Sterile surgical convenience kits recallZ-1431-2023 · ROi CPS LLC · initiated Mar 22, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from ROi CPS LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The recalling firm was notified by their supplier that the light handle covers may separate from the light handle and fall off during use. The light handle covers are a kit component.
FDA root cause category: Nonconforming Material/Component
What the company is doing
The recalling firm issued letters dated 3/22/2023 via email the same day. The letter explained the issue, identified the affected product, and listed the actions to be taken by the customer. The actions included checking their inventory for the affected product, segregate and quarantine all affected kits in their inventory, and use the attached label template to print warning labels to be affixed to all affected kits in their inventory. The label should be applied to each affected kit and outer product case in a prominent location to end users of the kits. The location chosen should not cover critical product information found on the existing label. The label instructs users to remove and discard the STERIS Light Handle Cover(s) when the kit is opened for use and to pull a sterile replacement from stock. A reply form is included with the letter for completion to indicate the consignee's actions in response to the notice. The response form is to be returned to the recalling firm via email. If the direct account consignee further distributed or forwarded the affected product, they are to immediately notify those customers. A copy of the ROi CPS recall letter, the letter from STERIS, and the attached label template may be included.
Recall details
- Recall number
- Z-1431-2023
- Status
- Open, Classified
- Initiated by the company
- Mar 22, 2023
- Posted by the FDA
- Apr 25, 2023
- Recalling firm
- ROi CPS LLC
- Quantity
- 278 kits
- Product code
- FDE
- Distribution
- US Nationwide distribution in the states of FL and MO. There was no foreign/military/government distribution.
Product description
Sterile surgical convenience kits: 1. regard Item Number: 880405003, GS01011C - General Laparoscopy; 2. regard Item Number: 880424004, GS01031D - Thoracic.
Code information (models, lots, serial numbers)
1. Lot #95475, exp. 10/29/2024, UDI (01)10194717112211(17)241029(10)95475. 2. Lot #95493, exp. 9/7/2024, UDI (01)10194717112327(17)240907(10)95493.
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Regard™ GS01011CRoi Cps, LLC · Laparoscopy Kit1 recall0 reports
- Regard™ GS01031DRoi Cps, LLC · Laparoscopy Kit1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 91984).
- Z-1432-2023 · Open, ClassifiedSterile surgical convenience kits: 1. regard Item Number: 880353009, GS00797I - Laparotomy Pk; 2. regard Item Number: 880429002, GS01036B - Universal Custom ; 3. regard Item …
- Z-1433-2023 · Open, ClassifiedSterile surgical convenience kit: regard Item Number: 800718003, GY00923C - WPH Major Vaginal Pack.
- Z-1434-2023 · Open, ClassifiedSterile surgical convenience kit: regard Item Number: 880393003, HT00997C - Cardiac Cath.
- Z-1435-2023 · Open, ClassifiedSterile surgical convenience kit: regard Item Number: 800753005, GY00958E - WPH GYN Laparoscopy Pack.
- Z-1436-2023 · Open, ClassifiedSterile surgical convenience kit: regard Item Number 800706005, LD0091E - OH C-Section Pack.
- Z-1437-2023 · Open, ClassifiedSterile surgical convenience kits: 1. regard Item Number: 800756005, NU00961E - OH Laminectomy Cust Pack; 2. regard Item Number: 880411008, NU01018H - Lumbar Universal ; 3.…
- Z-1438-2023 · Open, ClassifiedSterile surgical convenience kits: 1. regard Item Number: 880473001, OR01105A - Total Shoulder/Hip; 2. regard Item Number: 880426004, OR01033D - Total Hip; 3. regard Item N…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page