Chromic Gut recallZ-1388-2023 · Covidien, LP · initiated Mar 1, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Covidien, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medtronic was made aware of a supplier calibration issue during the contract sterilization process that may lead to unsterile product. Use of affected product may lead to an unspecified infection or, in the case of sutures used in the eye, pose a loss of vision risk.
FDA root cause category: Process control
What the company is doing
An "URGENT PRODUCT RECALL" letter dated 3/1/23 was sent to customers. Actions to be Taken: " Immediately identify and quarantine all unused affected product (Refer to Attachment A for identifying affected product). " Please complete and return the enclosed Customer Confirmation Form, even if you have no product to return and email to [email protected]. " Product purchased directly from Medtronic please contact [email protected] for Return Goods Authorization (RGA). Your local Medtronic Representative can assist you as necessary in initiating the return of this product. " Credit for the returned affected product will be issued based on the RGA number. " Return all unused affected product(s) in your inventory to: Medtronic, Attn: Field Returns Dept. 195 McDermott Road, North Haven, CT 06473 USA " If purchased from a distributor, contact your distributor directly to arrange for the return of the product back to your distributor. " Forward this notification to any entities within your organization or to any other organization where the potentially affected product has been transferred or further distributed. Local contact details: Adverse reactions or quality problems experienced with this product should be reported to FDA and Medtronic: " Online at http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm (form available to fax or mail), or Call FDA (800) FDA-1088 " Contact Medtronic Customer Service Support at [email protected] or call 800-962-9888, option 2. We regret any inconvenience this may cause. We are committed to patient safety and appreciate your prompt attention to this matter. If you have any questions regarding this communication, please contact your Medtronic Representative Customer Service at 800-962-9888, option 2.
Recall details
- Recall number
- Z-1388-2023
- Status
- Open, Classified
- Initiated by the company
- Mar 1, 2023
- Posted by the FDA
- Apr 12, 2023
- Recalling firm
- Covidien, LP
- Quantity
- 36204 units
- Product code
- GAL
- FDA submissions
- K142656
- Distribution
- Worldwide Distribution: US (nationwide) to States of: AR, CA, FL, GA, LA, MA, NC, NY, OH, WA; and countries of: Canada, Colombia, El Salvador, Indonesia, Nepal, Puerto Rico, Taiwan, and Thailand.
Product description
Chromic Gut Absorbable Sutures, SURGIGUT* 0 CHR GS-25 90CM X36 (CG-904), SURGIGUT* 1 CHR GS-24 90CM X36 (CG-915), CHROMIC GUT 4-0 75CM V20 (GG-121), CHROMIC GUT 3-0 75CM V20 X36 (GG-122), CHROMIC GUT 2-0 75CM V20 X36 (GG-123), CHROMIC GUT 2-0 75CM V26 (GG-127), CHROMIC GUT 4-0 45CM P12 X36 (SG-5637), CHROMIC GUT 5-0 45CM P-13 X12 (SG-5687G), CHROMIC GUT 5-0 75CM CV23 (UG-202), CHROMIC GUT 4-0 75CM CV23 X36 (UG-203)
Code information (models, lots, serial numbers)
GTIN: 20884521054001, 20884521054056, 20884521054889, 20884521054896, 20884521054896, 20884521054902, 20884521054902, 10884521054905, 20884521054940, 20884521055688, 20884521055770, 20884521056135, 10884521056145, 20884521056142, 20884521056142, 20884521056142; Lot: D2H1593ZY D2H1022ZY D2F1106ZY D2H1607ZY D2H1609ZY D2H0819ZY D2H2053ZY D2H2053ZY D2H2055ZY D2H1435FZY D2H0829FZY D2H0830FZY D2H0381ZY D2H1523ZY D2H1584ZY D2H1584ZY D2H1587ZY D2H1589ZY
Devices named in this recall
10 devices in the FDA device registry whose identifier appears in the recall's code information.
- Chromic Gut GG-122Covidien LP · SUTURE, ABSORBABLE, NATURAL1 recall12 reports
- Chromic Gut UG-202Covidien LP · SUTURE, ABSORBABLE, NATURAL1 recall7 reports
- Chromic Gut GG-123Covidien LP · SUTURE, ABSORBABLE, NATURAL1 recall6 reports
- Chromic Gut SG-5687GCovidien LP · SUTURE, ABSORBABLE, NATURAL1 recall6 reports
- Chromic Gut UG-203Covidien LP · SUTURE, ABSORBABLE, NATURAL1 recall6 reports
- Chromic Gut GG-121Covidien LP · SUTURE, ABSORBABLE, NATURAL1 recall3 reports
- Chromic Gut GG-127Covidien LP · SUTURE, ABSORBABLE, NATURAL1 recall3 reports
- Chromic Gut CG-904Covidien LP · SUTURE, ABSORBABLE, NATURAL1 recall0 reports
- Chromic Gut CG-915Covidien LP · SUTURE, ABSORBABLE, NATURAL1 recall0 reports
- Chromic Gut SG-5637Covidien LP · SUTURE, ABSORBABLE, NATURAL1 recall0 reports
Product lines
Other devices under the same FDA submission
38 devices cleared under K142656 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Chromic Gut G-1780KCovidien LP · SUTURE, ABSORBABLE, NATURAL1 recall8 reports
- Plain Gut SG-5686Covidien LP · SUTURE, ABSORBABLE, NATURAL1 recall6 reports
- Chromic Gut G-1766KCovidien LP · SUTURE, ABSORBABLE, NATURAL1 recall0 reports
- McKesson S635GXMckesson Medical-Surgical Inc. · Suture, Absorbable, Natural13 reports
- Chromic Gut SG-5644GCovidien LP · SUTURE, ABSORBABLE, NATURAL11 reports
- Mild Chromic Gut G-3810Covidien LP · SUTURE, ABSORBABLE, NATURAL10 reports
- Plain Gut G-1735-KCovidien LP · SUTURE, ABSORBABLE, NATURAL7 reports
- Chromic Gut CG-905Covidien LP · SUTURE, ABSORBABLE, NATURAL6 reports
- Chromic Gut CG-811Covidien LP · SUTURE, ABSORBABLE, NATURAL4 reports
- Chromic Gut SG-5687Covidien LP · SUTURE, ABSORBABLE, NATURAL4 reports
- Chromic Gut SG-635Covidien LP · SUTURE, ABSORBABLE, NATURAL4 reports
- Plain Gut CG-172Covidien LP · SUTURE, ABSORBABLE, NATURAL3 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 91873).
- Z-1387-2023 · Open, ClassifiedStep Bladeless Trocars, 10 mm cannula and dilator with radially expandable sleeve and 5 - 10 mm reducer, Product Number S101010
- Z-1389-2023 · Open, ClassifiedMild Chromic Gut Absorbable Sutures, CH GUT 5-0 12 UD SS-14 SS-14 (G-1766-K), CHROMIC GUT 6-0 45CM HE-1 X12 (G-1780K)
- Z-1390-2023 · Open, ClassifiedPlain Gut Absorbable Sutures, PLAIN GUT 5/0 18 P-13 X12 (SG-5686), PLAIN GUT 4-0 18 UD SC-11 DA (SG828-2)
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page