Preat Implant Prosthetics recallZ-1285-2023 · Preat Corp · initiated Feb 21, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Preat Corp or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Engaging titanium base contains a dimensional condition that allows vertical movement of the titanium base after the prosthetic screw is torqued to the recommended value, which if undetected and installed on the implant, may lead to abutment/screw fracture, which could result in the abutment/screw/crown being ingested by the patient.

FDA root cause category: Process design

What the company is doing

On 2/21/23, recall notices were mailed to customers who were asked to discard or return affected devices to the recalling firm. In addition, in the event affected product was further distributed, customers were asked to inform their customers, or to provide a customer list so the recalling firm could notify these additional customers. The customer service number was provided: 800-232-7732

Recall details

Recall number
Z-1285-2023
Status
Open, Classified
Initiated by the company
Feb 21, 2023
Posted by the FDA
Mar 24, 2023
Recalling firm
Preat Corp
Quantity
26
Product code
NHA
FDA submissions
K220823
Distribution
US Nationwide distribution in the states of TX, OR, UT, NY.

Product description

Neodent GM X 6mm Engaging Titanium Base, REF: 9007162

Code information (models, lots, serial numbers)

UDI: 00842092161326, Lots: 211891 and 211912

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

4 devices cleared under K220823 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.