Skip to content

ambIT* PIB-PCA v1.5 System w/Spike Filter Cass 220574Recalls & reported problems

Avanos Medical, Inc. · ambIT* PIB-PCA v1.5 System w/Spike Filter Cass · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00193494001308), received between Apr 9, 2024 and Apr 10, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Excess Flow or Over-Infusion” (2 reports).

5 other recalls cover devices cleared under the same FDA submission (K162165); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Excess Flow or Over-Infusion2
  • Improper Flow or Infusion1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 10, 2025Excess Flow or Over-Infusion; Excess Flow or Over-Infusion

    FILL VOLUME: 500ML, FLOW RATE: 4CC/HR, PROCEDURE: TOTAL LEFT KNEE, CATHPLACE: UNKNOWN. IT WAS REPORTED "WHEN PUMP WAS TURNED ON IT SHOWED 20. GAVE MEDICINE VERY QUICKLY AND ACTED LIKE IT WASN'T GOING TO STOP- WENT PAST 4 ML. TURNED OFF/ON, STILL RAN 'FAST', TRIED TO REPROGRAM. [S]TARTED AT 25. STILL GIVING MEDICINE 'FAST' AND NOT STOP. CHANGED PUMP." THE PATIENT WAS REPORTED TO BE DOING "GOOD." THERE WAS NO REPORTED INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21822153 ↗
  • MalfunctionDec 19, 2024Excess Flow or Over-Infusion; Excess Flow or Over-Infusion

    FILL VOLUME: UNKNOWN. FLOW RATE: 15 ML/HR. PROCEDURE: TOTAL KNEE. CATHPLACE: UNKNOWN. IT WAS REPORTED THE NURSE REPORTED THE "PUMP [WAS] GIVING OVER BOLUS AMOUNT (PROGRAMMED FOR 15ML). WHEN DOSE REACHED 15 ML NURSE DISCONNECTED FROM PATIENT AND PUMP CONTINUED TO RUN UNTIL IT REACHED 20 ML. RECHECKED SETTINGS AND PROGRAMMED FOR 15 ML." THER WAS NO REPORTED INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20981824 ↗
  • MalfunctionOct 16, 2024Improper Flow or Infusion; Improper Flow or Infusion

    IT WAS REPORTED VIA FDA MEDWATCH MEDSUN REPORT (B)(4), ¿THE (RN) SET UP AN AMBIT PERIPHERAL NERVE CATHETER PUMP FOR A PATIENT. THE PUMP WAS PROGRAMMED TO DELIVER AN INTERMITTENT BOLUS OF 8ML EVERY 2 HOURS. HOWEVER, AFTER THE INITIAL BOLUS WAS STARTED, THE PUMP MALFUNCTIONED AND WAS ABOUT TO ADMINISTER MORE THAN THE INTENDED DOSE. THE RN QUICKLY DISCONNECTED THE PUMP FROM THE PATIENT'S CATHETER AND OBSERVED THAT IT HAD REACHED 9.6ML BEFORE TURNING IT OFF. THE RN THEN DISCUSSED THE ISSUE WITH THE PRIMARY MEDICAL TEAM AND WAS INSTRUCTED TO USE EXISTING ORDERS AND OBTAIN A NEW AMBIT PUMP. THE INITIAL 8ML BOLUS WAS WASTED TO ENSURE THAT THE PATIENT DOES NOT RECEIVE A DOUBLE DOSE OF MEDICATION. THERAPIES BEING USED ON THE PATIENT AT THE TIME OF THE EVENT THAT MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT: NOT KNOWN;¿ THE INCIDENT OCCURRED IN MAY 2024.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20464324 ↗
  • MalfunctionApr 9, 2024

    WHILE ATTEMPTING TO START AMBIT PUMP, THERE WAS ALREADY A BOLUS GIVEN WHICH HAD NOT BEEN PROGRAMMED INTO THE PUMP YET.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19070460 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K162165. One submission often covers many models, so check the product and lot details in each record.

  • Z-1685-2023 · Apr 24, 2023 · Terminated
    AVNS ambIT System - PIB/PCA W 5in Cath, Product Code: 220507
  • Z-1686-2023 · Apr 24, 2023 · Terminated
    SMMT ambIT Kit Basic Filter Luer Cass, Product Code: 220537
  • Z-1688-2023 · Apr 24, 2023 · Terminated
    ambIT Cassette Filter Male Luer, Product Code: 220266
  • Z-1687-2023 · Apr 24, 2023 · Terminated
    SMMT ambIT Reusable Program - Luer Cass, Product Code: 220568
  • Z-1684-2023 · Apr 24, 2023 · Terminated
    AVNS ambIT System - PIB/PCA W 2.5in Cath, Product Code: 220509

Other Pump, infusion devices

All Pump, infusion devices →

Sources and updates