Plum 1425128Recalls & reported problems
ICU Medical, Inc. · LIFESHIELD Primary PLUM Set Piggyback with Backcheck Valve, 2 CLAVE Y-Sites, Secure Lock, 103 Inch · Class 2
All models of Plum (26)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10887787006480), received between Mar 21, 2024 and Mar 21, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “No Flow” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
11 other recalls cover devices cleared under the same FDA submission (K141789); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- May 12, 2026Open, Classified
Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall. The presence of discoloration and particulate matter in the fluid path has not been ruled out.
Z-3156-2026 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- No Flow1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 21, 2024No Flow; No Flow
THE EVENT OCCURRED ON AN UNSPECIFIED DATE AND INVOLVED A PRIMARY PLUMSET, PIGGYBACK, BACKCHECK VALVE, 2 CLAVE Y-SITES, SECURE LOCK, 103 INCH WHERE INFUSION SETS WERE SHOWING MULTIPLE PROXIMAL OCCLUSION ALARMS. THERE WAS UNKNOWN PATIENT INVOLVED AND UNKNOWN HUMAN HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18949399 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K141789. One submission often covers many models, so check the product and lot details in each record.
- Z-3162-2026 · May 12, 2026 · Open, ClassifiedIV Sets with drip chambers
- Z-2447-2024 · May 7, 2024 · Open, ClassifiedReplacement Battery List Number SUB0000864, found in Plum 360 Infusion System, List Number 30010. Product description is PLUM 360 Infuser. OUS units marketed with the following de
- Z-2430-2023 · Jul 13, 2023 · Open, ClassifiedPlum 360 Infusion System, List Numbers: 300100405, 300100409, 300100413, 300100414, 300100415, 300100418, 300100451, 300100452, 300100453, 300100481, 300100483, 300100613, 3001010
- Z-1565-2023 · Mar 22, 2023 · Open, ClassifiedReplacement Battery List Number SUB0000864, found in Plum 360 Infusion System, List Number 30010
- Z-0128-2020 · Jul 29, 2019 · TerminatedInfusion Pump I.V. Administration, specifically Plum Microbore Infusion Sets with Inline Filters labeled as Primary PLUM Set Yellow Striped Tubing, Distal Microbore Tubing, 0.2 Mic
- Z-0127-2020 · Jul 29, 2019 · TerminatedInfusion Pump I.V. Administration, specifically Plum Microbore Infusion Sets with Inline Filters labeled as Primary PLUM Set 0.2 Micron Filter, Prepierced Y-Site, Polyethylene Line
- Z-0145-2020 · Jul 29, 2019 · TerminatedInfusion Pump I.V. Administration, specifically Plum Microbore Infusion Sets with Inline Filters labeled as PLUM Epidural Set with NRFit Connector Yellow-Striped Tubing, Distal Mic
- Z-3031-2017 · Jun 15, 2017 · TerminatedPlum 360 Infusion System
- Z-1682-2017 · Dec 30, 2016 · TerminatedPlum 360 Infusion Pump, Software Version 15.02. The infusion pump is capable of delivering fluids for a variety of therapies such as parenteral, enteral, or epidural infusions
- Z-1833-2016 · Mar 25, 2016 · TerminatedThe Plum A+ is a dual-line volumetric infusion system designed to meet the growing demand for hospital-wide, as well as alternate site and home healthcare, standardization. With it
- Z-1832-2016 · Mar 25, 2016 · TerminatedThe Plum 360 is a large volume infuser capable of delivering fluids for a variety of therapies such as parenteral, enteral, or epidural infusions. The Plum 360 infuser can deliver
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page