Skip to content

JULIET Ti LL recallZ-1426-2023 · Spineart SA · initiated Mar 16, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Spineart SA or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

JULIET Lateral Lumbar Ti-Life Plate, a component of the JULIET Ti Lateral Lumbar Cage, may disassemble during the preparation stage before implantation, or during surgery when removing the implant holder, which could cause surgical delay.

FDA root cause category: Process change control

What the company is doing

On 3/16/23, recall notices were distributed to customers who were told the following: Plate replacement: If the issue occurs, the surgeon should proceed as follows: 1- Unscrew the Plate screw of the cage with the screwdriver, 2- Scrap the current plate assembly, 3- Use a new set of pre-assembled plate/screw (a minimum of 2 sets of pre-assembled plates/screws per size are part of the device listing). The cages are to be used with the corresponding plate or higher plate which means that a plate with the same height than the cage can be used but also a plate with higher height. 4- In the case of 0-degree and 8-degree cages, a pre-assembled plate/screw is not mandatory as described in the surgical technique Acknowledge receipt of this letter by emailing to [email protected] Discontinue use and return any product on hand. Customers with questions can contact firm at 949-313-6400 Monday through Friday, 8:00am to 5:00pm, Pacific, or email at [email protected]

Recall details

Recall number
Z-1426-2023
Status
Open, Classified
Initiated by the company
Mar 16, 2023
Posted by the FDA
Apr 24, 2023
Recalling firm
Spineart SA
Quantity
544
Product code
OVD
FDA submissions
K190877
Distribution
US Nationwide distribution in the states of NH, TX, CA, VA, IL, RI, NC, AZ.

Product description

JULIET Ti LL, LATERAL LUMBAR TI-LIFE PLATE SMALL, REF: JLT-PL 02 08-S, JLT-PL 02 10-S, JLT-PL 02 12-S, JLT-PL 02 14-S, JLT-PL 02 16-S; JULIET Ti LL, LATERAL LUMBAR TI-LIFE PLATE LARGE, REF: JLT-PL 04 08-S, JLT-PL 04 10-S, JLT-PL 04 12-S, JLT-PL 04 14-S, JLT-PL 04 16-S, JLT-PL 04 18-S

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

REF/UDI-DI/Lot: JLT-PL 02 08-S/07640178987388/4-8826, 5-4142, 5-5377, 5-0643, 5-2431; JLT-PL 02 10-S/07640178987395/6-5488, 6-6556, 5-0644, 6-4734, 6-4735, 5-5378, 6-3211; JLT-PL 02 12-S/07640178987401/6-5489, 6-6558, 6-7645, 6-8734, 6-4732, 4-8828, 5-5379, 6-4733; JLT-PL 02 14-S/07640178987418/5-8439, 4-8829, 6-0714, 5-5380; JLT-PL 02 16-S/07640178987425/4-8830, 5-4145, 6-3217, 6-0726, 6-5490; JLT-PL 04 08-S/07640178987456/4-8845, 5-4146, 6-0725; JLT-PL 04 10-S/07640178987463/4-8831, 5-4147, 6-0722; JLT-PL 04 12-S/07640178987470/4-8832, 5-4148, 5-8438, 6-0721, 6-3209; JLT-PL 04 14-S/07640178987487/4-3861, 4-8833, 5-4149; JLT-PL 04 16-S/07640178987494/5-4150; JLT-PL 04 18-S/07640178987500/5-4151

Devices named in this recall

11 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page