Skip to content

VITROS recallZ-1396-2023 · Ortho-Clinical Diagnostics, Inc. · initiated Mar 9, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ortho-Clinical Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Incorrect Number of Calibrator Levels (2) Set for the VITROS Immunodiagnostics Products Anti-SARS-CoV-2 Total N Antibody Calibrators in Assay Data Disks (ADD), resulting in delayed patients results

FDA root cause category: Employee error

What the company is doing

Ortho Clinical Diagnostics issued Important Product Correction Notification letter on 3/9/23. Letter states reason for recall, health risk and action to take: To calibrate all lots currently within expiring date of Anti-SARS-CoV-2 Total N, load ADD 6219 or higher on your VITROS System. " Complete the enclosed Confirmation of Receipt form no later than March 17, 2023. " Please forward this notification and ADD if the affected product was distributed outside of your facility. " If your laboratory has experienced the issue with this product and you have not already done so, please report the occurrence to your local Ortho Care" Technical Solutions Center. If you have further questions, please contact Ortho Care Technical Solutions Center at 1-800-421-3311.

Recall details

Recall number
Z-1396-2023
Status
Open, Classified
Initiated by the company
Mar 9, 2023
Posted by the FDA
Apr 18, 2023
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Quantity
256 units
Product code
QKO
Distribution
US Nationwide distribution in the states of AL, AZ, CA, GA, NY.

Product description

VITROS Immunodiagnostic Products Anti-SARSCoV- 2 Total N Antibody Calibrators programmed onto the Assay Data Disks (ADD) Data Release Versions (DRV): DRVs 6213 through 6218 supporting VITROS Calibrators. Product Code: 6199976

Code information (models, lots, serial numbers)

UDI: 10758750034581 Lot # /Expiration Date: 0230 22-Feb-2023; 0240 22-Mar-2023; 0250 26-Apr-2023.