Incisive CT Power (China)-Computed Tomography X-Ray System recallZ-2260-2023 · Philips North America · initiated Apr 25, 2023
Class II Open, Classified All recalls posted in July 2023
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips North America or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Metal mounting box on the rotating scanner on rotor (heat exchanger box) located within the Incisive CT system may become expelled and make contact with other components and cause injury to user or if detached a loud noise will be emitted and the system will shut down
FDA root cause category: Under Investigation by firm
What the company is doing
Phiips issued Urgent Medical Device Correction Letters mailed to US consignees via Certified United States Postal Service (USPS) with delivery confirmation on 4/25/23 International global markets were notified of the issue and instructed to distribute the FSN to affected consignees in accordance with local requirements. Letter states reason for recall, health risk and action to take: Please continue to use your system in accordance with its intended use. If you need to remain in the room for the duration of a scan, Philips recommends using protective eyewear. " Circulate this Urgent Medical Device Correction Letter to all users of this device so that they are aware of the issue. Please retain this letter with your system(s) until a solution is installed on your system; ensure the letter is in a place likely to be seen/viewed. " Place this URGENT Medical Device Correction Letter with your system documentation. " Please complete and return the attached response form to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction Letter, understanding of the issue, and required actions to be taken. Philips will contact you to schedule a time for a Field Service Engineer (FSE) to visit your site and install a solution to address the issue (reference FCO72800798). If you need additional information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377).
Recall details
- Recall number
- Z-2260-2023
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Apr 25, 2023
- Posted by the FDA
- Jul 28, 2023
- Recalling firm
- Philips North America
- Quantity
- 53 units OUS
- Product code
- JAK
- FDA submissions
- K180015
- Distribution
- Nationwide Foreign: China Algeria Argentina Australia Austria Azerbaijan Bahrain Bangladesh Belarus Belgium Bolivia Brazil Bulgaria Canada Chile China Colombia Czech Republic Ecuador Egypt France Germany Greece India Indonesia Iran Iraq Israel Italy Japan Jordan Korea, Republic of Kosovo Latvia Macedonia Mexico Moldova Netherlands Nigeria Pakistan Panama Philippines Poland Romania Russian Federation Saint Lucia Saint Pierre and Miquelon Saudi Arabia Serbia Slovakia Spain Sweden Taiwan Thailand Tunisia Turkey Turkmenistan United Kingdom Uzbekistan Viet Nam
Product description
Incisive CT Power (China)-Computed Tomography X-Ray System Model: 728148
Code information (models, lots, serial numbers)
UDI-DI: (01)00884838103467 (21) Serial Numbers: 530456 530338 530360 530489 530370 530363 530389 530369 530481 530350 530464 530500 530436 530301 530501 530372 530466 530368 530476 530358 530294 530421 530435 530480 530458 530342 530432 530419 530460 530502 530349 530475 530482 530431 530335 530378 530494 530492 530470 530345 530461 530374 530384 530388 530357 530404 530329 530373 530409 530462 530420 530452 530307
Other devices under the same FDA submission
1 device cleared under K180015 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · System, x-ray, tomography, computed13 recalls5 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 92202).
- Z-2258-2023 · Open, ClassifiedIncisive CT-Computed Tomography X-Ray System Model: 728143
- Z-2259-2023 · Open, ClassifiedIncisive CT for Brazil SKD-Computed Tomography X-Ray System Model: 728146
- Z-2261-2023 · Open, ClassifiedIncisive CT Plus-Computed Tomography X-Ray System Model: 728149
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page