Skip to content

BRONCHOFIBERSCOPE recallZ-2012-2023 · Olympus Corporation of the Americas · initiated Jun 8, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There have been complaints of endobronchial combustion during therapeutic laser procedures with the bronchoscope.

FDA root cause category: Under Investigation by firm

What the company is doing

An "URGENT MEDICAL DEVICE CORRECTIVE ACTION" notification letter dated 6/8/23 was sent to customers. Actions to be taken by the company: The labelling will be updated to include specificity about laser compatibility, improved instructions regarding patient preparation, and warnings about patient injury and death resulting from incompatible laser use. Actions to be taken by the end user: 1. Inspect your inventory for the referenced devices and identify any device with the model names. Please check all areas of the hospital to determine if any of these devices remain in inventory. 2. Carefully read the content of this Medical Device Correction Action as well as the attached Addendum . The addendum provides compatible laser type. 3. Ensure all personnel are completely knowledgeable and thoroughly aware that Olympus laser compatible bronchoscopes are compatible only with Nd: YAG laser or 810 nm diode lasers. 4. Olympus requests that you acknowledge receipt of this letter through the Olympus recall portal. a. Go to https://olympusamerica.com/recall b. Enter the file (recall) number: 0427 5. If you have further distributed this product, identify your customers, forward them this notification, and appropriately document your notification process. Olympus requests that you report complaints, including any injuries associated with laser procedures with Olympus bronchoscopes, to our Technical Assistance Center (TAC) at 1-800-848-9024 (option 1). Adverse events experienced with the use of this product may also be reported to the Food and Drug Administration MedWatch Adverse Event Reporting program either online, by regular mail or by fax. Olympus regrets any inconvenience caused and fully appreciates your cooperation in this matter. Please do not hesitate to call (647) 999-3203 or email: at [email protected] for any additional information or support concerning this matter.

Recall details

Recall number
Z-2012-2023
Status
Open, Classified
Initiated by the company
Jun 8, 2023
Posted by the FDA
Jul 17, 2023
Recalling firm
Olympus Corporation of the Americas
Quantity
4 units
Product code
EOQ
FDA submissions
K023984
Distribution
US Nationwide Distribution.

Product description

Olympus BRONCHOFIBERSCOPE BF Type, Model Numbers BF-PE2, BF-TE2

Code information (models, lots, serial numbers)

All serial numbers BF-PE2 BRONCHOFIBERSCOPE OLYMPUS BF TYPE PE2 UDI: 04953170339974 BF-TE2 BRONCHOFIBERSCOPE OLYMPUS BF TYPE TE2 UDI: 04953170339998

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 92362).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page