Crome VR MRI SureScan recallZ-1823-2023 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · initiated May 10, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
FDA root cause category: Under Investigation by firm
What the company is doing
Medtronic issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to its consignees on 05/10/2023 by UPS 2-day delivery or email. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. The notice explained the problem and the risk and requested the following patient management recommendations: - Prophylactic device replacement is NOT recommended. - Program all HV therapy pathways B>AX in all therapy zones to minimize the risk for this issue. - Prioritize reprogramming patients who have both a history of HV therapy and Rx1 programmed AX>B. - For remaining patients with AX>B programming in any HV therapy sequence, schedule (with appropriate discretion) the next follow-up for in-clinic reprogramming to minimize potential for reduced- or no-energy HV therapies to occur. - Contact Medtronic Technical Services (1-800-929-4043) or your local representative if one of the following is observed as these may be an indication of either a device or lead-related issue: o Reduced- or no-energy HV therapy is displayed in Episode Text (regardless of programmed pathway) o A persistent drop of approximately 50% in RA, RV and LV pacing lead impedance measurements as this may be an indication of increased potential for a future reduced- or no-energy therapy. Update 10/10/2023: Beginning 09/27/2023, Medtronic issued communication via letter to consignees regarding the availability of the revised IFU and SmartSync Software. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. Consignees are asked to complete and return a confirmation certificate (or equivalent record) confirming they have received the notification.
Recall details
- Recall number
- Z-1823-2023
- Status
- Open, Classified
- Initiated by the company
- May 10, 2023
- Posted by the FDA
- Jun 16, 2023
- Recalling firm
- Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
- Quantity
- 2031 units
- Product code
- LWS
- FDA submissions
- P980016S713
- Distribution
- US Nationwide - Worldwide Distribution
Product description
ICD CROME VR MRI DF4, Model Number DVPC3D4; Implantable Cardioverter Defibrillators
Code information (models, lots, serial numbers)
Show the full code information (24k characters)
GTIN 00763000178550. Lot Serial Numbers: RSH600068S, RSH600173S, RSH600239S, RSH601126S, RSH601127S, RSH601883S, RSH601888S, RSH601444S, RSH601463S, RSH601537S, RSH601538S, RSH600064S, RSH600099S, RSH600174S, RSH600175S, RSH600228S, RSH600275S, RSH600276S, RSH600889S, RSH600903S, RSH601415S, RSH601522S, RSH601873S, RSH600071S, RSH600906S, RSH600922S, RSH601034S, RSH601041S, RSH601123S, RSH600007S, RSH600082S, RSH600110S, RSH601765S, RSH601894S, RSH600236S, RSH600264S, RSH601531S, RSH600057S, RSH600124S, RSH600181S, RSH600211S, RSH600827S, RSH600887S, RSH601024S, RSH601036S, RSH601366S, RSH600902S, RSH601519S, RSH601520S, RSH601751S, RSH601753S, RSH601754S, RSH601805S, RSH600814S, RSH601139S, RSH601355S, RSH601407S, RSH600147S, RSH600101S, RSH601443S, RSH601460S, RSH601462S, RSH601535S, RSH601536S, RSH601539S, RSH601542S, RSH600041S, RSH600045S, RSH600055S, RSH600066S, RSH600067S, RSH600086S, RSH600097S, RSH600102S, RSH600103S, RSH600108S, RSH600109S, RSH600117S, RSH600126S, RSH600130S, RSH600142S, RSH600145S, RSH600157S, RSH600188S, RSH600199S, RSH600204S, RSH600207S, RSH600222S, RSH600225S, RSH600240S, RSH600829S, RSH601023S, RSH601058S, RSH600169S, RSH600171S, RSH600172S, RSH600815S, RSH601129S, RSH600238S, RSH601431S, RSH601540S, RSH601761S, RSH600135S, RSH600219S, RSH600821S, RSH600849S, RSH600921S, RSH601749S, RSH600132S, RSH600262S, RSH600911S, RSH601030S, RSH601742S, RSH600205S, RSH601728S, RSH601730S, RSH601796S, RSH601863S, RSH600170S, RSH600820S, RSH601914S, RSH601915S, RSH601922S, RSH601923S, RSH600196S, RSH601423S, RSH601527S, RSH601755S, RSH601756S, RSH600046S, RSH600047S, RSH600048S, RSH600049S, RSH600056S, RSH600065S, RSH600074S, RSH600143S, RSH600153S, RSH600158S, RSH601026S, RSH601027S, RSH601098S, RSH601105S, RSH600061S, RSH600168S, RSH601546S, RSH601547S, RSH601548S, RSH601554S, RSH601564S, RSH601566S, RSH601750S, RSH600080S, RSH600183S, RSH601130S, RSH601107S, RSH601051S, RSH601925S, RSH601927S, RSH600165S, RSH600092S, RSH600150S, RSH600221S, RSH600093S, RSH600063S, RSH600141S, RSH600197S, RSH600282S, RSH600824S, RSH600195S, RSH600281S, RSH601523S, RSH601740S, RSH600078S, RSH600177S, RSH600823S, RSH601031S, RSH600259S, RSH601521S, RSH600830S, RSH601386S, RSH600901S, RSH601106S, RSH601131S, RSH600161S, RSH601022S, RSH601735S, RSH601416S, RSH601417S, RSH601524S, RSH601525S, RSH601758S, RSH600036S, RSH600070S, RSH600091S, RSH600105S, RSH600107S, RSH600122S, RSH600123S, RSH600128S, RSH600129S, RSH600140S, RSH600148S, RSH600152S, RSH600159S, RSH600164S, RSH601025S, RSH601039S, RSH601118S, RSH601121S, RSH601122S, RSH601766S, RSH601532S, RSH601726S, RSH601771S, RSH601772S, RSH601802S, RSH601115S, RSH600184S, RSH600151S, RSH600833S, RSH601569S, RSH601395S, RSH600162S, RSH600506S, RSH601109S, RSH601110S, RSH601113S, RSH601138S, RSH601354S, RSH600111S, RSH600208S, RSH600825S, RSH600831S, RSH600832S, RSH600879S, RSH600895S, RSH600908S, RSH601140S, RSH601060S, RSH601061S, RSH601052S, RSH601567S, RSH601744S, RSH600069S, RSH600062S, RSH601108S, RSH600119S, RSH600120S, RSH600125S, RSH600149S, RSH601420S, RSH600054S, RSH601369S, RSH601425S, RSH601517S, RSH601533S, RSH601534S, RSH601778S, RSH601895S, RSH601897S, RSH600058S, RSH600280S, RSH600822S, RSH601033S, RSH601764S, RSH600076S, RSH600084S, RSH600112S, RSH600113S, RSH601401S, RSH601528S, RSH600060S, RSH600828S, RSH600182S, RSH600230S, RSH600231S, RSH600253S, RSH600257S, RSH601035S, RSH600072S, RSH600258S, RSH600882S, RSH601132S, RSH600059S, RSH601541S, RSH601543S, RSH601544S, RSH600138S, RSH600920S, RSH601124S, RSH601125S, RSH600053S, RSH600176S, RSH600166S, RSH600819S, RSH600194S, RSH600039S, RSH600818S, RSH601032S, RSH601553S, RSH600134S, RSH600215S, RSH600260S, RSH601028S, RSH600094S, RSH600100S, RSH600114S, RSH600139S, RSH600826S, RSH600037S, RSH601770S, RSH601529S, RSH601530S, RSH601545S, RSH601549S, RSH601550S, RSH601552S, RSH601746S, RSH601748S, RSH600277S, RSH600912S, RSH601133S, RSH601134S, RSH601135S, RSH601136S, RSH601137S, RSH601388S, RSH601389S, RSH600043S, RSH601461S, RSH600278S, RSH600891S, RSH601410S, RSH601807S, RSH601808S, RSH601810S, RSH601893S, RSH601813S, RSH601819S; GTIN 00763000178567, Lot Serial Numbers: RSH600544S, RSH602201S, RSH602205S, RSH602187S, RSH602191S, RSH602043S, RSH602058S, RSH602229S, RSH602230S, RSH602008S, RSH602010S, RSH602059S, RSH601390S, RSH601312S, RSH601994S, RSH602288S, RSH602472S, RSH601278S, RSH602470S, RSH602305S, RSH601371S, RSH601372S, RSH600644S, RSH600640S, RSH600793S, RSH601383S, RSH601716S, RSH602301S, RSH602159S, RSH601868S, RSH600589S, RSH600762S, RSH600639S, RSH600651S, RSH600659S, RSH600692S, RSH600694S, RSH600766S, RSH600975S, RSH600648S, RSH600763S, RSH600929S, RSH600930S, RSH600948S, RSH600959S, RSH601245S, RSH601303S, RSH601373S, RSH601398S, RSH601247S, RSH601253S, RSH601255S, RSH601292S, RSH601402S, RSH601404S, RSH601405S, RSH601760S, RSH600864S, RSH600942S, RSH601231S, RSH601809S, RSH601850S, RSH602066S, RSH602104S, RSH601811S, RSH600758S, RSH600430S, RSH600574S, RSH600637S, RSH600667S, RSH600714S, RSH600722S, RSH600726S, RSH600742S, RSH600995S, RSH601089S, RSH601100S, RSH601104S, RSH601116S, RSH601299S, RSH601300S, RSH601301S, RSH601309S, RSH601376S, RSH601403S, RSH601752S, RSH601784S, RSH601826S, RSH601827S, RSH601975S, RSH602126S, RSH602131S, RSH602284S, RSH602302S, RSH602317S, RSH600801S, RSH601096S, RSH601304S, RSH601305S, RSH601306S, RSH601307S, RSH600654S, RSH600666S, RSH600817S, RSH600642S, RSH600951S, RSH601379S, RSH601556S, RSH600782S, RSH600634S, RSH600641S, RSH600675S, RSH600685S, RSH600689S, RSH600854S, RSH600932S, RSH600935S, RSH601072S, RSH601099S, RSH601147S, RSH601235S, RSH601298S, RSH601367S, RSH601378S, RSH601380S, RSH601381S, RSH601585S, RSH601817S, RSH601831S, RSH601832S, RSH601833S, RSH601834S, RSH601835S, RSH601839S, RSH601842S, RSH602021S, RSH602123S, RSH602138S, RSH602139S, RSH602141S, RSH600030S, RSH600936S, RSH602455S, RSH601717S, RSH601731S, RSH602100S, RSH600673S, RSH600710S, RSH600788S, RSH600806S, RSH601409S, RSH601964S, RSH601004S, RSH601854S, RSH600553S, RSH600653S, RSH600787S, RSH600800S, RSH601269S, RSH600676S, RSH601587S, RSH600283S, RSH600962S, RSH601148S, RSH601588S, RSH601259S, RSH600406S, RSH600412S, RSH600414S, RSH600415S, RSH600418S, RSH600190S, RSH600270S, RSH600359S, RSH600452S, RSH600537S, RSH600548S, RSH600601S, RSH600650S, RSH600664S, RSH600993S, RSH601221S, RSH601412S, RSH601700S, RSH601705S, RSH600401S, RSH601078S, RSH600583S, RSH601319S, RSH600584S, RSH600645S, RSH600871S, RSH601070S, RSH601242S, RSH601374S, RSH601421S, RSH600646S, RSH600631S, RSH600647S, RSH600671S, RSH601088S, RSH601090S, RSH601097S, RSH601101S, RSH601228S, RSH601314S, RSH601316S, RSH601362S, RSH601737S, RSH600636S, RSH601009S, RSH601011S, RSH601738S, RSH601989S, RSH602112S, RSH600635S, RSH600655S, RSH600656S, RSH600657S, RSH600658S, RSH600923S, RSH601114S, RSH601295S, RSH601297S, RSH601321S, RSH600674S, RSH600698S, RSH601313S, RSH600210S, RSH600396S, RSH600812S, RSH601174S, RSH601794S, RSH600979S, RSH600983S, RSH601843S, RSH602073S, RSH602315S, RSH602113S, RSH600633S, RSH601736S, RSH600377S, RSH600378S, RSH601057S, RSH601859S, RSH602077S, RSH602081S, RSH602121S, RSH602122S, RSH602179S, RSH600643S, RSH601272S, RSH602501S, RSH602502S, RSH600858S, RSH601014S, RSH601359S, RSH601555S, RSH601575S, RSH601612S, RSH601745S, RSH601795S, RSH601818S, RSH601862S, RSH601870S, RSH601877S, RSH601878S, RSH602040S, RSH602046S, RSH602163S, RSH602177S, RSH602183S, RSH602194S, RSH602282S, RSH602283S, RSH602333S, RSH602471S, RSH601062S, RSH601217S, RSH601233S, RSH601234S, RSH601384S, RSH601571S, RSH601578S, RSH601579S, RSH601580S, RSH602064S, RSH602393S, RSH602196S, RSH602197S, RSH602209S, RSH602210S, RSH602211S, RSH602212S, RSH602213S, RSH602214S, RSH602215S, RSH602216S, RSH602217S, RSH602224S, RSH602226S, RSH602232S, RSH602233S, RSH602234S, RSH602235S, RSH602236S, RSH602237S, RSH602238S, RSH602239S, RSH602240S, RSH602241S, RSH602242S, RSH602243S, RSH602244S, RSH602245S, RSH602246S, RSH602247S, RSH602248S, RSH602249S, RSH602250S, RSH602251S, RSH602252S, RSH602253S, RSH602254S, RSH602255S, RSH602256S, RSH602257S, RSH602258S, RSH602259S, RSH602260S, RSH602261S, RSH602262S, RSH602263S, RSH602264S, RSH602265S, RSH602266S, RSH602267S, RSH602268S, RSH602269S, RSH602270S, RSH602271S, RSH602272S, RSH602273S, RSH602274S, RSH602275S, RSH602276S, RSH602277S, RSH602278S, RSH600006S, RSH600357S, RSH600373S, RSH600374S, RSH600419S, RSH600426S, RSH601422S, RSH601591S, RSH601592S, RSH601593S, RSH601594S, RSH601762S, RSH601828S, RSH601829S, RSH602025S, RSH602096S, RSH602097S, RSH602099S, RSH602102S, RSH602109S, RSH602129S, RSH602029S, RSH602039S, RSH602280S, RSH602281S, RSH602320S, RSH602321S, RSH602322S, RSH602390S, RSH602391S, RSH602063S, RSH600383S, RSH601317S, RSH601345S, RSH601346S, RSH601385S, RSH601392S, RSH601393S, RSH601394S, RSH601582S, RSH601583S, RSH601825S, RSH602089S, RSH602125S, RSH602304S, RSH602308S, RSH602314S, RSH602404S, RSH602415S, RSH602439S, RSH601322S, RSH601268S, RSH601763S, RSH602107S, RSH602110S, RSH602117S, RSH602142S, RSH602172S, RSH602173S, RSH602175S, RSH602313S, RSH601729S, RSH601776S, RSH601782S, RSH601847S, RSH601852S, RSH601858S, RSH602072S, RSH602108S, RSH602130S, RSH602169S, RSH602170S, RSH602189S, RSH602300S, RSH601993S, RSH602448S, RSH602451S, RSH602464S, RSH602465S, RSH600797S, RSH600619S, RSH600632S, RSH600638S, RSH600683S, RSH600773S, RSH600934S, RSH600996S, RSH601093S, RSH601175S, RSH601177S, RSH601220S, RSH601375S, RSH601576S, RSH601581S, RSH601613S, RSH601741S, RSH601747S, RSH601821S, RSH601848S, RSH601966S, RSH601970S, RSH602083S, RSH602124S, RSH602165S, RSH602166S, RSH602285S, RSH602342S, RSH602405S, RSH602434S, RSH602453S, RSH601270S, RSH601602S, RSH601603S, RSH601604S, RSH601605S, RSH601606S, RSH601607S, RSH601608S, RSH601609S, RSH601610S, RSH601437S, RSH601440S, RSH601441S, RSH601466S, RSH601467S, RSH601468S, RSH601469S, RSH601470S, RSH601471S, RSH601472S, RSH601473S, RSH601474S, RSH601475S, RSH601476S, RSH601477S, RSH601478S, RSH601479S, RSH601480S, RSH601481S, RSH601482S, RSH601483S, RSH601484S, RSH601485S, RSH601486S, RSH601487S, RSH601488S, RSH601489S, RSH601490S, RSH601491S, RSH601492S, RSH601493S, RSH601494S, RSH601495S, RSH601496S, RSH601497S, RSH601498S, RSH601499S, RSH601500S, RSH601501S, RSH601502S, RSH601503S, RSH601504S, RSH601505S, RSH601506S, RSH601507S, RSH601508S, RSH601509S, RSH601510S, RSH601511S, RSH601512S, RSH601223S, RSH600024S, RSH600033S, RSH600497S, RSH600503S, RSH600721S, RSH600861S, RSH601018S, RSH600382S, RSH600572S, RSH601325S, RSH600202S, RSH600273S, RSH600279S, RSH600413S, RSH600454S, RSH600542S, RSH600557S, RSH600753S, RSH600804S, RSH600839S, RSH600227S, RSH600245S, RSH600355S, RSH600398S, RSH600463S, RSH600530S, RSH600538S, RSH600554S, RSH600610S, RSH600723S, RSH600865S, RSH601172S, RSH602339S, RSH600388S, RSH600842S, RSH600012S, RSH600453S, RSH600606S, RSH600899S, RSH601280S, RSH601734S, RSH600021S, RSH600023S, RSH600404S, RSH600408S, RSH600534S, RSH600543S, RSH600564S, RSH600596S, RSH600743S, RSH600795S, RSH600857S, RSH600881S, RSH600913S, RSH600027S, RSH600011S, RSH600028S, RSH600032S, RSH600200S, RSH600242S, RSH600324S, RSH600385S, RSH600456S, RSH600465S, RSH600531S, RSH600546S, RSH600547S, RSH600561S, RSH600599S, RSH600615S, RSH600807S, RSH600813S, RSH600838S, RSH600868S, RSH600957S, RSH601005S, RSH601017S, RSH601037S, RSH601143S, RSH601145S, RSH601163S, RSH601164S, RSH601166S, RSH601219S, RSH601328S, RSH601358S, RSH601361S, RSH601430S, RSH601595S, RSH601599S, RSH601601S, RSH601611S, RSH601712S, RSH601713S, RSH601851S, RSH601856S, RSH602085S, RSH600217S, RSH600459S, RSH600545S, RSH600551S, RSH600607S, RSH600668S, RSH600896S, RSH601360S, RSH600690S, RSH600696S, RSH600697S, RSH600852S, RSH600853S, RSH600931S, RSH600933S, RSH600937S, RSH600944S, RSH601111S, RSH601112S, RSH601150S, RSH601244S, RSH601293S, RSH601408S, RSH602178S, RSH600035S, RSH600233S, RSH600354S, RSH600402S, RSH600428S, RSH600455S, RSH600500S, RSH600729S, RSH600738S, RSH600786S, RSH600834S, RSH600840S, RSH600916S, RSH601169S, RSH601178S, RSH602330S, RSH602119S, RSH600025S, RSH600566S, RSH601015S, RSH600622S, RSH600214S, RSH600540S, RSH600568S, RSH600784S, RSH601363S, RSH601225S, RSH600627S, RSH600859S, RSH600877S, RSH600890S, RSH600953S, RSH601120S, RSH601559S, RSH601589S, RSH600883S, RSH600886S, RSH600888S, RSH600909S, RSH600947S, RSH600958S, RSH601012S, RSH601016S, RSH601029S, RSH601042S, RSH601043S, RSH601048S, RSH601081S, RSH601082S, RSH601170S, RSH601176S, RSH601327S, RSH601364S, RSH601429S, RSH601570S, RSH601573S, RSH601691S, RSH601692S, RSH602192S, RSH600034S, RSH600212S, RSH600246S, RSH600247S, RSH600248S, RSH600272S, RSH600362S, RSH600363S, RSH600368S, RSH600371S, RSH600376S, RSH600379S, RSH600390S, RSH600400S, RSH600451S, RSH600457S, RSH600458S, RSH600462S, RSH600464S, RSH600489S, RSH600505S, RSH600529S, RSH600532S, RSH600535S, RSH600536S, RSH600541S, RSH600563S, RSH600580S, RSH600590S, RSH600616S, RSH600618S, RSH600625S, RSH600663S, RSH600670S, RSH600732S, RSH600759S, RSH600761S, RSH600785S, RSH600791S, RSH600837S, RSH600850S, RSH600862S, RSH600869S, RSH600874S, RSH600892S, RSH600894S, RSH600904S, RSH600938S, RSH600952S, RSH600954S, RSH601040S, RSH601053S, RSH601073S, RSH601103S, RSH601119S, RSH601146S, RSH601596S, RSH602087S, RSH600016S, RSH600017S, RSH600020S, RSH600185S, RSH600237S, RSH600409S, RSH600410S, RSH600496S, RSH600582S, RSH600624S, RSH600660S, RSH600699S, RSH600461S, RSH600501S, RSH600504S, RSH600559S, RSH600613S, RSH600780S, RSH600836S, RSH600848S, RSH600873S, RSH600876S, RSH600878S, RSH600918S, RSH601055S, RSH600244S, RSH600384S, RSH600502S, RSH600552S, RSH600567S, RSH600600S, RSH600811S, RSH600915S, RSH600939S, RSH600941S, RSH601092S, RSH601249S, RSH600649S, RSH601074S, RSH601216S, RSH602344S, RSH601427S, RSH601075S, RSH601426S, RSH601428S, RSH601526S, RSH602161S, RSH602332S, RSH600914S, RSH601010S, RSH601171S, RSH601702S, RSH602181S, RSH600254S, RSH600271S, RSH600628S, RSH601068S, RSH601254S, RSH601282S, RSH601287S, RSH601558S, RSH601598S, RSH601714S, RSH601759S, RSH602075S, RSH602088S, RSH602118S, RSH602160S, RSH602341S, RSH602431S, RSH600471S, RSH600472S, RSH600473S, RSH600474S, RSH600475S, RSH600476S, RSH600477S, RSH600478S, RSH600479S, RSH600481S, RSH600482S, RSH600483S, RSH600484S, RSH600485S, RSH600490S, RSH600491S, RSH600493S, RSH600494S, RSH600507S, RSH600509S, RSH600677S, RSH600681S, RSH600684S, RSH600718S, RSH600719S, RSH600720S, RSH600724S, RSH600727S, RSH600731S, RSH600733S, RSH600734S, RSH600735S, RSH600736S, RSH600737S, RSH600739S, RSH600740S, RSH600741S, RSH600744S, RSH600745S, RSH600747S, RSH600765S, RSH600769S, RSH600777S, RSH600963S, RSH600972S, RSH600973S, RSH600981S, RSH600984S, RSH600986S, RSH600997S, RSH600998S, RSH601044S, RSH601054S, RSH601181S, RSH601187S, RSH601191S, RSH601210S, RSH601211S, RSH601227S, RSH601237S, RSH601248S, RSH601252S, RSH601256S, RSH601258S, RSH601289S, RSH601291S, RSH601330S, RSH601333S, RSH601337S, RSH601340S, RSH601349S, RSH601445S, RSH601451S, RSH601453S, RSH601633S, RSH601639S, RSH601649S, RSH601653S, RSH601654S, RSH601658S, RSH601660S, RSH601664S, RSH601665S, RSH601674S, RSH601676S, RSH601677S, RSH601679S, RSH601682S, RSH601683S, RSH601686S, RSH601688S, RSH601695S, RSH601707S, RSH601822S, RSH601823S, RSH601849S, RSH601861S, RSH601869S, RSH601872S, RSH601874S, RSH601884S, RSH601891S, RSH601892S, RSH601899S, RSH601900S, RSH601902S, RSH601910S, RSH601912S, RSH601917S, RSH601918S, RSH601921S, RSH601962S, RSH601984S, RSH602002S, RSH602033S, RSH602034S, RSH602051S, RSH602052S, RSH602060S, RSH602186S, RSH602190S, RSH602200S, RSH602218S, RSH602219S, RSH600470S, RSH600774S, RSH601326S, RSH601643S, RSH601672S, RSH601684S, RSH601685S, RSH601941S, RSH601949S, RSH600703S, RSH600704S, RSH600705S, RSH600706S, RSH600950S, RSH600969S, RSH600970S, RSH601076S, RSH601182S, RSH601183S, RSH601296S, RSH601655S, RSH601656S, RSH601657S, RSH601696S, RSH601697S, RSH601901S, RSH601907S, RSH601926S, RSH601985S, RSH601978S, RSH600693S, RSH600695S, RSH600700S, RSH600701S, RSH600702S, RSH601001S, RSH601047S, RSH601050S, RSH601069S, RSH601083S, RSH601184S, RSH601185S, RSH601188S, RSH601215S, RSH601240S, RSH601250S, RSH601260S, RSH601261S, RSH601263S, RSH601271S, RSH601274S, RSH601277S, RSH601281S, RSH601283S, RSH601351S, RSH601457S, RSH601632S, RSH601636S, RSH601641S, RSH601663S, RSH601666S, RSH601667S, RSH601668S, RSH601669S, RSH601670S, RSH601671S, RSH601687S, RSH601689S, RSH601916S, RSH601919S, RSH601924S, RSH601937S, RSH601938S, RSH601965S, RSH601987S, RSH601990S, RSH602028S, RSH602030S, RSH602032S, RSH601898S, RSH602045S, RSH600686S, RSH600687S, RSH600688S, RSH600754S, RSH600985S, RSH600994S, RSH601002S, RSH601045S, RSH601046S, RSH601077S, RSH601189S, RSH601194S, RSH601229S, RSH601236S, RSH601332S, RSH601452S, RSH601631S, RSH601635S, RSH601648S, RSH601680S, RSH602228S, RSH600717S, RSH600748S, RSH600749S, RSH600751S, RSH600752S, RSH600755S, RSH600756S, RSH600757S, RSH600760S, RSH600767S, RSH600768S, RSH600965S, RSH600967S, RSH600976S, RSH600980S, RSH600987S, RSH600989S, RSH600990S, RSH600991S, RSH600992S, RSH601013S, RSH601038S, RSH601080S, RSH601091S, RSH601186S, RSH601190S, RSH601218S, RSH601286S, RSH601320S, RSH601334S, RSH601347S, RSH601438S, RSH601449S, RSH601450S, RSH601577S, RSH601651S, RSH601652S, RSH601659S, RSH601698S, RSH601703S, RSH601709S, RSH601711S, RSH601715S, RSH601976S, RSH601983S, RSH602026S, RSH602050S, RSH602053S, RSH601294S, RSH601885S, RSH601931S, RSH600678S, RSH600680S, RSH601455S, RSH601644S, RSH601645S, RSH601678S, RSH601944S, RSH601954S, RSH602061S, RSH600403S, RSH600560S, RSH600662S, RSH601769S, RSH600550S, RSH600558S, RSH600796S, RSH600863S, RSH600209S, RSH600381S, RSH600395S, RSH600397S, RSH600466S, RSH600467S, RSH600562S, RSH600565S, RSH600585S, RSH600587S, RSH600602S, RSH600626S, RSH600629S, RSH600665S, RSH600669S, RSH600779S, RSH600808S, RSH600851S, RSH600855S, RSH600880S, RSH600905S, RSH600940S, RSH600943S, RSH601007S, RSH601165S, RSH601173S, RSH601265S, RSH601273S, RSH601285S, RSH601329S, RSH601350S, RSH601597S, RSH601775S, RSH602445S, RSH602227S, RSH600588S, RSH600441S, RSH600444S, RSH600581S, RSH600591S, RSH600592S, RSH600597S, RSH600605S, RSH600608S, RSH600612S, RSH601434S, RSH601435S, RSH600252S, RSH600422S, RSH600423S, RSH600436S, RSH600445S, RSH600449S, RSH601357S, RSH601433S, RSH601617S, RSH601624S, RSH601880S, RSH600434S, RSH600437S, RSH600614S, RSH601439S, RSH601812S, RSH602198S, RSH601724S, RSH600925S, RSH600949S, RSH600974S, RSH600978S, RSH600927S, RSH600928S, RSH602397S, RSH600269S, RSH600420S, RSH600421S, RSH600440S, RSH600446S, RSH600448S, RSH600450S, RSH600569S, RSH600573S, RSH600577S, RSH600578S, RSH600598S, RSH601432S, RSH601563S, RSH601620S, RSH601621S, RSH601627S, RSH601708S, RSH602206S, RSH602207S, RSH600255S, RSH600256S, RSH600424S, RSH600425S, RSH600432S, RSH600443S, RSH600571S, RSH600579S, RSH600586S, RSH600603S, RSH600604S, RSH601614S, RSH601615S, RSH601616S, RSH601618S, RSH601619S, RSH601626S, RSH601719S, RSH600435S, RSH601622S, RSH600593S, RSH600594S, RSH600781S, RSH601128S, RSH600495S, RSH600844S, RSH600846S, RSH600847S, RSH601049S, RSH601066S, RSH601323S, RSH601557S, RSH601780S, RSH601845S, RSH602044S, RSH602146S, RSH600018S, RSH600019S, RSH601336S, RSH601418S, RSH601706S, RSH602093S, RSH600429S, RSH600438S, RSH600447S, RSH600480S, RSH600508S, RSH600725S, RSH600770S, RSH600771S, RSH600982S, RSH600999S, RSH601000S, RSH601212S, RSH601213S, RSH601288S, RSH601343S, RSH601638S, RSH601640S, RSH601661S, RSH601690S, RSH601718S, RSH601876S, RSH602012S, RSH602092S, RSH602303S, RSH600008S, RSH600015S, RSH600029S, RSH600213S, RSH600411S, RSH600498S, RSH600533S, RSH600539S, RSH600609S, RSH600623S, RSH600870S, RSH600898S, RSH601600S, RSH602094S, RSH601623S, RSH600460S, RSH600620S, RSH600789S, RSH600810S, RSH601264S, RSH601275S, RSH601276S, RSH601352S, RSH601786S, RSH601787S, RSH601791S, RSH601857S, RSH602069S, RSH600728S, RSH600730S, RSH600468S, RSH600469S, RSH600487S, RSH600488S, RSH601193S, RSH601424S, RSH601458S, RSH601647S, RSH601853S, RSH601889S, RSH600009S, RSH600206S, RSH600226S, RSH600243S, RSH600399S, RSH600556S, RSH600794S, RSH600803S, RSH601142S, RSH601335S, RSH600389S, RSH602356S, RSH600405S, RSH600549S, RSH600709S, RSH600805S, RSH600956S, RSH600961S, RSH601085S, RSH601226S, RSH601338S, RSH601341S, RSH601344S, RSH601396S, RSH601397S, RSH601400S, RSH601586S, RSH601701S, RSH600708S, RSH600711S, RSH600713S, RSH600715S, RSH600716S, RSH600764S, RSH600964S, RSH600971S, RSH600977S, RSH600988S, RSH601020S, RSH601056S, RSH601059S, RSH601079S, RSH601102S, RSH601214S, RSH601230S, RSH601246S, RSH601251S, RSH601290S, RSH601315S, RSH601454S, RSH601456S, RSH601630S, RSH601634S, RSH601642S, RSH601650S, RSH601673S, RSH601694S, RSH601841S, RSH601875S, RSH601886S, RSH601887S, RSH601913S, RSH601943S, RSH601958S, RSH601963S, RSH601977S, RSH602027S, RSH602055S, RSH602056S, RSH602062S, RSH600884S, RSH601003S, RSH600772S, RSH600775S, RSH600776S, RSH600966S, RSH601192S, RSH601232S, RSH601637S, RSH601646S, RSH601662S, RSH601896S, RSH601960S, RSH601442S, RSH601446S, RSH601447S, RSH601448S, RSH601513S, RSH601514S, RSH601560S, RSH601561S, RSH601562S, RSH601572S, RSH601574S, RSH601628S, RSH601629S, RSH601720S, RSH601806S, RSH602176S, RSH602406S, RSH602407S, RSH602408S, RSH602409S, RSH602410S, RSH602411S, RSH602412S, RSH601370S, RSH600380S, RSH600576S, RSH600652S, RSH601094S, RSH600431S, RSH600679S, RSH600555S, RSH600575S, RSH600746S, RSH600750S, RSH600798S, RSH600860S, RSH600875S, RSH601006S, RSH601019S, RSH601071S, RSH601095S, RSH601152S, RSH601158S, RSH601159S, RSH601160S, RSH601161S, RSH601162S, RSH601179S, RSH601243S, RSH601266S, RSH601284S, RSH601342S, RSH601693S, RSH601699S, RSH601710S, RSH601777S, RSH601814S, RSH601815S, RSH601816S, RSH601837S, RSH601846S, RSH601867S, RSH602294S, RSH602297S, RSH601866S, RSH601871S, RSH600856S, RSH600630S, RSH600778S, RSH600792S, RSH600809S, RSH600841S, RSH600843S, RSH600845S, RSH600866S, RSH600872S, RSH600897S, RSH600945S, RSH600946S, RSH601064S, RSH601084S, RSH601087S, RSH601153S, RSH601154S, RSH601155S, RSH601156S, RSH601180S, RSH601310S, RSH601339S, RSH601356S, RSH600499S, RSH600783S, RSH600799S, RSH600835S, RSH600867S, RSH600885S, RSH600893S, RSH600900S, RSH601086S, RSH601141S, RSH601144S, RSH601167S, RSH601168S, RSH601222S, RSH601241S, RSH601257S, RSH601353S, RSH601413S, RSH601732S, RSH602015S, RSH602090S, RSH602289S, RSH602338S, RSH601239S, RSH601302S, RSH601308S, RSH601311S, RSH601318S, RSH601377S, RSH601391S, RSH601459S, RSH601584S, RSH601625S, RSH601681S, RSH601722S, RSH601723S, RSH601725S, RSH601727S, RSH601767S, RSH601793S, RSH601844S, RSH601905S, RSH601942S, RSH601959S, RSH601968S, RSH601969S, RSH601973S, RSH601974S, RSH601979S, RSH601980S, RSH601981S, RSH601982S, RSH601986S, RSH602004S, RSH602013S, RSH602018S, RSH602023S, RSH602031S, RSH602035S, RSH602037S, RSH602038S, RSH602042S, RSH602048S, RSH602049S, RSH602054S, RSH602067S, RSH602068S, RSH602070S, RSH602071S, RSH602074S, RSH602076S, RSH602078S, RSH602079S, RSH602080S, RSH602084S, RSH602086S, RSH602111S, RSH602116S, RSH602120S, RSH602127S, RSH602128S, RSH602133S, RSH602143S, RSH602145S, RSH602147S, RSH602148S, RSH602150S, RSH602155S, RSH602162S, RSH602164S, RSH602167S, RSH602168S, RSH602174S, RSH602180S, RSH602184S, RSH602185S, RSH602199S, RSH602204S, RSH602220S, RSH602221S, RSH602222S, RSH602223S, RSH602225S, RSH602231S, RSH602286S, RSH602287S, RSH602311S, RSH602319S, RSH602353S, RSH602372S, RSH602373S, RSH602374S, RSH602375S, RSH602376S, RSH602377S, RSH602384S, RSH602388S, RSH602398S, RSH602433S, RSH602473S, RSH602478S, RSH602479S, RSH602491S, RSH602492S, RSH602493S, RSH602494S, RSH602495S, RSH602496S, RSH602497S, RSH602498S, RSH602499S, RSH602500S, RSH602503S, RSH602504S, RSH602505S, RSH602506S, RSH602507S, RSH602508S, RSH602509S, RSH602510S, RSH602511S, RSH602512S, RSH602513S, RSH602514S, RSH602515S, RSH602516S, RSH602517S, RSH602518S, RSH602519S, RSH602520S, RSH602521S, RSH602522S, RSH602523S, RSH602524S, RSH602525S, RSH602526S, RSH602527S, RSH602528S, RSH602529S, RSH602530S, RSH602531S, RSH602532S, RSH602533S, RSH602534S, RSH602535S, RSH602536S, RSH602537S, RSH602538S, RSH602539S, RSH602540S, RSH602541S, RSH602542S, RSH602543S, RSH602545S, RSH602547S, RSH602555S, RSH601067S, RSH602550S, RSH602551S, RSH602552S, RSH602553S, RSH602557S, RSH601279S, RSH601368S, RSH601733S, RSH601757S, RSH601939S, RSH601972S, RSH602006S, RSH602036S, RSH602041S, RSH602047S, RSH602057S, RSH602208S, RSH602318S, RSH602323S, RSH602355S, RSH602392S, RSH602548S
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Crome™ VR MRI SureScan™ DVPC3D4Medtronic, Inc. · Pulse generator, permanent, implantable3 recalls18 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 92328).
- Z-1704-2023 · Open, ClassifiedICD-DR DDBB1D1 EVERA XT IS1/DF1 US, Model Number DDBB1D1; Implantable Cardioverter Defibrillators
- Z-1705-2023 · Open, ClassifiedICD-DR DDBB1D4 EVERA XT IS1/DF4 US, Model Number DDBB1D4; Implantable Cardioverter Defibrillators
- Z-1709-2023 · Open, ClassifiedICD-DR DDBC3D1 EVERA S IS1/DF1 GLOB, Model Number DDBC3D1; Implantable Cardioverter Defibrillators
- Z-1710-2023 · Open, ClassifiedICD-DR DDBC3D4 EVERA S IS1/DF4 GLOB, Model Number DDBC3D4; Implantable Cardioverter Defibrillators
- Z-1711-2023 · Open, ClassifiedICD-DR DDMB1D1 EVERA MRI XT US DF1, Model Number DDMB1D1; Implantable Cardioverter Defibrillators
- Z-1712-2023 · Open, ClassifiedICD-DR DDMB1D4 EVERA MRI XT IS-1/DF4 US, Model Number DDMB1D4; Implantable Cardioverter Defibrillators
- Z-1715-2023 · Open, ClassifiedICD-DR DDMC3D1 EVERA MRI S OUS/US DF1, Model Number DDMC3D1; Implantable Cardioverter Defibrillators
- Z-1716-2023 · Open, ClassifiedICD-DR DDMC3D4 EVERA MRI S IS-1/DF4 GLOB, Model Number DDMC3D4; Implantable Cardioverter Defibrillators
- Z-1717-2023 · Open, ClassifiedICD-DR DDMD3D1 PRIMO MRI, Model Number DDMD3D1; Implantable Cardioverter Defibrillators
- Z-1718-2023 · Open, ClassifiedICD-DR DDMD3D4 PRIMO MRI, Model Number DDMD3D4; Implantable Cardioverter Defibrillators
- Z-1721-2023 · Open, ClassifiedICD COBALT XT DR MRI IS1 DF1, Model Number DDPA2D1; Implantable Cardioverter Defibrillators
- Z-1723-2023 · Open, ClassifiedICD COBALT XT DR MRI IS1 DF4, Model Number DDPA2D4; Implantable Cardioverter Defibrillators
- Z-1725-2023 · Open, ClassifiedICD COBALT DR MRI IS1 DF1, Model Number DDPB3D1; Implantable Cardioverter Defibrillators
- Z-1726-2023 · Open, ClassifiedICD COBALT DR MRI IS1 DF4, Model Number DDPB3D4; Implantable Cardioverter Defibrillators
- Z-1727-2023 · Open, ClassifiedICD CROME DR MRI IS1 DF1, Model Number DDPC3D1; Implantable Cardioverter Defibrillators
- Z-1728-2023 · Open, ClassifiedICD CROME DR MRI IS1 DF4, Model Number DDPC3D4; Implantable Cardioverter Defibrillators
- Z-1729-2023 · Open, ClassifiedCRT-D DTBA1D1 VIVA XT IS1/DF1 US, Model Number DTBA1D1; Implantable Cardioverter Defibrillators
- Z-1731-2023 · Open, ClassifiedCRT-D DTBA1D4 VIVA XT IS1/DF4 US, Model Number DTBA1D4; Implantable Cardioverter Defibrillators
- Z-1732-2023 · Open, ClassifiedCRT-D DTBA1Q1 VIVA QUAD XT IS4/DF1 US, Model Number DTBA1Q1; Implantable Cardioverter Defibrillators
- Z-1733-2023 · Open, ClassifiedCRT-D DTBA1QQ VIVA QUAD XT IS4/DF4 US, Model Number DTBA1QQ; Implantable Cardioverter Defibrillators
- Z-1741-2023 · Open, ClassifiedCRT-D DTBB1D1 VIVA S IS1/DF1 US, Model Number DTBB1D1; Implantable Cardioverter Defibrillators
- Z-1742-2023 · Open, ClassifiedCRT-D DTBB1D4 VIVA S IS1/DF4 US, Model Number DTBB1D4; Implantable Cardioverter Defibrillators
- Z-1743-2023 · Open, ClassifiedCRT-D DTBB1Q1 VIVA QUAD S IS4/DF1 US, Model Number DTBB1Q1; Implantable Cardioverter Defibrillators
- Z-1744-2023 · Open, ClassifiedCRT-D DTBB1QQ VIVA QUAD S IS4/DF4 US, Model Number DTBB1QQ; Implantable Cardioverter Defibrillators
- Z-1753-2023 · Open, ClassifiedCRTD DTMA1D1 CLARIA MRI US DF1, Model Number DTMA1D1; Implantable Cardioverter Defibrillators
- Z-1754-2023 · Open, ClassifiedCRTD DTMA1D4 CLARIA MRI US DF4, Model Number DTMA1D4; Implantable Cardioverter Defibrillators
- Z-1755-2023 · Open, ClassifiedCRTD DTMA1Q1 CLARIA MRI QUAD US DF1, Model Number DTMA1Q1; Implantable Cardioverter Defibrillators
- Z-1756-2023 · Open, ClassifiedCRTD DTMA1QQ CLARIA MRI QUAD US DF4, Model Number DTMA1QQ; Implantable Cardioverter Defibrillators
- Z-1761-2023 · Open, ClassifiedCRTD DTMB1D1 AMPLIA MRI US DF1, Model Number DTMB1D1; Implantable Cardioverter Defibrillators
- Z-1762-2023 · Open, ClassifiedCRTD DTMB1D4 AMPLIA MRI US DF4, Model Number DTMB1D4; Implantable Cardioverter Defibrillators
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page