Skip to content

QUADROX-i recallZ-1659-2023 · Maquet Medical Systems USA · initiated May 18, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Maquet Medical Systems USA or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Product packaging may exhibit small pinholes which may not be visible to the unaided eye, compromising the sterile barrier of the packaging. Exposure to non-sterile device may result in inflammation, infection, sepsis, and ischemia.

FDA root cause category: Under Investigation by firm

What the company is doing

On May 18, 2023, the firm notified customers of the recall via an Urgent Medical Device Removal letter. Actions to be taken by the customer: If an affected device is already in use, please use according to normal practices. Monitor the patient for signs and symptoms of inflammation, infection, sepsis, and ischemia. If detected, treat according to clinical protocols. Please remove any affected product in inventory from areas of use. Unopened and unexpired affected product is eligible for credit upon return. Product Return: Please contact Getinge Customer Service at 1-888-943-8872 option 2 between the hours of 6:00 AM and 5:00 PM Pacific Standard Time to request a return authorization (RMA) and shipping instructions to return any affected product. Distributors who have shipped affected products to customers should forward the notification to customers' attention for appropriate action. Maquet Cardiopulmonary GmbH is working with all possible urgency on the redesign and revalidation of a packaging solution, which is anticipated to be available by Q4 2023.

Recall details

Recall number
Z-1659-2023
Status
Open, Classified
Initiated by the company
May 18, 2023
Posted by the FDA
May 30, 2023
Recalling firm
Maquet Medical Systems USA
Quantity
0 (US)
Product code
DTZ
FDA submissions
K082117
Distribution
Nationwide domestic distribution

Product description

QUADROX-i Adult, Oxygenator used during cardiopulmonary bypass Model Item No. HMO 70000-USA 701067820 HMO 71000-USA 701067823 BEQ-HMO 71000-USA 701067829

Code information (models, lots, serial numbers)

Model Item No. UDI HMO 70000-USA 701067820 4058863019147 HMO 71000-USA 701067823 4058863017341 BEQ-HMO 71000-USA 701067829 4058863017372 All lots

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 92006).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page