Radiation therapy software, RayStation standalone software… recallZ-2039-2023 · RaySearch America Inc · initiated Jan 22, 2021

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from RaySearch America Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A software bug affecting results when using deep learning (DL) segmentation followed by geometrical functions: Create wall function, Expand Contract function or ROI algebra function, the results are incorrect. If dose statistics for an incorrectly created ROI are used to determine if a plan is acceptable for treatment, this could potentially lead to an inappropriate plan being approved. The error could lead to a more conservative or more liberal plan than intended.

FDA root cause category: Software design

What the company is doing

Field Safety Notice 73474 was emailed January 22, 2021 to affected customers. The notice identifies affected RayStation models: 8B, 8B SP1, 8B SP2, 9A, 9B, 9B SP1, 10A, 10A SP1 and 10B running software versions RS 8.1, 9.0, 9.1, 9.2, 10.0, 10.1. Users are instructed not to use Create wall, Expand/Contract, or ROI algebra functions directly on ROIs created by the DL segmentation. Instead, convert the ROIs to contour or mesh representation before using the affected functions. Users are to confirm they read and understood the notice by replying to the email. Error fixed in RS 10B Service Pack 1 and corrected in the next version of RayStation.

Recall details

Recall number
Z-2039-2023
Status
Terminated
Initiated by the company
Jan 22, 2021
Posted by the FDA
Jun 29, 2023
Recalling firm
RaySearch America Inc
Quantity
5 systems
Product code
MUJ
FDA submissions
K200569
Distribution
US: AZ, CA, GA & WA

Product description

Radiation therapy software, RayStation standalone software treatment planning system. Model numbers: 8B, 8B SP1, 8B SP2, 9A, 9B, 9B SP1, 10A, 10A SP1 and 10B. Software versions: 8.1, 9.0, 9.1, 9.2, 10.0, 10.1.

Code information (models, lots, serial numbers)

UDI-DI: 07350002010129 UDI-DI: 07350002010204 UDI-DI: 07350002010235 UDI-DI: 07350002010174 UDI-DI: 07350002010266 UDI-DI: 07350002010297 UDI-DI: 07350002010303 UDI-DI: 07350002010365 UDI-DI: 07350002010310

Devices named in this recall

8 devices in the FDA device registry whose identifier appears in the recall's code information.