Centricity recallZ-1501-2023 · GE Healthcare · initiated Mar 28, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Healthcare or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
GE HealthCare has become aware of an issue where information from two different patients can be mismatched when correcting patient or study information.
FDA root cause category: Other
What the company is doing
GE issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 03/28/2023 by mail (traceable means). The notice explained the issue and the risk, and provideed safety instructions for continued use of the device. GE HealthCare will address this issue through a software correction.
Recall details
- Recall number
- Z-1501-2023
- Status
- Open, Classified
- Initiated by the company
- Mar 28, 2023
- Posted by the FDA
- May 2, 2023
- Recalling firm
- GE Healthcare
- Quantity
- 329 units
- Product code
- LLZ
- FDA submissions
- K150420
- Distribution
- Worldwide Distribution
Product description
Centricity Universal Viewer, Centricity Universal Viewer 6.0, Model/Catalog Numbers: a) 2088026-003, b) 2088026-026, c) 2088026-032, d) 2088026-043, e) 2088026-044, f) 2088026-115, g) 2088026-132, h) 2088026-306, i) 2088026-308, j) 2088026-406, k) 2088026-506, l) 2088026-613, m) 2088026-713, n) 2088026-714, o) 2088026-723, p) 2099714-001, q) 2100169-001, r) 2100491-006, s) 2102675-010, t) 2102676-001, u) 2104867-044, v) 2104867-045; Radiological Image Processing System
Code information (models, lots, serial numbers)
Show the full code information (16k characters)
a) 2088026-003, No UDI/DI, Product ID numbers: 233637-Centricity Universal Viewer-15773152; b) 2088026-026, No UDI/DI, Product ID numbers: 4446-2-Centricity Universal Viewer-00558643, 4491-1-Centricity Universal Viewer-00558690, 100256-3-Centricity Universal Viewer-00558700, 100256-3-Centricity Universal Viewer-01810788, 102143-1-Centricity Universal Viewer-00558685, 101044-1-Centricity Universal Viewer-00558694, 13356-2-Centricity Universal Viewer-00558681, 102335-1-Centricity Universal Viewer-00558675, 13228-1-Centricity Universal Viewer-00558678, 101274-1-Centricity Universal Viewer-00558679, 101274-1-Centricity Universal Viewer-20279342, 13125-1-Centricity Universal Viewer-02268546, 102359-1-Centricity Universal Viewer-00558696, 4578-1-Centricity Universal Viewer-00558340, A54120-Centricity Universal Viewer-00040449, KW0000-Centricity Universal Viewer-01416913, BON01-Centricity Universal Viewer-00044131, PL3277-Centricity Universal Viewer-00038831; c) 2088026-032, No UDI/DI, Product ID numbers: 1027-1-Centricity Universal Viewer-00558706, 1027-1-Centricity Universal Viewer-20759776, 124774-Centricity Universal Viewer-00000208; d) 2088026-043, No UDI/DI, Product ID numbers: 4441-1-Centricity Universal Viewer-00558677, PL3349-Centricity Universal Viewer-00038811, RU3866-Centricity Universal Viewer-01445731, RU8945-Centricity Universal Viewer-01445740, RU2580-Centricity Universal Viewer-01445743, RU7175-Centricity Universal Viewer-01445748, RU1646-Centricity Universal Viewer-01445763, RU3453-Centricity Universal Viewer-01273249, TBD-Centricity Universal Viewer-01273292; e) 2088026-044, No UDI/DI, Product ID numbers: 237870-Centricity Universal Viewer-01444278; f) 2088026-115, No UDI/DI, Product ID numbers: 4413-2-Centricity Universal Viewer-01300927, 4413-2-Centricity Universal Viewer-01300929, 100837-1-Centricity Universal Viewer-00558435, 100837-1-Centricity Universal Viewer-17594527, 13240-2-Centricity Universal Viewer-00558103, 11042-1-Centricity Universal Viewer-02285888, 11042-1-Centricity Universal Viewer-19603374; g) 2088026-132, No UDI/DI, Product ID numbers: A91901-Centricity Universal Viewer-00044549, CC4105-Centricity Universal Viewer-01282278, HC4105-Centricity Universal Viewer-00001393, 651948-Centricity Universal Viewer-01813739; h) 2088026-306, UDI/DI840682103800, Product ID numbers: 11438-1-Centricity Universal Viewer-00558464, 101280-1-Centricity Universal Viewer-19630915, ZA2541-Centricity Universal Viewer-01985829, ZA2472-Centricity Universal Viewer-00116565; i) 2088026-308, UDI/DI 840682103800, Product ID numbers: 4267-5-Centricity Universal Viewer-19618406, 100763-1-Centricity Universal Viewer-00558366, 101280-1-Centricity Universal Viewer-19630914, 101280-1-Centricity Universal Viewer-19630916, 101280-1-Centricity Universal Viewer-20553950, 4451-1-Centricity Universal Viewer-00558107, 101170-1-Centricity Universal Viewer-00558305, 4483-1-Centricity Universal Viewer-00558457, 4483-1-Centricity Universal Viewer-20583582, PL2975-Centricity Universal Viewer-00162605, 318055-Centricity Universal Viewer-01457316, 4267-5-Centricity Universal Viewer-20884363, 4267-5-Centricity Universal Viewer-20884364, 4267-5-Centricity Universal Viewer-20884365; j) 2088026-406, UDI/DI 840682103800, Product ID numbers: 4552-1-Centricity Universal Viewer-01294074, 100029-1-Centricity Universal Viewer-02256808, 100717-1-Centricity Universal Viewer-01293807, 100717-1-Centricity Universal Viewer-01293808, 13345-1-Centricity Universal Viewer-00602707, 8746-1-Centricity Universal Viewer-01448329, 4683-1-Centricity Universal Viewer-00558456, 4487-1-Centricity Universal Viewer-01291589, 4487-1-Centricity Universal Viewer-02097167, 571064-Centricity Universal Viewer-01293639, 786787-Centricity Universal Viewer-01283171; k) 2088026-506, UDI/DI 840682103800, Product ID numbers: 9094-1-Centricity Universal Viewer-01766830, 100770-1-Centricity Universal Viewer-01804127, 313741-Centricity Universal Viewer-02097256, 268034-Centricity Universal Viewer-18862276, RU1107-Centricity Universal Viewer-01813331, RU5388-Centricity Universal Viewer-01719556, 16218680-Centricity Universal Viewer-02266272, ZA2267-Centricity Universal Viewer-02001877; l) 2088026-613, UDI/DI 840682103800, Product ID numbers: 1011-1-Centricity Universal Viewer-00558109, 1011-1-Centricity Universal Viewer-02097166, 1011-1-Centricity Universal Viewer-02097165, 100159-1-Centricity Universal Viewer-00558324, 101598-1-Centricity Universal Viewer-00667138, 104491-1-Centricity Universal Viewer-01288014; m) 2088026-713, No UDI/DI, Product ID numbers: 2403-7-Centricity Universal Viewer-00558347, RU1107-Centricity Universal Viewer-01273207, MPX352-Centricity Universal Viewer-19629107, DP7603-Centricity Universal Viewer-20093544, FP5028-Centricity Universal Viewer-00110911; UDI/DI 840682103800, Product ID numbers: 4362-1-Centricity Universal Viewer-01299095, 4362-1-Centricity Universal Viewer-19732280, HC0399-Centricity Universal Viewer-01282972, 509585-Centricity Universal Viewer-01826272, 81111-Centricity Universal Viewer-01286673, AMS39-Centricity Universal Viewer-00545606, 85021-Centricity Universal Viewer-01426096; n) 2088026-714, No UDI/DI, Product ID numbers: 4379-1-Centricity Universal Viewer-02184891; UDI/DI 840682103800, Product ID numbers: 101690-1-Centricity Universal Viewer-20091140, 531344-Centricity Universal Viewer-00000707; o) 2088026-723, UDI/DI 840682103800, Product ID numbers: 100666-1-Centricity Universal Viewer-14194768, 100666-1-Centricity Universal Viewer-20084199, 4415-1-Centricity Universal Viewer-01973510, 4318-1-Centricity Universal Viewer-00558703, 4318-1-Centricity Universal Viewer-20080694, 9768-1-Centricity Universal Viewer-00558661, 9768-1-Centricity Universal Viewer-01972254, 4353-2-Centricity Universal Viewer-01589220, 4698-1-Centricity Universal Viewer-00558313, 101725-2-Centricity Universal Viewer-00558642, 4687-1-Centricity Universal Viewer-00558334, 100366-1-Centricity Universal Viewer-00558522, 100366-1-Centricity Universal Viewer-14777071, 100718-2-Centricity Universal Viewer-02350739, 100718-1-Centricity Universal Viewer-02373799, 12968-1-Centricity Universal Viewer-02277843, 12968-1-Centricity Universal Viewer-02278145, 9852-2-Centricity Universal Viewer-00558644, 9852-2-Centricity Universal Viewer-02267839, 9852-2-Centricity Universal Viewer-01271417, 4391-1-Centricity Universal Viewer-02205222, 4391-3-Centricity Universal Viewer-01459013, 4698-11-Centricity Universal Viewer-19667347, 4698-11-Centricity Universal Viewer-15961473, 4382-1-Centricity Universal Viewer-02268547, 4433-1-Centricity Universal Viewer-00558704, 4433-1-Centricity Universal Viewer-17737529, 4669-1-Centricity Universal Viewer-00558691, 4669-1-Centricity Universal Viewer-14768587, 12961-1-Centricity Universal Viewer-00558236, 4306-3-Centricity Universal Viewer-00558573, 4306-1-Centricity Universal Viewer-02263351, 4306-3-Centricity Universal Viewer-02345023, 4306-3-Centricity Universal Viewer-02288781, 100092-1-Centricity Universal Viewer-15531177, 100092-1-Centricity Universal Viewer-14767331, 100092-1-Centricity Universal Viewer-00558648, 100092-1-Centricity Universal Viewer-00558697, 1057-1-Centricity Universal Viewer-02180910, 1057-1-Centricity Universal Viewer-00558402, 1057-1-Centricity Universal Viewer-14961014, 101690-1-Centricity Universal Viewer-00558680, 9796-1-Centricity Universal Viewer-00558571, 9752-1-Centricity Universal Viewer-02393053, 9752-1-Centricity Universal Viewer-00558454, 100468-1-Centricity Universal Viewer-02287985, 100468-1-Centricity Universal Viewer-02288300, 101310-1-Centricity Universal Viewer-02039946, AP6264-Centricity Universal Viewer-00000180, 313741-Centricity Universal Viewer-01422997, 313741-Centricity Universal Viewer-15591262, 000282202-Centricity Universal Viewer-02291736, HBON01-Centricity Universal Viewer-00260801, 691749-Centricity Universal Viewer-01539092, 509940-Centricity Universal Viewer-01813036, HC0529-Centricity Universal Viewer-01364776, HC0245-Centricity Universal Viewer-00059319, 600791-Centricity Universal Viewer-02099049, HC1414-Centricity Universal Viewer-01814335, PC0023-Centricity Universal Viewer-01282279, CC0154-Centricity Universal Viewer-01813738, 601860-Centricity Universal Viewer-01373156, HC0565-Centricity Universal Viewer-00545595, HC4357-Centricity Universal Viewer-02090320, PL1190-Centricity Universal Viewer-21245580, RU8414-Centricity Universal Viewer-01445730, RU7519-Centricity Universal Viewer-01816694, RU4255-Centricity Universal Viewer-02331160, RU0003-Centricity Universal Viewer-02335931, RU9059-Centricity Universal Viewer-01445746, RU9327-Centricity Universal Viewer-02337466, RU1011-Centricity Universal Viewer-02330918, RU1112-Centricity Universal Viewer-01453917, RU1112-Centricity Universal Viewer-01977813, RU1112-Centricity Universal Viewer-01980259, RU1516-Centricity Universal Viewer-02331146, RU1717-Centricity Universal Viewer-01476628, RU0472-Centricity Universal Viewer-02092677, RU3052-Centricity Universal Viewer-02330856, 600101-Centricity Universal Viewer-01273190, RU1544-Centricity Universal Viewer-02330871, RU2526-Centricity Universal Viewer-02331246, RU0994-Centricity Universal Viewer-02331254, RU3794-Centricity Universal Viewer-19673911, RU9226-Centricity Universal Viewer-19673912, RU8356-Centricity Universal Viewer-01445768, 600105-Centricity Universal Viewer-02331152, RU9697-Centricity Universal Viewer-19673913, 2422939-Centricity Universal Viewer-01509505, RU3998-Centricity Universal Viewer-01553790, 482896-Centricity Universal Viewer-00545574, 941804-Centricity Universal Viewer-02287043, 4305-7-Centricity Universal Viewer-16204605, 4305-7-Centricity Universal Viewer-16204606, 4305-7-Centricity Universal Viewer-16204607, 4305-7-Centricity Universal Viewer-16204608, 4305-7-Centricity Universal Viewer-16204609, 4305-7-Centricity Universal Viewer-16204612, 4305-7-Centricity Universal Viewer-16204613; p) 2099714-001, UDI/DI 840682103800, Product ID numbers: 4449-1-Centricity Universal Viewer-00558276, 4449-1-Centricity Universal Viewer-20689478; q) 2100169-001, UDI/DI 840682103800, Product ID numbers: 102123-1-Centricity Universal Viewer-00558570, 102123-1-Centricity Universal Viewer-02157634, 4415-1-Centricity Universal Viewer-00558359, 104287-2-Centricity Universal Viewer-00555505, 4583-1-Centricity Universal Viewer-01367198, 2337-5-Centricity Universal Viewer-01810490, 2337-5-Centricity Universal Viewer-19670358, 12902-1-Centricity Universal Viewer-00602638, 4585-3-Centricity Universal Viewer-01448365, 4585-3-Centricity Universal Viewer-20502096, 1530-1-Centricity Universal Viewer-00558491, 1530-12-Centricity Universal Viewer-01301357, 13295-1-Centricity Universal Viewer-01292684, 13295-1-Centricity Universal Viewer-19609993, 13295-1-Centricity Universal Viewer-19611189, 100971-1-Centricity Universal Viewer-00558469, 4637-1-Centricity Universal Viewer-00558372, 4637-1-Centricity Universal Viewer-02176556, 8746-1-Centricity Universal Viewer-00558308, 2403-7-Centricity Universal Viewer-01973509, 4347-1-Centricity Universal Viewer-00558520; r) 2100491-006, UDI/DI 840682103800, Product ID numbers: 13162-8-Centricity Universal Viewer-01452133, 13162-8-Centricity Universal Viewer-14776447, 13162-8-Centricity Universal Viewer-14776449, 3245-1-Centricity Universal Viewer-01451973, 100864-1-Centricity Universal Viewer-00558400, 100769-1-Centricity Universal Viewer-01298556, 100769-1-Centricity Universal Viewer-02395799, 12966-1-Centricity Universal Viewer-01283020, 12966-1-Centricity Universal Viewer-20025013, RU8366-Centricity Universal Viewer-18045776, RU9595-Centricity Universal Viewer-01812133, RU9546-Centricity Universal Viewer-01812747; s) 2102675-010, UDI/DI 840682103800, Product ID numbers: PC1695-Centricity Universal Viewer-01447701, PC1961-Centricity Universal Viewer-00001720, SA1339-Centricity Universal Viewer-00116444; t) 2102676-001, UDI/DI 840682103800, Product ID numbers: 100104-1-Centricity Universal Viewer-16204735, 100104-1-Centricity Universal Viewer-01299102, 104336-1-Centricity Universal Viewer-00346858, M41980-Centricity Universal Viewer-00044515, 427990-Centricity Universal Viewer-01813160, 485135-Centricity Universal Viewer-01447706, 2336375-Centricity Universal Viewer-01444341, A59927-Centricity Universal Viewer-00040382, PL3673-Centricity Universal Viewer-00038810, PL1494-Centricity Universal Viewer-00038799, PL2970-Centricity Universal Viewer-00038797, PL1124-Centricity Universal Viewer-00038795, PL1499-Centricity Universal Viewer-00038800, PL1501-Centricity Universal Viewer-00038802, PL2624-Centricity Universal Viewer-00038804, PL1005-Centricity Universal Viewer-02336290, 600062-Centricity Universal Viewer-02095166, PL0326-Centricity Universal Viewer-02040892, PL2971-Centricity Universal Viewer-00038808, RU6968-Centricity Universal Viewer-02335840, 4601-1-Centricity Universal Viewer-00558699; u) 2104867-044, UDI/DI 840682103800, Product ID numbers: 9789-1-Centricity Universal Viewer-00558645, 4413-2-Centricity Universal Viewer-02288447, 101171-1-Centricity Universal Viewer-02398531, 104535-3-Centricity Universal Viewer-02396026, 104535-3-Centricity Universal Viewer-02396027, 104535-3-Centricity Universal Viewer-02396028, 104535-3-Centricity Universal Viewer-02396029, 104535-4-Centricity Universal Viewer-02396030, 104535-4-Centricity Universal Viewer-01806096, 4308-1-Centricity Universal Viewer-00558676, 4048-1-Centricity Universal Viewer-00558695, 4048-1-Centricity Universal Viewer-14767809, 4048-1-Centricity Universal Viewer-14767808, 4048-1-Centricity Universal Viewer-14767807, 4048-1-Centricity Universal Viewer-14767806, 4048-1-Centricity Universal Viewer-02256595, 4460-1-Centricity Universal Viewer-02255583, 4460-1-Centricity Universal Viewer-01819618, 4460-7-Centricity Universal Viewer-02255586, 4460-7-Centricity Universal Viewer-02255587, 100019-1-Centricity Universal Viewer-00558666, 100019-1-Centricity Universal Viewer-01477448, 100019-1-Centricity Universal Viewer-01477447, 100019-1-Centricity Universal Viewer-15993728, 100489-1-Centricity Universal Viewer-01819402, 100489-1-Centricity Universal Viewer-20791988, 104249-1-Centricity Universal Viewer-00558667, 4742-2-Centricity Universal Viewer-01766912, 101039-1-Centricity Universal Viewer-01373579, 101039-1-Centricity Universal Viewer-14580319, 9764-1-Centricity Universal Viewer-01722652, 102768-1-Centricity Universal Viewer-02179353, 102768-1-Centricity Universal Viewer-02255613, 102768-1-Centricity Universal Viewer-16174584, 100636-1-Centricity Universal Viewer-01815255, 9094-1-Centricity Universal Viewer-01977629, 4682-1-Centricity Universal Viewer-02274840, 4440-1-Centricity Universal Viewer-00558353, 4440-1-Centricity Universal Viewer-02332746, 100581-1-Centricity Universal Viewer-01814001, 13125-1-Centricity Universal Viewer-00558110, 13125-1-Centricity Universal Viewer-20497227, 4351-1-Centricity Universal Viewer-01627896, 4351-1-Centricity Universal Viewer-16023960, 4742-3-Centricity Universal Viewer-02182462, 13225-1-Centricity Universal Viewer-02032334, 613493-Centricity Universal Viewer-02175791, 683198-Centricity Universal Viewer-02263377, TP0575-Centricity Universal Viewer-01825777, A61971-Centricity Universal Viewer-01891850, PL1961-Centricity Universal Viewer-01981010, PL3362-Centricity Universal Viewer-20631670, RU9255-Centricity Universal Viewer-19727239, RU9216-Centricity Universal Viewer-01982768, RU0120-Centricity Universal Viewer-02335932, RU6164-Centricity Universal Viewer-19720917, SI4014-Centricity Universal Viewer-01889278, 101171-1-Centricity Universal Viewer-02398531, 4308-1-Centricity Universal Viewer-21200827, 4308-1-Centricity Universal Viewer-21200830, 6579-11-Centricity Universal Viewer-02339896, 4742-2-Centricity Universal Viewer-20584004, 4742-2-Centricity Universal Viewer-20685485; v) 2104867-045, UDI/DI 840682103800, Product ID numbers: AH4325-Centricity Universal Viewer-00160202, AH5291-Centricity Universal Viewer-00394902, AP6469-Centricity Universal Viewer-01816421, AH0116-Centricity Universal Viewer-00160203, AH0106-Centricity Universal Viewer-00160204, A54852-Centricity Universal Viewer-00043560.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- CENTRICITY UNIVERSAL VIEWER 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page