Skip to content

ORTHOLOC 3Di recallZ-1466-2023 · Wright Medical Technology, Inc. · initiated Apr 10, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Wright Medical Technology, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Two lots of Ortholoc 3Di Fusion plates contain incorrect dorsiflexion descriptions laser marked on the parts.

FDA root cause category: Employee error

What the company is doing

The firm issued an URGENT: MEDICAL DEVICE RECALL notice to its sales representatives and distributors on 04/10/2023 by email and followed with letters disseminated to the remaining consignees via tracked FedEx mailing on Wednesday, April 12. The notice explained the issue and the risk and requested the following actions be taken: Identify all affected products and isolate/quarantine to prevent accidental use. All parties that further distributed the product were directed to notify anyone to the recipients of the affected products. The firm is seeking return of the products.

Recall details

Recall number
Z-1466-2023
Status
Open, Classified
Initiated by the company
Apr 10, 2023
Posted by the FDA
Apr 26, 2023
Recalling firm
Wright Medical Technology, Inc.
Quantity
41 units
Product code
HRS
FDA submissions
K120359
Distribution
US Nationwide distribution.

Product description

stryker ORTHOLOC 3Di Fusion Plate MTP Right, REF 58A220RT, bone stabilization and fixation

Code information (models, lots, serial numbers)

UDI/DI 00889797101011, Lot Number 1736809

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 92078).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page