 2022121907 2022121908
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Color Cuff 5921-218-235NSStryker Corporation · TOURNIQUET, PNEUMATIC1 recall2 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 91849).
- Z-1339-2023 · Open, ClassifiedStryker Color Cuff Non-Sterile Disposable Tourniquet Cuff -NS DISP 30X4,1BLA,1PRT QUICK Catalog number: 5921-030-135NS Indicated for use in surgical procedures that require the tem…
- Z-1340-2023 · Open, ClassifiedStryker Color Cuff Non-Sterile Disposable Tourniquet Cuff -NS DISP 30X4,1BLA,2PRT QUICK Catalog number: 5921-030-235NS Indicated for use in surgical procedures that require the te…
- Z-1341-2023 · Open, ClassifiedStryker Color Cuff Non-Sterile Disposable Tourniquet Cuff -NS DISP 34X4,1BLA,1PRT QUICK Catalog number: 5921-034-135NS Indicated for use in surgical procedures that require the te…
- Z-1342-2023 · Open, ClassifiedStryker Color Cuff Non-Sterile Disposable Tourniquet Cuff -NS DISP 34X4,1BLA,2PRT QUICK Catalog number: 5921-034-235NS Indicated for use in surgical procedures that require the te…
- Z-1343-2023 · Open, ClassifiedStryker Color Cuff Non-Sterile Disposable Tourniquet Cuff -NS DISP 18X4,1 BLA,1PRT QUICK Catalog number: 5921-218-135NS
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page