Skip to content

EMPOWR Porous Knee Patella Over Drill, REF: 802-05-104 recallZ-1384-2023 · Encore Medical, LP · initiated Feb 28, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Encore Medical, LP or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Patella drills are marked with incorrect diameter size of 5.6mm, when they should be 5.4mm. Porous patella overdrills are marked with incorrect diameter size of 5.4 mm, when they should be 5.6 mm. If the incorrect drill is selected during surgery, there is risk of the implant loosening, implant not properly fitting, and/or fractured bone.

FDA root cause category: Process change control

What the company is doing

On 2/28/23, field safety notices were emailed to surgical sales agents who were asked to document the quantity of affected devices complete and return the acknowledgement form. Customer service contact, [email protected], was provided to report failures or to request replacements. Customers with questions can phone (914) 450-3931, or email [email protected].

Recall details

Recall number
Z-1384-2023
Status
Open, Classified
Initiated by the company
Feb 28, 2023
Posted by the FDA
Apr 12, 2023
Recalling firm
Encore Medical, LP
Quantity
29
Product code
MBH
FDA submissions
K212941
Distribution
US: TX, IN, MD, CA, KS, AL, HI, WA

Product description

EMPOWR Porous Knee Patella Over Drill, REF: 802-05-104; EMPOWR Knee Patella Drill, REF: 802-05-108

Code information (models, lots, serial numbers)

REF/UDI-DI/Lot: 802-05-104/00190446238843/388620L01; 802-05-108/00190446365907/388620L03

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

5 devices cleared under K212941 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates