EarlyVue VS30 EarlyVue VS30 Vitals monitorRecalls & reported problems

Philips Medizin Systeme Böblingen GmbH · EarlyVue VS30 Vitals Signs Monitor · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Philips EarlyVue VS30 vitals monitor measures vital signs. Of its 164 adverse event reports, 162 are malfunctions and 2 are injuries, with no deaths. Some describe blood pressure readings that were too high, and others describe incorrect measurements. Reports rose from 13 in the prior 12 months to 101 in the last 12 months. Two open recalls name this device, over software that stops breathing-rate alarms and a missing CO2 calibration alarm.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
167
Deaths reported
0
Injuries reported
2
Recalls naming it
2

Summary

The FDA's MAUDE database contains 167 adverse event reports that list this exact device (DI 00884838091412), received between Aug 25, 2020 and Sep 29, 2026 — classified as 2 injuries, 165 malfunctions.

The most frequently reported specific device problem is “Incorrect Measurement” (46 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 2 reportsMar 2024: 1 reportsApr 2024: 2 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 3 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 4 reportsSep 2025: 1 reportsOct 2025: 2 reportsNov 2025: 2 reportsDec 2025: 15 reportsJan 2026: 8 reportsFeb 2026: 22 reportsMar 2026: 16 reportsApr 2026: 10 reportsMay 2026: 13 reportsJun 2026: 11 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 5 reportsOct 2026: 0 reports20242025202622
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 31, 2023Class IIOpen, Classified

    A software issue prevents the monitor from alarming when the high/low acoustic respiration rate (RRa) limits are violated.

    Z-2654-2023 · Root cause: Software design · Recall details · FDA record ↗

  2. Mar 6, 2023Class IIOpen, Classified

    EarlyVue VS30 Vital Signs Monitor and EarlyVue VS30 Vitals monitor Missing Calibration Alarm, a risk of an inaccurate CO2 measurement which may lead to a failure to recognize a change in patient condition.

    Z-1645-2023 · Root cause: Software design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Incorrect Measurement46
  • High Readings36
  • No Audible Prompt/Feedback14
  • No Audible Alarm13
  • Detachment of Device or Device Component10
  • Device Fell8
  • Imprecise or Erratic Results7
  • Incorrect, Inadequate or Imprecise Result or Readings5

Patient problems

  • No Clinical Signs, Symptoms or Conditions161
  • Insufficient Information2
  • No Patient Involvement2
  • Bruise/Contusion1
  • Laceration(s)1
  • Numbness1
  • Pain1

Recent adverse event reports

The 25 most recent of 167 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Incorrect Measurement

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50931172 ↗
  • MalfunctionSep 28, 2026Imprecise or Erratic Results

    Philips received a complaint on an EarlyVue VS30 monitor indicating that the SpO2 was not reading correctly. The device was in clinical use at the time the issue was discovered. There was no adverse event or patient harm reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50918153 ↗
  • MalfunctionSep 27, 2026Device Fell

    Patient problems: Insufficient Information; Laceration(s)

    MAUDE report 50910385 ↗
  • MalfunctionSep 24, 2026High Readings

    The customer reported that the monitor's diastolic value was higher than expected. The device was not in use on a patient at the time of event, there was no adverse event reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50877226 ↗
  • MalfunctionSep 10, 2026High Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50722083 ↗
  • MalfunctionAug 27, 2026Imprecise or Erratic Results

    It was reported that the device is experiencing an accuracy issue involving a low SPO2 error. It is unknown if the device was in use on a patient at the time of the event. There was no adverse event reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50476020 ↗
  • MalfunctionAug 13, 2026Low Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50235392 ↗
  • MalfunctionJun 30, 2026High Readings

    PHILIPS RECEIVED A COMPLAINT ON THE EARLYVUE VS30 VITALS MONITOR INDICATING THAT THE NON-INVASIVE BLOOD PRESSURE (NIBP) WAS SHOWING HIGH IN THE SIMULATOR. THE DEVICE WAS NOT IN USE ON A PATIENT AT THE TIME OF THE EVENT, SO THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25671439 ↗
  • MalfunctionJun 30, 2026High Readings

    PHILIPS RECEIVED A COMPLAINT ON AN EARLYVUE VS30 INDICATING THAT THE NON-INVASIVE BLOOD PRESSURE (NIBP) READINGS WERE HIGH. THE DEVICE WAS IN USE ON A PATIENT AT THE TIME OF EVENT, NO ADVERSE EVENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25664983 ↗
  • MalfunctionJun 25, 2026High Readings

    PHILIPS RECEIVED A COMPLAINT ON AN EARLYVUE VS30 MONITOR INDICATING THAT THE NON-INVASIVE BLOOD PRESSURE (NIBP) DIASTOLIC IS READING INACCURATE AND VERY HIGH 90. THE DEVICE WAS NOT IN CLINICAL USE AT THE TIME THE ISSUE WAS DISCOVERED. THERE WAS NO ADVERSE EVENT OR HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25627274 ↗
  • MalfunctionJun 18, 2026High Readings

    PHILIPS RECEIVED A COMPLAINT ON THE INDICATING THAT NBP READINGS WERE HIGHER THAN THOSE FROM MANUAL TESTING. THE DEVICE WAS IN CLINICAL USE ON A PATIENT AT THE TIME OF THE EVENT. NO ADVERSE EVENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25545806 ↗
  • MalfunctionJun 16, 2026Incorrect Measurement

    PHILIPS RECEIVED A COMPLAINT ON THE EARLYVUE VS30 VITAL SIGNS MONITOR INDICATING THAT THE BLOOD PRESSURE WAS NOT WORKING. THE DEVICE WAS IN CLINICAL USE ON A PATIENT AT THE TIME OF THE EVENT. NO ADVERSE EVENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25527035 ↗
  • MalfunctionJun 16, 2026Incorrect Measurement

    PHILIPS RECEIVED A COMPLAINT ON THE EARLYVUE VS30 VITAL SIGNS MONITOR INDICATING THAT THE BLOOD PRESSURE WAS NOT WORKING. THE DEVICE WAS IN CLINICAL USE ON A PATIENT AT THE TIME OF THE EVENT. NO ADVERSE EVENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25526966 ↗
  • MalfunctionJun 9, 2026Incorrect Measurement

    THE CUSTOMER REPORTED THAT THE MAP IS NOT ACCURATE ON THE DEVICE. THE DEVICE WAS IN USE ON A PATIENT AT THE TIME OF THE EVENT, THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25439450 ↗
  • MalfunctionJun 8, 2026No Audible Prompt/Feedback

    PHILIPS RECEIVED A COMPLAINT ON AN EARLYVUE VS30 INDICATING THAT AFTER REPLACING THE MAINBOARD FOR AN UNRELATED ISSUE, A SPEAKER MALFUNCTION ERROR WAS DISPLAYED. IT IS UNKNOWN IF THERE WAS AUDIBLE SOUND PRODUCED OR NOT. THE DEVICE WAS NOT IN USE ON PATIENT AT TIME OF EVENT, THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25435838 ↗
  • MalfunctionJun 1, 2026High Readings

    IT WAS REPORTED THAT THE NBP BOTTOM NUMBER (DIASTOLIC) IS 10 POINTS TOO HIGH COMPARED TO OTHER DEVICES. THE DEVICE WAS IN USE ON A PATIENT AT THE TIME OF THE EVENT. THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25354154 ↗
  • MalfunctionJun 1, 2026No Audible Alarm

    IT WAS REPORTED THAT THE DEVICE IS DISPLAYING AN ERROR CODE FOR A SPEAKER MALFUNCTION AND CONFIRMED NO SOUND. THE DEVICE WAS IN USE ON A PATIENT AT THE TIME OF THE EVENT. THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25354137 ↗
  • MalfunctionJun 1, 2026Incorrect Measurement

    PHILIPS RECEIVED A COMPLAINT ON THE EARLYVUE VS30 INDICATING THAT THAT CLINICAL STAFF WERE QUESTIONING THE DIFFERENCE BETWEEN THE VS30 MONITOR'S MEASURED MAP (MEAN ARTERIAL PRESSURE) AND MANUALLY CALCULATED VALUES. THE DEVICE WAS IN USE ON PATIENT AT TIME OF EVENT, THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25348172 ↗
  • MalfunctionMay 29, 2026High Readings

    PHILIPS RECEIVED A COMPLAINT ON AN EARLYVUE VS30 INDICATING THE DEVICE HAS A HIGH DIASTOLIC BP AS COMPARED TO OTHER VS MONITORING DEVICES. IT WAS ALSO NOTED THAT THE DEVICE INFLATES TO 200 AND DEFLATES DOWN TO 2 AND THEN CYCLES. THE DEVICE WAS IN USE ON A PATIENT AT TIME OF EVENT, THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25333071 ↗
  • MalfunctionMay 28, 2026Incorrect Measurement

    CUSTOMER REPORTED THAT THE UNIT IS STILL NOT REPORTING CORRECT VITALS. EVERY TEMPERATURE SAYS 102 AND SYSTOLIC BLOOD PRESSURE (BP) READS 20 POINTS HIGHER. THE DEVICE WAS NOT IN USE ON A PATIENT AT THE TIME OF EVENT, THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25326203 ↗
  • MalfunctionMay 21, 2026Low Readings

    PHILIPS RECEIVED A COMPLAINT ON AN EARLYVUE VS30 DEVICE INDICATING THAT THE DIASTOLIC BLOOD PRESSURE IS MEASURING LOW. WHEN NON-INVASIVE BLOOD PRESSURE (NIBP) IS TAKEN IT MANUALLY THEY GET A MUCH DIFFERENT READING. THE DEVICE WAS IN USE ON A PATIENT AT TIME OF EVENT, THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25241283 ↗
  • MalfunctionMay 21, 2026Incorrect Measurement

    THE CUSTOMER REPORTED THAT BLOOD PRESSURE NOT READING WHEN SET TO LOW MEASUREMENTS. THE DEVICE WAS NOT IN USE ON A PATIENT AT THE TIME OF EVENT, THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25238224 ↗
  • MalfunctionMay 20, 2026High Readings

    IT WAS REPORTED THAT THE DEVICE IS PROVIDING A DIASTOLIC VALUE THAT IS TOO HIGH. THE DEVICE WAS IN USE ON A PATIENT AT THE TIME OF THE EVENT. THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25229686 ↗
  • MalfunctionMay 20, 2026Incorrect Measurement

    THE CUSTOMER REPORTED INCORRECT BLOOD PRESSURE READINGS ON THE DEVICE, OCCURRING RANDOMLY OUT OF 5 CHECKS. THE DEVICE WAS IN USE ON A PATIENT AT THE TIME OF THE EVENT, THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25226311 ↗
  • MalfunctionMay 19, 2026High Readings

    IT WAS REPORTED THAT THE BLOOD PRESSURE (BP) READINGS PROVIDED ARE HIGH. THE DEVICE WAS IN USE ON A PATIENT AT THE TIME OF THE EVENT. THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25215597 ↗

Other Oximeter devices