EarlyVue VS30 Vitals monitor with Software Versions A.00.02… recallZ-1645-2023 · Philips North America Llc · initiated Mar 6, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips North America Llc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
EarlyVue VS30 Vital Signs Monitor and EarlyVue VS30 Vitals monitor Missing Calibration Alarm, a risk of an inaccurate CO2 measurement which may lead to a failure to recognize a change in patient condition.
FDA root cause category: Software design
What the company is doing
Philips issued Urgent Field Safety Notice Letter on 3/3/23. Letter states reason for recall, health risk and action to take: 1. Identify the last calibration date of the device and review your hospital maintenance plan. Determine if your device has E01 End Tidal CO2 and has not been calibrated within one year of the production calibration. If needed adjust the maintenance plan and incorporate the calibration of your affected device. a. The device should not be used until the calibration or maintenance plan has been reviewed. b. The calibration date can be accessed via the following steps on the device: Settings > Admin > Diagnostics > Page 2 of Diagnostics > Maintenance > 129 > CO2 Test. The image below shows the example of the screen to see the last calibration date of your device: 2. Review this URGENT Field Safety Notice in its entirety and pass this notice to all those who need to be aware within your organization or to any organizations where the potentially affected devices have been transferred. (If appropriate). 3. Complete and return attached form to Philips promptly to confirm receipt of the URGENT Field Safety Notice Letter, understanding of the issue, and required actions to be taken. 4. Actions planned by Philips Hospital Patient Monitoring to correct the problem A Philips representative will reach out to you to arrange a software upgrade to your monitor(s). Philips will start to schedule the upgrade upon formal release of the field action, which is planned for a six month implementation timeline. If you need any further information or support concerning this issue, please contact your local Philips representative
Recall details
- Recall number
- Z-1645-2023
- Status
- Open, Classified
- Initiated by the company
- Mar 6, 2023
- Posted by the FDA
- May 26, 2023
- Recalling firm
- Philips North America Llc
- Quantity
- 1421 units US; 3 units OUS
- Product code
- DSJ
- FDA submissions
- K190624
- Distribution
- Nationwide Foreign: Belgium Brazil Canada Denmark Germany Guam Hong Kong Indonesia Italy Norway Poland Singapore Sweden Switzerland United Kingdom
Product description
EarlyVue VS30 Vitals monitor with Software Versions A.00.02 and A.00.01 Product Number: 863380
Code information (models, lots, serial numbers)
UDI: (01)00884838091412 Software Version A.00.02 and A.00.01
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- EarlyVue VS30 EarlyVue VS30 Vitals monitorPhilips Medizin Systeme Böblingen GmbH · Oximeter2 recalls167 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page