RUSCH IPN041501Recalls & reported problems
Teleflex Incorporated · ENDOTRACHEAL TUBE - WITH CUFF - MURPHY - STERILE PACK · Class 2
All models of RUSCH (220)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 14026704341419), received between Aug 30, 2024 and Aug 30, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Gas/Air Leak” (1 report).
2 FDA recall records name this device's identifier, 2 still open.
34 other recalls cover devices cleared under the same FDA submission (K961837); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- May 25, 2023Open, Classified
Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
Z-1887-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Dec 8, 2022Open, Classified
Potential for pilot balloon non-inflation or cuff non-deflation.
Z-1011-2023 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Gas/Air Leak1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 30, 2024Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED BY CLINICAL STAFF AT TE HOSPITAL THAT: "PATIENT PRESENTED BRADYPNEA AND APNEA, WITH BRADYCARDIA AND SUBSEQUENT CARDI ORESPIRATORY ARREST DUE TO HYPOXIA, REQUIRING SUPPORT WITH VENTILATION AND CARDIOPULMONARY RESUSCITATION, INTUBATED AT THE FIRST ATTEMPT, BUT DUE TO AIR LEAKAGE WAS CHANGED TO A 6.5 CANNULA WITH FIXED BALLOON IN 16". ADDITIONAL INFOMRATION REPORTED BY THE CLINICAL STAFF: "DEATH (B)(6) 2023. CARDIORESPIRATORY ARREST UNRELATED TO THE DEVICE INCIDENT. THE ISSUE WAS RESOLVED CHANGING THE CANNULA WITH ANOTHER CANNULA OF THE SAME CALIBER AND BRAND. THE AIR LEAK WAS AFTER DE ARREST. THE AIR LEAK, DIDN'T DELAY THE HEALTH CARE. THE PATIENT DIDN'T SUFFER HYPOXIA WHEN THE CANNULA WAS CHANGED."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20107521 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K961837. One submission often covers many models, so check the product and lot details in each record.
- Z-1847-2023 · May 25, 2023 · Open, ClassifiedFlexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504560
- Z-1893-2023 · May 25, 2023 · Open, ClassifiedEndotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff, REF 112082095
- Z-1894-2023 · May 25, 2023 · Open, ClassifiedEndotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff, REF 112082100
- Z-1884-2023 · May 25, 2023 · Open, ClassifiedEndotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff, REF 112082050
- Z-1851-2023 · May 25, 2023 · Open, ClassifiedFlexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504580
- Z-1962-2023 · May 25, 2023 · Open, ClassifiedEndotracheal Tube oral/nasal Magill, High Volume, Low Pressure Cuff, REF 112080075
- Z-1882-2023 · May 25, 2023 · Open, ClassifiedEndotracheal Tube oral/nasal Magill, High Volume, Low Pressure Cuff, REF 112080095
- Z-1853-2023 · May 25, 2023 · Open, ClassifiedFlexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504590
- Z-1890-2023 · May 25, 2023 · Open, ClassifiedEndotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff, REF 112082080
- Z-1881-2023 · May 25, 2023 · Open, ClassifiedEndotracheal Tube oral/nasal Magill, High Volume, Low Pressure Cuff, REF 112080055
- Z-1899-2023 · May 25, 2023 · Open, ClassifiedRUSCHELIT Safety Clear Tracheal Tube, oral/nasal, Magill, REF 112480045
- Z-1848-2023 · May 25, 2023 · Open, ClassifiedFlexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504565
- Z-1849-2023 · May 25, 2023 · Open, ClassifiedFlexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504570
- Z-1846-2023 · May 25, 2023 · Open, ClassifiedFlexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504555
- Z-1958-2023 · May 25, 2023 · Open, ClassifiedEndotracheal Tube oral/nasal Magill, High Volume, Low Pressure Cuff, REF 112080050
…and 19 more.
Other Tube, tracheal (w/wo connector) devices
- Shiley 18875Covidien LP · Tube, tracheal (w/wo connector)4 deaths570 reports
- RUSCH IPN041476Teleflex Incorporated · Tube, tracheal (w/wo connector)10 recalls0 reports
- Shiley 18775Covidien LP · Tube, tracheal (w/wo connector)489 reports
- Shiley 18770Covidien LP · Tube, tracheal (w/wo connector)2 deaths434 reports
- Shiley 18880Covidien LP · Tube, tracheal (w/wo connector)2 deaths370 reports
- Shiley 18780Covidien LP · Tube, tracheal (w/wo connector)3 deaths286 reports
- Shiley 18770SCovidien LP · Tube, tracheal (w/wo connector)273 reports
- Shiley 18780SCovidien LP · Tube, tracheal (w/wo connector)251 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page