RUSCH recallZ-1893-2023 · Teleflex LLC · initiated May 25, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Teleflex LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
FDA root cause category: Under Investigation by firm
What the company is doing
Teleflex issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 05/25/2012 by FedEx. The notice explained the problem, the risk, and requested the following: Users should cease use and distribution of affected product and immediately quarantine the affected product. Distributors were directed to cease use, distribution, and quarantine the product, and to notify their customers. Teleflex is seeking the return of the affected product. For further information: Contact: Customer Service, Telephone: 1-866-396-2111 FAX: 1-855-419-8507, Email: [email protected]
Recall details
- Recall number
- Z-1893-2023
- Status
- Open, Classified
- Initiated by the company
- May 25, 2023
- Posted by the FDA
- Jun 29, 2023
- Recalling firm
- Teleflex LLC
- Quantity
- 18620 units
- Product code
- BTR
- FDA submissions
- K961837
- Distribution
- US Nationwide distribution including Puerto Rico.
Product description
Endotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff, REF 112082095
Code information (models, lots, serial numbers)
UDI/DI 14026704341471, Batch Numbers: 18HG08, 18HT04, 18IG29, 18IT04, 18JT16, 18KG37, 18KT27, 18LT16, 19BG02, 19CG02, 19CT04, 19CT72, 19GT25, 19HT19, 19HT47, 19JT05, 19KT55, 19LT08, 20AT13, 20CT02, 20CT29, 20DT32, 20GT23, 20GT44, KME20H1486, KME20J1899, KME20K2166, KME20L1592, KME20L2462, KME20L2557, KME20M2115, KME20M2198, KME21A2346, KME21E1393, KME21K1818, KME22B1259, KME22B1420, KME22B1854, KME22C0677, KME22C0713, KME22C0844, KME22C1612, KME22C2606, KME22C2757, KME22D1593, KME22D1778, KME22E1285, KME22E2362, KME22F1401, KME22F2452, KME23A0557, KME23A0763
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- RUSCH IPN041507Teleflex Incorporated · Tube, tracheal (w/wo connector)2 recalls0 reports
Product lines
Other devices under the same FDA submission
21 devices cleared under K961837 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- RUSCH IPN041476Teleflex Incorporated · Tube, tracheal (w/wo connector)10 recalls0 reports
- RUSCH IPN041502Teleflex Incorporated · Tube, tracheal (w/wo connector)2 recalls30 reports
- RUSCH IPN041504Teleflex Incorporated · Tube, tracheal (w/wo connector)2 recalls8 reports
- RUSCH IPN041503Teleflex Incorporated · Tube, tracheal (w/wo connector)2 recalls4 reports
- RUSCH IPN041498Teleflex Incorporated · Tube, tracheal (w/wo connector)2 recalls2 reports
- RUSCH IPN044687Teleflex Incorporated · Tube, tracheal (w/wo connector)2 recalls1 report
- RUSCH IPN041501Teleflex Incorporated · Tube, tracheal (w/wo connector)2 recalls1 report
- RUSCH IPN041505Teleflex Incorporated · Tube, tracheal (w/wo connector)2 recalls1 report
- RUSCH IPN044681Teleflex Incorporated · Tube, tracheal (w/wo connector)2 recalls0 reports
- RUSCH IPN044683Teleflex Incorporated · Tube, tracheal (w/wo connector)2 recalls0 reports
- RUSCH IPN044688Teleflex Incorporated · Tube, tracheal (w/wo connector)2 recalls0 reports
- RUSCH IPN044689Teleflex Incorporated · Tube, tracheal (w/wo connector)2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 92373).
- Z-1827-2023 · Open, ClassifiedSlick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170050
- Z-1828-2023 · Open, ClassifiedSlick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170055
- Z-1829-2023 · Open, ClassifiedSlick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170060
- Z-1830-2023 · Open, ClassifiedSlick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170065
- Z-1831-2023 · Open, ClassifiedSlick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170070
- Z-1832-2023 · Open, ClassifiedSlick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170075
- Z-1833-2023 · Open, ClassifiedSlick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170080
- Z-1834-2023 · Open, ClassifiedSlick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170085
- Z-1835-2023 · Open, ClassifiedSlick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170090
- Z-1836-2023 · Open, ClassifiedSlick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170095
- Z-1837-2023 · Open, ClassifiedSlick Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170120
- Z-1838-2023 · Open, ClassifiedSlick Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170125
- Z-1839-2023 · Open, ClassifiedSlick Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170130
- Z-1840-2023 · Open, ClassifiedSlick Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170135
- Z-1841-2023 · Open, ClassifiedSlick Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170140
- Z-1842-2023 · Open, ClassifiedSlick Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170145
- Z-1843-2023 · Open, ClassifiedSlick Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170150
- Z-1844-2023 · Open, ClassifiedSlick Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170155
- Z-1845-2023 · Open, ClassifiedFlexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504550
- Z-1846-2023 · Open, ClassifiedFlexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504555
- Z-1847-2023 · Open, ClassifiedFlexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504560
- Z-1848-2023 · Open, ClassifiedFlexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504565
- Z-1849-2023 · Open, ClassifiedFlexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504570
- Z-1850-2023 · Open, ClassifiedFlexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504575
- Z-1851-2023 · Open, ClassifiedFlexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504580
- Z-1852-2023 · Open, ClassifiedFlexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504585
- Z-1853-2023 · Open, ClassifiedFlexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504590
- Z-1854-2023 · Open, ClassifiedFlexi-Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 506525
- Z-1855-2023 · Open, ClassifiedFlexi-Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 506530
- Z-1856-2023 · Open, ClassifiedFlexi-Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 506535
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page