Magic3 Go Pediatric Catheter with SureGrip recallZ-1700-2023 · C.R. Bard Inc · initiated Apr 20, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from C.R. Bard Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Some catheters had a manufacturing defect where the clear resealing label separating from the product foil pouch, resulting in a possible opening into the product packaging that could impact product sterility.
FDA root cause category: Under Investigation by firm
What the company is doing
The firm issued an URGENT MEDICAL DEVICE RECALL to its consignee beginning on 04/20/2023 by Fed Ex and email. The notice explained the problem, risk, and requested the following actions be taken: Discontinue distribution of the product and dispose of all affected product, Distributors are directed to notify their customers. For further assistance please contact: 1-844-8BD- LIFE (1-844-823-5433) Say "Recall" when prompted; M-F 8am - 5pm CT
Recall details
- Recall number
- Z-1700-2023
- Status
- Open, Classified
- Initiated by the company
- Apr 20, 2023
- Posted by the FDA
- Jun 8, 2023
- Recalling firm
- C.R. Bard Inc
- Quantity
- 5760 devices
- Product code
- EZD
- FDA submissions
- K172247
- Distribution
- Worldwide
Product description
Magic3 Go Intermittent Urinary Catheters, pediatric Catalog Numbers: a) REF 52810G
Code information (models, lots, serial numbers)
a) REF 52810G, UDI/DI 00801741136870, Lot Numbers: JUGW1445
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Magic3 Go Pediatric Catheter with SureGrip 52810GC. R. Bard, Inc. · CATHETER, STRAIGHT1 recall3 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 92082).
- Z-1698-2023 · Open, ClassifiedMagic3 Go Intermittent Urinary Catheters, Female Catalog Numbers: a) REF 51808, b) REF 51810, c) REF 51812, d) REF 51814, e) REF 51816, f) REF 51818
- Z-1699-2023 · Open, ClassifiedMagic3 Go Intermittent Urinary Catheters, male Catalog Numbers: a) REF 50810G, b) REF 50812G, c) REF 50814G, d) REF 50816G, e) REF 50818G, f) REF 53810G, g) REF 53812G, h) REF 5381…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.