Evera S VR recallZ-1799-2023 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · initiated May 10, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
FDA root cause category: Under Investigation by firm
What the company is doing
Medtronic issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to its consignees on 05/10/2023 by UPS 2-day delivery or email. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. The notice explained the problem and the risk and requested the following patient management recommendations: - Prophylactic device replacement is NOT recommended. - Program all HV therapy pathways B>AX in all therapy zones to minimize the risk for this issue. - Prioritize reprogramming patients who have both a history of HV therapy and Rx1 programmed AX>B. - For remaining patients with AX>B programming in any HV therapy sequence, schedule (with appropriate discretion) the next follow-up for in-clinic reprogramming to minimize potential for reduced- or no-energy HV therapies to occur. - Contact Medtronic Technical Services (1-800-929-4043) or your local representative if one of the following is observed as these may be an indication of either a device or lead-related issue: o Reduced- or no-energy HV therapy is displayed in Episode Text (regardless of programmed pathway) o A persistent drop of approximately 50% in RA, RV and LV pacing lead impedance measurements as this may be an indication of increased potential for a future reduced- or no-energy therapy. Update 10/10/2023: Beginning 09/27/2023, Medtronic issued communication via letter to consignees regarding the availability of the revised IFU and SmartSync Software. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. Consignees are asked to complete and return a confirmation certificate (or equivalent record) confirming they have received the notification.
Recall details
- Recall number
- Z-1799-2023
- Status
- Open, Classified
- Initiated by the company
- May 10, 2023
- Posted by the FDA
- Jun 16, 2023
- Recalling firm
- Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
- Quantity
- 6821 units
- Product code
- LWS
- FDA submissions
- P980016S536
- Distribution
- US Nationwide - Worldwide Distribution
Product description
ICD-VR DVBC3D1 EVERA S IS1/DF1 GLOB, Model Number DVBC3D1; Implantable Cardioverter Defibrillators
Code information (models, lots, serial numbers)
Show the full code information (82k characters)
GTIN 00643169720473, Lot Serial Numbers: BWM204909H, BWM204998H, BWM205011H, BWM205017H, BWM204963H, BWM204962H, BWM205000H, BWM204927H, BWM204911H, BWM204921H, BWM204926H, BWM204932H, BWM204936H, BWM204944H, BWM204954H, BWM204955H, BWM204959H, BWM204983H, BWM204984H, BWM204989H, BWM204990H, BWM205004H, BWM205006H, BWM205008H, BWM205013H, BWM205014H, BWM205015H, BWM205016H, BWM205019H, BWM205021H, BWM205023H, BWM205024H, BWM205027H, BWM205033H, BWM205036H, BWM205037H, BWM204792H, BWM204938H, BWM204961H, BWM204916H, BWM204922H, BWM204950H, BWM204957H, BWM204993H, BWM205009H, BWM204923H, BWM204924H, BWM204987H, BWM205028H, BWM204956H, BWM204953H, BWM204789H, BWM204790H, BWM204935H, BWM204905H, BWM204906H, BWM204960H, BWM204980H, BWM204991H, BWM204914H, BWM204933H, BWM204985H, BWM204997H, BWM204948H, BWM204971H, BWM204992H, BWM205007H, BWM205034H, BWM204793H, BWM204904H, BWM204910H, BWM204970H, BWM204977H, BWM204982H, BWM204928H, BWM204930H, BWM205002H, BWM205035H, BWM204967H, BWM204974H, BWM204966H, BWM204978H, BWM204946H; GTIN 00643169720480, Lot Serial Numbers: BWM606690S, BWM606723S, BWM608186S, BWM607102S, BWM606494S, BWM607083S, BWM606927S, BWM607005S, BWM607272S, BWM606644S, BWM606971S, BWM607093S, BWM606920S, BWM606879S, BWM606579S, BWM606943S, BWM607045S, BWM607027S, BWM606500S, BWM606861S, BWM606975S, BWM607158S, BWM606473S, BWM608141S, BWM606616S, BWM606706S, BWM606946S, BWM607013S, BWM606719S, BWM606720S, BWM607007S, BWM607154S, BWM607170S, BWM607176S, BWM608142S, BWM608146S, BWM608951S, BWM606685S, BWM606728S, BWM606911S, BWM606921S, BWM606930S, BWM606931S, BWM606932S, BWM606986S, BWM607014S, BWM607407S, BWM606903S, BWM607018S, BWM606681S, BWM606907S, BWM607024S, BWM606178S, BWM606475S, BWM606478S, BWM606852S, BWM606853S, BWM607114S, BWM607119S, BWM606686S, BWM606963S, BWM607173S, BWM607175S, BWM607264S, BWM606863S, BWM606913S, BWM606914S, BWM606915S, BWM606916S, BWM606934S, BWM607094S, BWM606864S, BWM606987S, BWM607156S, BWM607019S, BWM606929S, BWM606627S, BWM606721S, BWM606925S, BWM606935S, BWM607110S, BWM607406S, BWM606933S, BWM606973S, BWM606684S, BWM607369S, BWM606739S, BWM607120S, BWM606503S, BWM606961S, BWM607085S, BWM607086S, BWM607087S, BWM607104S, BWM607090S, BWM607117S, BWM606641S, BWM606840S, BWM606512S, BWM607255S, BWM606623S, BWM606850S, BWM606988S, BWM607012S, BWM608139S, BWM608909S, BWM607022S, BWM606729S, BWM606510S, BWM608140S, BWM606505S, BWM606504S, BWM606738S, BWM607025S, BWM607354S, BWM606948S, BWM606951S, BWM606737S, BWM607163S, BWM608829S, BWM609061S, BWM607257S, BWM606528S, BWM606904S, BWM607118S, BWM608911S, BWM607164S, BWM606631S, BWM607084S, BWM606926S, BWM606949S, BWM606950S, BWM606955S, BWM607105S, BWM607017S, BWM607004S, BWM606629S, BWM607092S, BWM607115S, BWM606680S, BWM606717S, BWM606466S, BWM606493S, BWM606716S, BWM606989S, BWM607044S, BWM606824S, BWM606836S, BWM606838S, BWM606909S, BWM606917S, BWM607020S, BWM606643S, BWM607026S, BWM607404S, BWM606724S, BWM606702S, BWM607917S, BWM606994S, BWM606648S, BWM606908S, BWM606839S, BWM606906S, BWM607157S, BWM608818S, BWM609152S, BWM606462S, BWM606507S, BWM606508S, BWM606633S, BWM606634S, BWM606635S, BWM606636S, BWM606637S, BWM606638S, BWM606639S, BWM606780S, BWM606781S, BWM606782S, BWM606783S, BWM606788S, BWM606792S, BWM606794S, BWM606795S, BWM606800S, BWM606810S, BWM606812S, BWM606813S, BWM606816S, BWM606817S, BWM606900S, BWM606969S, BWM606970S, BWM606972S, BWM606974S, BWM606976S, BWM606977S, BWM606688S, BWM606689S, BWM606962S, BWM606912S, BWM606461S, BWM606626S, BWM607155S, BWM607166S, BWM606924S, BWM607379S, BWM606947S, BWM607382S, BWM607171S, BWM606600S, BWM606910S, BWM607139S, BWM606646S, BWM607003S, BWM606727S, BWM607052S, BWM606658S, BWM606659S, BWM606662S, BWM606665S, BWM606666S, BWM606669S, BWM606670S, BWM606671S, BWM606672S, BWM606673S, BWM606881S, BWM606883S, BWM606884S, BWM606885S, BWM606886S, BWM606887S, BWM606888S, BWM606889S, BWM606890S, BWM606923S, BWM607371S, BWM607378S, BWM607381S, BWM607383S, BWM607386S, BWM607387S, BWM607389S, BWM607390S, BWM607392S, BWM607393S, BWM606919S, BWM606621S, BWM606657S, BWM606683S, BWM606708S, BWM606713S, BWM606859S, BWM606877S, BWM606958S, BWM607088S, BWM607089S, BWM607420S, BWM607438S, BWM607535S, BWM607856S, BWM607903S, BWM608110S, BWM608143S, BWM608152S, BWM608253S, BWM608255S, BWM608323S, BWM608487S, BWM606640S, BWM606660S, BWM606622S, BWM606624S, BWM606628S, BWM606661S, BWM606667S, BWM606714S, BWM606770S, BWM606799S, BWM606821S, BWM606866S, BWM606870S, BWM606918S, BWM607099S, BWM607112S, BWM607124S, BWM607159S, BWM607161S, BWM607162S, BWM607384S, BWM607415S, BWM607423S, BWM607445S, BWM607450S, BWM607497S, BWM607539S, BWM607725S, BWM607842S, BWM607845S, BWM607848S, BWM607853S, BWM607858S, BWM607875S, BWM607906S, BWM607953S, BWM608148S, BWM608242S, BWM608248S, BWM608252S, BWM608259S, BWM608271S, BWM608274S, BWM608286S, BWM608320S, BWM608326S, BWM608352S, BWM608353S, BWM608357S, BWM608358S, BWM608360S, BWM608361S, BWM608362S, BWM608498S, BWM608514S, BWM608520S, BWM606725S, BWM606746S, BWM607103S, BWM607376S, BWM607442S, BWM607537S, BWM607860S, BWM607861S, BWM607966S, BWM608098S, BWM608102S, BWM608106S, BWM608153S, BWM608250S, BWM608257S, BWM608309S, BWM608311S, BWM608317S, BWM608453S, BWM608524S, BWM607167S, BWM606695S, BWM606945S, BWM606632S, BWM606518S, BWM607262S, BWM606764S, BWM606768S, BWM606777S, BWM606957S, BWM606960S, BWM606965S, BWM606983S, BWM607253S, BWM607261S, BWM607263S, BWM607265S, BWM607267S, BWM607268S, BWM607270S, BWM607515S, BWM606773S, BWM606847S, BWM606953S, BWM606954S, BWM606984S, BWM606574S, BWM606575S, BWM606576S, BWM606578S, BWM606589S, BWM606593S, BWM606594S, BWM606595S, BWM606956S, BWM606964S, BWM606517S, BWM606588S, BWM606472S, BWM606474S, BWM606476S, BWM606482S, BWM606485S, BWM606541S, BWM606542S, BWM606553S, BWM606554S, BWM606556S, BWM606557S, BWM606558S, BWM606559S, BWM606560S, BWM606561S, BWM606562S, BWM606563S, BWM606564S, BWM606565S, BWM606581S, BWM606587S, BWM606590S, BWM606591S, BWM606599S, BWM606825S, BWM606826S, BWM606829S, BWM606830S, BWM606831S, BWM606833S, BWM606854S, BWM606868S, BWM606869S, BWM607049S, BWM607051S, BWM607056S, BWM607058S, BWM607060S, BWM607061S, BWM607065S, BWM607069S, BWM607073S, BWM607307S, BWM607308S, BWM607309S, BWM607310S, BWM607311S, BWM607312S, BWM607345S, BWM607351S, BWM607352S, BWM607355S, BWM607365S, BWM607416S, BWM607419S, BWM607451S, BWM607452S, BWM607453S, BWM607462S, BWM607463S, BWM607464S, BWM607466S, BWM607467S, BWM607468S, BWM607469S, BWM607470S, BWM607471S, BWM607472S, BWM607603S, BWM607604S, BWM607605S, BWM607687S, BWM607688S, BWM607689S, BWM607760S, BWM607761S, BWM607766S, BWM607767S, BWM607768S, BWM607769S, BWM607770S, BWM607777S, BWM607778S, BWM607851S, BWM607852S, BWM607855S, BWM607901S, BWM607908S, BWM607916S, BWM607939S, BWM607940S, BWM607941S, BWM607942S, BWM607943S, BWM607968S, BWM608030S, BWM608031S, BWM608032S, BWM608033S, BWM608062S, BWM608145S, BWM608147S, BWM608183S, BWM608184S, BWM608237S, BWM608238S, BWM608240S, BWM608243S, BWM608258S, BWM608260S, BWM608261S, BWM608263S, BWM608264S, BWM608314S, BWM608316S, BWM608318S, BWM608407S, BWM608408S, BWM608409S, BWM608410S, BWM608411S, BWM608435S, BWM608436S, BWM608437S, BWM608438S, BWM608439S, BWM608569S, BWM608570S, BWM608647S, BWM608648S, BWM608649S, BWM608650S, BWM608651S, BWM608652S, BWM608699S, BWM608700S, BWM608701S, BWM608702S, BWM608703S, BWM608777S, BWM608778S, BWM608779S, BWM608814S, BWM608815S, BWM608819S, BWM608820S, BWM608821S, BWM608833S, BWM608834S, BWM608835S, BWM608836S, BWM608837S, BWM608857S, BWM608859S, BWM608875S, BWM608876S, BWM608877S, BWM608893S, BWM608908S, BWM608919S, BWM608920S, BWM608921S, BWM608926S, BWM608927S, BWM608938S, BWM609044S, BWM609071S, BWM609072S, BWM609073S, BWM609076S, BWM609086S, BWM609103S, BWM609151S, BWM609171S, BWM609172S, BWM606463S, BWM606703S, BWM607047S, BWM607048S, BWM607260S, BWM607269S, BWM607303S, BWM607320S, BWM607321S, BWM607322S, BWM607346S, BWM607348S, BWM607349S, BWM607350S, BWM607434S, BWM607436S, BWM607437S, BWM607439S, BWM607440S, BWM607446S, BWM607454S, BWM607455S, BWM607460S, BWM607478S, BWM607480S, BWM607536S, BWM607540S, BWM607541S, BWM607587S, BWM607588S, BWM607589S, BWM607590S, BWM607593S, BWM607594S, BWM607595S, BWM607596S, BWM607598S, BWM607600S, BWM607602S, BWM607611S, BWM607666S, BWM607712S, BWM607713S, BWM607714S, BWM607716S, BWM607717S, BWM607718S, BWM607719S, BWM607721S, BWM607722S, BWM607723S, BWM607724S, BWM607771S, BWM607772S, BWM607773S, BWM607774S, BWM607775S, BWM607844S, BWM607867S, BWM607870S, BWM607871S, BWM607885S, BWM607890S, BWM607894S, BWM607898S, BWM607930S, BWM608025S, BWM608026S, BWM608027S, BWM608028S, BWM608057S, BWM608058S, BWM608059S, BWM608060S, BWM608075S, BWM608076S, BWM608097S, BWM608099S, BWM608100S, BWM608101S, BWM608103S, BWM608104S, BWM608137S, BWM608185S, BWM608187S, BWM608188S, BWM608190S, BWM608191S, BWM608202S, BWM608204S, BWM608205S, BWM608207S, BWM608208S, BWM608232S, BWM608235S, BWM608245S, BWM608246S, BWM608247S, BWM608249S, BWM608251S, BWM608254S, BWM608256S, BWM608265S, BWM608266S, BWM608270S, BWM608272S, BWM608273S, BWM608275S, BWM608276S, BWM608277S, BWM608278S, BWM608279S, BWM608441S, BWM608450S, BWM608451S, BWM608452S, BWM608454S, BWM608455S, BWM608456S, BWM608457S, BWM608458S, BWM608459S, BWM608460S, BWM608461S, BWM608462S, BWM608463S, BWM608465S, BWM608466S, BWM608467S, BWM608468S, BWM608510S, BWM608511S, BWM608512S, BWM608513S, BWM608515S, BWM608517S, BWM608521S, BWM608522S, BWM608523S, BWM608525S, BWM608561S, BWM608562S, BWM608704S, BWM608705S, BWM608706S, BWM608707S, BWM608708S, BWM608709S, BWM608710S, BWM608711S, BWM608712S, BWM608713S, BWM608740S, BWM608741S, BWM608742S, BWM608743S, BWM608745S, BWM608748S, BWM608750S, BWM608751S, BWM608753S, BWM608754S, BWM608791S, BWM608794S, BWM608796S, BWM608839S, BWM608840S, BWM608841S, BWM608842S, BWM608843S, BWM608844S, BWM608845S, BWM608846S, BWM608847S, BWM608848S, BWM608849S, BWM606468S, BWM606523S, BWM606180S, BWM606465S, BWM606524S, BWM606525S, BWM606527S, BWM606617S, BWM606692S, BWM606693S, BWM606700S, BWM606522S, BWM606469S, BWM607913S, BWM606705S, BWM606625S, BWM606905S, BWM607023S, BWM606577S, BWM606580S, BWM606760S, BWM606531S, BWM606601S, BWM606602S, BWM606603S, BWM606604S, BWM606605S, BWM606675S, BWM606678S, BWM606679S, BWM606968S, BWM607266S, BWM606179S, BWM606694S, BWM606841S, BWM606842S, BWM607410S, BWM607414S, BWM607417S, BWM607418S, BWM606928S, BWM606959S, BWM606997S, BWM607006S, BWM607008S, BWM607096S, BWM607107S, BWM606515S, BWM606536S, BWM606538S, BWM607010S, BWM606514S, BWM606819S, BWM606526S, BWM606530S, BWM606731S, BWM606743S, BWM606745S, BWM607062S, BWM607256S, BWM607691S, BWM607780S, BWM608109S, BWM608111S, BWM608267S, BWM608737S, BWM608738S, BWM608767S, BWM608768S, BWM608769S, BWM606966S, BWM606572S, BWM606573S, BWM606828S, BWM606875S, BWM606876S, BWM607405S, BWM607409S, BWM607411S, BWM607473S, BWM607474S, BWM607476S, BWM607477S, BWM607583S, BWM607584S, BWM607586S, BWM607726S, BWM607727S, BWM607729S, BWM607730S, BWM607731S, BWM607757S, BWM607758S, BWM607762S, BWM607763S, BWM607764S, BWM607765S, BWM607846S, BWM607849S, BWM607850S, BWM607857S, BWM607862S, BWM607904S, BWM607905S, BWM607907S, BWM607909S, BWM607910S, BWM607911S, BWM607912S, BWM607932S, BWM607934S, BWM607937S, BWM607967S, BWM607969S, BWM608034S, BWM608035S, BWM608036S, BWM608105S, BWM608107S, BWM608108S, BWM608179S, BWM608180S, BWM608182S, BWM608209S, BWM608210S, BWM608211S, BWM608212S, BWM608213S, BWM608214S, BWM608280S, BWM608281S, BWM608282S, BWM608283S, BWM608284S, BWM608288S, BWM608394S, BWM608415S, BWM608417S, BWM608418S, BWM608419S, BWM608420S, BWM608421S, BWM608422S, BWM608423S, BWM608433S, BWM608434S, BWM608440S, BWM608442S, BWM608443S, BWM608444S, BWM608445S, BWM608448S, BWM608449S, BWM608469S, BWM608478S, BWM608546S, BWM608547S, BWM608548S, BWM608549S, BWM608550S, BWM608556S, BWM608558S, BWM608559S, BWM608735S, BWM608736S, BWM608756S, BWM608757S, BWM608795S, BWM608797S, BWM608798S, BWM608800S, BWM608801S, BWM608802S, BWM608803S, BWM608804S, BWM608805S, BWM608806S, BWM608807S, BWM608854S, BWM608855S, BWM608894S, BWM608897S, BWM606697S, BWM606699S, BWM606642S, BWM607095S, BWM607306S, BWM608268S, BWM608269S, BWM606592S, BWM606596S, BWM606597S, BWM606598S, BWM607178S, BWM607179S, BWM607180S, BWM607181S, BWM607193S, BWM607091S, BWM607508S, BWM607498S, BWM607502S, BWM606520S, BWM606663S, BWM606664S, BWM606718S, BWM606747S, BWM606820S, BWM606832S, BWM606851S, BWM606860S, BWM606874S, BWM607101S, BWM607109S, BWM607116S, BWM607126S, BWM607160S, BWM607364S, BWM607391S, BWM607413S, BWM607457S, BWM607538S, BWM607690S, BWM607715S, BWM607720S, BWM607728S, BWM607841S, BWM607843S, BWM607874S, BWM607893S, BWM607896S, BWM607900S, BWM607915S, BWM607950S, BWM608239S, BWM608241S, BWM608262S, BWM608312S, BWM608315S, BWM608327S, BWM608328S, BWM608349S, BWM608350S, BWM608351S, BWM608354S, BWM608365S, BWM608464S, BWM608499S, BWM608500S, BWM608503S, BWM608516S, BWM608519S, BWM606668S, BWM606674S, BWM606827S, BWM606880S, BWM606882S, BWM607071S, BWM607318S, BWM607319S, BWM607465S, BWM607475S, BWM607753S, BWM607755S, BWM607756S, BWM607781S, BWM607782S, BWM607783S, BWM608073S, BWM608162S, BWM608181S, BWM608189S, BWM608233S, BWM608234S, BWM608355S, BWM608405S, BWM608518S, BWM608568S, BWM608936S, BWM608937S, BWM608952S, BWM609143S, BWM609144S, BWM609145S, BWM609158S, BWM609159S, BWM609169S, BWM607505S, BWM606467S, BWM606803S, BWM606804S, BWM606805S, BWM606823S, BWM607304S, BWM607305S, BWM607357S, BWM607403S, BWM607412S, BWM607447S, BWM607448S, BWM607864S, BWM607872S, BWM607873S, BWM607876S, BWM607883S, BWM607914S, BWM607927S, BWM607963S, BWM607965S, BWM607970S, BWM608024S, BWM608029S, BWM608061S, BWM608074S, BWM608144S, BWM608163S, BWM608200S, BWM608201S, BWM608244S, BWM608310S, BWM608313S, BWM608363S, BWM608364S, BWM608366S, BWM608367S, BWM608389S, BWM608390S, BWM608391S, BWM608392S, BWM608393S, BWM608406S, BWM608424S, BWM608425S, BWM608426S, BWM608427S, BWM608428S, BWM608446S, BWM608472S, BWM608473S, BWM608474S, BWM608475S, BWM608476S, BWM608477S, BWM608480S, BWM608481S, BWM608563S, BWM608564S, BWM608565S, BWM608567S, BWM608646S, BWM608669S, BWM608670S, BWM608694S, BWM608770S, BWM608775S, BWM608776S, BWM608781S, BWM608782S, BWM608799S, BWM608812S, BWM608813S, BWM608838S, BWM608850S, BWM608851S, BWM608852S, BWM608853S, BWM608856S, BWM608874S, BWM608890S, BWM608891S, BWM608892S, BWM608895S, BWM608896S, BWM608899S, BWM608900S, BWM608902S, BWM608903S, BWM608928S, BWM608929S, BWM608930S, BWM608932S, BWM608933S, BWM609087S, BWM606776S, BWM607080S, BWM607501S, BWM606543S, BWM606802S, BWM606993S, BWM607070S, BWM607962S, BWM608231S, BWM608285S, BWM608319S, BWM608322S, BWM608479S, BWM609074S, BWM609162S, BWM606691S, BWM606871S, BWM606872S, BWM606873S, BWM607499S, BWM607511S, BWM607513S, BWM607504S, BWM606499S, BWM606501S, BWM606529S, BWM606484S, BWM606487S, BWM606618S, BWM606619S, BWM606620S, BWM606630S, BWM606711S, BWM607514S, BWM606477S, BWM606479S, BWM606480S, BWM606481S, BWM606483S, BWM606486S, BWM606488S, BWM606489S, BWM606496S, BWM606497S, BWM606498S, BWM606534S, BWM606535S, BWM606537S, BWM606539S, BWM606540S, BWM606544S, BWM606545S, BWM606546S, BWM606547S, BWM606548S, BWM606549S, BWM606550S, BWM606551S, BWM606552S, BWM606566S, BWM606567S, BWM606568S, BWM606582S, BWM606583S, BWM606584S, BWM606585S, BWM606586S, BWM606650S, BWM606651S, BWM606676S, BWM606677S, BWM606806S, BWM606807S, BWM606811S, BWM606835S, BWM606837S, BWM606843S, BWM606844S, BWM606845S, BWM606846S, BWM606848S, BWM606849S, BWM606862S, BWM606865S, BWM606867S, BWM606995S, BWM606999S, BWM607001S, BWM607002S, BWM607009S, BWM607016S, BWM607021S, BWM607031S, BWM607035S, BWM607038S, BWM607040S, BWM607041S, BWM607097S, BWM607098S, BWM607100S, BWM607254S, BWM607356S, BWM607360S, BWM607366S, BWM607479S, BWM607516S, BWM607517S, BWM607582S, BWM607682S, BWM606834S, BWM607343S, BWM607426S, BWM607432S, BWM607533S, BWM607534S, BWM607847S, BWM608116S, BWM608149S, BWM608287S, BWM608321S, BWM608368S, BWM608905S, BWM608906S, BWM609078S, BWM609118S, BWM609148S, BWM607500S, BWM606464S, BWM606893S, BWM606894S, BWM606941S, BWM607036S, BWM607053S, BWM607072S, BWM607121S, BWM607122S, BWM607127S, BWM607143S, BWM607347S, BWM607585S, BWM607591S, BWM607791S, BWM607854S, BWM607923S, BWM607113S, BWM606757S, BWM606765S, BWM606766S, BWM606767S, BWM606771S, BWM606772S, BWM606774S, BWM606936S, BWM606937S, BWM606938S, BWM606998S, BWM607000S, BWM607028S, BWM607067S, BWM607068S, BWM607123S, BWM607125S, BWM607149S, BWM607150S, BWM607151S, BWM607153S, BWM607259S, BWM607435S, BWM607441S, BWM607443S, BWM607444S, BWM607449S, BWM607456S, BWM607458S, BWM607459S, BWM607461S, BWM607759S, BWM607878S, BWM607889S, BWM607892S, BWM607895S, BWM607897S, BWM606611S, BWM606612S, BWM606613S, BWM606614S, BWM606615S, BWM606990S, BWM607029S, BWM607054S, BWM607055S, BWM607592S, BWM607597S, BWM607865S, BWM607931S, BWM607933S, BWM606715S, BWM606726S, BWM607147S, BWM607396S, BWM607397S, BWM607790S, BWM607792S, BWM607794S, BWM607879S, BWM607880S, BWM607935S, BWM607936S, BWM606655S, BWM606656S, BWM606707S, BWM606902S, BWM607037S, BWM607074S, BWM607145S, BWM607146S, BWM607148S, BWM607372S, BWM607380S, BWM607599S, BWM607601S, BWM607886S, BWM607887S, BWM607888S, BWM607921S, BWM607922S, BWM607924S, BWM607427S, BWM606734S, BWM606735S, BWM606736S, BWM606740S, BWM606741S, BWM606742S, BWM606744S, BWM606748S, BWM606749S, BWM606814S, BWM606815S, BWM606818S, BWM606822S, BWM606899S, BWM606901S, BWM607030S, BWM607032S, BWM607033S, BWM607078S, BWM607108S, BWM607135S, BWM607136S, BWM607137S, BWM607138S, BWM607140S, BWM607141S, BWM607142S, BWM607144S, BWM607402S, BWM607408S, BWM607422S, BWM607425S, BWM607899S, BWM607902S, BWM607928S, BWM607929S, BWM606606S, BWM606607S, BWM606608S, BWM606609S, BWM606610S, BWM606649S, BWM606652S, BWM606653S, BWM606654S, BWM606709S, BWM606710S, BWM606712S, BWM606730S, BWM606732S, BWM606733S, BWM606785S, BWM606786S, BWM606787S, BWM606797S, BWM606798S, BWM606808S, BWM606809S, BWM606895S, BWM606896S, BWM606897S, BWM606898S, BWM606940S, BWM606942S, BWM606980S, BWM606982S, BWM606985S, BWM607039S, BWM607042S, BWM607043S, BWM607046S, BWM607050S, BWM607059S, BWM607063S, BWM607064S, BWM607066S, BWM607079S, BWM607128S, BWM607129S, BWM607130S, BWM607131S, BWM607132S, BWM607133S, BWM607134S, BWM607394S, BWM607395S, BWM607398S, BWM607399S, BWM607400S, BWM607401S, BWM607752S, BWM607776S, BWM607859S, BWM607863S, BWM607866S, BWM607868S, BWM607869S, BWM607881S, BWM607882S, BWM607884S, BWM607919S, BWM607920S, BWM607925S, BWM607926S, BWM606891S, BWM606892S, BWM607082S, BWM607034S, BWM607429S, BWM607433S, BWM606939S, BWM607057S, BWM607152S, BWM607877S, BWM607779S, BWM606495S, BWM606502S, BWM606506S, BWM606511S, BWM606519S, BWM606521S, BWM606532S, BWM606533S, BWM606555S, BWM606569S, BWM606570S, BWM606571S, BWM606645S, BWM606647S, BWM606791S, BWM606793S, BWM606796S, BWM606801S, BWM606855S, BWM606856S, BWM606857S, BWM606878S, BWM606944S, BWM606952S, BWM606978S, BWM606981S, BWM606991S, BWM606992S, BWM607075S, BWM607076S, BWM607077S, BWM607081S, BWM607258S, BWM607344S, BWM607353S, BWM607358S, BWM607359S, BWM607361S, BWM607362S, BWM607363S, BWM607367S, BWM607368S, BWM607370S, BWM607373S, BWM607374S, BWM607375S, BWM607377S, BWM607385S, BWM607388S, BWM607421S, BWM607424S, BWM607428S, BWM607430S, BWM607431S, BWM607681S, BWM607784S, BWM607785S, BWM607786S, BWM607787S, BWM607788S, BWM607789S, BWM607793S, BWM607795S, BWM607796S, BWM607797S, BWM607798S, BWM607799S, BWM608324S, BWM608325S, BWM608356S, BWM608359S, BWM608412S, BWM608413S, BWM608414S, BWM608431S, BWM608432S, BWM608470S, BWM608497S, BWM608560S, BWM608773S, BWM608774S, BWM608780S, BWM608816S, BWM608817S, BWM607111S, BWM606753S, BWM606754S, BWM606755S, BWM606756S, BWM606758S, BWM606759S, BWM606761S, BWM606762S, BWM606763S, BWM606769S, BWM606775S, BWM606779S, BWM606784S, BWM606789S, BWM606790S, BWM607313S, BWM607314S, BWM607315S, BWM607316S, BWM607317S, BWM607481S, BWM607606S, BWM607607S, BWM607608S, BWM607609S, BWM607610S, BWM607652S, BWM607653S, BWM607654S, BWM607655S, BWM607656S, BWM607657S, BWM607658S, BWM607659S, BWM607660S, BWM607661S, BWM607662S, BWM607663S, BWM607664S, BWM607665S, BWM607667S, BWM607668S, BWM607669S, BWM607670S, BWM607671S, BWM607672S, BWM607673S, BWM607674S, BWM607675S, BWM607676S, BWM607677S, BWM607678S, BWM607679S, BWM607680S, BWM607683S, BWM607684S, BWM607685S, BWM607686S, BWM607800S, BWM607801S, BWM607891S, BWM607938S, BWM607944S, BWM607945S, BWM607946S, BWM607947S, BWM607948S, BWM607949S, BWM607951S, BWM607952S, BWM607954S, BWM607955S, BWM607956S, BWM607957S, BWM607958S, BWM607959S, BWM607960S, BWM607961S, BWM607964S, BWM607971S, BWM607972S, BWM607973S, BWM607974S, BWM607975S, BWM607976S, BWM607977S, BWM608017S, BWM608018S, BWM608020S, BWM608021S, BWM608022S, BWM608023S, BWM608063S, BWM608064S, BWM608065S, BWM608066S, BWM608067S, BWM608068S, BWM608069S, BWM608070S, BWM608071S, BWM608072S, BWM608112S, BWM608113S, BWM608114S, BWM608115S, BWM608192S, BWM608193S, BWM608194S, BWM608195S, BWM608196S, BWM608197S, BWM608198S, BWM608199S, BWM608215S, BWM608216S, BWM608217S, BWM608218S, BWM608219S, BWM608220S, BWM608221S, BWM608222S, BWM608223S, BWM608224S, BWM608225S, BWM608226S, BWM608227S, BWM608228S, BWM608230S, BWM608395S, BWM608396S, BWM608397S, BWM608398S, BWM608399S, BWM608400S, BWM608401S, BWM608402S, BWM608403S, BWM608404S, BWM608429S, BWM608430S, BWM608471S, BWM608482S, BWM608483S, BWM608484S, BWM608485S, BWM608486S, BWM608488S, BWM608489S, BWM608490S, BWM608491S, BWM608492S, BWM608493S, BWM608494S, BWM608495S, BWM608496S, BWM608501S, BWM608502S, BWM608504S, BWM608505S, BWM608506S, BWM608507S, BWM608508S, BWM608509S, BWM608551S, BWM608552S, BWM608553S, BWM608554S, BWM608555S, BWM608571S, BWM608572S, BWM608573S, BWM608574S, BWM608575S, BWM608653S, BWM608654S, BWM608655S, BWM608656S, BWM608657S, BWM608658S, BWM608659S, BWM608660S, BWM608661S, BWM608662S, BWM608663S, BWM608664S, BWM608665S, BWM608666S, BWM608667S, BWM608668S, BWM608671S, BWM608672S, BWM608673S, BWM608674S, BWM608675S, BWM608695S, BWM608696S, BWM608697S, BWM608698S, BWM608759S, BWM608760S, BWM608761S, BWM608763S, BWM608764S, BWM608783S, BWM608784S, BWM608785S, BWM608786S, BWM608787S, BWM608788S, BWM608789S, BWM608790S, BWM608792S, BWM608793S, BWM608822S, BWM608823S, BWM608824S, BWM608825S, BWM608826S, BWM608827S, BWM608828S, BWM608830S, BWM608831S, BWM608832S, BWM608860S, BWM608861S, BWM608862S, BWM608863S, BWM608864S, BWM608865S, BWM608866S, BWM608867S, BWM608868S, BWM608869S, BWM608888S, BWM608889S, BWM608898S, BWM608901S, BWM608904S, BWM608907S, BWM608910S, BWM608912S, BWM608913S, BWM608914S, BWM608915S, BWM608916S, BWM608917S, BWM608918S, BWM608922S, BWM608923S, BWM608924S, BWM608925S, BWM608934S, BWM608935S, BWM608941S, BWM608942S, BWM608943S, BWM608944S, BWM608945S, BWM608953S, BWM608954S, BWM609085S, BWM609088S, BWM609089S, BWM609090S, BWM609091S, BWM609092S, BWM609093S, BWM609094S, BWM609095S, BWM609096S, BWM609097S, BWM609098S, BWM609099S, BWM609100S, BWM609101S, BWM609104S, BWM609105S, BWM609106S, BWM609107S, BWM609113S, BWM609149S, BWM609150S, BWM609163S, BWM609164S, BWM609165S, BWM609166S, BWM609167S, BWM607271S, BWM608739S, BWM608744S, BWM608808S, BWM608809S, BWM608810S, BWM608811S, BWM608939S, BWM608940S, BWM609080S, BWM609081S, BWM609082S, BWM609083S, BWM609084S, BWM609108S, BWM609109S, BWM609110S, BWM609111S, BWM609112S, BWM609114S, BWM609115S, BWM609116S, BWM609117S, BWM609122S, BWM609123S, BWM609146S, BWM608416S, BWM608447S, BWM608746S, BWM608747S, BWM608749S, BWM608752S, BWM608755S, BWM608758S, BWM608762S, BWM608765S, BWM608766S, BWM608771S, BWM608772S, BWM608878S, BWM608879S, BWM608880S, BWM608881S, BWM608882S, BWM608883S, BWM608884S, BWM608885S, BWM608886S, BWM608887S, BWM608931S, BWM608946S, BWM608947S, BWM608948S, BWM609045S, BWM609046S, BWM609047S, BWM609048S, BWM609049S, BWM609050S, BWM609051S, BWM609052S, BWM609053S, BWM609054S, BWM609055S, BWM609056S, BWM609057S, BWM609059S, BWM609060S, BWM609062S, BWM609063S, BWM609064S, BWM609065S, BWM609066S, BWM609124S, BWM609125S, BWM609126S, BWM609127S, BWM609128S, BWM609129S, BWM609130S, BWM609132S, BWM609133S, BWM609134S, BWM609135S, BWM609136S, BWM609137S, BWM609138S, BWM609141S, BWM609142S, BWM609160S, BWM609161S, BWM608870S, BWM608871S, BWM608872S, BWM608873S, BWM608949S, BWM608950S, BWM609067S, BWM609068S, BWM609069S, BWM609070S, BWM609075S, BWM609077S, BWM609079S, BWM609102S, BWM609119S, BWM609120S, BWM609121S, BWM609139S, BWM609140S, BWM609147S, BWM609153S, BWM609154S, BWM609155S, BWM609156S, BWM609157S, BWM609168S, BWM609170S, BWM609173S; GTIN 00643169980266, Lot Serial Numbers: BWM612848S, BWM612852S, BWM609327S, BWM607837S, BWM608082S, BWM607702S, BWM607334S, BWM607816S, BWM607198S, BWM608330S, BWM607562S, BWM608007S, BWM608295S, BWM608383S, BWM608334S, BWM607287S, BWM607618S, BWM607629S, BWM607996S, BWM609218S, BWM607340S, BWM607342S, BWM608292S, BWM611257S, BWM608000S, BWM608373S, BWM607333S, BWM607496S, BWM609220S, BWM607523S, BWM607615S, BWM607810S, BWM609201S, BWM607805S, BWM608290S, BWM607833S, BWM611419S, BWM607199S, BWM607208S, BWM607279S, BWM607241S, BWM607555S, BWM611587S, BWM609810S, BWM607551S, BWM607732S, BWM607994S, BWM610739S, BWM607221S, BWM607617S, BWM607636S, BWM607990S, BWM608054S, BWM610647S, BWM607978S, BWM609356S, BWM607341S, BWM607206S, BWM607710S, BWM607556S, BWM607630S, BWM607581S, BWM607526S, BWM607695S, BWM607213S, BWM607645S, BWM608968S, BWM608971S, BWM608980S, BWM608981S, BWM607204S, BWM607212S, BWM607217S, BWM607245S, BWM607246S, BWM607288S, BWM607292S, BWM607488S, BWM607492S, BWM607518S, BWM607522S, BWM607530S, BWM607561S, BWM607697S, BWM607698S, BWM607809S, BWM607817S, BWM607985S, BWM607992S, BWM608291S, BWM608369S, BWM608376S, BWM608972S, BWM608982S, BWM609038S, BWM609179S, BWM609180S, BWM609200S, BWM609297S, BWM609582S, BWM609586S, BWM609621S, BWM609802S, BWM609817S, BWM609820S, BWM610596S, BWM610634S, BWM610717S, BWM610741S, BWM610959S, BWM611429S, BWM611625S, BWM607209S, BWM607580S, BWM607285S, BWM607326S, BWM607519S, BWM607570S, BWM607812S, BWM607815S, BWM610651S, BWM612730S, BWM607201S, BWM607203S, BWM607331S, BWM607332S, BWM607490S, BWM607491S, BWM607291S, BWM607493S, BWM607494S, BWM607620S, BWM607637S, BWM608160S, BWM608293S, BWM610748S, BWM608158S, BWM607196S, BWM607625S, BWM607804S, BWM607211S, BWM607579S, BWM607979S, BWM607549S, BWM607482S, BWM607613S, BWM608002S, BWM607998S, BWM607207S, BWM607210S, BWM608118S, BWM608131S, BWM608133S, BWM608370S, BWM608955S, BWM609381S, BWM608379S, BWM607619S, BWM607329S, BWM607520S, BWM607545S, BWM607575S, BWM607741S, BWM611604S, BWM609806S, BWM609807S, BWM609811S, BWM607243S, BWM608130S, BWM608132S, BWM607252S, BWM607205S, BWM610779S, BWM608958S, BWM608961S, BWM609354S, BWM607997S, BWM607483S, BWM607218S, BWM607239S, BWM607275S, BWM607542S, BWM607634S, BWM607640S, BWM607709S, BWM607983S, BWM607835S, BWM607622S, BWM607197S, BWM609389S, BWM609467S, BWM607229S, BWM607231S, BWM607633S, BWM607736S, BWM609383S, BWM607705S, BWM608123S, BWM607338S, BWM607230S, BWM607335S, BWM607336S, BWM607484S, BWM607553S, BWM607649S, BWM607693S, BWM607696S, BWM607706S, BWM607708S, BWM607739S, BWM608956S, BWM608959S, BWM608960S, BWM608964S, BWM608965S, BWM608966S, BWM608967S, BWM608979S, BWM609742S, BWM610376S, BWM610575S, BWM610686S, BWM607238S, BWM607567S, BWM607626S, BWM607984S, BWM608008S, BWM608331S, BWM608973S, BWM609184S, BWM607325S, BWM607612S, BWM610650S, BWM610656S, BWM610660S, BWM611590S, BWM607646S, BWM612764S, BWM609696S, BWM609757S, BWM610749S, BWM610857S, BWM611407S, BWM611599S, BWM612735S, BWM612737S, BWM609385S, BWM611427S, BWM612734S, BWM609695S, BWM609470S, BWM609471S, BWM609577S, BWM609809S, BWM610738S, BWM610958S, BWM611202S, BWM611433S, BWM611581S, BWM611585S, BWM611636S, BWM612728S, BWM608079S, BWM611600S, BWM611459S, BWM612769S, BWM611634S, BWM609466S, BWM607651S, BWM608092S, BWM609393S, BWM608296S, BWM609815S, BWM606751S, BWM607280S, BWM607281S, BWM607295S, BWM606752S, BWM609257S, BWM607298S, BWM607302S, BWM607995S, BWM609037S, BWM609681S, BWM607216S, BWM607301S, BWM608984S, BWM609256S, BWM609378S, BWM612760S, BWM607999S, BWM607235S, BWM611398S, BWM607543S, BWM607703S, BWM608983S, BWM609039S, BWM607814S, BWM607802S, BWM607194S, BWM607289S, BWM607489S, BWM607529S, BWM607330S, BWM607525S, BWM609196S, BWM609469S, BWM611428S, BWM608042S, BWM607300S, BWM607628S, BWM607244S, BWM607247S, BWM607297S, BWM607643S, BWM607692S, BWM607699S, BWM607711S, BWM607811S, BWM608001S, BWM608335S, BWM608382S, BWM608963S, BWM609219S, BWM609667S, BWM609814S, BWM609818S, BWM610716S, BWM608091S, BWM608004S, BWM607641S, BWM607503S, BWM607290S, BWM607734S, BWM607694S, BWM608121S, BWM607635S, BWM608122S, BWM609694S, BWM607616S, BWM608976S, BWM608977S, BWM609202S, BWM609203S, BWM607273S, BWM607222S, BWM607223S, BWM607485S, BWM607527S, BWM607569S, BWM607621S, BWM607623S, BWM607624S, BWM607819S, BWM608038S, BWM608124S, BWM608537S, BWM608540S, BWM609183S, BWM609370S, BWM609484S, BWM608969S, BWM609460S, BWM607546S, BWM607547S, BWM607548S, BWM607559S, BWM607560S, BWM607735S, BWM607827S, BWM607829S, BWM608039S, BWM608127S, BWM608128S, BWM608159S, BWM608289S, BWM608332S, BWM608333S, BWM608371S, BWM608377S, BWM608386S, BWM608676S, BWM608677S, BWM608998S, BWM608999S, BWM609177S, BWM609187S, BWM609210S, BWM609222S, BWM609362S, BWM609365S, BWM609384S, BWM609386S, BWM609388S, BWM609394S, BWM609395S, BWM609478S, BWM609480S, BWM609622S, BWM609669S, BWM609670S, BWM609680S, BWM609697S, BWM609698S, BWM609699S, BWM609745S, BWM609746S, BWM609753S, BWM609756S, BWM609766S, BWM609782S, BWM609784S, BWM609797S, BWM609798S, BWM609801S, BWM609822S, BWM609825S, BWM609835S, BWM609853S, BWM610355S, BWM610393S, BWM610662S, BWM610712S, BWM610745S, BWM610747S, BWM610784S, BWM610808S, BWM610829S, BWM610830S, BWM610910S, BWM610916S, BWM610926S, BWM610971S, BWM610974S, BWM610995S, BWM611145S, BWM611190S, BWM611204S, BWM611232S, BWM611244S, BWM611252S, BWM611306S, BWM611309S, BWM611400S, BWM611405S, BWM611421S, BWM611424S, BWM611435S, BWM611561S, BWM611564S, BWM611572S, BWM611588S, BWM611627S, BWM611629S, BWM611630S, BWM611633S, BWM612478S, BWM612568S, BWM612627S, BWM612683S, BWM612691S, BWM612726S, BWM612740S, BWM612743S, BWM612745S, BWM612746S, BWM612749S, BWM612773S, BWM612774S, BWM612775S, BWM612776S, BWM607803S, BWM608161S, BWM607236S, BWM607240S, BWM607747S, BWM609020S, BWM609628S, BWM611199S, BWM608041S, BWM608043S, BWM608051S, BWM608117S, BWM608527S, BWM608528S, BWM608530S, BWM608531S, BWM608532S, BWM609415S, BWM609416S, BWM609590S, BWM609596S, BWM609655S, BWM609656S, BWM609660S, BWM609684S, BWM609741S, BWM609748S, BWM609761S, BWM610419S, BWM610641S, BWM607227S, BWM607228S, BWM607232S, BWM607233S, BWM607234S, BWM607237S, BWM607242S, BWM607248S, BWM607250S, BWM607274S, BWM607276S, BWM607277S, BWM607278S, BWM607195S, BWM607832S, BWM607323S, BWM607572S, BWM607577S, BWM607200S, BWM607219S, BWM607224S, BWM607226S, BWM607327S, BWM607339S, BWM607988S, BWM607495S, BWM607808S, BWM608125S, BWM608962S, BWM607614S, BWM607337S, BWM608385S, BWM607986S, BWM609359S, BWM611236S, BWM609821S, BWM607987S, BWM609221S, BWM607528S, BWM607220S, BWM609040S, BWM609182S, BWM607225S, BWM607573S, BWM609041S, BWM607249S, BWM607487S, BWM607521S, BWM607531S, BWM607251S, BWM607324S, BWM607578S, BWM607486S, BWM607557S, BWM607550S, BWM607813S, BWM607642S, BWM607644S, BWM607647S, BWM607648S, BWM607650S, BWM607743S, BWM607749S, BWM607750S, BWM607820S, BWM607821S, BWM607822S, BWM607823S, BWM607824S, BWM607825S, BWM607828S, BWM607830S, BWM607831S, BWM607839S, BWM607918S, BWM608009S, BWM608010S, BWM608011S, BWM608012S, BWM608013S, BWM608014S, BWM608015S, BWM608016S, BWM608093S, BWM608094S, BWM608095S, BWM608096S, BWM608337S, BWM608338S, BWM608339S, BWM608340S, BWM608341S, BWM608342S, BWM608343S, BWM608344S, BWM608345S, BWM608346S, BWM608576S, BWM608577S, BWM608578S, BWM608579S, BWM608580S, BWM608957S, BWM608993S, BWM608995S, BWM608996S, BWM609006S, BWM609189S, BWM609190S, BWM609191S, BWM609192S, BWM609193S, BWM609194S, BWM609215S, BWM607989S, BWM610955S, BWM611416S, BWM611420S, BWM612731S, BWM609468S, BWM608526S, BWM609339S, BWM608544S, BWM608545S, BWM608589S, BWM608685S, BWM608994S, BWM609003S, BWM609011S, BWM609012S, BWM609015S, BWM609230S, BWM609231S, BWM609245S, BWM609271S, BWM609325S, BWM609535S, BWM609679S, BWM609750S, BWM609803S, BWM610470S, BWM610839S, BWM608601S, BWM608608S, BWM608610S, BWM608620S, BWM608627S, BWM608630S, BWM608631S, BWM608637S, BWM608683S, BWM608688S, BWM608978S, BWM609008S, BWM609009S, BWM609010S, BWM609013S, BWM609014S, BWM609016S, BWM609042S, BWM609223S, BWM609229S, BWM609234S, BWM609242S, BWM609244S, BWM609246S, BWM609302S, BWM609309S, BWM609313S, BWM609320S, BWM609345S, BWM609357S, BWM609360S, BWM609363S, BWM609364S, BWM609371S, BWM609380S, BWM609422S, BWM609425S, BWM609426S, BWM609427S, BWM609537S, BWM609588S, BWM609639S, BWM609645S, BWM609651S, BWM609653S, BWM609668S, BWM609671S, BWM609682S, BWM609700S, BWM609713S, BWM609716S, BWM609719S, BWM609722S, BWM609737S, BWM609743S, BWM609747S, BWM609754S, BWM609759S, BWM609765S, BWM609776S, BWM609786S, BWM609838S, BWM609840S, BWM609842S, BWM609847S, BWM609856S, BWM609884S, BWM609913S, BWM609926S, BWM610151S, BWM610183S, BWM610186S, BWM610373S, BWM610511S, BWM610595S, BWM610610S, BWM608607S, BWM608611S, BWM608988S, BWM609239S, BWM609240S, BWM609241S, BWM609251S, BWM609252S, BWM609376S, BWM609666S, BWM607704S, BWM607576S, BWM607639S, BWM607993S, BWM608003S, BWM607638S, BWM607748S, BWM607751S, BWM609043S, BWM609382S, BWM609400S, BWM609404S, BWM609410S, BWM609411S, BWM609420S, BWM610714S, BWM610734S, BWM611574S, BWM611653S, BWM612771S, BWM612772S, BWM609617S, BWM609625S, BWM609626S, BWM609623S, BWM608599S, BWM608600S, BWM608602S, BWM608605S, BWM609343S, BWM609348S, BWM609350S, BWM609412S, BWM609413S, BWM609500S, BWM609541S, BWM609564S, BWM609604S, BWM608622S, BWM608987S, BWM608989S, BWM608990S, BWM608991S, BWM610645S, BWM611251S, BWM608609S, BWM609001S, BWM609417S, BWM609418S, BWM609494S, BWM609496S, BWM609627S, BWM609646S, BWM609647S, BWM609650S, BWM609652S, BWM609654S, BWM609648S, BWM609649S, BWM607826S, BWM609236S, BWM609237S, BWM609255S, BWM609258S, BWM609268S, BWM609279S, BWM609281S, BWM609287S, BWM609288S, BWM609289S, BWM609290S, BWM609291S, BWM609322S, BWM609331S, BWM609332S, BWM609347S, BWM609349S, BWM609351S, BWM609352S, BWM609353S, BWM609581S, BWM609607S, BWM609609S, BWM609209S, BWM610248S, BWM609482S, BWM609483S, BWM609556S, BWM607701S, BWM607700S, BWM609414S, BWM610966S, BWM611005S, BWM611605S, BWM611580S, BWM608633S, BWM608634S, BWM608635S, BWM608636S, BWM608640S, BWM608641S, BWM608643S, BWM608644S, BWM610108S, BWM610109S, BWM610111S, BWM610112S, BWM610113S, BWM610115S, BWM610116S, BWM610118S, BWM610121S, BWM610122S, BWM610123S, BWM610126S, BWM610127S, BWM610128S, BWM610129S, BWM610130S, BWM610131S, BWM610134S, BWM610135S, BWM610137S, BWM611411S, BWM611876S, BWM607568S, BWM607563S, BWM607564S, BWM607565S, BWM607566S, BWM607282S, BWM607283S, BWM607284S, BWM607632S, BWM609823S, BWM609824S, BWM609833S, BWM609834S, BWM610375S, BWM610381S, BWM610387S, BWM610415S, BWM610451S, BWM610455S, BWM610480S, BWM610481S, BWM610488S, BWM610495S, BWM610498S, BWM610663S, BWM610665S, BWM610666S, BWM610667S, BWM610669S, BWM610672S, BWM610673S, BWM610674S, BWM610675S, BWM610676S, BWM610677S, BWM610950S, BWM610951S, BWM609760S, BWM609789S, BWM609808S, BWM607215S, BWM609387S, BWM609986S, BWM613068S, BWM613071S, BWM613072S, BWM613234S, BWM613237S, BWM613238S, BWM613475S, BWM613476S, BWM613477S, BWM613478S, BWM609175S, BWM609176S, BWM609398S, BWM609431S, BWM609432S, BWM609434S, BWM609436S, BWM609442S, BWM609445S, BWM609446S, BWM609448S, BWM609454S, BWM609457S, BWM609459S, BWM609486S, BWM609488S, BWM609489S, BWM609491S, BWM609492S, BWM609493S, BWM609619S, BWM609661S, BWM609662S, BWM609663S, BWM609690S, BWM610362S, BWM610369S, BWM610371S, BWM610927S, BWM610928S, BWM610989S, BWM610991S, BWM610993S, BWM610994S, BWM607554S, BWM607558S, BWM609207S, BWM609744S, BWM610894S, BWM610661S, BWM611436S, BWM607981S, BWM608005S, BWM608006S, BWM608678S, BWM608679S, BWM608680S, BWM608681S, BWM609178S, BWM609204S, BWM609205S, BWM609464S, BWM609472S, BWM609861S, BWM609891S, BWM610038S, BWM610040S, BWM610050S, BWM610176S, BWM610297S, BWM610805S, BWM610846S, BWM611074S, BWM611113S, BWM611120S, BWM611121S, BWM611379S, BWM611500S, BWM611685S, BWM611754S, BWM611755S, BWM611879S, BWM611880S, BWM611881S, BWM611882S, BWM611883S, BWM611978S, BWM611979S, BWM611980S, BWM611981S, BWM611982S, BWM612186S, BWM612188S, BWM612390S, BWM612396S, BWM612525S, BWM612595S, BWM612617S, BWM612626S, BWM612628S, BWM612673S, BWM612930S, BWM613066S, BWM613067S, BWM613166S, BWM613167S, BWM613216S, BWM613217S, BWM613383S, BWM613384S, BWM613385S, BWM613393S, BWM613394S, BWM610156S, BWM610627S, BWM610139S, BWM608307S, BWM609703S, BWM610756S, BWM609724S, BWM609597S, BWM609603S, BWM607742S, BWM609640S, BWM609644S, BWM609634S, BWM609598S, BWM609602S, BWM609629S, BWM609631S, BWM609633S, BWM609707S, BWM607740S, BWM609546S, BWM609547S, BWM609552S, BWM609554S, BWM607294S, BWM607296S, BWM607836S, BWM607840S, BWM608529S, BWM608533S, BWM609265S, BWM608590S, BWM608598S, BWM608604S, BWM608606S, BWM608614S, BWM608619S, BWM608624S, BWM608625S, BWM608626S, BWM608628S, BWM608638S, BWM608639S, BWM608642S, BWM608645S, BWM609017S, BWM609217S, BWM609225S, BWM609227S, BWM609235S, BWM609243S, BWM609248S, BWM609314S, BWM609368S, BWM609424S, BWM609428S, BWM609429S, BWM609557S, BWM609600S, BWM609606S, BWM609608S, BWM609610S, BWM609614S, BWM609630S, BWM609677S, BWM609758S, BWM609769S, BWM609785S, BWM609792S, BWM609794S, BWM609795S, BWM609799S, BWM609804S, BWM609864S, BWM609877S, BWM609899S, BWM609906S, BWM610707S, BWM610721S, BWM610761S, BWM609329S, BWM609658S, BWM609731S, BWM609197S, BWM609198S, BWM609199S, BWM609475S, BWM609476S, BWM609673S, BWM610351S, BWM609007S, BWM609018S, BWM609019S, BWM609224S, BWM609226S, BWM609232S, BWM609264S, BWM609342S, BWM609344S, BWM609346S, BWM609355S, BWM609402S, BWM609461S, BWM609462S, BWM609513S, BWM609514S, BWM609550S, BWM609828S, BWM609841S, BWM609882S, BWM609883S, BWM609890S, BWM609894S, BWM609900S, BWM609901S, BWM609902S, BWM609932S, BWM609943S, BWM609944S, BWM609945S, BWM609947S, BWM609948S, BWM609950S, BWM609953S, BWM609955S, BWM609956S, BWM609957S, BWM609958S, BWM609959S, BWM609960S, BWM609964S, BWM609965S, BWM609978S, BWM609985S, BWM609996S, BWM610009S, BWM610013S, BWM610014S, BWM610051S, BWM610075S, BWM610098S, BWM610099S, BWM610100S, BWM610105S, BWM610133S, BWM610136S, BWM610138S, BWM610149S, BWM610152S, BWM610161S, BWM610179S, BWM610205S, BWM610217S, BWM610221S, BWM610229S, BWM610241S, BWM610242S, BWM610243S, BWM610253S, BWM610256S, BWM610258S, BWM610279S, BWM610280S, BWM610282S, BWM610292S, BWM610301S, BWM610302S, BWM610360S, BWM610361S, BWM610363S, BWM610364S, BWM610367S, BWM610395S, BWM610396S, BWM610435S, BWM610436S, BWM610439S, BWM610459S, BWM610474S, BWM610475S, BWM610482S, BWM610483S, BWM610492S, BWM610493S, BWM610507S, BWM610508S, BWM610509S, BWM610521S, BWM610522S, BWM610528S, BWM610529S, BWM610537S, BWM610539S, BWM610541S, BWM610550S, BWM610554S, BWM610555S, BWM610598S, BWM610735S, BWM610771S, BWM610810S, BWM610831S, BWM610832S, BWM610833S, BWM610834S, BWM610845S, BWM610918S, BWM610921S, BWM610922S, BWM610938S, BWM611025S, BWM611026S, BWM611028S, BWM611034S, BWM611035S, BWM611046S, BWM611047S, BWM611048S, BWM611049S, BWM611050S, BWM611051S, BWM611052S, BWM611075S, BWM611103S, BWM611118S, BWM611128S, BWM611129S, BWM611147S, BWM611148S, BWM611149S, BWM611189S, BWM611191S, BWM611192S, BWM611193S, BWM611197S, BWM611209S, BWM611210S, BWM611211S, BWM611212S, BWM611214S, BWM611215S, BWM611219S, BWM611225S, BWM611283S, BWM611284S, BWM611286S, BWM611323S, BWM611324S, BWM611325S, BWM611344S, BWM611354S, BWM611356S, BWM611357S, BWM611361S, BWM611362S, BWM611363S, BWM611372S, BWM611373S, BWM611392S, BWM611394S, BWM611425S, BWM611426S, BWM611430S, BWM611431S, BWM611450S, BWM611451S, BWM611458S, BWM611460S, BWM611461S, BWM611497S, BWM611498S, BWM611524S, BWM611525S, BWM611526S, BWM611527S, BWM611528S, BWM611529S, BWM611550S, BWM611552S, BWM611553S, BWM611554S, BWM611555S, BWM611556S, BWM611563S, BWM611565S, BWM611566S, BWM611589S, BWM611591S, BWM611594S, BWM611595S, BWM611596S, BWM611597S, BWM611628S, BWM611631S, BWM611632S, BWM611635S, BWM611637S, BWM611639S, BWM611647S, BWM611648S, BWM611663S, BWM611664S, BWM611665S, BWM611666S, BWM611667S, BWM611686S, BWM611687S, BWM611688S, BWM611689S, BWM611690S, BWM611693S, BWM611721S, BWM611723S, BWM611724S, BWM611725S, BWM611726S, BWM611727S, BWM611735S, BWM611741S, BWM611742S, BWM611743S, BWM611744S, BWM611745S, BWM611756S, BWM611757S, BWM611758S, BWM611759S, BWM611760S, BWM611761S, BWM611762S, BWM611763S, BWM611766S, BWM611767S, BWM611768S, BWM611769S, BWM611791S, BWM611792S, BWM611793S, BWM611794S, BWM611808S, BWM611809S, BWM611810S, BWM611811S, BWM611812S, BWM611813S, BWM611814S, BWM611815S, BWM611816S, BWM611817S, BWM611854S, BWM611855S, BWM611856S, BWM611857S, BWM611858S, BWM611859S, BWM611860S, BWM611861S, BWM611894S, BWM611942S, BWM611943S, BWM611944S, BWM611945S, BWM611946S, BWM611947S, BWM611948S, BWM611983S, BWM611984S, BWM611985S, BWM612052S, BWM612054S, BWM612055S, BWM612056S, BWM612057S, BWM612068S, BWM612069S, BWM612070S, BWM612071S, BWM612072S, BWM612078S, BWM612090S, BWM612095S, BWM612096S, BWM612098S, BWM612109S, BWM612110S, BWM612112S, BWM612113S, BWM612114S, BWM612116S, BWM612117S, BWM612118S, BWM612119S, BWM612123S, BWM612131S, BWM612132S, BWM612139S, BWM612176S, BWM612182S, BWM612183S, BWM612185S, BWM612187S, BWM612189S, BWM612192S, BWM612210S, BWM612211S, BWM612212S, BWM612213S, BWM612214S, BWM612215S, BWM612219S, BWM612220S, BWM612222S, BWM612223S, BWM612224S, BWM612225S, BWM612226S, BWM612228S, BWM612230S, BWM612231S, BWM612236S, BWM612239S, BWM612241S, BWM612247S, BWM612255S, BWM612258S, BWM612260S, BWM612262S, BWM612265S, BWM612300S, BWM612302S, BWM612303S, BWM612305S, BWM612306S, BWM612308S, BWM612309S, BWM612312S, BWM612314S, BWM612327S, BWM612328S, BWM612329S, BWM612330S, BWM612331S, BWM612332S, BWM612333S, BWM612349S, BWM612350S, BWM612351S, BWM612355S, BWM612357S, BWM612359S, BWM612360S, BWM612361S, BWM612362S, BWM612363S, BWM612377S, BWM612378S, BWM612379S, BWM612380S, BWM612382S, BWM612383S, BWM612384S, BWM612385S, BWM612386S, BWM612387S, BWM612391S, BWM612397S, BWM612398S, BWM612399S, BWM612400S, BWM612411S, BWM612447S, BWM612449S, BWM612450S, BWM612451S, BWM612452S, BWM612454S, BWM612455S, BWM612457S, BWM612458S, BWM612522S, BWM612557S, BWM612580S, BWM612585S, BWM612625S, BWM612686S, BWM612689S, BWM612693S, BWM612697S, BWM612702S, BWM612703S, BWM612704S, BWM612705S, BWM612706S, BWM612707S, BWM612708S, BWM612709S, BWM612710S, BWM612711S, BWM612712S, BWM612713S, BWM612714S, BWM612715S, BWM612716S, BWM612717S, BWM612718S, BWM612719S, BWM612720S, BWM612721S, BWM612722S, BWM612723S, BWM612741S, BWM612766S, BWM612837S, BWM612838S, BWM612843S, BWM612845S, BWM612846S, BWM612847S, BWM612849S, BWM612853S, BWM612855S, BWM612869S, BWM612870S, BWM612871S, BWM612872S, BWM612873S, BWM612874S, BWM612875S, BWM612915S, BWM612916S, BWM612917S, BWM612918S, BWM612919S, BWM612921S, BWM612934S, BWM612972S, BWM612973S, BWM612974S, BWM612975S, BWM612976S, BWM612977S, BWM612978S, BWM612983S, BWM612984S, BWM612985S, BWM612986S, BWM612990S, BWM612991S, BWM613002S, BWM613003S, BWM613004S, BWM613005S, BWM613006S, BWM613007S, BWM613008S, BWM613009S, BWM613010S, BWM613011S, BWM613012S, BWM613020S, BWM613021S, BWM613065S, BWM613069S, BWM613070S, BWM613074S, BWM613075S, BWM613082S, BWM613083S, BWM613084S, BWM613085S, BWM613162S, BWM613243S, BWM613244S, BWM613245S, BWM613246S, BWM613247S, BWM613252S, BWM613253S, BWM613254S, BWM613255S, BWM613256S, BWM613257S, BWM613258S, BWM613259S, BWM613260S, BWM613261S, BWM613291S, BWM613292S, BWM613305S, BWM613307S, BWM613408S, BWM613409S, BWM613410S, BWM613411S, BWM613412S, BWM613413S, BWM613414S, BWM613415S, BWM613416S, BWM613469S, BWM613483S, BWM609253S, BWM609254S, BWM609430S, BWM609438S, BWM609657S, BWM609850S, BWM609962S, BWM609991S, BWM610124S, BWM610250S, BWM610251S, BWM610257S, BWM610293S, BWM610341S, BWM610437S, BWM610441S, BWM610456S, BWM610458S, BWM610484S, BWM610527S, BWM610606S, BWM610643S, BWM610891S, BWM610932S, BWM610937S, BWM610940S, BWM610949S, BWM611023S, BWM611084S, BWM611119S, BWM611127S, BWM611198S, BWM611200S, BWM611267S, BWM611313S, BWM611314S, BWM611339S, BWM611350S, BWM611351S, BWM611399S, BWM611502S, BWM611567S, BWM611617S, BWM611659S, BWM611764S, BWM611780S, BWM611852S, BWM611853S, BWM611864S, BWM611865S, BWM611871S, BWM611872S, BWM611884S, BWM611885S, BWM611888S, BWM611891S, BWM612120S, BWM612135S, BWM612136S, BWM612267S, BWM612271S, BWM612356S, BWM612358S, BWM612366S, BWM612371S, BWM612372S, BWM612373S, BWM612374S, BWM612376S, BWM612392S, BWM612393S, BWM612394S, BWM612421S, BWM612446S, BWM612462S, BWM612466S, BWM612511S, BWM612521S, BWM612596S, BWM612601S, BWM612798S, BWM612799S, BWM612839S, BWM612840S, BWM612841S, BWM612842S, BWM612935S, BWM612994S, BWM612995S, BWM612996S, BWM613046S, BWM613048S, BWM613056S, BWM613076S, BWM613077S, BWM613078S, BWM613079S, BWM613080S, BWM613081S, BWM613202S, BWM613203S, BWM613224S, BWM613233S, BWM613239S, BWM613317S, BWM613318S, BWM613386S, BWM613387S, BWM613471S, BWM613472S, BWM613503S, BWM613506S, BWM608348S, BWM608997S, BWM609000S, BWM609002S, BWM609004S, BWM609005S, BWM609195S, BWM609545S, BWM609563S, BWM609613S, BWM609615S, BWM609616S, BWM609685S, BWM609701S, BWM609805S, BWM609857S, BWM610506S, BWM610546S, BWM610551S, BWM610559S, BWM610590S, BWM610597S, BWM610599S, BWM610607S, BWM610622S, BWM610625S, BWM610708S, BWM610727S, BWM610728S, BWM610731S, BWM610954S, BWM607744S, BWM607746S, BWM609632S, BWM607214S, BWM607524S, BWM611577S, BWM609543S, BWM609599S, BWM608992S, BWM610394S, BWM608539S, BWM608603S, BWM608616S, BWM608975S, BWM609715S, BWM609865S, BWM609866S, BWM609481S, BWM609764S, BWM611592S, BWM607738S, BWM609605S, BWM608045S, BWM608077S, BWM608078S, BWM608297S, BWM608298S, BWM608299S, BWM608300S, BWM608388S, BWM608534S, BWM608535S, BWM608536S, BWM608543S, BWM608581S, BWM608582S, BWM608612S, BWM608613S, BWM608682S, BWM608684S, BWM608690S, BWM608691S, BWM608692S, BWM608723S, BWM608724S, BWM608164S, BWM608165S, BWM608166S, BWM608167S, BWM608168S, BWM608301S, BWM608303S, BWM608304S, BWM608305S, BWM608374S, BWM608375S, BWM608387S, BWM608586S, BWM608587S, BWM608588S, BWM608596S, BWM608597S, BWM608615S, BWM608617S, BWM608629S, BWM608632S, BWM608718S, BWM608719S, BWM608720S, BWM608721S, BWM608722S, BWM608728S, BWM608729S, BWM608730S, BWM608731S, BWM608732S, BWM608080S, BWM608081S, BWM608083S, BWM608084S, BWM608085S, BWM608727S, BWM608591S, BWM608594S, BWM608087S, BWM608134S, BWM608541S, BWM608542S, BWM608693S, BWM608717S, BWM608726S, BWM609213S, BWM609214S, BWM609228S, BWM609233S, BWM608046S, BWM608047S, BWM608048S, BWM608049S, BWM608050S, BWM608169S, BWM608170S, BWM608171S, BWM608172S, BWM608173S, BWM608174S, BWM608380S, BWM608381S, BWM608592S, BWM608593S, BWM608595S, BWM608686S, BWM608687S, BWM608052S, BWM608053S, BWM608055S, BWM608056S, BWM608135S, BWM608136S, BWM608175S, BWM608176S, BWM608177S, BWM608178S, BWM608714S, BWM608715S, BWM608716S, BWM608725S, BWM608618S, BWM608623S, BWM608689S, BWM608044S, BWM608306S, BWM608308S, BWM608583S, BWM608584S, BWM608733S, BWM608734S, BWM608372S, BWM608086S, BWM608585S, BWM609212S, BWM611123S, BWM612778S, BWM611395S, BWM611462S, BWM611208S, BWM610836S, BWM610840S, BWM610853S, BWM610924S, BWM610953S, BWM609542S, BWM610377S, BWM609247S, BWM609249S, BWM609250S, BWM609266S, BWM609267S, BWM609269S, BWM609270S, BWM609272S, BWM609273S, BWM609274S, BWM609275S, BWM609276S, BWM609278S, BWM609282S, BWM609283S, BWM609284S, BWM609296S, BWM609298S, BWM609299S, BWM609300S, BWM609301S, BWM609303S, BWM609304S, BWM609305S, BWM609306S, BWM609307S, BWM609308S, BWM609310S, BWM609358S, BWM609361S, BWM609374S, BWM609375S, BWM609377S, BWM609403S, BWM609405S, BWM609406S, BWM609439S, BWM609440S, BWM609441S, BWM609443S, BWM609444S, BWM609452S, BWM609453S, BWM609455S, BWM609456S, BWM609458S, BWM609487S, BWM609490S, BWM609530S, BWM609531S, BWM609532S, BWM609533S, BWM609534S, BWM609536S, BWM609538S, BWM609539S, BWM609540S, BWM609548S, BWM609551S, BWM609553S, BWM609555S, BWM609558S, BWM609559S, BWM609560S, BWM609561S, BWM609562S, BWM609573S, BWM609574S, BWM609578S, BWM609584S, BWM609585S, BWM609593S, BWM609611S, BWM609612S, BWM609659S, BWM609664S, BWM609672S, BWM609738S, BWM609767S, BWM609787S, BWM609793S, BWM609800S, BWM609844S, BWM609845S, BWM609862S, BWM609863S, BWM609870S, BWM609871S, BWM609872S, BWM609873S, BWM609874S, BWM609875S, BWM609923S, BWM609924S, BWM609925S, BWM609927S, BWM609928S, BWM609929S, BWM609930S, BWM609931S, BWM609934S, BWM609935S, BWM609936S, BWM609951S, BWM609952S, BWM609966S, BWM609967S, BWM609971S, BWM609972S, BWM609974S, BWM609994S, BWM610004S, BWM610005S, BWM610006S, BWM610008S, BWM610023S, BWM610024S, BWM610025S, BWM610026S, BWM610027S, BWM610029S, BWM610039S, BWM610041S, BWM610048S, BWM610052S, BWM610053S, BWM610054S, BWM610055S, BWM610056S, BWM610066S, BWM610069S, BWM610070S, BWM610071S, BWM610077S, BWM610078S, BWM610081S, BWM610103S, BWM610104S, BWM610125S, BWM610140S, BWM610141S, BWM610142S, BWM610143S, BWM610144S, BWM610146S, BWM610147S, BWM610148S, BWM610171S, BWM610172S, BWM610173S, BWM610174S, BWM610175S, BWM610177S, BWM610213S, BWM610214S, BWM610215S, BWM610216S, BWM610218S, BWM610220S, BWM610228S, BWM610232S, BWM610233S, BWM610234S, BWM610239S, BWM610240S, BWM610267S, BWM610268S, BWM610269S, BWM610270S, BWM610271S, BWM610304S, BWM610305S, BWM610306S, BWM610307S, BWM610308S, BWM610309S, BWM610310S, BWM610311S, BWM610332S, BWM610333S, BWM610335S, BWM610336S, BWM610337S, BWM610338S, BWM610339S, BWM610340S, BWM610354S, BWM610383S, BWM610384S, BWM610386S, BWM610388S, BWM610389S, BWM610390S, BWM610392S, BWM610397S, BWM610398S, BWM610399S, BWM610400S, BWM610401S, BWM610403S, BWM610404S, BWM610405S, BWM610406S, BWM610407S, BWM610408S, BWM610409S, BWM610410S, BWM610411S, BWM610412S, BWM610414S, BWM610416S, BWM610421S, BWM610423S, BWM610444S, BWM610449S, BWM610450S, BWM610452S, BWM610453S, BWM610467S, BWM610468S, BWM610471S, BWM610573S, BWM610574S, BWM610576S, BWM610577S, BWM610578S, BWM610579S, BWM610580S, BWM610581S, BWM610582S, BWM610583S, BWM610584S, BWM610585S, BWM610586S, BWM610644S, BWM610646S, BWM610648S, BWM610649S, BWM610652S, BWM610655S, BWM610685S, BWM610705S, BWM610709S, BWM610710S, BWM610711S, BWM610713S, BWM610720S, BWM610722S, BWM610726S, BWM610729S, BWM610730S, BWM610736S, BWM610737S, BWM610740S, BWM610742S, BWM610743S, BWM610744S, BWM610746S, BWM610751S, BWM610758S, BWM610762S, BWM610763S, BWM610764S, BWM610765S, BWM610768S, BWM610769S, BWM610775S, BWM610776S, BWM610778S, BWM610781S, BWM610818S, BWM610821S, BWM610822S, BWM610825S, BWM610826S, BWM610866S, BWM610867S, BWM610868S, BWM610876S, BWM610881S, BWM610882S, BWM610883S, BWM610898S, BWM610899S, BWM610900S, BWM610901S, BWM610902S, BWM610903S, BWM610904S, BWM610933S, BWM610934S, BWM610935S, BWM610936S, BWM610946S, BWM610947S, BWM610948S, BWM610973S, BWM610977S, BWM611002S, BWM611003S, BWM611004S, BWM611012S, BWM611014S, BWM611024S, BWM611036S, BWM611037S, BWM611038S, BWM611053S, BWM611054S, BWM611063S, BWM611064S, BWM611065S, BWM611070S, BWM611071S, BWM611072S, BWM611073S, BWM611079S, BWM611080S, BWM611081S, BWM611082S, BWM611083S, BWM611087S, BWM611088S, BWM611089S, BWM611090S, BWM611091S, BWM611098S, BWM611100S, BWM611108S, BWM611109S, BWM611111S, BWM611112S, BWM611122S, BWM611124S, BWM611125S, BWM611126S, BWM611132S, BWM611133S, BWM611134S, BWM611135S, BWM611138S, BWM611143S, BWM611163S, BWM611172S, BWM611173S, BWM611229S, BWM611235S, BWM611237S, BWM611254S, BWM611255S, BWM611256S, BWM611258S, BWM611259S, BWM611260S, BWM611261S, BWM611262S, BWM611263S, BWM611304S, BWM611305S, BWM611307S, BWM611308S, BWM611310S, BWM611326S, BWM611327S, BWM611328S, BWM611329S, BWM611330S, BWM611340S, BWM611353S, BWM611355S, BWM611377S, BWM611384S, BWM611385S, BWM611415S, BWM611417S, BWM611418S, BWM611422S, BWM611423S, BWM611464S, BWM611465S, BWM611466S, BWM611467S, BWM611468S, BWM611469S, BWM611470S, BWM611471S, BWM611472S, BWM611473S, BWM611474S, BWM611475S, BWM611489S, BWM611490S, BWM611499S, BWM611505S, BWM611506S, BWM611512S, BWM611513S, BWM611514S, BWM611515S, BWM611516S, BWM611517S, BWM611518S, BWM611519S, BWM611531S, BWM611532S, BWM611533S, BWM611534S, BWM611535S, BWM611536S, BWM611537S, BWM611538S, BWM611568S, BWM611569S, BWM611570S, BWM611571S, BWM611598S, BWM611601S, BWM611602S, BWM611603S, BWM611606S, BWM611607S, BWM611608S, BWM611609S, BWM611610S, BWM611611S, BWM611781S, BWM611782S, BWM611783S, BWM611797S, BWM611798S, BWM611799S, BWM611800S, BWM611801S, BWM611840S, BWM611841S, BWM611873S, BWM611874S, BWM611875S, BWM611877S, BWM611878S, BWM611897S, BWM611898S, BWM611899S, BWM611900S, BWM611901S, BWM611902S, BWM611903S, BWM611904S, BWM611905S, BWM611906S, BWM611949S, BWM611950S, BWM611951S, BWM611952S, BWM611953S, BWM611954S, BWM611955S, BWM611957S, BWM611958S, BWM611959S, BWM611960S, BWM611961S, BWM611962S, BWM611963S, BWM611964S, BWM611965S, BWM611966S, BWM611967S, BWM611968S, BWM611969S, BWM611971S, BWM611974S, BWM611975S, BWM612005S, BWM612006S, BWM612007S, BWM612009S, BWM612010S, BWM612011S, BWM612012S, BWM612013S, BWM612014S, BWM612015S, BWM612017S, BWM612018S, BWM612028S, BWM612030S, BWM612034S, BWM612036S, BWM612044S, BWM612045S, BWM612046S, BWM612047S, BWM612048S, BWM612049S, BWM612050S, BWM612053S, BWM612058S, BWM612059S, BWM612066S, BWM612074S, BWM612075S, BWM612076S, BWM612077S, BWM612080S, BWM612081S, BWM612083S, BWM612085S, BWM612086S, BWM612087S, BWM612089S, BWM612091S, BWM612099S, BWM612100S, BWM612134S, BWM612137S, BWM612138S, BWM612142S, BWM612143S, BWM612144S, BWM612145S, BWM612147S, BWM612148S, BWM612149S, BWM612151S, BWM612152S, BWM612153S, BWM612181S, BWM612184S, BWM612190S, BWM612193S, BWM612194S, BWM612195S, BWM612196S, BWM612197S, BWM612198S, BWM612199S, BWM612200S, BWM612201S, BWM612202S, BWM612246S, BWM612248S, BWM612249S, BWM612250S, BWM612251S, BWM612252S, BWM612253S, BWM612254S, BWM612256S, BWM612257S, BWM612259S, BWM612261S, BWM612263S, BWM612264S, BWM612266S, BWM612268S, BWM612269S, BWM612270S, BWM612272S, BWM612339S, BWM612340S, BWM612341S, BWM612342S, BWM612343S, BWM612345S, BWM612352S, BWM612354S, BWM612375S, BWM612401S, BWM612402S, BWM612403S, BWM612405S, BWM612406S, BWM612407S, BWM612408S, BWM612409S, BWM612410S, BWM612412S, BWM612413S, BWM612414S, BWM612415S, BWM612419S, BWM612420S, BWM612423S, BWM612425S, BWM612426S, BWM612429S, BWM612430S, BWM612432S, BWM612456S, BWM612463S, BWM612470S, BWM612476S, BWM612491S, BWM612492S, BWM612493S, BWM612494S, BWM612497S, BWM612503S, BWM612509S, BWM612512S, BWM612513S, BWM612514S, BWM612515S, BWM612516S, BWM612517S, BWM612520S, BWM612527S, BWM612528S, BWM612529S, BWM612530S, BWM612532S, BWM612533S, BWM612534S, BWM612537S, BWM612538S, BWM612539S, BWM612639S, BWM612640S, BWM612641S, BWM612642S, BWM612643S, BWM612644S, BWM612645S, BWM612647S, BWM612649S, BWM612650S, BWM612652S, BWM612653S, BWM612767S, BWM612768S, BWM612788S, BWM612789S, BWM612790S, BWM612804S, BWM612805S, BWM612806S, BWM612807S, BWM612808S, BWM612809S, BWM612810S, BWM612811S, BWM612812S, BWM612813S, BWM612814S, BWM612815S, BWM612816S, BWM612879S, BWM612881S, BWM612884S, BWM612885S, BWM612886S, BWM612888S, BWM612889S, BWM612891S, BWM612893S, BWM612895S, BWM612905S, BWM612907S, BWM612909S, BWM612913S, BWM612922S, BWM612936S, BWM612937S, BWM612940S, BWM612941S, BWM612942S, BWM612943S, BWM612944S, BWM612945S, BWM612946S, BWM612947S, BWM612979S, BWM612980S, BWM613013S, BWM613014S, BWM613015S, BWM613016S, BWM613017S, BWM613018S, BWM613019S, BWM613033S, BWM613034S, BWM613035S, BWM613036S, BWM613037S, BWM613038S, BWM613039S, BWM613040S, BWM613041S, BWM613119S, BWM613120S, BWM613121S, BWM613122S, BWM613123S, BWM613124S, BWM613125S, BWM613126S, BWM613127S, BWM613128S, BWM613129S, BWM613130S, BWM613131S, BWM613132S, BWM613133S, BWM613134S, BWM613135S, BWM613136S, BWM613137S, BWM613138S, BWM613139S, BWM613140S, BWM613141S, BWM613142S, BWM613143S, BWM613144S, BWM613145S, BWM613146S, BWM613147S, BWM613395S, BWM613417S, BWM609781S, BWM610391S, BWM607631S, BWM608378S, BWM608538S, BWM609174S, BWM609665S, BWM609693S, BWM609702S, BWM609749S, BWM609751S, BWM609752S, BWM609762S, BWM609763S, BWM609778S, BWM609779S, BWM609780S, BWM609791S, BWM609796S, BWM609846S, BWM609849S, BWM609852S, BWM610145S, BWM610168S, BWM610190S, BWM610194S, BWM610210S, BWM610219S, BWM610222S, BWM610225S, BWM610254S, BWM610255S, BWM610291S, BWM610298S, BWM610323S, BWM610334S, BWM610344S, BWM610352S, BWM610620S, BWM610635S, BWM610680S, BWM610681S, BWM610683S, BWM610750S, BWM610752S, BWM610753S, BWM610759S, BWM610760S, BWM610766S, BWM610772S, BWM610773S, BWM610774S, BWM610777S, BWM610788S, BWM610789S, BWM610790S, BWM610791S, BWM610794S, BWM610795S, BWM610796S, BWM610798S, BWM610799S, BWM610801S, BWM610819S, BWM610820S, BWM610978S, BWM610979S, BWM610980S, BWM610981S, BWM610982S, BWM610983S, BWM610984S, BWM610985S, BWM610986S, BWM610988S, BWM611218S, BWM611224S, BWM611226S, BWM611285S, BWM611292S, BWM611301S, BWM609708S, BWM610931S, BWM611220S, BWM609940S, BWM609323S, BWM609324S, BWM609326S, BWM609407S, BWM609408S, BWM609435S, BWM609437S, BWM609499S, BWM609506S, BWM609507S, BWM609526S, BWM609570S, BWM609571S, BWM609572S, BWM609768S, BWM609770S, BWM609771S, BWM609773S, BWM609827S, BWM609837S, BWM609839S, BWM609851S, BWM609855S, BWM609939S, BWM609987S, BWM609995S, BWM609997S, BWM609998S, BWM610002S, BWM610015S, BWM610016S, BWM610018S, BWM610019S, BWM610028S, BWM610033S, BWM610034S, BWM610035S, BWM610036S, BWM610037S, BWM610049S, BWM610067S, BWM610068S, BWM610074S, BWM610083S, BWM610119S, BWM610132S, BWM610150S, BWM610153S, BWM610158S, BWM610159S, BWM610178S, BWM610180S, BWM610181S, BWM610182S, BWM610184S, BWM610185S, BWM610187S, BWM610188S, BWM610192S, BWM610207S, BWM610208S, BWM610249S, BWM610252S, BWM610273S, BWM610274S, BWM610276S, BWM610277S, BWM610278S, BWM610283S, BWM610284S, BWM610285S, BWM610286S, BWM610318S, BWM610358S, BWM610359S, BWM610365S, BWM610366S, BWM610425S, BWM610426S, BWM610429S, BWM610465S, BWM610466S, BWM610469S, BWM610472S, BWM610489S, BWM610490S, BWM610502S, BWM610515S, BWM610516S, BWM610519S, BWM610520S, BWM610524S, BWM610525S, BWM610526S, BWM610534S, BWM610545S, BWM610547S, BWM610548S, BWM610552S, BWM610553S, BWM610600S, BWM610601S, BWM610624S, BWM610626S, BWM610628S, BWM610629S, BWM610671S, BWM610678S, BWM610679S, BWM610682S, BWM610684S, BWM610687S, BWM610689S, BWM610701S, BWM610702S, BWM610703S, BWM610704S, BWM610718S, BWM610719S, BWM610732S, BWM610733S, BWM610806S, BWM610809S, BWM610823S, BWM610824S, BWM610835S, BWM610838S, BWM610849S, BWM610850S, BWM610884S, BWM610885S, BWM610892S, BWM610895S, BWM610896S, BWM610897S, BWM610907S, BWM610909S, BWM610919S, BWM610920S, BWM610987S, BWM610990S, BWM611001S, BWM611021S, BWM611029S, BWM611030S, BWM611031S, BWM611032S, BWM611033S, BWM611055S, BWM611078S, BWM611085S, BWM611086S, BWM611102S, BWM611151S, BWM611152S, BWM611153S, BWM611154S, BWM611156S, BWM611157S, BWM611158S, BWM611159S, BWM611160S, BWM611178S, BWM611179S, BWM611180S, BWM611216S, BWM611227S, BWM611231S, BWM611233S, BWM611238S, BWM611239S, BWM611243S, BWM611245S, BWM611246S, BWM611247S, BWM611248S, BWM611249S, BWM611250S, BWM611253S, BWM611264S, BWM611279S, BWM611289S, BWM611290S, BWM611331S, BWM611332S, BWM611333S, BWM611334S, BWM611335S, BWM611336S, BWM611352S, BWM611364S, BWM611365S, BWM611366S, BWM611367S, BWM611413S, BWM611414S, BWM611501S, BWM611503S, BWM611504S, BWM611542S, BWM611543S, BWM611578S, BWM611579S, BWM611582S, BWM611583S, BWM611584S, BWM611612S, BWM611613S, BWM611614S, BWM611615S, BWM611624S, BWM611626S, BWM611654S, BWM611655S, BWM611720S, BWM611733S, BWM611734S, BWM611740S, BWM611746S, BWM611747S, BWM611784S, BWM611785S, BWM611786S, BWM611787S, BWM611788S, BWM611789S, BWM611790S, BWM611832S, BWM611833S, BWM611834S, BWM611835S, BWM611836S, BWM611837S, BWM611838S, BWM611839S, BWM611842S, BWM611843S, BWM611844S, BWM611845S, BWM611846S, BWM611847S, BWM611848S, BWM611849S, BWM611850S, BWM611851S, BWM611866S, BWM611867S, BWM611868S, BWM611869S, BWM611870S, BWM611926S, BWM611927S, BWM611928S, BWM611929S, BWM611930S, BWM611931S, BWM611932S, BWM611933S, BWM611934S, BWM611935S, BWM611936S, BWM611956S, BWM611988S, BWM611989S, BWM611991S, BWM611992S, BWM612008S, BWM612016S, BWM612019S, BWM612020S, BWM612021S, BWM612022S, BWM612023S, BWM612024S, BWM612025S, BWM612026S, BWM612027S, BWM612029S, BWM612031S, BWM612032S, BWM612033S, BWM612035S, BWM612037S, BWM612038S, BWM612039S, BWM612041S, BWM612062S, BWM612063S, BWM612067S, BWM612073S, BWM612092S, BWM612093S, BWM612154S, BWM612156S, BWM612157S, BWM612158S, BWM612159S, BWM612160S, BWM612161S, BWM612162S, BWM612164S, BWM612166S, BWM612487S, BWM612488S, BWM612498S, BWM612499S, BWM612500S, BWM612501S, BWM612502S, BWM612504S, BWM612505S, BWM612506S, BWM612507S, BWM612508S, BWM612510S, BWM612518S, BWM612519S, BWM612523S, BWM612524S, BWM612526S, BWM612545S, BWM612547S, BWM612551S, BWM612552S, BWM612553S, BWM612554S, BWM612555S, BWM612556S, BWM612558S, BWM612560S, BWM612561S, BWM612570S, BWM612572S, BWM612593S, BWM612618S, BWM612619S, BWM612620S, BWM612621S, BWM612622S, BWM612623S, BWM612624S, BWM612632S, BWM612633S, BWM612634S, BWM612636S, BWM612637S, BWM612646S, BWM612648S, BWM612651S, BWM612655S, BWM612659S, BWM612661S, BWM612663S, BWM612664S, BWM612672S, BWM612687S, BWM612688S, BWM612696S, BWM612759S, BWM612793S, BWM612794S, BWM612796S, BWM612797S, BWM612817S, BWM612821S, BWM612948S, BWM612949S, BWM612951S, BWM612952S, BWM612953S, BWM612997S, BWM612998S, BWM612999S, BWM613042S, BWM613044S, BWM613045S, BWM613047S, BWM613051S, BWM613052S, BWM613053S, BWM613054S, BWM613055S, BWM613057S, BWM613058S, BWM613059S, BWM613060S, BWM613061S, BWM613062S, BWM613063S, BWM613064S, BWM613184S, BWM613185S, BWM613186S, BWM613187S, BWM613211S, BWM613212S, BWM613213S, BWM613215S, BWM613231S, BWM613232S, BWM613235S, BWM613236S, BWM613240S, BWM613322S, BWM613323S, BWM613418S, BWM613419S, BWM613420S, BWM613421S, BWM613422S, BWM613423S, BWM613424S, BWM613425S, BWM613426S, BWM613427S, BWM613428S, BWM613429S, BWM613430S, BWM613470S, BWM613482S, BWM613496S, BWM613507S, BWM613508S, BWM609259S, BWM609260S, BWM609261S, BWM609262S, BWM609263S, BWM609286S, BWM609366S, BWM609367S, BWM609369S, BWM609372S, BWM609390S, BWM609391S, BWM609397S, BWM609399S, BWM609401S, BWM609409S, BWM609419S, BWM609421S, BWM609433S, BWM609447S, BWM609449S, BWM609450S, BWM609451S, BWM609473S, BWM609474S, BWM609477S, BWM609479S, BWM609485S, BWM609495S, BWM609497S, BWM609498S, BWM609501S, BWM609502S, BWM609503S, BWM609504S, BWM609505S, BWM609508S, BWM609509S, BWM609510S, BWM609511S, BWM609512S, BWM609515S, BWM609516S, BWM609517S, BWM609518S, BWM609519S, BWM609520S, BWM609521S, BWM609522S, BWM609524S, BWM609525S, BWM609527S, BWM609528S, BWM609529S, BWM609544S, BWM609549S, BWM609566S, BWM609567S, BWM609568S, BWM609569S, BWM609579S, BWM609580S, BWM609587S, BWM609594S, BWM609595S, BWM609601S, BWM609618S, BWM609620S, BWM609624S, BWM609635S, BWM609675S, BWM609729S, BWM609777S, BWM609783S, BWM609788S, BWM609812S, BWM609843S, BWM609848S, BWM609867S, BWM609885S, BWM609886S, BWM609887S, BWM609888S, BWM609889S, BWM609897S, BWM609898S, BWM609904S, BWM609905S, BWM609907S, BWM609908S, BWM609909S, BWM609910S, BWM609911S, BWM609912S, BWM609914S, BWM609915S, BWM609916S, BWM609917S, BWM609918S, BWM609919S, BWM609920S, BWM609921S, BWM609922S, BWM609937S, BWM609941S, BWM609975S, BWM609976S, BWM609977S, BWM609988S, BWM609989S, BWM609990S, BWM609992S, BWM610010S, BWM610020S, BWM610021S, BWM610022S, BWM610042S, BWM610058S, BWM610059S, BWM610060S, BWM610061S, BWM610062S, BWM610064S, BWM610065S, BWM610076S, BWM610084S, BWM610085S, BWM610086S, BWM610087S, BWM610101S, BWM610102S, BWM610154S, BWM610155S, BWM610162S, BWM610163S, BWM610164S, BWM610165S, BWM610166S, BWM610195S, BWM610196S, BWM610197S, BWM610206S, BWM610209S, BWM610211S, BWM610227S, BWM610236S, BWM610237S, BWM610238S, BWM610245S, BWM610247S, BWM610259S, BWM610260S, BWM610261S, BWM610262S, BWM610263S, BWM610264S, BWM610265S, BWM610266S, BWM610287S, BWM610288S, BWM610289S, BWM610290S, BWM610296S, BWM610299S, BWM610315S, BWM610316S, BWM610317S, BWM610320S, BWM610321S, BWM610322S, BWM610324S, BWM610325S, BWM610326S, BWM610327S, BWM610328S, BWM610329S, BWM610331S, BWM610342S, BWM610356S, BWM610357S, BWM610370S, BWM610372S, BWM610374S, BWM610380S, BWM610382S, BWM610418S, BWM610420S, BWM610434S, BWM610440S, BWM610460S, BWM610461S, BWM610462S, BWM610463S, BWM610464S, BWM610476S, BWM610477S, BWM610478S, BWM610485S, BWM610486S, BWM610494S, BWM610496S, BWM610510S, BWM610512S, BWM610513S, BWM610530S, BWM610538S, BWM610542S, BWM610549S, BWM610556S, BWM610557S, BWM610558S, BWM610560S, BWM610564S, BWM610565S, BWM610566S, BWM610567S, BWM610591S, BWM610592S, BWM610593S, BWM610604S, BWM610605S, BWM610616S, BWM610617S, BWM610618S, BWM610621S, BWM610623S, BWM610657S, BWM610658S, BWM610659S, BWM610664S, BWM610694S, BWM610695S, BWM610782S, BWM610783S, BWM610785S, BWM610803S, BWM610804S, BWM610811S, BWM610812S, BWM610813S, BWM610841S, BWM610844S, BWM610847S, BWM610848S, BWM610851S, BWM610852S, BWM610854S, BWM610855S, BWM610856S, BWM610858S, BWM610859S, BWM610860S, BWM610861S, BWM610862S, BWM610863S, BWM610864S, BWM610865S, BWM610869S, BWM610870S, BWM610871S, BWM610879S, BWM610880S, BWM610911S, BWM610912S, BWM610913S, BWM610972S, BWM611006S, BWM611007S, BWM611008S, BWM611009S, BWM611010S, BWM611011S, BWM611020S, BWM611022S, BWM611076S, BWM611077S, BWM611093S, BWM611095S, BWM611096S, BWM611097S, BWM611110S, BWM611114S, BWM611115S, BWM611136S, BWM611137S, BWM611150S, BWM611161S, BWM611182S, BWM611184S, BWM611213S, BWM611222S, BWM611223S, BWM611240S, BWM611241S, BWM611242S, BWM611276S, BWM611277S, BWM611278S, BWM611287S, BWM611293S, BWM611294S, BWM611295S, BWM611296S, BWM611297S, BWM611298S, BWM611299S, BWM611315S, BWM611317S, BWM611337S, BWM611338S, BWM611345S, BWM611349S, BWM611368S, BWM611369S, BWM611370S, BWM611371S, BWM611378S, BWM611380S, BWM611388S, BWM611393S, BWM611397S, BWM611401S, BWM611402S, BWM611403S, BWM611404S, BWM611406S, BWM611408S, BWM611409S, BWM611410S, BWM611412S, BWM611447S, BWM611448S, BWM611449S, BWM611476S, BWM611477S, BWM611478S, BWM611479S, BWM611480S, BWM611481S, BWM611482S, BWM611483S, BWM611507S, BWM611508S, BWM611509S, BWM611510S, BWM611511S, BWM611573S, BWM611618S, BWM611619S, BWM611620S, BWM611621S, BWM611622S, BWM611623S, BWM611668S, BWM611676S, BWM611681S, BWM611682S, BWM611683S, BWM611697S, BWM611703S, BWM611705S, BWM611716S, BWM611717S, BWM611718S, BWM611722S, BWM611748S, BWM611749S, BWM611795S, BWM611796S, BWM611802S, BWM611803S, BWM611804S, BWM611805S, BWM611806S, BWM611807S, BWM611822S, BWM611823S, BWM611824S, BWM611825S, BWM611829S, BWM611892S, BWM611893S, BWM611907S, BWM611908S, BWM611909S, BWM611910S, BWM611912S, BWM611913S, BWM611914S, BWM611915S, BWM611916S, BWM611917S, BWM611918S, BWM611919S, BWM611920S, BWM611937S, BWM611938S, BWM611939S, BWM611940S, BWM611941S, BWM612060S, BWM612061S, BWM612064S, BWM612065S, BWM612079S, BWM612082S, BWM612084S, BWM612088S, BWM612094S, BWM612097S, BWM612101S, BWM612102S, BWM612103S, BWM612104S, BWM612105S, BWM612106S, BWM612107S, BWM612108S, BWM612111S, BWM612115S, BWM612121S, BWM612122S, BWM612124S, BWM612125S, BWM612126S, BWM612140S, BWM612141S, BWM612146S, BWM612150S, BWM612155S, BWM612163S, BWM612165S, BWM612167S, BWM612169S, BWM612172S, BWM612173S, BWM612175S, BWM612203S, BWM612204S, BWM612205S, BWM612206S, BWM612207S, BWM612208S, BWM612209S, BWM612227S, BWM612229S, BWM612232S, BWM612233S, BWM612234S, BWM612235S, BWM612237S, BWM612238S, BWM612240S, BWM612242S, BWM612243S, BWM612245S, BWM612273S, BWM612275S, BWM612276S, BWM612277S, BWM612278S, BWM612279S, BWM612280S, BWM612281S, BWM612282S, BWM612283S, BWM612299S, BWM612307S, BWM612310S, BWM612311S, BWM612313S, BWM612315S, BWM612317S, BWM612318S, BWM612319S, BWM612320S, BWM612381S, BWM612388S, BWM612395S, BWM612404S, BWM612416S, BWM612417S, BWM612418S, BWM612422S, BWM612424S, BWM612427S, BWM612428S, BWM612431S, BWM612433S, BWM612434S, BWM612435S, BWM612436S, BWM612438S, BWM612439S, BWM612440S, BWM612441S, BWM612443S, BWM612444S, BWM612445S, BWM612467S, BWM612468S, BWM612472S, BWM612474S, BWM612475S, BWM612477S, BWM612480S, BWM612483S, BWM612485S, BWM612486S, BWM612531S, BWM612544S, BWM612546S, BWM612548S, BWM612549S, BWM612550S, BWM612573S, BWM612574S, BWM612575S, BWM612576S, BWM612577S, BWM612579S, BWM612609S, BWM612610S, BWM612611S, BWM612613S, BWM612614S, BWM612670S, BWM612671S, BWM612674S, BWM612675S, BWM612676S, BWM612677S, BWM612680S, BWM612698S, BWM612747S, BWM612748S, BWM612754S, BWM612755S, BWM612756S, BWM612757S, BWM612758S, BWM612761S, BWM612762S, BWM612763S, BWM612765S, BWM612770S, BWM612777S, BWM612779S, BWM612780S, BWM612781S, BWM612782S, BWM612783S, BWM612786S, BWM612787S, BWM612791S, BWM612792S, BWM612818S, BWM612819S, BWM612820S, BWM612822S, BWM612823S, BWM612824S, BWM612825S, BWM612826S, BWM612827S, BWM612828S, BWM612844S, BWM612851S, BWM612857S, BWM612863S, BWM612864S, BWM612867S, BWM612868S, BWM612876S, BWM612877S, BWM612878S, BWM612924S, BWM612925S, BWM612928S, BWM612931S, BWM612933S, BWM612981S, BWM612982S, BWM612987S, BWM612988S, BWM612989S, BWM613022S, BWM613023S, BWM613024S, BWM613025S, BWM613026S, BWM613027S, BWM613028S, BWM613029S, BWM613030S, BWM613031S, BWM613032S, BWM613148S, BWM613149S, BWM613150S, BWM613151S, BWM613152S, BWM613153S, BWM613154S, BWM613155S, BWM613156S, BWM613157S, BWM613158S, BWM613159S, BWM613160S, BWM613161S, BWM613163S, BWM613164S, BWM613172S, BWM613173S, BWM613174S, BWM613175S, BWM613176S, BWM613177S, BWM613178S, BWM613179S, BWM613180S, BWM613181S, BWM613190S, BWM613191S, BWM613192S, BWM613197S, BWM613198S, BWM613199S, BWM613248S, BWM613249S, BWM613250S, BWM613251S, BWM613263S, BWM613264S, BWM613265S, BWM613266S, BWM613267S, BWM613268S, BWM613269S, BWM613270S, BWM613271S, BWM613272S, BWM613273S, BWM613274S, BWM613275S, BWM613276S, BWM613277S, BWM613278S, BWM613279S, BWM613281S, BWM613309S, BWM613310S, BWM613311S, BWM613312S, BWM613313S, BWM613315S, BWM613316S, BWM613349S, BWM613350S, BWM613351S, BWM613352S, BWM613374S, BWM613375S, BWM613376S, BWM613377S, BWM613378S, BWM613389S, BWM613390S, BWM613391S, BWM613392S, BWM613473S, BWM613474S, BWM613479S, BWM613480S, BWM613481S, BWM613497S, BWM613499S, BWM613500S, BWM613501S, BWM613502S, BWM613504S, BWM613505S, BWM609293S, BWM609294S, BWM609295S, BWM609311S, BWM609312S, BWM609315S, BWM609317S, BWM609318S, BWM609319S, BWM609321S, BWM609333S, BWM609334S, BWM609335S, BWM609336S, BWM609337S, BWM609338S, BWM609340S, BWM609341S, BWM609565S, BWM609583S, BWM609589S, BWM609591S, BWM609592S, BWM609683S, BWM609691S, BWM609692S, BWM609709S, BWM609714S, BWM609733S, BWM609774S, BWM609775S, BWM609829S, BWM609830S, BWM609831S, BWM609832S, BWM609836S, BWM609854S, BWM609858S, BWM609859S, BWM609860S, BWM609868S, BWM609869S, BWM609876S, BWM609878S, BWM609879S, BWM609880S, BWM609881S, BWM609892S, BWM609895S, BWM609896S, BWM609903S, BWM609933S, BWM609938S, BWM609942S, BWM609946S, BWM609949S, BWM609954S, BWM609961S, BWM609963S, BWM609968S, BWM609969S, BWM609970S, BWM609973S, BWM609979S, BWM609980S, BWM609981S, BWM609982S, BWM609983S, BWM609984S, BWM609993S, BWM609999S, BWM610000S, BWM610001S, BWM610003S, BWM610007S, BWM610011S, BWM610012S, BWM610017S, BWM610030S, BWM610031S, BWM610032S, BWM610043S, BWM610044S, BWM610046S, BWM610047S, BWM610057S, BWM610063S, BWM610072S, BWM610073S, BWM610079S, BWM610080S, BWM610082S, BWM610088S, BWM610089S, BWM610090S, BWM610091S, BWM610092S, BWM610093S, BWM610094S, BWM610095S, BWM610096S, BWM610097S, BWM610106S, BWM610107S, BWM610110S, BWM610114S, BWM610117S, BWM610120S, BWM610157S, BWM610160S, BWM610167S, BWM610169S, BWM610170S, BWM610189S, BWM610193S, BWM610198S, BWM610199S, BWM610200S, BWM610201S, BWM610202S, BWM610203S, BWM610204S, BWM610212S, BWM610223S, BWM610224S, BWM610226S, BWM610230S, BWM610231S, BWM610235S, BWM610244S, BWM610246S, BWM610272S, BWM610294S, BWM610295S, BWM610300S, BWM610303S, BWM610312S, BWM610313S, BWM610314S, BWM610319S, BWM610343S, BWM610345S, BWM610346S, BWM610347S, BWM610348S, BWM610349S, BWM610350S, BWM610353S, BWM610417S, BWM610424S, BWM610427S, BWM610428S, BWM610430S, BWM610431S, BWM610432S, BWM610433S, BWM610438S, BWM610442S, BWM610443S, BWM610445S, BWM610446S, BWM610447S, BWM610448S, BWM610454S, BWM610457S, BWM610473S, BWM610479S, BWM610487S, BWM610491S, BWM610497S, BWM610499S, BWM610500S, BWM610501S, BWM610503S, BWM610504S, BWM610505S, BWM610517S, BWM610518S, BWM610523S, BWM610531S, BWM610532S, BWM610533S, BWM610535S, BWM610536S, BWM610540S, BWM610543S, BWM610544S, BWM610561S, BWM610562S, BWM610563S, BWM610568S, BWM610569S, BWM610570S, BWM610571S, BWM610572S, BWM610587S, BWM610588S, BWM610589S, BWM610594S, BWM610602S, BWM610603S, BWM610609S, BWM610611S, BWM610612S, BWM610613S, BWM610614S, BWM610615S, BWM610631S, BWM610632S, BWM610633S, BWM610636S, BWM610637S, BWM610638S, BWM610640S, BWM610642S, BWM610668S, BWM610670S, BWM610688S, BWM610690S, BWM610692S, BWM610693S, BWM610696S, BWM610697S, BWM610698S, BWM610699S, BWM610700S, BWM610770S, BWM610786S, BWM610787S, BWM610792S, BWM610793S, BWM610797S, BWM610800S, BWM610814S, BWM610815S, BWM610816S, BWM610817S, BWM610827S, BWM610828S, BWM610842S, BWM610843S, BWM610872S, BWM610873S, BWM610874S, BWM610875S, BWM610877S, BWM610878S, BWM610886S, BWM610887S, BWM610888S, BWM610889S, BWM610890S, BWM610906S, BWM610914S, BWM610915S, BWM610917S, BWM610925S, BWM610929S, BWM610930S, BWM610939S, BWM610941S, BWM610942S, BWM610943S, BWM610944S, BWM610945S, BWM610968S, BWM610969S, BWM610970S, BWM611015S, BWM611016S, BWM611018S, BWM611019S, BWM611039S, BWM611040S, BWM611041S, BWM611042S, BWM611044S, BWM611045S, BWM611056S, BWM611057S, BWM611058S, BWM611059S, BWM611060S, BWM611061S, BWM611062S, BWM611067S, BWM611068S, BWM611069S, BWM611101S, BWM611104S, BWM611105S, BWM611106S, BWM611116S, BWM611117S, BWM611130S, BWM611131S, BWM611139S, BWM611140S, BWM611142S, BWM611144S, BWM611146S, BWM611162S, BWM611164S, BWM611165S, BWM611166S, BWM611167S, BWM611168S, BWM611169S, BWM611170S, BWM611171S, BWM611174S, BWM611175S, BWM611176S, BWM611177S, BWM611181S, BWM611185S, BWM611186S, BWM611187S, BWM611188S, BWM611194S, BWM611195S, BWM611196S, BWM611201S, BWM611203S, BWM611205S, BWM611206S, BWM611207S, BWM611217S, BWM611221S, BWM611228S, BWM611265S, BWM611266S, BWM611268S, BWM611269S, BWM611270S, BWM611271S, BWM611272S, BWM611273S, BWM611274S, BWM611275S, BWM611280S, BWM611281S, BWM611282S, BWM611288S, BWM611300S, BWM611303S, BWM611311S, BWM611312S, BWM611318S, BWM611319S, BWM611320S, BWM611321S, BWM611322S, BWM611341S, BWM611342S, BWM611343S, BWM611346S, BWM611347S, BWM611348S, BWM611358S, BWM611359S, BWM611360S, BWM611374S, BWM611375S, BWM611376S, BWM611381S, BWM611382S, BWM611383S, BWM611389S, BWM611390S, BWM611391S, BWM611396S, BWM611432S, BWM611434S, BWM611437S, BWM611438S, BWM611439S, BWM611440S, BWM611441S, BWM611442S, BWM611443S, BWM611444S, BWM611445S, BWM611446S, BWM611452S, BWM611453S, BWM611454S, BWM611455S, BWM611456S, BWM611457S, BWM611484S, BWM611485S, BWM611486S, BWM611487S, BWM611488S, BWM611491S, BWM611492S, BWM611493S, BWM611494S, BWM611495S, BWM611496S, BWM611520S, BWM611521S, BWM611522S, BWM611523S, BWM611530S, BWM611539S, BWM611540S, BWM611544S, BWM611546S, BWM611547S, BWM611548S, BWM611549S, BWM611551S, BWM611557S, BWM611559S, BWM611560S, BWM611562S, BWM611586S, BWM611640S, BWM611641S, BWM611642S, BWM611643S, BWM611645S, BWM611646S, BWM611649S, BWM611650S, BWM611651S, BWM611652S, BWM611656S, BWM611657S, BWM611658S, BWM611660S, BWM611661S, BWM611662S, BWM611669S, BWM611670S, BWM611671S, BWM611672S, BWM611673S, BWM611674S, BWM611675S, BWM611677S, BWM611678S, BWM611679S, BWM611680S, BWM611691S, BWM611692S, BWM611694S, BWM611695S, BWM611696S, BWM611698S, BWM611699S, BWM611700S, BWM611701S, BWM611702S, BWM611704S, BWM611706S, BWM611707S, BWM611708S, BWM611709S, BWM611710S, BWM611711S, BWM611712S, BWM611713S, BWM611714S, BWM611715S, BWM611719S, BWM611728S, BWM611729S, BWM611730S, BWM611731S, BWM611732S, BWM611736S, BWM611737S, BWM611738S, BWM611739S, BWM611750S, BWM611751S, BWM611752S, BWM611753S, BWM611765S, BWM611770S, BWM611771S, BWM611772S, BWM611773S, BWM611774S, BWM611775S, BWM611776S, BWM611777S, BWM611778S, BWM611779S, BWM611818S, BWM611819S, BWM611820S, BWM611821S, BWM611826S, BWM611827S, BWM611828S, BWM611830S, BWM611831S, BWM611886S, BWM611887S, BWM611889S, BWM611890S, BWM611895S, BWM611896S, BWM611921S, BWM611922S, BWM611923S, BWM611924S, BWM611925S, BWM611970S, BWM611972S, BWM611973S, BWM611976S, BWM611977S, BWM611986S, BWM611987S, BWM611990S, BWM611993S, BWM611994S, BWM611995S, BWM611996S, BWM611997S, BWM611998S, BWM611999S, BWM612000S, BWM612001S, BWM612002S, BWM612003S, BWM612004S, BWM612040S, BWM612043S, BWM612127S, BWM612128S, BWM612129S, BWM612130S, BWM612133S, BWM612168S, BWM612170S, BWM612171S, BWM612174S, BWM612177S, BWM612178S, BWM612179S, BWM612180S, BWM612244S, BWM612274S, BWM612284S, BWM612285S, BWM612286S, BWM612287S, BWM612288S, BWM612289S, BWM612290S, BWM612291S, BWM612292S, BWM612293S, BWM612294S, BWM612295S, BWM612296S, BWM612297S, BWM612298S, BWM612301S, BWM612304S, BWM612316S, BWM612321S, BWM612322S, BWM612323S, BWM612324S, BWM612325S, BWM612326S, BWM612334S, BWM612335S, BWM612336S, BWM612337S, BWM612338S, BWM612344S, BWM612346S, BWM612347S, BWM612348S, BWM612364S, BWM612365S, BWM612367S, BWM612368S, BWM612369S, BWM612370S, BWM612389S, BWM612437S, BWM612442S, BWM612448S, BWM612453S, BWM612459S, BWM612460S, BWM612461S, BWM612464S, BWM612465S, BWM612469S, BWM612471S, BWM612473S, BWM612479S, BWM612481S, BWM612482S, BWM612484S, BWM612489S, BWM612490S, BWM612495S, BWM612496S, BWM612535S, BWM612540S, BWM612541S, BWM612542S, BWM612543S, BWM612559S, BWM612562S, BWM612563S, BWM612564S, BWM612565S, BWM612566S, BWM612567S, BWM612569S, BWM612571S, BWM612578S, BWM612581S, BWM612582S, BWM612583S, BWM612584S, BWM612586S, BWM612587S, BWM612588S, BWM612589S, BWM612590S, BWM612591S, BWM612592S, BWM612594S, BWM612597S, BWM612598S, BWM612599S, BWM612600S, BWM612602S, BWM612603S, BWM612604S, BWM612605S, BWM612606S, BWM612607S, BWM612608S, BWM612612S, BWM612615S, BWM612616S, BWM612629S, BWM612630S, BWM612631S, BWM612635S, BWM612638S, BWM612654S, BWM612656S, BWM612657S, BWM612658S, BWM612660S, BWM612662S, BWM612665S, BWM612666S, BWM612667S, BWM612668S, BWM612669S, BWM612678S, BWM612679S, BWM612682S, BWM612684S, BWM612685S, BWM612690S, BWM612692S, BWM612694S, BWM612695S, BWM612699S, BWM612700S, BWM612701S, BWM612724S, BWM612725S, BWM612727S, BWM612729S, BWM612732S, BWM612733S, BWM612736S, BWM612739S, BWM612742S, BWM612744S, BWM612750S, BWM612751S, BWM612752S, BWM612753S, BWM612800S, BWM612801S, BWM612829S, BWM612830S, BWM612831S, BWM612832S, BWM612833S, BWM612834S, BWM612835S, BWM612836S, BWM612850S, BWM612954S, BWM612955S, BWM612957S, BWM612958S, BWM612959S, BWM612960S, BWM612961S, BWM612962S, BWM612963S, BWM612964S, BWM612965S, BWM612966S, BWM612967S, BWM612968S, BWM612969S, BWM612970S, BWM612971S, BWM612992S, BWM612993S, BWM613000S, BWM613001S, BWM613049S, BWM613050S, BWM613073S, BWM613086S, BWM613087S, BWM613088S, BWM613089S, BWM613090S, BWM613091S, BWM613092S, BWM613093S, BWM613094S, BWM613095S, BWM613096S, BWM613097S, BWM613098S, BWM613099S, BWM613100S, BWM613101S, BWM613102S, BWM613104S, BWM613105S, BWM613106S, BWM613107S, BWM613108S, BWM613109S, BWM613110S, BWM613111S, BWM613112S, BWM613113S, BWM613114S, BWM613115S, BWM613116S, BWM613117S, BWM613118S, BWM613165S, BWM613168S, BWM613169S, BWM613170S, BWM613171S, BWM613182S, BWM613183S, BWM613188S, BWM613189S, BWM613193S, BWM613194S, BWM613195S, BWM613196S, BWM613200S, BWM613201S, BWM613204S, BWM613205S, BWM613206S, BWM613207S, BWM613208S, BWM613209S, BWM613214S, BWM613218S, BWM613219S, BWM613220S, BWM613221S, BWM613222S, BWM613223S, BWM613225S, BWM613226S, BWM613227S, BWM613228S, BWM613229S, BWM613230S, BWM613241S, BWM613242S, BWM613262S, BWM613282S, BWM613283S, BWM613284S, BWM613285S, BWM613286S, BWM613287S, BWM613288S, BWM613289S, BWM613290S, BWM613293S, BWM613294S, BWM613295S, BWM613296S, BWM613297S, BWM613298S, BWM613299S, BWM613300S, BWM613301S, BWM613302S, BWM613303S, BWM613304S, BWM613306S, BWM613308S, BWM613319S, BWM613320S, BWM613321S, BWM613332S, BWM613333S, BWM613334S, BWM613335S, BWM613336S, BWM613337S, BWM613338S, BWM613339S, BWM613340S, BWM613341S, BWM613342S, BWM613343S, BWM613344S, BWM613345S, BWM613346S, BWM613347S, BWM613348S, BWM613353S, BWM613354S, BWM613355S, BWM613356S, BWM613357S, BWM613358S, BWM613359S, BWM613360S, BWM613361S, BWM613362S, BWM613363S, BWM613364S, BWM613365S, BWM613366S, BWM613367S, BWM613368S, BWM613369S, BWM613370S, BWM613371S, BWM613372S, BWM613373S, BWM613379S, BWM613380S, BWM613381S, BWM613382S, BWM613388S, BWM613396S, BWM613397S, BWM613398S, BWM613399S, BWM613400S, BWM613401S, BWM613402S, BWM613403S, BWM613404S, BWM613405S, BWM613406S, BWM613407S, BWM608621S, BWM609674S, BWM609676S, BWM610608S, BWM610619S, BWM610780S, BWM610802S, BWM609642S, BWM610281S, BWM610691S, BWM610715S, BWM610723S, BWM610724S, BWM609637S, BWM609638S, BWM609641S, BWM609643S, BWM609687S, BWM609688S, BWM609689S, BWM609704S, BWM609705S, BWM609706S, BWM609710S, BWM609711S, BWM609717S, BWM609718S, BWM609720S, BWM609721S, BWM609723S, BWM609725S, BWM609726S, BWM609727S, BWM609728S, BWM609732S, BWM609734S, BWM609735S, BWM609739S, BWM609740S, BWM610379S, BWM610639S, BWM610754S, BWM610755S, BWM610757S, BWM610767S, BWM610807S, BWM610837S, BWM611593S, BWM611575S, BWM611576S, BWM612738S, BWM611066S, BWM611092S, BWM611230S, BWM611234S, BWM611386S, BWM611387S, BWM611541S, BWM611616S, BWM611862S, BWM611863S, BWM612042S, BWM612051S, BWM612216S, BWM612217S, BWM612218S, BWM612221S, BWM613210S, BWM612795S, BWM612866S, BWM612880S, BWM613461S, BWM613462S, BWM613463S, BWM613464S, BWM613466S, BWM613467S, BWM613468S, BWM613484S, BWM613485S, BWM613486S, BWM613487S, BWM613488S, BWM613489S, BWM613490S, BWM613491S, BWM613492S, BWM613493S, BWM612191S
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Evera™ S VR DVBC3D1Medtronic, Inc. · Implantable cardioverter defibrillator (non-CRT)1 recall8 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 92328).
- Z-1704-2023 · Open, ClassifiedICD-DR DDBB1D1 EVERA XT IS1/DF1 US, Model Number DDBB1D1; Implantable Cardioverter Defibrillators
- Z-1705-2023 · Open, ClassifiedICD-DR DDBB1D4 EVERA XT IS1/DF4 US, Model Number DDBB1D4; Implantable Cardioverter Defibrillators
- Z-1709-2023 · Open, ClassifiedICD-DR DDBC3D1 EVERA S IS1/DF1 GLOB, Model Number DDBC3D1; Implantable Cardioverter Defibrillators
- Z-1710-2023 · Open, ClassifiedICD-DR DDBC3D4 EVERA S IS1/DF4 GLOB, Model Number DDBC3D4; Implantable Cardioverter Defibrillators
- Z-1711-2023 · Open, ClassifiedICD-DR DDMB1D1 EVERA MRI XT US DF1, Model Number DDMB1D1; Implantable Cardioverter Defibrillators
- Z-1712-2023 · Open, ClassifiedICD-DR DDMB1D4 EVERA MRI XT IS-1/DF4 US, Model Number DDMB1D4; Implantable Cardioverter Defibrillators
- Z-1715-2023 · Open, ClassifiedICD-DR DDMC3D1 EVERA MRI S OUS/US DF1, Model Number DDMC3D1; Implantable Cardioverter Defibrillators
- Z-1716-2023 · Open, ClassifiedICD-DR DDMC3D4 EVERA MRI S IS-1/DF4 GLOB, Model Number DDMC3D4; Implantable Cardioverter Defibrillators
- Z-1717-2023 · Open, ClassifiedICD-DR DDMD3D1 PRIMO MRI, Model Number DDMD3D1; Implantable Cardioverter Defibrillators
- Z-1718-2023 · Open, ClassifiedICD-DR DDMD3D4 PRIMO MRI, Model Number DDMD3D4; Implantable Cardioverter Defibrillators
- Z-1721-2023 · Open, ClassifiedICD COBALT XT DR MRI IS1 DF1, Model Number DDPA2D1; Implantable Cardioverter Defibrillators
- Z-1723-2023 · Open, ClassifiedICD COBALT XT DR MRI IS1 DF4, Model Number DDPA2D4; Implantable Cardioverter Defibrillators
- Z-1725-2023 · Open, ClassifiedICD COBALT DR MRI IS1 DF1, Model Number DDPB3D1; Implantable Cardioverter Defibrillators
- Z-1726-2023 · Open, ClassifiedICD COBALT DR MRI IS1 DF4, Model Number DDPB3D4; Implantable Cardioverter Defibrillators
- Z-1727-2023 · Open, ClassifiedICD CROME DR MRI IS1 DF1, Model Number DDPC3D1; Implantable Cardioverter Defibrillators
- Z-1728-2023 · Open, ClassifiedICD CROME DR MRI IS1 DF4, Model Number DDPC3D4; Implantable Cardioverter Defibrillators
- Z-1729-2023 · Open, ClassifiedCRT-D DTBA1D1 VIVA XT IS1/DF1 US, Model Number DTBA1D1; Implantable Cardioverter Defibrillators
- Z-1731-2023 · Open, ClassifiedCRT-D DTBA1D4 VIVA XT IS1/DF4 US, Model Number DTBA1D4; Implantable Cardioverter Defibrillators
- Z-1732-2023 · Open, ClassifiedCRT-D DTBA1Q1 VIVA QUAD XT IS4/DF1 US, Model Number DTBA1Q1; Implantable Cardioverter Defibrillators
- Z-1733-2023 · Open, ClassifiedCRT-D DTBA1QQ VIVA QUAD XT IS4/DF4 US, Model Number DTBA1QQ; Implantable Cardioverter Defibrillators
- Z-1741-2023 · Open, ClassifiedCRT-D DTBB1D1 VIVA S IS1/DF1 US, Model Number DTBB1D1; Implantable Cardioverter Defibrillators
- Z-1742-2023 · Open, ClassifiedCRT-D DTBB1D4 VIVA S IS1/DF4 US, Model Number DTBB1D4; Implantable Cardioverter Defibrillators
- Z-1743-2023 · Open, ClassifiedCRT-D DTBB1Q1 VIVA QUAD S IS4/DF1 US, Model Number DTBB1Q1; Implantable Cardioverter Defibrillators
- Z-1744-2023 · Open, ClassifiedCRT-D DTBB1QQ VIVA QUAD S IS4/DF4 US, Model Number DTBB1QQ; Implantable Cardioverter Defibrillators
- Z-1753-2023 · Open, ClassifiedCRTD DTMA1D1 CLARIA MRI US DF1, Model Number DTMA1D1; Implantable Cardioverter Defibrillators
- Z-1754-2023 · Open, ClassifiedCRTD DTMA1D4 CLARIA MRI US DF4, Model Number DTMA1D4; Implantable Cardioverter Defibrillators
- Z-1755-2023 · Open, ClassifiedCRTD DTMA1Q1 CLARIA MRI QUAD US DF1, Model Number DTMA1Q1; Implantable Cardioverter Defibrillators
- Z-1756-2023 · Open, ClassifiedCRTD DTMA1QQ CLARIA MRI QUAD US DF4, Model Number DTMA1QQ; Implantable Cardioverter Defibrillators
- Z-1761-2023 · Open, ClassifiedCRTD DTMB1D1 AMPLIA MRI US DF1, Model Number DTMB1D1; Implantable Cardioverter Defibrillators
- Z-1762-2023 · Open, ClassifiedCRTD DTMB1D4 AMPLIA MRI US DF4, Model Number DTMB1D4; Implantable Cardioverter Defibrillators
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page