Trackmaster recallZ-2052-2023 · Full Vision Inc · initiated May 12, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Full Vision Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Treadmill malfunction: drive PCB can misread the signals and stop or reverse and then accelerate forward.
FDA root cause category: Other
What the company is doing
Trackmaster issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees beginning on 05/12/2023 by letter. The notice explained the problem and the risk during use, and requested the following: Immediate Actions to be taken by Customer/User: These steps are the recommendation of the Manufacturer to prevent injury and allow use of the device. Step 1: Instruct patient to straddle the belt and hold on to the handrails during initialization of the treadmill before starting the test Step 2: Send start/run command to the treadmill with patient still straddling the belt, wait 30 seconds to ensure there is no uncommanded motion Step 3: Instruct patient to carefully step on the belt while holding on to the handrails, begin walking and continue with the test Step 4: Patients unable to perform steps 1-3 must not use the treadmill Step 5: Print Appendix B and attach to the treadmill until corrections are made and ensure it is visible to the clinician and user
Recall details
- Recall number
- Z-2052-2023
- Status
- Open, Classified
- Initiated by the company
- May 12, 2023
- Posted by the FDA
- Jun 30, 2023
- Recalling firm
- Full Vision Inc
- Quantity
- 565 units
- Product code
- IOL
- Distribution
- USA (Nationwide), Australia, Bahrain, Canada, Chile, Colombia, Costa Rica, Dominican Republic, FINLAND, Guatemala, HONG KONG, India, Indonesia, Israel, Mexico, NEW ZEALAND, Nicaragua, Panama, PARAQUAY, Peru, Philippines, Saudi Arabia, Singapore, SOUTH KOREA, Thailand, Ukraine, United Arab Emirates, United Kingdom, Uruguay
Product description
TRACKMASTER, TREADMILL TMX428 110V, MODEL NUMBER 317-07926
Code information (models, lots, serial numbers)
Show the full code information (6k characters)
UDI/DI 00860176000606, Serial Numbers: FVDC-7585, FVDC-7586, FVDC-7587, FVDC-7588, FVDC-7589, FVDC-7590, FVDC-7591, FVDC-7592, FVDC-7593, FVDC-7594, FVDC-7595, FVDC-7596, FVDC-7597, FVDC-7598, FVDC-7599, FVDC-7600, FVDC-7601, FVDC-7603, FVDC-7604, FVDC-7605, FVDC-7606, FVDC-7607, FVDC-7608, FVDC-7609, FVDC-7610, FVDC-7611, FVDC-7612, FVDC-7613, FVDC-7614, FVDC-7615, FVDC-7616, FVDC-7617, FVDC-7618, FVDC-7619, FVDC-7620, FVDC-7621, FVDC-7622, FVDC-7623, FVDC-7624, FVDC-7625, FVDC-7626, FVDC-7627, FVDC-7628, FVDC-7629, FVDC-7640, FVDC-7666, FVDC-7667, FVDC-7668, FVDC-7669, FVDC-7670, FVDC-7671, FVDC-7672, FVDC-7673, FVDC-7674, FVDC-7675, FVDC-7676, FVDC-7677, FVDC-7678, FVDC-7679, FVDC-7680, FVDC-7681, FVDC-7682, FVDC-7683, FVDC-7684, FVDC-7685, FVDC-7687, FVDC-7688, FVDC-7689, FVDC-7690, FVDC-7691, FVDC-7692, FVDC-7693, FVDC-7694, FVDC-7695, FVDC-7696, FVDC-7697, FVDC-7698, FVDC-7699, FVDC-7700, FVDC-7701, FVDC-7702, FVDC-7703, FVDC-7704, FVDC-7705, FVDC-7706, FVDC-7717, FVDC-7719, FVDC-7720, FVDC-7721, FVDC-7722, FVDC-7723, FVDC-7724, FVDC-7725, FVDC-7726, FVDC-7727, FVDC-7728, FVDC-7729, FVDC-7730, FVDC-7731, FVDC-7732, FVDC-7733, FVDC-7734, FVDC-7735, FVDC-7736, FVDC-7737, FVDC-7738, FVDC-7782, FVDC-7783, FVDC-7785, FVDC-7787, FVDC-7788, FVDC-7789, FVDC-7790, FVDC-7791, FVDC-7792, FVDC-7793, FVDC-7794, FVDC-7795, FVDC-7796, FVDC-7797, FVDC-7798, FVDC-7799, FVDC-7800, FVDC-7811, FVDC-7812, FVDC-7813, FVDC-7814, FVDC-7815, FVDC-7816, FVDC-7817, FVDC-7818, FVDC-7819, FVDC-7836, FVDC-7837, FVDC-7838, FVDC-7839, FVDC-7840, FVDC-7841, FVDC-7842, FVDC-7843, FVDC-7844, FVDC-7845, FVDC-7846, FVDC-7847, FVDC-7848, FVDC-7849, FVDC-7850, FVDC-7864, FVDC-7865, FVDC-7866, FVDC-7867, FVDC-7868, FVDC-7869, FVDC-7870, FVDC-7871, FVDC-7872, FVDC-7873, FVDC-7874, FVDC-7875, FVDC-7876, FVDC-7877, FVDC-7878, FVDC-7879, FVDC-7880, FVDC-7881, FVDC-7882, FVDC-7883, FVDC-7884, FVDC-7885, FVDC-7886, FVDC-7887, FVDC-7889, FVDC-7890, FVDC-7899, FVDC-7915, FVDC-7916, FVDC-7917, FVDC-7918, FVDC-7919, FVDC-7920, FVDC-7921, FVDC-7922, FVDC-7923, FVDC-7924, FVDC-7925, FVDC-7926, FVDC-7943, FVDC-7944, FVDC-7945, FVDC-7946, FVDC-7947, FVDC-7964, FVDC-7965, FVDC-7966, FVDC-7983, FVDC-7984, FVDC-7985, FVDC-7986, FVDC-7987, FVDC-7988, FVDC-7989, FVDC-7990, FVDC-7991, FVDC-7992, FVDC-7993, FVDC-7994, FVDC-7995, FVDC-8001, FVDC-8002, FVDC-8003, FVDC-8004, FVDC-8005, FVDC-8006, FVDC-8007, FVDC-8008, FVDC-8009, FVDC-8010, FVDC-8011, FVDC-8012, FVDC-8013, FVDC-8014, FVDC-8015, FVDC-8016, FVDC-8017, FVDC-8018, FVDC-8019, FVDC-8020, FVDC-8021, FVDC-8022, FVDC-8023, FVDC-8024, FVDC-8025, FVDC-8029, FVDC-8030, FVDC-8031, FVDC-8032, FVDC-8033, FVDC-8034, FVDC-8035, FVDC-8036, FVDC-8037, FVDC-8038, FVDC-8039, FVDC-8040, FVDC-8041, FVDC-8042, FVDC-8043, FVDC-8044, FVDC-8045, FVDC-8046, FVDC-8047, FVDC-8048, FVDC-8049, FVDC-8050, FVDC-8051, FVDC-8052, FVDC-8053, FVDC-8054, FVDC-8055, FVDC-8056, FVDC-8110, FVDC-8111, FVDC-8112, FVDC-8113, FVDC-8114, FVDC-8115, FVDC-8116, FVDC-8117, FVDC-8118, FVDC-8119, FVDC-8120, FVDC-8121, FVDC-8122, FVDC-8123, FVDC-8124, FVDC-8125, FVDC-8127, FVDC-8128, FVDC-8129, FVDC-8130, FVDC-8131, FVDC-8132, FVDC-8133, FVDC-8134, FVDC-8135, FVDC-8136, FVDC-8137, FVDC-8138, FVDC-8139, FVDC-8140, FVDC-8141, FVDC-8142, FVDC-8143, FVDC-8144, FVDC-8145, FVDC-8146, FVDC-8147, FVDC-8148, FVDC-8149, FVDC-8150, FVDC-8151, FVDC-8152, FVDC-8153, FVDC-8154, FVDC-8155, FVDC-8156, FVDC-8193, FVDC-8194, FVDC-8195, FVDC-8196, FVDC-8197, FVDC-8198, FVDC-8199, FVDC-8200, FVDC-8201, FVDC-8202, FVDC-8213, FVDC-8214, FVDC-8215, FVDC-8216, FVDC-8217, FVDC-8218, FVDC-8219, FVDC-8220, FVDC-8221, FVDC-8222, FVDC-8223, FVDC-8224, FVDC-8225, FVDC-8226, FVDC-8227, FVDC-8228, FVDC-8229, FVDC-8230, FVDC-8231, FVDC-8232, FVDC-8233, FVDC-8234, FVDC-8245, FVDC-8246, FVDC-8247, FVDC-8248, FVDC-8249, FVDC-8250, FVDC-8251, FVDC-8252, FVDC-8253, FVDC-8254, FVDC-8256, FVDC-8257, FVDC-8258, FVDC-8259, FVDC-8260, FVDC-8261, FVDC-8262, FVDC-8263, FVDC-8264, FVDC-8265, FVDC-8266, FVDC-8269.FVDC-8273, FVDC-8274, FVDC-8275, FVDC-8276, FVDC-8277, FVDC-8278, FVDC-8279, FVDC-8280, FVDC-8281, FVDC-8282, FVDC-8283, FVDC-8284, FVDC-8285, FVDC-8286, FVDC-8287, FVDC-8288, FVDC-8289, FVDC-8290, FVDC-8291, FVDC-8292, FVDC-8293, FVDC-8294, FVDC-8295, FVDC-8296, FVDC-8297, FVDC-8298, FVDC-8299, FVDC-8300, FVDC-8301, FVDC-8302, FVDC-8303, FVDC-8304, FVDC-8305, FVDC-8306, FVDC-8307, FVDC-8308, FVDC-8309, FVDC-8326, FVDC-8327, FVDC-8328, FVDC-8329, FVDC-8330, FVDC-8331, FVDC-8332, FVDC-8333, FVDC-8334, FVDC-8335, FVDC-8336, FVDC-8337, FVDC-8338, FVDC-8339, FVDC-8340, FVDC-8341, FVDC-8342, FVDC-8343, FVDC-8344, FVDC-8345, FVDC-8346, FVDC-8347, FVDC-8348, FVDC-8349, FVDC-8350, FVDC-8351, FVDC-8352, FVDC-8353, FVDC-8354, FVDC-8355, FVDC-8396, FVDC-8397, FVDC-8398, FVDC-8399, FVDC-8426, FVDC-8427, FVDC-8428, FVDC-8429, FVDC-8430, FVDC-8431, FVDC-8432, FVDC-8433, FVDC-8434, FVDC-8435, FVDC-8456, FVDC-8457, FVDC-8458, FVDC-8459, FVDC-8460, FVDC-8461, FVDC-8462, FVDC-8463, FVDC-8464, FVDC-8465, FVDC-8494, FVDC-8495, FVDC-8496, FVDC-8497, FVDC-8498, FVDC-8499, FVDC-8500, FVDC-8501, FVDC-8502, FVDC-8503, FVDC-8504, FVDC-8505, FVDC-8506, FVDC-8507, FVDC-8508, FVDC-8514, FVDC-8515, FVDC-8516, FVDC-8517, FVDC-8518, FVDC-8519, FVDC-8520, FVDC-8521, FVDC-8522, FVDC-8523, FVDC-8559, FVDC-8560, FVDC-8561, FVDC-8562, FVDC-8563, FVDC-8564, FVDC-8565, FVDC-8566, FVDC-8567, FVDC-8568, FVDC-8569, FVDC-8570, FVDC-8571, FVDC-8572, FVDC-8573, FVDC-8599, FVDC-8600, FVDC-8601, FVDC-8602, FVDC-8603, FVDC-8604, FVDC-8605, FVDC-8606, FVDC-8607, FVDC-8608, FVDC-8609, FVDC-8610, FVDC-8611, FVDC-8612, FVDC-8613, FVDC-8614, FVDC-8615, FVDC-8616, FVDC-8629, FVDC-8630, FVDC-8648, FVDC-8654, FVDC-8655, FVDC-8656, FVDC-8657, FVDC-8658, FVDC-8659, FVDC-8660, FVDC-8661, FVDC-8662, FVDC-8663, FVDC-8674, FVDC-8675, FVDC-8676, FVDC-8677, FVDC-8678, FVDC-8679, FVDC-8680, FVDC-8681, FVDC-8682, FVDC-8683, FVDC-8697, FVDC-8698, FVDC-8699, FVDC-8700, FVDC-8701, FVDC-8757, FVDC-8758, FVDC-8759, FVDC-8760, FVDC-8761, FVDC-8762, FVDC-8763, FVDC-8814, FVDC-8819, FVDC-8820, FVDC-8821, FVDC-8822, FVDC-8823, FVDC-8824, FVDC-8825, FVDC-8826, FVDC-8827, FVDC-8828, FVDC-8860, FVDC-8861, FVDC-8862, FVDC-8863, FVDC-8864, FVDC-8865, FVDC-8866, FVDC-8867, FVDC-8868, FVDC-8869
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Trackmaster TMX428 110VFull Vision, Inc. · Treadmill, Powered2 recalls0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 92382).
- Z-2053-2023 · Open, ClassifiedTRACKMASTER, TREADMILL TMX428 220V, MODEL NUMBER 317-07927
- Z-2054-2023 · Open, ClassifiedTRACKMASTER, TREADMILL TMX428CP 110V, MODEL NUMBER 317-07928
- Z-2055-2023 · Open, ClassifiedTRACKMASTER, TREADMILL TMX428CP 220V, MODEL NUMBER 317-07929
- Z-2056-2023 · Open, ClassifiedTRACKMASTER, TREADMILL GE T2100-ST1 110VV, MODEL NUMBER 317-07926GE
- Z-2057-2023 · Open, ClassifiedTRACKMASTER, TREADMILL GE T2100-ST2 220V, MODEL NUMBER 317-07927GE
- Z-2058-2023 · Open, ClassifiedTRACKMASTER, TREADMILL GE T2100-ST2 220V CHINA, MODEL NUMBER 317-07927GE CHINA
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page