ambIT* PIB-PCA v1.5 System w/Spike Filter Cass 220574Recalls & reported problems
Avanos Medical, Inc. · ambIT* PIB-PCA v1.5 System w/Spike Filter Cass · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00193494001308), received between Apr 9, 2024 and Apr 10, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Excess Flow or Over-Infusion” (2 reports).
5 other recalls cover devices cleared under the same FDA submission (K162165); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Excess Flow or Over-Infusion2
- Improper Flow or Infusion1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 10, 2025Excess Flow or Over-Infusion; Excess Flow or Over-Infusion
FILL VOLUME: 500ML, FLOW RATE: 4CC/HR, PROCEDURE: TOTAL LEFT KNEE, CATHPLACE: UNKNOWN. IT WAS REPORTED "WHEN PUMP WAS TURNED ON IT SHOWED 20. GAVE MEDICINE VERY QUICKLY AND ACTED LIKE IT WASN'T GOING TO STOP- WENT PAST 4 ML. TURNED OFF/ON, STILL RAN 'FAST', TRIED TO REPROGRAM. [S]TARTED AT 25. STILL GIVING MEDICINE 'FAST' AND NOT STOP. CHANGED PUMP." THE PATIENT WAS REPORTED TO BE DOING "GOOD." THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21822153 ↗MalfunctionDec 19, 2024Excess Flow or Over-Infusion; Excess Flow or Over-Infusion
FILL VOLUME: UNKNOWN. FLOW RATE: 15 ML/HR. PROCEDURE: TOTAL KNEE. CATHPLACE: UNKNOWN. IT WAS REPORTED THE NURSE REPORTED THE "PUMP [WAS] GIVING OVER BOLUS AMOUNT (PROGRAMMED FOR 15ML). WHEN DOSE REACHED 15 ML NURSE DISCONNECTED FROM PATIENT AND PUMP CONTINUED TO RUN UNTIL IT REACHED 20 ML. RECHECKED SETTINGS AND PROGRAMMED FOR 15 ML." THER WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20981824 ↗MalfunctionOct 16, 2024Improper Flow or Infusion; Improper Flow or Infusion
IT WAS REPORTED VIA FDA MEDWATCH MEDSUN REPORT (B)(4), ¿THE (RN) SET UP AN AMBIT PERIPHERAL NERVE CATHETER PUMP FOR A PATIENT. THE PUMP WAS PROGRAMMED TO DELIVER AN INTERMITTENT BOLUS OF 8ML EVERY 2 HOURS. HOWEVER, AFTER THE INITIAL BOLUS WAS STARTED, THE PUMP MALFUNCTIONED AND WAS ABOUT TO ADMINISTER MORE THAN THE INTENDED DOSE. THE RN QUICKLY DISCONNECTED THE PUMP FROM THE PATIENT'S CATHETER AND OBSERVED THAT IT HAD REACHED 9.6ML BEFORE TURNING IT OFF. THE RN THEN DISCUSSED THE ISSUE WITH THE PRIMARY MEDICAL TEAM AND WAS INSTRUCTED TO USE EXISTING ORDERS AND OBTAIN A NEW AMBIT PUMP. THE INITIAL 8ML BOLUS WAS WASTED TO ENSURE THAT THE PATIENT DOES NOT RECEIVE A DOUBLE DOSE OF MEDICATION. THERAPIES BEING USED ON THE PATIENT AT THE TIME OF THE EVENT THAT MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT: NOT KNOWN;¿ THE INCIDENT OCCURRED IN MAY 2024.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20464324 ↗MalfunctionApr 9, 2024
WHILE ATTEMPTING TO START AMBIT PUMP, THERE WAS ALREADY A BOLUS GIVEN WHICH HAD NOT BEEN PROGRAMMED INTO THE PUMP YET.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19070460 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K162165. One submission often covers many models, so check the product and lot details in each record.
- Z-1685-2023 · Apr 24, 2023 · TerminatedAVNS ambIT System - PIB/PCA W 5in Cath, Product Code: 220507
- Z-1686-2023 · Apr 24, 2023 · TerminatedSMMT ambIT Kit Basic Filter Luer Cass, Product Code: 220537
- Z-1688-2023 · Apr 24, 2023 · TerminatedambIT Cassette Filter Male Luer, Product Code: 220266
- Z-1687-2023 · Apr 24, 2023 · TerminatedSMMT ambIT Reusable Program - Luer Cass, Product Code: 220568
- Z-1684-2023 · Apr 24, 2023 · TerminatedAVNS ambIT System - PIB/PCA W 2.5in Cath, Product Code: 220509
Other Pump, infusion devices
- Alaris 8100Carefusion 303, Inc. · Pump, infusion2 recalls2 deaths650,218 reports
- Alaris 8100Carefusion 303, Inc. · Pump, infusion4 recalls78 deaths344,514 reports
- Alaris 8015Carefusion 303, Inc. · Pump, infusion2 recalls45 deaths148,093 reports
- BD® Alaris™ Pump Module, Model 8100Carefusion 303, Inc. · Pump, infusion5 recalls30 deaths51,256 reports
- Alaris 8110Carefusion 303, Inc. · Pump, infusion2 recalls4 deaths21,956 reports
- Alaris 8110Carefusion 303, Inc. · Pump, infusion2 recalls2 deaths15,547 reports
- Alaris 8015Carefusion 303, Inc. · Pump, infusion3 recalls6 deaths13,022 reports
- CADD 21-6500-51Icu Medical, Inc. · Pump, infusion2 deaths8,231 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page