DigitalDiagnost C90 recallZ-2042-2023 · Philips North America Llc · initiated May 19, 2023

Class II Open, Classified All recalls posted in June 2023

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips North America Llc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is a software login in issue that may prevent the user from logging in.

FDA root cause category: Software change control

What the company is doing

Consignees in the United States were notified via Urgent Medical Device Correction Letter that was distributed on May 26, 2023. US consignees were initially notified via email. Follow-up notifications via FedEx will follow. Philips Market Organizations are responsible for distributing letters outside of the US in accordance with local requirements. A Philips Field Service Engineer will visit impacted customer sites to perform a software upgrade to prevent the issue from recurring. Philips has requested customers perform the following: a) To prevent the system from becoming inoperable, do not reboot or log out of your system until Philips has implemented the software correction; b) Circulate this notice to all users of the device so that they are aware of the issue; and c) Complete and return the response form to Philips promptly and no later than 30 days from receipt.

Recall details

Recall number
Z-2042-2023
Recall class
Class II
Status
Open, Classified
Initiated by the company
May 19, 2023
Posted by the FDA
Jun 30, 2023
Recalling firm
Philips North America Llc
Quantity
58
Product code
MQB
FDA submissions
K202564
Distribution
Distribution in US: AZ, GA, IN, IA, MN, NE, NH, OH, TX OUS distribution to Australia, Chile, China, Malta, Namibia, New Zealand, Oman, Panama, Qatar, Suriname, United Kingdom

Product description

DigitalDiagnost C90, Model No. 712034 and 712035

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

1) UDI-DI (01)00884838090699, Model 712034, serial numbers: 10000074, 10000083, 10000163, 10000188, 10001057, 10001071, 10001072, 10001079, 10001089, 10001106, 10001133, 10001136, 10001137, 10001142, 10001146, 10001169, 10001180, 10001181, 10001186, 10001203, 10001211, 10001229, 10001234, 10001276, 19861179, 19861365, 21860026, 21860084, 21861001, 21861233, 21861386, 21861415, 21861416, 21861493, 21861528, 22860017, 22860114, 22860115, 22860448, 22860454, SN19000006, SN20861092, SN20861107, SN20861108, SN20861115, SN20861125, SN20861126, and SN20861128 2) UDI-DI (01)00884838090705, Model 712035, serial numbers: 10000034, 10000098, 10001051, 10001113, 10001130, 21860040, 22860052, SN20862021, SN20862022, and SN20862024

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page