EndoWrist recallZ-2059-2023 · Intuitive Surgical, Inc. · initiated Jun 12, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Intuitive Surgical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Fenestrated Graspers have a manufacturing issue has the potential to cause breakage on both upper and/or lower grips, which could lead to stainless steel fragments falling into the patient that may be either detected or undetected.

FDA root cause category: Process control

What the company is doing

On 6/12/23 Field Safety Notices were mailed and emailed to customers asking them to do the following: 1) Locate and return affected devices. 2) Notify all surgeons and users of the device about the contents of the notice. 3) Complete and return the acknowledgement form to [email protected] If you need further information or support concerning this Medical Device Notification, please contact your Clinical Sales Representative or contact Customer Service at: - North and South America: (800) 876-1310, Option 3 (4 AM to 5 PM PST) or mail: [email protected]. - Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 AM to 6 PM CET) or [email protected] - South Korea: 02-3271-3200 (9 AM to 6 PM KSTJ) - Japan: 0120-56-5635 or 03-5575-1362 (9 AM to 6 PM JST) - India: +1-800-103-6952 (9 AM to 6 PM IT) - Taiwan: +0800-86-8181 (9 AM to 6 PM CT)

Recall details

Recall number
Z-2059-2023
Status
Open, Classified
Initiated by the company
Jun 12, 2023
Posted by the FDA
Jun 30, 2023
Recalling firm
Intuitive Surgical, Inc.
Quantity
46,033
Product code
NAY
FDA submissions
K131861
Distribution
Worldwide - US Nationwide distribution in the states of PR, FL, MN, PA, AZ, OH, IL, NC, GA, KS, CA, HI, MD, OR, SC, AK, NY, NE, IA, WA, OK, ND, UT, MS, AR, WI, MI, TX, TN, IN, NJ, VA, SD, CO, KY, AL, MO, LA, DE, MA, MT, WV, ID, CT, NV, NH, ME, WY, DC, RI, NM, VT and the countries of Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Cyprus, Czech Republic, Denmark, Dominican Republic, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Luxembourg, Mexico, Monaco, Netherlands, Norway, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, San Marino, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom.

Product description

da Vinci Xi/X Tip-Up Fenestrated Grasper, REF: 470347, VER: 12

Code information (models, lots, serial numbers)

UDI-DI: 00886874112496, All Lots.