EndoWrist 470347Recalls & reported problems

Intuitive Surgical, Inc. · TIP-UP FENESTRATED GRASPER · Class 2

All models of EndoWrist (62)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This EndoWrist grasper is used with the da Vinci surgical robot. Its 2,111 adverse event reports mostly describe frayed, torn or broken parts, and some mention foreign bodies in patients. The reports include no deaths and 85 injuries. There are 2 open recalls: one from 2023 for a manufacturing issue that could cause grip breakage and steel fragments in patients, and one from 2024 for frayed or broken grip cables.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
2,115
Deaths reported
0
Injuries reported
86
Recalls naming it
2

Summary

The FDA's MAUDE database contains 2,115 adverse event reports that list this exact device (DI 00886874112496), received between Feb 28, 2017 and Sep 30, 2026 — classified as 86 injuries, 2,029 malfunctions.

The most frequently reported specific device problem is “Material Split, Cut or Torn” (841 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 2 reportsJan 2024: 6 reportsFeb 2024: 2 reportsMar 2024: 3 reportsApr 2024: 3 reportsMay 2024: 6 reportsJun 2024: 7 reportsJul 2024: 9 reportsAug 2024: 0 reportsSep 2024: 6 reportsOct 2024: 9 reportsNov 2024: 6 reportsDec 2024: 7 reportsJan 2025: 4 reportsFeb 2025: 4 reportsMar 2025: 172 reportsApr 2025: 87 reportsMay 2025: 128 reportsJun 2025: 114 reportsJul 2025: 98 reportsAug 2025: 92 reportsSep 2025: 63 reportsOct 2025: 60 reportsNov 2025: 106 reportsDec 2025: 94 reportsJan 2026: 99 reportsFeb 2026: 263 reportsMar 2026: 150 reportsApr 2026: 114 reportsMay 2026: 61 reportsJun 2026: 0 reportsJul 2026: 2 reportsAug 2026: 38 reportsSep 2026: 54 reportsOct 2026: 0 reportsNov 2026: 0 reports202420252026263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 19, 2024Class IIOpen, Classified

    Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.

    Z-0869-2025 · Root cause: Process change control · Recall details · FDA record ↗

  2. Jun 12, 2023Class IIOpen, Classified

    Fenestrated Graspers have a manufacturing issue has the potential to cause breakage on both upper and/or lower grips, which could lead to stainless steel fragments falling into the patient that may be either detected or undetected.

    Z-2059-2023 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Split, Cut or Torn841
  • Material Frayed663
  • Detachment of Device or Device Component163
  • Break157
  • Material Fragmentation105
  • No Apparent Adverse Event67
  • Difficult to Remove35
  • Unintended Movement22

Patient problems

  • No Clinical Signs, Symptoms or Conditions1,423
  • Insufficient Information611
  • No Known Impact Or Consequence To Patient35
  • Foreign Body In Patient29
  • Unspecified Tissue Injury12
  • Hemorrhage/Blood Loss/Bleeding5
  • Bowel Perforation3
  • Injury3

Recent adverse event reports

The 25 most recent of 2,115 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954335 ↗
  • MalfunctionSep 30, 2026Material Frayed

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954064 ↗
  • MalfunctionSep 30, 2026Material Split, Cut or Torn

    It was reported that there was a broken cable with a da Vinci product. The customer reported event has no report of patient injury.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50942402 ↗
  • MalfunctionSep 28, 2026Material Frayed

    Patient problems: Insufficient Information

    MAUDE report 50926080 ↗
  • MalfunctionSep 28, 2026Material Frayed

    Patient problems: Insufficient Information

    MAUDE report 50926055 ↗
  • MalfunctionSep 28, 2026No Apparent Adverse Event

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50925127 ↗
  • MalfunctionSep 28, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50925070 ↗
  • MalfunctionSep 28, 2026Detachment of Device or Device Component

    It was reported that during a da Vinci-assisted surgical procedure, a Tip-Up Fenestrated Grasper instrument became stuck in the patient's chest because the distal end came out of the forceps sheath, the forceps jammed, making it impossible for the surgeon to manipulate it from the robot console. The procedure was completed with no reported injury. Intuitive Surgical, Inc. (ISI) followed up with the initial reporter and obtained the following additional information: The instrument was removed by hand. All the parts were still attached to the instrument. No fragment fell into the patient. The procedure was completed using a replacement instrument.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50910822 ↗
  • MalfunctionSep 25, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50905842 ↗
  • MalfunctionSep 25, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50905841 ↗
  • MalfunctionSep 25, 2026Material Frayed

    Patient problems: Insufficient Information

    MAUDE report 50905779 ↗
  • MalfunctionSep 25, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50905611 ↗
  • MalfunctionSep 25, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50905540 ↗
  • MalfunctionSep 25, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50905267 ↗
  • MalfunctionSep 25, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50905212 ↗
  • MalfunctionSep 25, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50905124 ↗
  • MalfunctionSep 25, 2026Material Frayed

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50905119 ↗
  • MalfunctionSep 25, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50905022 ↗
  • MalfunctionSep 25, 2026Material Frayed

    Patient problems: Insufficient Information

    MAUDE report 50905010 ↗
  • MalfunctionSep 25, 2026No Apparent Adverse Event

    Patient problems: Insufficient Information

    MAUDE report 50904899 ↗
  • MalfunctionSep 25, 2026No Apparent Adverse Event

    Patient problems: Insufficient Information

    MAUDE report 50904889 ↗
  • MalfunctionSep 25, 2026Unintended System Motion

    It was reported that during a da Vinci-assisted surgical procedure, a Tip-Up Fenestrated Grasper instrument was observed Moving erratically. The procedure was completed with no reported injury.

    Patient problems: Insufficient Information

    MAUDE report 50900400 ↗
  • MalfunctionSep 25, 2026Break

    It was reported that prior to the start of a da Vinci-assisted surgical procedure, a 8mm Tip-Up Fenestrated Grasper instrument was observed with the tip damaged.The procedure was completed as planned with no reported injury.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50893635 ↗
  • MalfunctionSep 25, 2026Unintended Movement

    It was reported that after a da Vinci-assisted surgical procedure, the Tip Up Fenestrated Grasper instrument would not close. The procedure was completed with no reported injury.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50893583 ↗
  • MalfunctionSep 24, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50891420 ↗

FDA clearance and classification

The FDA regulates product code NAY (system,surgical,computer controlled instrument) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the EndoWrist 470347

Has the EndoWrist 470347 been recalled?

Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-0869-2025 (Class II), began on Dec 19, 2024: Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the EndoWrist 470347?

The FDA's MAUDE database lists 2,115 adverse event reports for this device, received between Feb 28, 2017 and Sep 30, 2026. In those reports, the most common specific device problems are “Material Split, Cut or Torn” (841), “Material Frayed” (663) and “Detachment of Device or Device Component” (163); the most common patient problems are “Foreign Body In Patient” (29), “Unspecified Tissue Injury” (12) and “Hemorrhage/Blood Loss/Bleeding” (5). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the EndoWrist 470347?

None of the reports are classified as deaths, 86 as injuries and 2,029 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the EndoWrist 470347, and is it still sold?

The FDA's device database lists Intuitive Surgical, Inc. as the company responsible for it (brand EndoWrist, model 470347). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other System,surgical,computer controlled instrument devices