FilmArray Gastrointestinal (GI) Panel recallZ-1392-2023 · BioFire Diagnostics, LLC · initiated Mar 8, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from BioFire Diagnostics, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to manufacturing issue, Gastrointestinal (GI) Panel may result in false negative results.
FDA root cause category: Under Investigation by firm
What the company is doing
On March 8, 2023, Biomerieux ( Parent company of BioFire Diagnostics, LLC) issued an "Urgent: Medical Device Correction" notification to affected consignees via E-Mail. In addition, to notifying affected consignees about the recalled device, Biomerieux asked consignees to take the following actions: 1. Immediately examine your inventory for product identified in this recall (refer to Table 1). 2. Discontinue use and discard any remaining product in your possession that is subject to this recall. bioMerieux will replace the product at no charge in accordance with BIOFIRE s standard limited warranty. 3. Please complete the accompanying Acknowledgment of Receipt Form and confirm the number of pouches scrapped (if any). Return the Acknowledgment of Receipt to bioMerieux so that bioMerieux may acknowledge your receipt of this notification. 4. If you have further distributed this product, please identify any recipients, and notify them at once of this product recall. 5. For product replacement, as well as any other questions or concerns, please contact our customer support department at [email protected] or via telephone by dialing +1.800.736.6354 and selecting option 5 for Product Technical Support. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. " Complete and submit the report online. " Regular Mail or Fax: Download form or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form or submit by fax to 1-800-FDA-0178.
Recall details
- Recall number
- Z-1392-2023
- Status
- Open, Classified
- Initiated by the company
- Mar 8, 2023
- Posted by the FDA
- Apr 14, 2023
- Recalling firm
- BioFire Diagnostics, LLC
- Quantity
- 900 pouches (30 kits)
- Product code
- PCH
- FDA submissions
- K140407, K143005
- Distribution
- US Nationwide distribution in the states of AR, CA, DC, IL, KS, LA, MN, NC, NY, OK, WI.
Product description
FilmArray Gastrointestinal (GI) Panel for FILMARRAY Systems, REF: RFIT-ASY-0116, (30 pouches/kit)
Code information (models, lots, serial numbers)
Pouch Lot# 2N3R22 / Kit Lot# 2160422; UDI-DI: (00)815381020109
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- FilmArray® Gastrointestinal (GI) Panel RFIT-ASY-0116Biofire Diagnostics, LLC · Gastrointestinal Pathogen Panel Multiplex Nucleic Acid-Based Assay System6 recalls2 deaths24 reports
Product lines
Other devices under the same FDA submission
1 device cleared under K140407, K143005 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- FilmArray® Gastrointestinal (GI) Panel RFIT-ASY-0104Biofire Diagnostics, LLC · Gastrointestinal Pathogen Panel Multiplex Nucleic Acid-Based Assay System1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page