FujiFilm FDR AQRO (Model DR-XD1000)- A digital mobile X-ray… recallZ-1470-2023 · FUJIFILM Healthcare Americas Corporation · initiated Feb 27, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from FUJIFILM Healthcare Americas Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Arm which holds the tube head in place is susceptible to breaking, if the device breaks it could potentially cause serious harm or injury to the patient and/or user.

FDA root cause category: Device Design

What the company is doing

FUJIFILM Healthcare Americas Corporation (Fujifilm) issued Urgent Medical Device Recall letter on 2/27/23 via email to Distributors ad end-Users. Letter states reason for recall, health risk and action to take: A Fujifilm service engineer will contact your facility to arrange a visit to repair your affected unit(s). ACTIONS TO BE TAKEN BY THE CUSTOMER/END-USER 1)Perform an inspection according to the steps outlined below prior to using the product. a. If there are no abnormalities present during the pre-use inspection the product can continue to be used safely. b. Immediately remove the device from use if any abnormalities are observed and report the issueto your Fujifilm service engineer. 2)Coordinate with the Fujifilm service engineer to schedule the correction of your device whencontacted. 3)Please complete and return the Field Action Verification Form attached on page 6 of this communication. Responding with the information requested is essential for ensuring appropriate action is taken. Please contact our field action coordinator, Alicia Martinez, via phone at 973-459-9206 or email at [email protected], if you have any further questions regarding this field action.

Recall details

Recall number
Z-1470-2023
Status
Open, Classified
Initiated by the company
Feb 27, 2023
Posted by the FDA
Apr 27, 2023
Recalling firm
FUJIFILM Healthcare Americas Corporation
Quantity
77 units
Product code
IZL
FDA submissions
K170858
Distribution
US Nationwide distribution.

Product description

FujiFilm FDR AQRO (Model DR-XD1000)- A digital mobile X-ray system intended for use in general purpose radiography for generating radiographic images of human anatomy, including adult, pediatric, and neonatal exams

Code information (models, lots, serial numbers)

UDI: (01)04547410425505(11)200701. Serial numbers with the last 4 digits ending between ****0001 to ****1993.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.