Unicondylar Sled Prosthesis recallZ-1513-2023 · Linkbio Corp. · initiated Mar 21, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Linkbio Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
FDA root cause category: Labeling design
What the company is doing
LinkBio issued Urgent Medical Device Correction letter on March 21, 2023 to US Distributors. Letter states reason for recall, health risk and action to take: Please take note of the amendment / reinforcement of the Surgical Technique and participate in future offering of product training on the updated technique. Two warnings have been added. No product return is required. Please respond using the attached Distributor Reply form to indicate your understanding of this Device Correction and of the need to follow the updated surgical technique. Attendance in future product training on the LINK SLED is required for continued use of the system. No action is required for devices which have already been implanted; continue with regular follow-up activities for patients who have received the LINK SLED knee system. All product lots of the above item numbers still within the labeled expiration date are in scope for this device correction. Waldemar Link GmbH & Co. KG is updating the surgical technique to incorporate the new warnings. The company will also be providing product training, through the LinkBio Medical Education team. Please contact LinkBio at 973-625-1333 or [email protected] for any questions.
Recall details
- Recall number
- Z-1513-2023
- Status
- Open, Classified
- Initiated by the company
- Mar 21, 2023
- Posted by the FDA
- May 4, 2023
- Recalling firm
- Linkbio Corp.
- Quantity
- 13 units
- Product code
- HSX
- FDA submissions
- K954186
- Distribution
- NY, NV Foreign: Bulgaria, France, Italy, Norway, Poland, Sweden, Switzerland, Slovakia, Spain, United Kingdom, Germany, Austria, Argentina, Australia, China, India, Iraq, Jordan, Colombia, Mexico, Saudi Arabia, Ukraine, Vietnam, and Belarus
Product description
LINK SLED Knee System: ALL POLY TIBIAL COMPONENT 11MM X 45MM Item Number: 15-2028/03
Code information (models, lots, serial numbers)
UDI-DI: 04026575043958 All lot numbers within the labeled expiration date
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Unicondylar Sled Prosthesis 15-2028/03WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
Product lines
Other devices under the same FDA submission
31 devices cleared under K954186 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Unicondylar Sled Prosthesis 15-2020/52WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall2 reports
- Unicondylar Sled Prosthesis 15-2020/60WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall2 reports
- Unicondylar Sled Prosthesis 15-2030/06WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall1 report
- Unicondylar Sled Prosthesis 15-2030/10WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall1 report
- Unicondylar Sled Prosthesis 15-2020/46WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall1 report
- Unicondylar Sled Prosthesis 15-2028/13WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall1 report
- Unicondylar Sled Prosthesis 15-2030/13WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall1 report
- Unicondylar Sled Prosthesis 15-2030/14WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall1 report
- Unicondylar Sled Prosthesis 15-2020/40WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2028/01WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2028/02WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Unicondylar Sled Prosthesis 15-2028/04WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 92088).
- Z-1507-2023 · Open, ClassifiedLINK SLED Knee System: ENDO SLED KNEE FEM COMP, SMALL Item Number: 15-2020/40
- Z-1508-2023 · Open, ClassifiedLINK SLED Knee System: ENDO SLED KNEE FEM COMP, MEDIUM SMALL Item Number: 15-2020/46
- Z-1509-2023 · Open, ClassifiedLINK SLED Knee System: ENDO SLED KNEE FEM COMP, MEDIUM Item Number: 15-2020/52
- Z-1510-2023 · Open, ClassifiedLINK SLED Knee System: ENDO SLED KNEE FEM COMP, Large Item Number: 15-2020/60
- Z-1511-2023 · Open, ClassifiedLINK SLED Knee System: ALL POLY TIBIAL COMPONENT 7MM X 45MM Item Number: 15-2028/01
- Z-1512-2023 · Open, ClassifiedLINK SLED Knee System: ALL POLY TIBIAL COMPONENT 9MM X 45MM Item Number: 15-2028/02
- Z-1514-2023 · Open, ClassifiedLINK SLED Knee System: ALL POLY TIBIAL COMPONENT 13MM X 45MM Item Number: 15-2028/04
- Z-1515-2023 · Open, ClassifiedLINK SLED Knee System: ALL POLY TIBIAL COMPONENT 7MM X 50MM Item Number: 15-2028/05
- Z-1516-2023 · Open, ClassifiedLINK SLED Knee System: ALL POLY TIBIAL COMPONENT 9MM X 50MM Item Number: 15-2028/06
- Z-1517-2023 · Open, ClassifiedLINK SLED Knee System: ALL POLY TIBIAL COMPONENT 11MM X 50MM Item Number: 15-2028/07
- Z-1518-2023 · Open, ClassifiedLINK SLED Knee System: ALL POLY TIBIAL COMPONENT 13MM X 50MM Item Number: 15-2028/08
- Z-1519-2023 · Open, ClassifiedLINK SLED Knee System: ALL POLY TIBIAL COMPONENT 7MM X 55MM Item Number: 15-2028/09
- Z-1520-2023 · Open, ClassifiedLINK SLED Knee System: ALL POLY TIBIAL COMPONENT 9MM X 55MM Item Number: 15-2028/10
- Z-1521-2023 · Open, ClassifiedLINK SLED Knee System: ALL POLY TIBIAL COMPONENT 11MM X 55MM Item Number: 15-2028/11
- Z-1522-2023 · Open, ClassifiedLINK SLED Knee System: ALL POLY TIBIAL COMPONENT 13MM X 55MM Item Number: 15-2028/12
- Z-1523-2023 · Open, ClassifiedLINK SLED Knee System: ALL POLY TIBIAL COMPONENT 7MM X 58MM Item Number: 15-2028/13
- Z-1524-2023 · Open, ClassifiedLINK SLED Knee System: ALL POLY TIBIAL COMPONENT 9X58MM Item Number: 15-2028/14
- Z-1525-2023 · Open, ClassifiedLINK SLED Knee System: ALL POLY TIBIAL COMPONENT 11X58MM Item Number: 15-2028/15
- Z-1526-2023 · Open, ClassifiedLINK SLED Knee System: ALL POLY TIBIAL COMPONENT 13X58MM Item Number: 15-2028/16
- Z-1527-2023 · Open, ClassifiedLINK SLED Knee System: ENDO MODEL SLED TIBIAL COMP 9 MM X 45 MM Item Number: 15-2030/02
- Z-1528-2023 · Open, ClassifiedLINK SLED Knee System: ENDO MODEL SLED TIBIAL COMP 11 MM X 45 MM Item Number: 15-2030/03
- Z-1529-2023 · Open, ClassifiedLINK SLED Knee System ENDO MODEL SLED TIBIAL COMP 13 MM X 45 MM Item Number: 15-2030/04
- Z-1530-2023 · Open, ClassifiedLINK SLED Knee System ENDO MODEL SLED TIBIAL COMP 9 MM X 50 MM Item Number: 15-2030/06
- Z-1531-2023 · Open, ClassifiedLINK SLED Knee System ENDO MODEL SLED TIBIAL COMP 11 MM X 50 MM Item Number: 15-2030/07
- Z-1532-2023 · Open, ClassifiedLINK SLED Knee System ENDO MODEL SLED TIBIAL COMP 13 MM X 50 MM Item Number: 15-2030/08
- Z-1533-2023 · Open, ClassifiedLINK SLED Knee System ENDO MODEL SLED TIBIAL COMP 9 MM X 55 MM Item Number: 15-2030/10
- Z-1534-2023 · Open, ClassifiedLINK SLED Knee System ENDO MODEL SLED TIBIAL COMP 11 MM X 55 MM Item Number: 15-2030/11
- Z-1535-2023 · Open, ClassifiedLINK SLED Knee System ENDO MODEL SLED TIBIAL COMP 13 MM X 55 MM Item Number: 15-2030/12
- Z-1536-2023 · Open, ClassifiedLINK SLED Knee System ENDO MODEL SLED TIBIAL COMP 8 MM X 45 MM Item Number: 15-2030/13
- Z-1537-2023 · Open, ClassifiedLINK SLED Knee System ENDO MODEL SLED TIBIAL COMP 8 MM X 50 MM Item Number: 15-2030/14
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page