Beckman Coulter Access Free T3, Triiodothyronine, Free… recallZ-1419-2023 · Beckman Coulter, Inc. · initiated Mar 15, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Beckman Coulter, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The reagent lot produces elevated results (approximately >30%) in some (15-20%)patient samples that do not correlate with the results of other thyroid markers nor Free T3 results generated on other platforms. This introduces a risk of the positive bias not being detected by QC with the potential that clinical correlation with other thyroid function tests will be inconsistent requiring additional testing; however the clinical assessment and treatment decisions are not made in isolation on a single test result especially, in the setting of thyroid function testing. False positive test results could lead to a euthyroid patient being interpreted clinically as having elevated thyroid function (hyperthyroid) or a hypothyroid patient being within the normal range. Such misdiagnoses may lead to inappropriate treatment of a euthyroid patient with anti-thyroid medications or hypothyroid patient not getting sufficient treatment since their free T3 test result demonstrates them to be in the euthyroid range.
FDA root cause category: Process design
What the company is doing
The recalling firm issued letters dated 3/8/2023 via mail and email. The letter described the issue, impact, and actions to be taken. The actions included discontinuing use of the affected lot and discard all remaining reagent packs from this lot. The letter is to be shared with the consignee laboratory and/or medical director regarding the need to review previous patient test results. The consignee is to contact their local Beckman Coulter representative for replacement product requests. If the product has been forwarded to another laboratory, the consignee is requested to provide them a copy of the letter. A response was requested within 10 days via email or completion and return of the enclosed response form. The response form is to indicate the letter has been read and understood and all relevant personnel have been informed of its contents. The consignee is also to indicate whether or not they had the affected product.
Recall details
- Recall number
- Z-1419-2023
- Status
- Open, Classified
- Initiated by the company
- Mar 15, 2023
- Posted by the FDA
- Apr 20, 2023
- Recalling firm
- Beckman Coulter, Inc.
- Quantity
- 31,076 units
- Product code
- CDP
- Distribution
- Worldwide distribution - US Nationwide, including Puerto Rico and the countries of Bahrain, Brazil, Canada, China, Costa Rica, El Salvador, Georgia, Hong Kong, Malaysia, Mexico, Myanmar, New Zealand, Peru, Philippines, Qatar, Singapore, Taiwan, Thailand, Uruguay, and Viet Nam.
Product description
Beckman Coulter Access Free T3, Triiodothyronine, Free, Catalog #A13422, each box contains 2 reagent packs.
Code information (models, lots, serial numbers)
Lot #233968, exp. 2023-06-30, UDI (01)15099590201661(17)230630(11)220630(10)233968.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Access Free T3 A13422Beckman Coulter, Inc. · Radioimmunoassay, Total Triiodothyronine1 recall7 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page