Skip to content

BRONCHOFIBERSCOPE BF-TE2 (UDI 04953170339998)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus BF-TE2 bronchofiberscope, a flexible scope used to look into the airways, has 105 reports, no deaths and 5 injuries. Some reports describe dents, scratches, breaks, or tears, and a few describe inadequate cleaning. There are 3 recalls, all still open, over instructions for use with laser and other energy devices, a scope that can lodge in an endotracheal tube connector, and possible burning inside the airway during laser procedures.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
113
Deaths reported
0
Injuries reported
5
Recalls naming it
3

Summary

The FDA's MAUDE database contains 113 adverse event reports that list this exact device (DI 04953170339998), received between Sep 4, 2020 and Sep 29, 2026 — classified as 5 injuries, 108 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (24 reports).

3 FDA recall records name this device's identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 3 reportsJan 2024: 2 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 5 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 4 reportsSep 2024: 1 reportsOct 2024: 2 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 3 reportsSep 2025: 0 reportsOct 2025: 3 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 1 reportsFeb 2026: 4 reportsMar 2026: 6 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 8 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 11, 2025Open, Classified

    Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.

    Z-0064-2026 · Root cause: Device Design · FDA record ↗

  2. Jul 27, 2023Open, Classified

    Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.

    Z-2490-2023 · Root cause: Labeling Change Control · FDA record ↗

  3. Jun 8, 2023Open, Classified

    There have been complaints of endobronchial combustion during therapeutic laser procedures with the bronchoscope.

    Z-2012-2023 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Material Deformation24
  • Failure to Clean Adequately14
  • Break13
  • Scratched Material13
  • Material Split, Cut or Torn11
  • Corroded9
  • Peeled/Delaminated8
  • Adverse Event Without Identified Device or Use Problem5

Patient problems

  • No Clinical Signs, Symptoms or Conditions105
  • No Consequences Or Impact To Patient3
  • Pneumothorax3
  • Bronchospasm1
  • Hemorrhage/Blood Loss/Bleeding1
  • Hypoxia1

Recent adverse event reports

The 25 most recent of 113 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Material Deformation; Failure to Clean Adequately

    The customer returned the Bronchofiberscope to the service center for a separate issue. During the device analysis, the Olympus Service Repair Technician indicated that the Connecting Tube was sticky and the Forceps Channel Port was shaved. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50932277 ↗
  • MalfunctionSep 28, 2026Corroded

    The customer returned the Bronchofibervideoscope for a separate issue. During the initial evaluation of the returned device, a corroded forceps channel port was observed. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50917550 ↗
  • MalfunctionSep 25, 2026Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50903034 ↗
  • MalfunctionSep 22, 2026Corroded

    The customer returned the bronchofiberscope to the service center for a separate issue. During the device analysis, it was indicated that corrosion on the light guide cover glass unit was found. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50856744 ↗
  • MalfunctionSep 18, 2026Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50828727 ↗
  • MalfunctionSep 11, 2026Material Deformation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50734740 ↗
  • MalfunctionSep 8, 2026Scratched Material

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50672702 ↗
  • MalfunctionSep 2, 2026Corroded

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50573100 ↗
  • MalfunctionMay 6, 2026Material Deformation

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE BRONCHOFIBERSCOPE FORCEPS CHANNEL PORT HAD A DENT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25103871 ↗
  • MalfunctionApr 1, 2026Contamination; Dent in Material; Failure to Clean Adequately

    ADDITIONAL FINDINGS FROM THE OLYMPUS SERVICE CENTER: CONNECTING TUBE HAS DIRT DEPOSITION ON IT AND THE INTERNAL FORCEPS CHANNEL PORT HAS A DENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24753591 ↗
  • MalfunctionMar 25, 2026Corroded

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE BRONCHOFIBERSCOPE CHANNEL PORT CORRODED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24690567 ↗
  • MalfunctionMar 25, 2026Material Split, Cut or Torn; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE EVALUATION, THE BRONCHOVIDEOSCOPE EXHIBITED THAT THE GRIP, THE UNIVERSAL CORD, THE LIGHT GUIDE ARE ALL STICKY AND THE FORCEPS CHANNEL PORT IS SHAVED. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24688284 ↗
  • MalfunctionMar 25, 2026Dent in Material; Material Deformation

    IT WAS OBSERVED THAT DURING THE EVALUATION, THE BRONCHOFIBERSCOPE EXHIBITED THAT THE FORCEPS CHANNEL PORT HAS A DENT. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24688000 ↗
  • MalfunctionMar 17, 2026Device Reprocessing Problem; Corroded; Degraded

    DURING THE DEVICE EVALUATION, IT WAS FOUND THAT THE RETURNED BRONCHOFIBERSCOPE EXHIBITED FORCEPS CHANNEL PORT STICKY AND SHAVED, AND THE CONTROL UNIT DISPLAYED CORROSION DUE TO WATER LEAKAGE. NO PATIENTS WERE INVOLVED IN ASSOCIATION WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24616634 ↗
  • MalfunctionMar 12, 2026Corroded; Material Deformation

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE BRONCHOFIBERSCOPE FORCEPS CHANNEL PORT WAS SHAVED, LIGHT GUIDE COVER GLASS HAD CORROSION, AND THE ELECTRICAL CONTACT OF THE EYEPIECE HAD CORROSION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24578813 ↗
  • MalfunctionMar 11, 2026Material Deformation

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE BRONCHOFIBERSCOPE HAD A DENT ON THE FORCEPS CHANNEL PORT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24575360 ↗
  • MalfunctionFeb 27, 2026Corroded; Degraded

    DURING THE DEVICE EVALUATION, THE BRONCHOFIBERSCOPE WAS FOUND TO HAVE CORROSION ON THE SUCTION CYLINDER AND THE BIOPSY PORT WAS SHAVED OFF. NO PATIENTS WERE INVOLVED IN THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24474872 ↗
  • MalfunctionFeb 25, 2026Corroded

    DURING THE DEVICE EVALUATION, THE VIDEOSCOPE WAS FOUND TO HAVE CORROSION ON THE CONTROL UNIT DUE TO WATER LEAKAGE. NO PATIENTS WERE INVOLVED IN THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24442480 ↗
  • MalfunctionFeb 17, 2026Detachment of Device or Device Component; Material Deformation

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE BRONCHOFIBERSCOPE HAD FORCEPS CHANNEL PORT SHAVED OFF. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24370054 ↗
  • MalfunctionFeb 13, 2026Material Deformation

    IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE BRONCHOFIBERSCOPE EXHIBITED THE FORCEPS CHANNEL PORT HAD A DENT. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24338346 ↗
  • MalfunctionJan 30, 2026Degraded

    DURING THE DEVICE EVALUATION, IT WAS FOUND THAT THE FORCEPS CHANNEL PORT ON THE BRONCHOFIBERSCOPE WAS SHAVED. NO PATIENTS WERE INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24227188 ↗
  • MalfunctionDec 17, 2025Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE EVALUATION THAT BRONCHOFIBERVIDEOSCOPE FORCEPS CHANNEL PORT WAS STICKY. THERE WAS NO REPORT OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23835299 ↗
  • MalfunctionDec 5, 2025Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE EVALUATION, THE BRONCHOVIDEOSCOPE EXHIBITED THAT THE FORCEPS CHANNEL PORT IS STICKY. THERE WAS NO REPORT OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23729081 ↗
  • MalfunctionOct 27, 2025Peeled/Delaminated; Detachment of Device or Device Component; Peeled/Delaminated; Detachment of Device or Device Component

    DURING THE DEVICE EVALUATION, IT WAS FOUND THAT THE ADHESIVE ON THE BENDING SECTION COVER IS DETACHED AND THE CONNECTING TUBE HAS PEELING COATING. (1 MM 2 OR MORE) ON THE BRONCHOFIBERSCOPE. THERE WAS NO REPORT OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23398027 ↗
  • MalfunctionOct 15, 2025Scratched Material; Material Split, Cut or Torn; Scratched Material; Material Split, Cut or Torn

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE BRONCHOFIBERSCOPE EXHIBITED THE FORCEPS CHANNEL PORT WAS SHAVED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23285283 ↗

Other BRONCHOSCOPE (FLEXIBLE OR RIGID) devices

All BRONCHOSCOPE (FLEXIBLE OR RIGID) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.