StealthStation Cranial recallZ-1630-2023 · Medtronic Navigation, Inc. · initiated Apr 11, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Navigation, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
During non-axial/some axial exams, software anomalies occur during procedures affecting depth gauge graphic displays for cranial biopsy; causing displays to no longer synchronize with other navigational information showing inaccurate values, which may result in prolonged/additional procedure, tissue injury
FDA root cause category: Software design
What the company is doing
On 4/11/23, correction notices were sent to healthcare professionals that were advised of the following; 1) Do not utilize the following: a) Biopsy Depth Gauge Graphic in Cranial Biopsy procedures. In accordance with the IFU: Frequently confirm navigation accuracy and system responsiveness during live navigation. Use the navigation instrument to touch several bony anatomical landmarks and confirm that the locations identified on the images match the locations touched on the patient. If accuracy degrades, re-register the patient. Always use the Biopsy Needle Mechanical Depth Stop. If system navigation seems inaccurate and steps to restore accuracy are unsuccessful, abort use of the system. b) Distance to/past Target value for Tumor Resection, Shunt Placement, and DBS procedures. Per the IFU: "Warning: Frequently confirm navigation accuracy and system responsiveness during live navigation. Use the navigation instrument to touch several bony anatomical landmarks and confirm that the locations identified on the images match the locations touched on the patient. If accuracy degrades, re-register the patient." and "If system navigation seems inaccurate and steps to restore accuracy are unsuccessful, abort use of the system." 2) All physician users should review the correction notice. Pass on the notice to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Please maintain a copy of this notice in your records. 3) Complete and return the customer confirmation form that can be returned to [email protected]. Firm will provide warning and instructional placards to be applied to the affected systems. Firm is currently working on a software update and will communicate additional information when it becomes available. If you have any questions related to this issue, contact Technical Services for help at: 1-888-826-5603 or email at [email protected]
Recall details
- Recall number
- Z-1630-2023
- Status
- Open, Classified
- Initiated by the company
- Apr 11, 2023
- Posted by the FDA
- Jun 2, 2023
- Recalling firm
- Medtronic Navigation, Inc.
- Quantity
- 746
- Product code
- HAW
- FDA submissions
- K221087
- Distribution
- Worldwide Distibution: US (nationwide): OH, IL, CA, MI, FL, LA, OK, WA, AR, NJ, MN, MO, WI, MA, ND, NY, MD, OR, AK, NE, TN, AZ, IN, GA, GU, VA, PA, TX, SD, MT, NV, MS, KY, AL, CT, WV, KS, IA, NC, UT, CO, NH, DC, WY. OUS (foreign): ALBANIA, ARGENTINA, Australia, AUSTRIA, BANGLADESH, BARBADOS, BELGIUM, Bosnia and Herzegovina, BULGARIA, CANADA, CHILE, CROATIA, CZECH REPUBLIC, DENMARK, FRANCE, GREECE, INDIA, IRELAND, ITALY, JAPAN, LUXEMBOURG, MARTINIQUE, NEW ZEALAND, POLAND, PORTUGAL, Republic of Korea, ROMANIA, RUSSIAN FEDERATION, SINGAPORE, SLOVAKIA, SPAIN, SWITZERLAND, TAIWAN, Turkey, UNITED ARAB EMIRATES, UNITED KINGDOM
Product description
StealthStation Cranial Software, Models: 9735585, 9735586 (kit), 9735587 (kit), used with StealthStation S7/i7 systems
Code information (models, lots, serial numbers)
Model/UDI-DI; Software Version: 9735585/00763000631635, 9735586/00763000631765, 9735587/00763000631826; 3.1.4
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- StealthStation® Cranial 9735586Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1 recall0 reports
- StealthStation® Cranial 9735587Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1 recall0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page