Bellavista 301.100.030Recalls & reported problems

Vyaire Medical, Inc. · Adult and pediatric ventilator · Class 2

All models of Bellavista (2)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Vyaire Bellavista ventilator is a breathing machine for adults and children. Some reports describe screen and display problems, and others describe gas output failures. The 156 reports include no deaths and 10 injuries. The one recall, begun in May 2023 and now closed, concerned a touchscreen that could stop responding during use. Reports rose from 43 in the prior 12 months to 64 in the latest 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
149
Deaths reported
0
Injuries reported
10
Recalls naming it
1

Summary

The FDA's MAUDE database contains 149 adverse event reports that list this exact device (DI 10190752197664), received between Mar 5, 2024 and Jun 6, 2026 — classified as 10 injuries, 139 malfunctions.

The most frequently reported specific device problem is “Display or Visual Feedback Problem” (26 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 4 reportsApr 2024: 7 reportsMay 2024: 4 reportsJun 2024: 16 reportsJul 2024: 8 reportsAug 2024: 4 reportsSep 2024: 4 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 1 reportsApr 2025: 4 reportsMay 2025: 7 reportsJun 2025: 5 reportsJul 2025: 2 reportsAug 2025: 1 reportsSep 2025: 9 reportsOct 2025: 11 reportsNov 2025: 13 reportsDec 2025: 17 reportsJan 2026: 3 reportsFeb 2026: 7 reportsMar 2026: 7 reportsApr 2026: 8 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202617
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 5, 2023Class IICompleted

    Vyaire Medical identified two patient safety risks during the use of the bellavista 1000 and bellavista 1000e: 1. Under certain conditions of use, the touchscreen may become unresponsive and the device application stops responding (APP Hang). When this occurs, if the user continues to interact with the touchscreen, a message will appear that states DeviceSoftware.Application is not responding or a decommission screen will appear on the user interface. The ventilator will issue both an audible and visual alarm. Ventilation continues without interruption with the settings applied prior to the APP Hang error. 2. The potential for a use error has been identified if the operator applies the proposed settings without confirming the settings are suitable for the patient.

    Z-1966-2023 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Display or Visual Feedback Problem26
  • Gas Output Problem21
  • Failure to Deliver15
  • No Device Output14
  • Gas/Air Leak12
  • Incorrect Measurement11
  • No Display/Image8
  • Inaccurate Flow Rate7

Patient problems

  • No Clinical Signs, Symptoms or Conditions142
  • Insufficient Information8

Recent adverse event reports

The 25 most recent of 149 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 6, 2026Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE DEVICE EXPERIENCED INTERFACE ISSUES AND DISPLAYED A DECOMMISSION SCREEN. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THERE WAS NO ADVERSE EFFECT TO THE PATIENT AS A RESULT OF THE MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25418975 ↗
  • MalfunctionMay 4, 2026Gas Output Problem

    THE CUSTOMER REPORTED THAT THE VENTILATOR TRIGGERED A TECHNICAL FAILURE 300, 400, 545 ERROR. NO PATIENT INVOLVEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25072718 ↗
  • MalfunctionApr 21, 2026Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE DEVICE SHOWED A BLUE SCREEN. THERE WAS NO PATIENT INVOLVEMENT RELATED TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24937802 ↗
  • MalfunctionApr 16, 2026Failure to Deliver; Infusion or Flow Problem; Reset Problem

    IT WAS REPORTED THAT THE DEVICE EXPERIENCED A LONG PAUSE WITH INSPIRATORY AND EXPIRATION AND TECHNICAL FAILURE ALARMS 300, 316 AND 545. THERE WAS NO PATIENT INVOLVEMENT RELATED TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24897489 ↗
  • MalfunctionApr 13, 2026Inadequate User Interface

    IT WAS REPORTED THAT THE DEVICE SHOWED A BLUE SCREEN. THERE WAS NO PATIENT INVOLVEMENT RELATED TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24860900 ↗
  • MalfunctionApr 6, 2026Display or Visual Feedback Problem; Inadequate User Interface

    IT WAS REPORTED THAT WHILE VENT WAS IN USE ON A PATIENT, WHEN THE VENT SHOWED A BLUE SCREEN AND DID NOT APPEAR TO BE VENTILATING. NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24794226 ↗
  • MalfunctionApr 6, 2026Display or Visual Feedback Problem

    THE CUSTOMER REPORTED THAT THE VENTILATOR'S DISPLAY IS BLACK, AND UPON CONNECTION TO A POWER SOURCE, IT EXHIBITS A FLASHING RED LIGHT FOLLOWED BY YELLOW, ACCOMPANIED BY CONTINUOUS BEEPING. THERE WAS NO PATIENT INVOLVEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24792631 ↗
  • MalfunctionApr 6, 2026Gas/Air Leak

    IT WAS REPORTED THAT DURING USE ON A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE EXPERIENCED AN O2 LEAKAGE FAILURE. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24792625 ↗
  • MalfunctionApr 3, 2026Application Program Freezes or Fails to Launch

    COMPLAINANT REPORTED A FROZEN SCREEN AND BROKEN O2 FILTER ASSEMBLY. THERE WAS NO PATIENT INVOLVEMENT RELATED TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24783527 ↗
  • MalfunctionApr 2, 2026Circuit Failure

    IT WAS REPORTED THAT THE VENTILATOR FLOW SENSOR FAILED CIRCUIT TEST WHILE ON A PATIENT. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24769213 ↗
  • MalfunctionMar 25, 2026Display or Visual Feedback Problem; Failure to Deliver

    CUSTOMER REPORTED THE VENT WAS GIVING OFF TECHNICAL FAILURE ALARMS 300, 316, 545. THERE WAS NO PATIENT INVOLVEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24691397 ↗
  • MalfunctionMar 20, 2026Incorrect Measurement

    IT WAS REPORTED THAT THE DEVICE SHOWED INACCURATE OXYGEN READINGS. THERE WAS NO PATIENT INVOLVEMENT RELATED TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24651056 ↗
  • MalfunctionMar 20, 2026Incorrect Measurement

    IT WAS REPORTED THAT THE DEVICE WAS SHOWING INACCURATE OXYGEN READINGS. THERE WAS NO PATIENT INVOLVEMENT RELATED TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24651047 ↗
  • MalfunctionMar 14, 2026Gas Output Problem

    CUSTOMER REPORTED GETTING THE TECHNICAL FAILURE 271 AND 388 ALARMS ON THE UNIT. THERE WAS NO PATIENT INVOLVEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24600415 ↗
  • MalfunctionMar 12, 2026Gas Output Problem

    IT WAS REPORTED THAT THE DEVICE ALARMED TECHNICAL FAILURE 389, NO O2 DOSING POSSIBLE. THERE WAS NO PATIENT INVOLVEMENT RELATED TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24586395 ↗
  • MalfunctionMar 6, 2026Gas Output Problem

    IT WAS REPORTED THAT BEFORE USE, THE DEVICE EXPERIENCED A TECHNICAL FAILURE 379 - "NO O2 DOSING POSSIBLE". COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24532766 ↗
  • MalfunctionMar 5, 2026Improper Flow or Infusion

    IT WAS REPORTED THAT BEFORE USE, THE DEVICE DISPLAYED A TECHNICAL FAILURE 389 "NO O2 DOSING POSSIBLE". COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24520937 ↗
  • MalfunctionFeb 25, 2026Display or Visual Feedback Problem; Failure to Deliver

    CUSTOMER STATED THAT THE DEVICE SHOWED TF 542, 316, 300 ALARMS. THERE WAS NO PATIENT INVOLVEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24443341 ↗
  • MalfunctionFeb 23, 2026Inaccurate Flow Rate

    CUSTOMER STATED THE DEVICE WAS HAVING FIO2 MISMATCH ISSUES WHILE IN USE ON PATIENT. THERE WAS NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24420596 ↗
  • MalfunctionFeb 16, 2026Gas Output Problem

    THE CUSTOMER REPORTED THAT THE VENTILATOR TRIGGERED A TECHNICAL FAILURE 389 ALARM WHILE PERFORMING O2 GAS PATH TEST. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24345565 ↗
  • MalfunctionFeb 6, 2026Display or Visual Feedback Problem; No Flow

    COMPLAINANT REPORTED THAT WHILE ON A PATIENT, THE DEVICE STOPPED VENTILATING, ALARMED AND SHOWED AN INTERFACE ERROR. THE COMPLAINANT INDICATED ANOTHER DEVICE WAS OBTAINED TO CONTINUE TREATMENT, WITHOUT ANY PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24280362 ↗
  • MalfunctionFeb 3, 2026Erratic or Intermittent Display

    THE CUSTOMER REPORTED THAT THE VENTILATOR DISPLAYED A BLUE SCREEN. NO PATIENT INVOLVEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24251542 ↗
  • MalfunctionFeb 3, 2026Erratic or Intermittent Display

    THE CUSTOMER REPORTED THAT THE VENTILATOR SCREEN FLICKERS. NO PATIENT INVOLVEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24251526 ↗
  • MalfunctionFeb 2, 2026Device Alarm System; Gas Output Problem

    THE CUSTOMER REPORTED THAT THE VENTILATOR TRIGGERED AN O2 MISMATCH ALARM. NO PATIENT INVOLVEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24236679 ↗
  • MalfunctionJan 29, 2026Infusion or Flow Problem

    CUSTOMER STATED THAT THE OCCLUSION ERROR WILL NOT GO AWAY WHILE BEING USED ON A PATIENT. THERE WAS NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24210202 ↗