Skip to content

ambIT Cassette Filter Male Luer, Product Code: 220266 recallZ-1688-2023 · Avanos Medical, Inc. · initiated Apr 24, 2023

Class II Terminated All recalls posted in June 2023

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Avanos Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Some of the ambIT kits were potentially distributed without an air in-line filter.

FDA root cause category: Under Investigation by firm

What the company is doing

Avanos Medical notified customers on 04/24/2023 via email. Customers were instructed to evaluate their inventory of products to determine if any of the impacted product codes were present in their facility. The letter provided information identifying the affected product. It was indicated to segregate and contact Avanos to coordinate return/destruction of impacted product. Additionally, customers were instructed to complete and return the Acknowledgement Form.

Recall details

Recall number
Z-1688-2023
Recall class
Class II
Status
Terminated
Initiated by the company
Apr 24, 2023
Posted by the FDA
Jun 6, 2023
Recalling firm
Avanos Medical, Inc.
Quantity
1155 total, 280 individually packaged
Product code
FRN
FDA submissions
K162165
Distribution
Worldwide distribution - US Nationwide distribution in the states of Texas and Oregon. The country of Australia.

Product description

ambIT Cassette Filter Male Luer, Product Code: 220266

Code information (models, lots, serial numbers)

Lot Number (Expiration Date): F201701 (17 Jan 2025)

Other devices under the same FDA submission

4 devices cleared under K162165 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 92279).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page