JULIET Ti LL recallZ-1426-2023 · Spineart SA · initiated Mar 16, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Spineart SA or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
JULIET Lateral Lumbar Ti-Life Plate, a component of the JULIET Ti Lateral Lumbar Cage, may disassemble during the preparation stage before implantation, or during surgery when removing the implant holder, which could cause surgical delay.
FDA root cause category: Process change control
What the company is doing
On 3/16/23, recall notices were distributed to customers who were told the following: Plate replacement: If the issue occurs, the surgeon should proceed as follows: 1- Unscrew the Plate screw of the cage with the screwdriver, 2- Scrap the current plate assembly, 3- Use a new set of pre-assembled plate/screw (a minimum of 2 sets of pre-assembled plates/screws per size are part of the device listing). The cages are to be used with the corresponding plate or higher plate which means that a plate with the same height than the cage can be used but also a plate with higher height. 4- In the case of 0-degree and 8-degree cages, a pre-assembled plate/screw is not mandatory as described in the surgical technique Acknowledge receipt of this letter by emailing to [email protected] Discontinue use and return any product on hand. Customers with questions can contact firm at 949-313-6400 Monday through Friday, 8:00am to 5:00pm, Pacific, or email at [email protected]
Recall details
- Recall number
- Z-1426-2023
- Status
- Open, Classified
- Initiated by the company
- Mar 16, 2023
- Posted by the FDA
- Apr 24, 2023
- Recalling firm
- Spineart SA
- Quantity
- 544
- Product code
- OVD
- FDA submissions
- K190877
- Distribution
- US Nationwide distribution in the states of NH, TX, CA, VA, IL, RI, NC, AZ.
Product description
JULIET Ti LL, LATERAL LUMBAR TI-LIFE PLATE SMALL, REF: JLT-PL 02 08-S, JLT-PL 02 10-S, JLT-PL 02 12-S, JLT-PL 02 14-S, JLT-PL 02 16-S; JULIET Ti LL, LATERAL LUMBAR TI-LIFE PLATE LARGE, REF: JLT-PL 04 08-S, JLT-PL 04 10-S, JLT-PL 04 12-S, JLT-PL 04 14-S, JLT-PL 04 16-S, JLT-PL 04 18-S
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
REF/UDI-DI/Lot: JLT-PL 02 08-S/07640178987388/4-8826, 5-4142, 5-5377, 5-0643, 5-2431; JLT-PL 02 10-S/07640178987395/6-5488, 6-6556, 5-0644, 6-4734, 6-4735, 5-5378, 6-3211; JLT-PL 02 12-S/07640178987401/6-5489, 6-6558, 6-7645, 6-8734, 6-4732, 4-8828, 5-5379, 6-4733; JLT-PL 02 14-S/07640178987418/5-8439, 4-8829, 6-0714, 5-5380; JLT-PL 02 16-S/07640178987425/4-8830, 5-4145, 6-3217, 6-0726, 6-5490; JLT-PL 04 08-S/07640178987456/4-8845, 5-4146, 6-0725; JLT-PL 04 10-S/07640178987463/4-8831, 5-4147, 6-0722; JLT-PL 04 12-S/07640178987470/4-8832, 5-4148, 5-8438, 6-0721, 6-3209; JLT-PL 04 14-S/07640178987487/4-3861, 4-8833, 5-4149; JLT-PL 04 16-S/07640178987494/5-4150; JLT-PL 04 18-S/07640178987500/5-4151
Devices named in this recall
11 devices in the FDA device registry whose identifier appears in the recall's code information.
- JULIET ® Ti LL JLT-PL 02 08-SSpineArt SA · Intervertebral Fusion Device With Integrated Fixation, Lumbar1 recall0 reports
- JULIET ® Ti LL JLT-PL 02 10-SSpineArt SA · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- JULIET ® Ti LL JLT-PL 02 12-SSpineArt SA · Intervertebral Fusion Device With Integrated Fixation, Lumbar1 recall0 reports
- JULIET ® Ti LL JLT-PL 02 14-SSpineArt SA · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- JULIET ® Ti LL JLT-PL 02 16-SSpineArt SA · Intervertebral Fusion Device With Integrated Fixation, Lumbar1 recall0 reports
- JULIET ® Ti LL JLT-PL 04 08-SSpineArt SA · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- JULIET ® Ti LL JLT-PL 04 10-SSpineArt SA · Intervertebral Fusion Device With Integrated Fixation, Lumbar1 recall0 reports
- JULIET ® Ti LL JLT-PL 04 12-SSpineArt SA · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- JULIET ® Ti LL JLT-PL 04 14-SSpineArt SA · Intervertebral Fusion Device With Integrated Fixation, Lumbar1 recall0 reports
- JULIET ® Ti LL JLT-PL 04 16-SSpineArt SA · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- JULIET ® Ti LL JLT-PL 04 18-SSpineArt SA · Intervertebral Fusion Device With Integrated Fixation, Lumbar1 recall0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page