LinkSymphoKnee recallZ-2066-2023 · Waldemar Link GmbH & Co. KG (Mfg Site) · initiated Dec 23, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Waldemar Link GmbH & Co. KG (Mfg Site) or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Fixation screw was inserted through the wrong side of the augment (distal to proximal, as opposed to proximal to distal, which is the correct orientation).
FDA root cause category: Component change control
What the company is doing
LinkBio Corporation issued A Product Advisory Notice to consignees holding field inventory on 18-May-2023. The product is provided to the surgeon directly by the representative of the Distributor. Letter states reason for recall, health risk and action to take: Please ensure the surgical team is made aware of this issue before using one of the affected augments. While there is no safety impact due to this issue, there could be some confusion during assembly if the issue is not known in advance. No additional action is required by users beyond reinserting the screw in the correct orientation, and no product return is required at this time. Please contact your LinkBio Sales or Customer Service representative or the contact for any questions: Jennifer Lubrecht Senior Director, Quality Management LinkBio Corporation 69 King Street Dover, NJ 07801 Mobile: +1 862.477.0370 Email: [email protected]
Recall details
- Recall number
- Z-2066-2023
- Status
- Open, Classified
- Initiated by the company
- Dec 23, 2022
- Posted by the FDA
- Jul 3, 2023
- Recalling firm
- Waldemar Link GmbH & Co. KG (Mfg Site)
- Quantity
- 20 units
- Product code
- JWH
- FDA submissions
- K202924
- Distribution
- US Nationwide distribution in the states of AL, IL, IN, LA, GA, NV, TX.
Product description
LinkSymphoKnee Distal Femoral Augment - Tilastan - Size 7-8, Medial-Right/Lateral Left, 5mm- Intended for primary and revision total knee replacement in skeletally mature patients. Product Code: 880-307/11
Code information (models, lots, serial numbers)
UDI-DI: 04026575257379 Lot Number: 2123225
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- LinkSymphoKnee 880-307/11WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER2 recalls2 reports
Product lines
Other devices under the same FDA submission
116 devices cleared under K202924 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- LinkSymphoKnee 880-307/21WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER2 recalls2 reports
- LinkSymphoKnee 880-325/13WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER2 recalls1 report
- LinkSymphoKnee 880-100/50WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall1 report
- LinkSymphoKnee 880-305/11WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall1 report
- LinkSymphoKnee 880-305/22WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall1 report
- LinkSymphoKnee 880-313/11WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall1 report
- LinkSymphoKnee 880-313/22WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall1 report
- LinkSymphoKnee 880-315/21WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER1 recall1 report
- LinkSymphoKnee 880-300/21WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- LinkSymphoKnee 880-301/11WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- LinkSymphoKnee 880-301/21WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- LinkSymphoKnee 880-303/11WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 92496).
- Z-2067-2023 · Open, ClassifiedLinkSymphoKnee Distal Femoral Augment - Tilastan - Size. 7-8, Lateral-Right/Medial-Left, 5mm-Intended for primary and revision total knee replacement in skeletally mature patients.…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page