REF: 218-1604, 1.6mm x 4mm Profile Zero Auto-Drive Screw… recallZ-2107-2023 · Osteomed, LLC · initiated May 31, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Osteomed, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Screws for osteotomies have contain an undersize screwhead which has the potential to pass through the plate hole.
FDA root cause category: Nonconforming Material/Component
What the company is doing
On June 6, 2023, OsteoMed/Acumed issued an "Urgent: Medical Device Recall" notification to affected consignees via FedEx Ground. OsteoMed/Acumed asked consignees to take the following actions: IMMEDIATELY stop using and/or distributing the product and complete the following actions: 1) Perform a physical count of your inventory. Identify any inventory of the lot number listed above that is still in its packaging. Locate and remove the affected, unopened, unused packages of screws, such that they are out of service (i.e., quarantined). 2) Record this data (product and quantity) on the survey found at the end of this letter (the Outstanding Field Action Survey). 3) Complete the remaining, applicable fields in the Outstanding Field Action Survey and either return it via email to [email protected] or a physical copy can be mailed to: OsteoMed/Acumed, Attn: Lauren Freytag- R23-002-TX, 3885 Arapaho Rd, Addison, TX 75001 4) If you have product to return, please contact [email protected] so that an appropriate return number and return instructions can be provided. 5) If you do not have product to return, you are still required to complete the form below. 6) For patients implanted with these affected devices, OsteoMed/Acumed advises physicians to monitor patients.
Recall details
- Recall number
- Z-2107-2023
- Status
- Open, Classified
- Initiated by the company
- May 31, 2023
- Posted by the FDA
- Jul 7, 2023
- Recalling firm
- Osteomed, LLC
- Product code
- GWO
- FDA submissions
- K111176
- Distribution
- Worldwide - US Nationwide distribution in the states of AZ, CA, FL, GA, KS, KY, MD, MO, MT, NE, NY, TN, TX and the country of SG.
Product description
REF: 218-1604, 1.6mm x 4mm Profile Zero Auto-Drive Screw, Rx Only, Sterile
Code information (models, lots, serial numbers)
Lot # 1177027, 1177718, 1177812 / UDI-DI: (01)00845694034258
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- OsteoMed 218-1604Osteomed LLC · Plate, cranioplasty, preformed, alterable1 recall6 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 92479).
- Z-2108-2023 · Open, ClassifiedREF 218-1604-SP-10, Profile 0 4mm AutoDrive Screw Sterile Qty 10, Rx Only, Sterile
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page