Why it was recalled
Problem maintaining the stable placement of the battery may come loose from the housing and become damaged.
FDA root cause category: Device Design
What the company is doing
Tobii Dynavox issued Urgent Medical Device Recall Letter on 2/24/23. Letter states reason for recall, health risk and action to take: Tobii Dynavox will, at its full expense, provide you with a repair authorization number (RA#) and a shipping label to have the device inspected and repaired by adding more stability to the battery s position within the device. What should you do? There are numerous ways to contact the company to obtain a RA#: Online: https://us.tobiidynavox.com/pages/recall Toll Free: 800-344-1778, press 1, then press 2 (for the TD I-110 team) By Email: [email protected] We will provide you with your RA# and shipping labels via email, or by phone and mail where email is not possible. We will expedite all recalled battery repairs and return shipping to you, so you are without your device for as short a time as possible. You should obtain a RA# for the free repair to ensure that your device continues working as expected. You can continue to use your device and charge it as usual until you receive your RA# and shipping label. As with any Lithium-Ion battery, you should never puncture or vent a battery pack. You should not attempt to change the battery yourself or to remove the battery. This should only be done by a trained service technician. The device should not be stored in extremely hot conditions, as stated in the product manual, over 158F or 70C. Please use any of the methods above to tell us if your device is no longer in service. We will continue to contact all customers regarding this recall to be sure all customers are notified.
Recall details
- Recall number
- Z-1399-2023
- Status
- Open, Classified
- Initiated by the company
- Feb 24, 2023
- Posted by the FDA
- Apr 18, 2023
- Recalling firm
- Tobii Dynavox Llc
- Quantity
- 3924 units
- Product code
- ILQ
- Distribution
- Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Brazil, Canada, Croatia, Denmark, Estonia, Finland, France, Germany, Iceland Ireland, Italy, Latvia, Netherlands, Norway, Saudi Arabia, Slovakia, Slovenia, Sweden, United Arab Emirates, United Kingdom.
Product description
Tobii Dynavox TD I-110 Speech Device-tablet style speech generating device used for ambulatory people with disabilities. Catalog #: 750567
Code information (models, lots, serial numbers)
UDI: 7340074601714 Serial Numbers: TD110-211005000001 through TD110-220927004400
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Tobii Dynavox TD I-110Tobii Dynavox LLC · System, Communication, Powered1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page