Skip to content

QUADROX-iRecalls & reported problems across all models

Maquet Cardiopulmonary AG · 10 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

8 open recalls name at least one QUADROX-i model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls
10

We found 10 models of QUADROX-i by Maquet Cardiopulmonary AG in the FDA's device registry (GUDID). Together they are named in 1 adverse event report in the FDA's MAUDE database, received between Feb 27, 2025 and Feb 27, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Infusion or Flow Problem” (1 report).

10 FDA recall records name at least one of these models by its device identifier, 8 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

QUADROX-i recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. May 18, 2023Open, Classified

    Product packaging may exhibit small pinholes which may not be visible to the unaided eye, compromising the sterile barrier of the packaging. Exposure to non-sterile device may result in inflammation, infection, sepsis, and ischemia.

    Z-1659-2023 · Root cause: Under Investigation by firm · names 3 of these models · Recall details · FDA record ↗

  2. May 18, 2023Open, Classified

    Product packaging may exhibit small pinholes which may not be visible to the unaided eye, compromising the sterile barrier of the packaging. Exposure to non-sterile device may result in inflammation, infection, sepsis, and ischemia.

    Z-1658-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  3. May 18, 2023Open, Classified

    1. An accessory may be improperly placed during packaging, causing creases and damage to the accessory's sterile pouch. 2. Product packaging may exhibit holes, cracks, dents, and crushed areas. Both issues may compromise the sterile barrier of the packaging, and exposure to non-sterile product may result in inflammation, infection, sepsis, and ischemia.

    Z-1639-2023 · Root cause: Pending · Recall details · FDA record ↗

  4. May 18, 2023Open, Classified

    1. An accessory may be improperly placed during packaging, causing creases and damage to the accessory's sterile pouch. 2. Product packaging may exhibit holes, cracks, dents, and crushed areas. Both issues may compromise the sterile barrier of the packaging, and exposure to non-sterile product may result in inflammation, infection, sepsis, and ischemia.

    Z-1638-2023 · Root cause: Pending · names 2 of these models · Recall details · FDA record ↗

  5. May 18, 2023Open, Classified

    1. An accessory may be improperly placed during packaging, causing creases and damage to the accessory's sterile pouch. 2. Product packaging may exhibit holes, cracks, dents, and crushed areas. Both issues may compromise the sterile barrier of the packaging, and exposure to non-sterile product may result in inflammation, infection, sepsis, and ischemia.

    Z-1637-2023 · Root cause: Pending · names 2 of these models · Recall details · FDA record ↗

  6. May 18, 2023Open, Classified

    Product packaging may exhibit small pinholes which may not be visible to the unaided eye, compromising the sterile barrier of the packaging. Exposure to non-sterile device may result in inflammation, infection, sepsis, and ischemia.

    Z-1657-2023 · Root cause: Under Investigation by firm · names 3 of these models · Recall details · FDA record ↗

  7. May 18, 2023Open, Classified

    1. An accessory may be improperly placed during packaging, causing creases and damage to the accessory's sterile pouch. 2. Product packaging may exhibit holes, cracks, dents, and crushed areas. Both issues may compromise the sterile barrier of the packaging, and exposure to non-sterile product may result in inflammation, infection, sepsis, and ischemia.

    Z-1636-2023 · Root cause: Pending · Recall details · FDA record ↗

  8. Nov 13, 2019Open, Classified

    Sterile barrier system may be compromised

    Z-0677-2020 · Root cause: Package design/selection · Recall details · FDA record ↗

  9. Oct 25, 2018Terminated

    There is a potential for blood leakage at the blood inlet connector when using the reducing screw adapter 1/4 x 3/16 that may result in clinically relevant loss of prime before patient connection or blood loss during patient connection.

    Z-0437-2019 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

  10. Oct 25, 2018Terminated

    There is a potential for blood leakage at the blood inlet connector when using the reducing screw adapter 1/4 x 3/16 that may result in clinically relevant loss of prime before patient connection or blood loss during patient connection.

    Z-0436-2019 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported QUADROX-i problems

Device problems

  • Infusion or Flow Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

QUADROX-i models

10 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent QUADROX-i adverse event reports

The 1 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Maquet Cardiopulmonary AG products

All Maquet Cardiopulmonary AG devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page