Replacement Battery List Number SUB0000864, found in Plum… recallZ-1565-2023 · ICU Medical Inc · initiated Mar 22, 2023

Class I Open, Classified All recalls posted in May 2023

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from ICU Medical Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than expected.

FDA root cause category: Under Investigation by firm

What the company is doing

ICU medical issued an URGENT: MEDICAL DEVICE CORRECTION notice on 03/22/2023 to its direct and indirect consignees by mail. The notice explained the issue and the hazard and requested the following: Clinical Users: Whenever possible, keep the pump plugged into AC power. Before disconnecting the pump from AC power, please ensure that the battery is fully charged. Closely monitor the Battery Status Indicator while the pump is disconnected from AC power. Additionally, have a backup pump available when infusing critical medications. Biomedical Engineering: You may replace affected batteries with a new CSB battery until corrected batteries are available. Please do not use a replacement battery with corroded battery terminals. 1. Identify all affected batteries in your possession and ensure all users or potential users of these pumps are immediately made aware of this notification and proposed mitigations. 2. Complete and return the attached Response Form to [email protected] within ten days of receipt to acknowledge your understanding of this notification. 3. DISTRIBUTORS: If you have distributed potentially affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to [email protected].

Recall details

Recall number
Z-1565-2023
Recall class
Class I (most serious)
Status
Open, Classified
Initiated by the company
Mar 22, 2023
Posted by the FDA
May 16, 2023
Recalling firm
ICU Medical Inc
Quantity
28656 units
Product code
FRN
FDA submissions
K141789, K161469
Distribution
worldwide

Product description

Replacement Battery List Number SUB0000864, found in Plum 360 Infusion System, List Number 30010

Code information (models, lots, serial numbers)

All devices containing CSB BATTERY displaying the first two (2) characters as 22 or lower

Other devices under the same FDA submission

35 devices cleared under K141789, K161469 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.