Bellavista recallZ-1966-2023 · Vyaire Medical · initiated May 5, 2023

Completed

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Vyaire Medical or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Vyaire Medical identified two patient safety risks during the use of the bellavista 1000 and bellavista 1000e: 1. Under certain conditions of use, the touchscreen may become unresponsive and the device application stops responding (APP Hang). When this occurs, if the user continues to interact with the touchscreen, a message will appear that states DeviceSoftware.Application is not responding or a decommission screen will appear on the user interface. The ventilator will issue both an audible and visual alarm. Ventilation continues without interruption with the settings applied prior to the APP Hang error. 2. The potential for a use error has been identified if the operator applies the proposed settings without confirming the settings are suitable for the patient.

FDA root cause category: Nonconforming Material/Component

What the company is doing

The firm notified its consignees by email on 04/27/2023. The notice explained the issue, the risk to health and requested the following: ACTIONS TO BE TAKEN BY CONSUMERS/END USERS 1. Confirm receipt and thoroughly review the contents of the Customer Notification package. 2. If affected devices are transferred to another location or organization, forward the complete Customer Notification package to the respective users. 3. To reduce the change that the AppHang error will occur, Vyaire recommends that when the )2 suction maneuver is active, the end user should not attempt to activate any controls or screens until: The O2 suction maneuver has timed or The O2 suction maneuver has been aborted by selecting the O2 suction button again to turn it off. Vyaire further recommends the following: User re-configures the screen to remove the circuit monitoring tile, if applicable. User does not connect anything to either of the two USB ports on the left side of the ventilator during patient ventilation. For more detailed instructions, please refer to Appendix A. Following these steps will help to reduce the likelihood of the AppHang occurring. 4. When transitioning between ventilation modes, always check and confirm the proposed settings. The proposed settings are based on the current level of support provided in the active mode at the time of the mode change. Special attention should be given when transitioning back from a spontaneous breathing mode such as Pressure Support Ventilation (PSV).

Recall details

Recall number
Z-1966-2023
Status
Completed
Initiated by the company
May 5, 2023
Posted by the FDA
Jun 15, 2023
Recalling firm
Vyaire Medical
Quantity
910 units
Product code
CBK
FDA submissions
K183364
Distribution
US Nationwide Distribution

Product description

bellavista 1000 Ventilator, Catalogue Number 301.100.030; Continuous Use Ventilator

Code information (models, lots, serial numbers)

UDI/DI: 10190752197664 and 07640149381115, ALL serial numbers with Software Version 6.1.0.2 and 6.1.1

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

1 device cleared under K183364 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 92243).