Incisive CT (REF 728143)Recalls & reported problems
Philips Healthcare (Suzhou) Co., Ltd. · Class 2
All models of Incisive CT (2)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
10 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 13
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00884838085015), received between Feb 11, 2021 and Nov 25, 2025 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Use of Device Problem” (2 reports).
13 FDA recall records name this device's identifier, 10 still open.
11 other recalls cover devices cleared under the same FDA submission (K180015); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Jun 26, 2026Class IIOpen, Classified
Potential of failing to trigger clinical scan during Bolus Tracking.
Z-2988-2026 · Root cause: Radiation Control for Health and Safety Act · Recall details · FDA record ↗
- Oct 29, 2025Class IIOpen, Classified
If the screws of tube heat exchanger on the rotating scanner are not tightened well after replacing. This component may become detached and make contact with other components located within the Incisive CT systems during rotation. Other components could be damaged due to the contact with the tube heat exchanger.
Z-0881-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Sep 25, 2025Class IIOpen, Classified
Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axial scan with ORI (Dose Right index)/ DOM (Dose Modulation), the WED (Water Equivalent Diameter) value might set itself to zero, and this may lead to an insufficient dose setting after the surview. If operator misses the insufficient dose and the WED value in User Interface and continues with the subsequent helical/Axial scans, then the obtained images will be noisy due to low dose setting. Issue 3: Due to a software failure, the ECG wave file is not saved. As a result, the cardiac offline image reconstruction fails due to the missing ECG wave file, and the user might rescan the patient. Issue 4: This issue is software fault in which the patient orientation will be changed to NULL under specific random scenarios resulting in radiation being delivered from the wrong orientation (surview scan) or in the wrong position (clinical scan) and will generate incorrect and undiagnosable surview/clinical images. Patients may be rescanned for surview and/or clinical images. Issue 5: During scanning, the preview image display is not consistent, the images are not arriving at a fixed rate, skipping images from time to time during scanning. The obtained images might not support making the clinical diagnosis and the user might decide to re-scan the patient. Issue 6: If a user presses the left and middle/right mouse buttons together or afterwards in a short time, both commands are executed and an incorrect auto ROI is created. This ROI coordinates won t be assigned as the object is empty. As a result, the threshold of contrast level will not be reached in the UI and this will prevent the system from automatically triggering the subsequent clinical scan. There is potential safety risk identified for additionally tracker shots and/or rescan of a patient when the clinical scan is not triggered. This issue only affects the manual ROI process.
Z-0375-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Apr 15, 2024Class IIOpen, Classified
Metal mounting box on the rotating scanner on rotor (heat change box) located within the Incisive CT system, may become compromised, resulting in potential contact with other gantry components located within the Incisive CT system during rotation. Resulting in a fragment of a damaged component expelled at a low velocity
Z-1871-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Dec 7, 2023Class IIOpen, Classified
If the Patient Interface Monitor (PIM) Cable is plugged/unplugged frequently as part of daily operation, the connector and wire may disconnect from the welding point, which may result in intermittent loss of the ECG signal. Loss of ECG signal may lead to the decision to rescan the patient or delay in diagnosis during a clinical emergency.
Z-0719-2024 · Root cause: Component design/selection · Recall details · FDA record ↗
- Apr 25, 2023Class IIOpen, Classified
Metal mounting box on the rotating scanner on rotor (heat exchanger box) located within the Incisive CT system may become expelled and make contact with other components and cause injury to user or if detached a loud noise will be emitted and the system will shut down
Z-2258-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Apr 12, 2023Class IIOpen, Classified
Potential for Incorrect Image Orientation resulting images may be flipped or reversed result in misdiagnosis, incorrect treatment of a condition, or additional radiation exposure if a rescan is required
Z-1623-2023 · Root cause: Software design · Recall details · FDA record ↗
- Mar 20, 2023Class IIOpen, Classified
Multi-Function Foot Switch Unload Pedal Foot Entrapment-Pressing the Unload Pedal of the foot switch to perform the unload function may cause entrapment of the operator's foot and injury
Z-1586-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Dec 23, 2022Class IIOpen, Classified
Multiple software issues which have the potential to lead to misdiagnosis due to image artifacts or incorrect image orientation labels, or the need for a CT rescan. These include: 1. Artifacts may occur on perfusion images, 2. Orientation label reversed in 2D images, 3. Auto voice does not work during the cardiac scan, 4. Auto voice and operators voice are too quiet, patient cant hear, 5. Not all of the images are reconstructed in some cases, 6. No image displayed on the console monitor after enabling CCT Single Scan, 7. Ring artifact appears in 32x0.625 collimation of bone scan when soft tissue view is selected, 8. A design defect where the error couch motion signal of couch firmware causes couch to identify current position as target position in next protocol, and 9. calcium scoring study split into separate series after loading to ISP.
Z-0640-2024 · Root cause: Software design · Recall details · FDA record ↗
- Mar 29, 2022Class IICompleted
Three software issues affecting incorrect image display, error interpreting patient images due to image reporting, and patient exposure to incorrect image/function during clinical use.
Z-0576-2024 · Root cause: Software design · Recall details · FDA record ↗
- Feb 24, 2021Class IITerminated
Philips has internally detected an issue with the actuators used in the Incisive CT couch, which could result in the couch failure to self-lock, and unexpectedly moving slowly downward to its limit position.
Z-1273-2021 · Root cause: Process control · Recall details · FDA record ↗
- Feb 22, 2021Class IITerminated
Actuators in the Incisive gantry could result in the gantry failure to self-lock if both of the 2 sides actuators fail. In some cases in systems with affected components, the gantry may tilt forward slowly, in a worst case, reaching to the tilt limit angle +31degrees
Z-1265-2021 · Root cause: Process control · Recall details · FDA record ↗
- Dec 24, 2020Class IIOpen, Classified
Patient tabletop moved out to the home position during scan initialization, may cause operator/bystander staying by the rear of the table to collide with the moved tabletop and be injured
Z-1507-2024 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Use of Device Problem2
- Detachment of Device or Device Component1
- Insufficient Device Problem Information1
- Intermittent Program or Algorithm Execution1
- Program or Algorithm Execution Failure1
- Unintended System Motion1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Laceration(s)1
- No Patient Involvement1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryNov 25, 2025Use of Device Problem; Insufficient Device Problem Information; Use of Device Problem; Insufficient Device Problem Information
THIS COMPLAINT HAS BEEN EVALUATED BASED ON THE INFORMATION PROVIDED; THE ISSUE REPORTED WAS THAT A PATIENT WHO SUSTAINED A LACERATION ON THE LEG AFTER IT BECAME TRAPPED BETWEEN THE TABLE AND THE GANTRY WHILE THE TABLE WAS BEING MOVED. THE PATIENT WAS SEDATED AND INTUBATED AT THE TIME OF THE INCIDENT AND WAS BEING TREATED FOR GI BLEEDING. ADDITIONAL INFORMATION WAS RECEIVED AT A LATER TIME CONFIRMED THAT THE PATIENT HAD A LACERATION AROUND 5 INCHES LONG AND INTO THE MUSCLE. IT WAS UNABLE TO BE REPAIRED WITH SUTURES DUE TO THE SKIN BEING TOO DELICATE. BASED ON THE AVAILABLE INFORMATION, THIS DEVICE EVENT HAS BEEN CLASSIFIED AS A SERIOUS INJURY. THEREFORE, THIS EVENT HAS BEEN DETERMINED TO BE A REPORTABLE EVENT.
Patient problems: Laceration(s)
MAUDE report 23653241 ↗MalfunctionOct 24, 2023Intermittent Program or Algorithm Execution; Intermittent Program or Algorithm Execution
THIS COMPLAINT HAS BEEN EVALUATED BASED ON THE INFORMATION PROVIDED; THERE IS NO ALLEGATION OF DEATH OR SERIOUS INJURY. IT WAS REPORTED THAT SYSTEM HAS RECOGNIZED THE PATIENT POSITION FROM FACE UPWARD TO FACE DOWNWARD. ISSUE WAS IDENTIFIED BY THE USER AFTER THE SHOLE SCAN COMPLETION BUT BEFORE IMAGE DISTRIBUTION FOR DIAGNOSIS. THERE WAS NO ACTUAL INJURY OCCURRED. HOWEVER, IIA(ISSUE IMPACT ASSESSMENT) D001255537_REV_C WAS UPDATED WITH RISK ASSESSMENT RE-EVALUATED. AN INTERNAL RETROSPECTIVE REVIEW OF THE COMPLAINT HAS IDENTIFIED THIS COMPLAINT A SIMILAR MALFUNCTION OF A MDR EVENT WE HAVE REPORTED. THEREFORE, THIS ISSUE HAS BEEN DETERMINED TO BE A REPORTABLE EVENT. RISK EVALUATION CONCLUSION IN IIA: POH FOR S0 ¿ MISDIAGNOSIS IS ESTIMATED AT L3 - PROBABLE. POH FOR S1 ¿ MISDIAGNOSIS IS ESTIMATED AT L2 - REMOTE. POH FOR S2 ¿ MISDIAGNOSIS IS ESTIMATED AT L2 - REMOTE. POH FOR S3 ¿ MISDIAGNOSIS IS ESTIMATED AT L2 ¿ REMOTE. PHILIPS HAS COMPLETED THE INVESTIGATION OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17994028 ↗MalfunctionApr 4, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THE ISSUE REPORTED WAS DURING A PATIENT PROCEDURE, THE HEAT EXCHANGER CAME LOOSE/DETACHED WITHIN THE GANTRY CAUSING DAMAGE INTERNALLY. THERE WAS NO REPORT OF HARM. IF THIS ISSUE WERE TO RECUR WHERE THE HEAT EXCHANGER WERE TO COME LOOSE AND BE EXPELLED THROUGH THE GANTRY, THERE IS POTENTIAL FOR DEATH OR SERIOUS INJURY. BASED ON THE AVAILABLE INFORMATION, THIS ISSUE HAS BEEN DETERMINED TO BE A REPORTABLE EVENT. THIS EVENT IS CURRENTLY UNDER INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16672449 ↗MalfunctionDec 22, 2022Use of Device Problem; Program or Algorithm Execution Failure; Use of Device Problem; Program or Algorithm Execution Failure
THE ISSUE REPORTED WAS THE PREVIEW IMAGES FOR A PATIENT HEAD SCAN WERE FLIPPED; FACE UP POSITION TO FACE DOWN. A PHILIPS FIELD SERVICE ENGINEER (FSE) CONFIRMED THE FLIPPED/MISLABELED IMAGES WERE NOT INITIALLY RECOGNIZED WHICH RESULTED IN A MISDIAGNOSIS. AN ADDITIONAL SCAN WAS PERFORMED DURING A CLINICAL EXAMINATION OF THE PATIENT PRIOR TO SURGERY WHICH IDENTIFIED THAT THE IMAGES WERE FLIPPED/MISLABELED ON THE INITIAL SCAN. THE SURGERY WAS PERFORMED ON THE ACCURATE SIDE AND CORRECT MEDICAL TREATMENT WAS PROVIDED. THERE IS POTENTIAL FOR DEATH/SERIOUS INJURY IF THIS ISSUE WERE TO RECUR, THEREFORE THIS EVENT IS CONSIDERED REPORTABLE. THIS ISSUE IS CURRENTLY UNDER INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16035044 ↗MalfunctionFeb 11, 2021Unintended System Motion; Unintended System Motion
THIS COMPLAINT HAS BEEN EVALUATED BASED ON THE INFORMATION PROVIDED; THERE IS NO ALLEGATION OF DEATH OR SERIOUS INJURY. PHILIPS SUZHOU FACTORY DETECTED AN ISSUE WITH INCISIVE CT SYSTEMS WHERE THE COUCH AND GANTRY CANT SELF-LOCK RESULTING IN THE COUCH TO DESCEND DOWN / THE GANTRY TO TILT AUTOMATICALLY AT A VERY SLOW SPEED TO ITS LIMIT POSITION OR STOP WHEN REACH A FORCE BALANCE POINT. THIS HAS BEEN DETERMINED TO BE A REPORTABLE EVENT. THIS ISSUE IS CURRENTLY UNDER INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement
MAUDE report 11315961 ↗
FDA clearance and classification
Cleared through 510(k) K180015 on Mar 20, 2018 (decision: Substantially Equivalent). Submitted by Philips Healthcare (Suzhou) Co., Ltd. as “Philips Incisive CT”. The FDA review took 77 days. FDA record ↗
The FDA regulates product code JAK (system, x-ray, tomography, computed) under 21 CFR 892.1750 as a Class II device. 21 CFR 892.1750 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K180015. One submission often covers many models, so check the product and lot details in each record.
- Z-1003-2026 · Dec 3, 2025 · Open, ClassifiedPhilips Incisive CT
- Z-0376-2026 · Sep 25, 2025 · Open, ClassifiedModel: CT 5300; Product Code (REF): 728285; Software Versions: 5.1.0.X & 5.1.1.X;
- Z-0377-2026 · Sep 25, 2025 · Open, ClassifiedModel: Incisive CT for Brazil SKD; Product Code (REF): 728146; Software Versions: 5.1.0.X & 5.1.1.X;
- Z-2520-2023 · Aug 2, 2023 · Open, ClassifiedIncisive CT, software version 5.0
- Z-2261-2023 · Apr 25, 2023 · Open, ClassifiedIncisive CT Plus-Computed Tomography X-Ray System Model: 728149
- Z-2260-2023 · Apr 25, 2023 · Open, ClassifiedIncisive CT Power (China)-Computed Tomography X-Ray System Model: 728148
- Z-2259-2023 · Apr 25, 2023 · Open, ClassifiedIncisive CT for Brazil SKD-Computed Tomography X-Ray System Model: 728146
- Z-1625-2023 · Apr 12, 2023 · Open, ClassifiedIncisive CT Plus -Whole-body computed tomography (CT) X-Ray System Model Number: 728149 (OUS)
- Z-1622-2023 · Apr 12, 2023 · Open, ClassifiedIncisive CT- whole-body computed tomography (CT) X-Ray System Model Number: 728144
- Z-1624-2023 · Apr 12, 2023 · Open, ClassifiedIncisive CT Power- Whole-body computed tomography (CT) X-Ray System Model Number: 728148 (OUS)
- Z-1585-2023 · Mar 20, 2023 · Open, ClassifiedPhilips Incisive CT -Computed Tomography X-Ray System Model Number: 72814
Common questions about the Incisive CT
Has the Incisive CT been recalled?
Yes. 13 FDA recall records name this device's identifier: 10 still open, 1 completed by the company but not yet closed by the FDA and 2 closed (terminated) by the FDA. The most recent, Z-2988-2026 (Class II), began on Jun 26, 2026: Potential of failing to trigger clinical scan during Bolus Tracking. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Incisive CT?
The FDA's MAUDE database lists 5 adverse event reports for this device, received between Feb 11, 2021 and Nov 25, 2025. In those reports, the most common specific device problems are “Use of Device Problem” (2), “Detachment of Device or Device Component” (1) and “Intermittent Program or Algorithm Execution” (1); the most common patient problem is “Laceration(s)” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Incisive CT?
None of the reports are classified as deaths, 1 as an injury and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Incisive CT FDA approved?
It was cleared through 510(k) K180015 on Mar 20, 2018. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Incisive CT, and is it still sold?
The FDA's device database lists Philips Healthcare (Suzhou) Co., Ltd. as the company responsible for it (brand Incisive CT, model Incisive CT). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other System, x-ray, tomography, computed devices
- ARTIS icono biplane 11327600Siemens Healthcare GmbH · System, x-ray, tomography, computed15 recalls2 deaths49 reports
- ARTIS icono floor 11327700Siemens Healthcare GmbH · System, x-ray, tomography, computed15 recalls23 reports
- Spectral CTPhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed11 recalls3 reports
- Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · System, X-Ray, Tomography, Computed9 recalls20 reports
- Brilliance iCTPhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed8 recalls5 reports
- Revolution Apex EliteGE Medical Systems, L.L.C. · System, x-ray, tomography, computed8 recalls0 reports
- Brilliance CT Big BorePhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed7 recalls1 report
- Revolution ApexGE Medical Systems, L.L.C. · System, x-ray, tomography, computed6 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.