Power-PRO 2- A powered ambulance cot that consists of a… recallZ-2130-2023 · Stryker Medical Division of Stryker Corporation · initiated May 24, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Stryker Medical Division of Stryker Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Base leg assembly may bend and result in difficulty engaging/disengaging from fastener, difficulty loading/unloading from ambulance, difficulty raising or lowering the cot, cot tip, or erratic or unintended motion.
FDA root cause category: Process control
What the company is doing
Stryker issued Urgent Medical Device Correction Letter on May 24, 2023 via Fed'X. Letter states reason for recall, health risk and action to take: 1. Locate the units listed on the attached business reply form and identify the address where they can be serviced. 2. Review and follow the required customer actions set forth in Appendix A to ensure the unit continues to operate safely until the correction can be performed. If any of the issues listed are present, remove the unit from service and contact Tech Support to request a service visit. 3. Return the enclosed business reply form to confirm receipt of this notification by fax +1 269 488 8691 or email [email protected]. 4. Stryker will be initiating a correction for all affected Power-PRO 2 units and will contact customers to schedule device corrections once the correction is ready for implementation. The company will provide customers with updated information when it is available. 5. If you have loaned or sold any of the products listed in this letter, please forward a copy of this notice to the new users and advise us of their new location in the space provided on the business reply form. 6. If you have disposed of any of these units and they are no longer in use, please, advise us of their obsolescence by providing us with their serial number in the space provided on the business reply form. If you have any questions or concerns, contact Technical Support at 1 800 787 9537 option 2 Monday through Friday from 8:00 a.m. to 6:00 p.m. ET.
Recall details
- Recall number
- Z-2130-2023
- Status
- Open, Classified
- Initiated by the company
- May 24, 2023
- Posted by the FDA
- Jul 12, 2023
- Recalling firm
- Stryker Medical Division of Stryker Corporation
- Quantity
- 1,980 cots
- Product code
- FPO
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Canada, Netherlands, Switzerland, United Kingdom.
Product description
Power-PRO 2- A powered ambulance cot that consists of a platform, including a mattress, mounted on a wheeled, retractable X-frame that is designed to support and transport a maximum weight of 700 lb (318 kg). Model Number: 6507
Code information (models, lots, serial numbers)
Show the full code information (52k characters)
Model: 650700000000 UDI-DI: 07613327504460; Model: 650705550001 UDI-DI: 07613327559118; Model: 650705550002 UDI-DI: 07613327559125; Cots shipped between February 11, 2022 and February 1, 2023 Serial Numbers: 2112001485 2112001488 2112001489 2112001490 2112001491 2112001492 2112001493 2112001494 2112001495 2112001496 2112001501 2112001503 2112001504 2112001508 2112001509 2112001510 2112001511 2112001512 2112001513 2112001514 2112001515 2112001516 2112001517 2112001518 2112001519 2112001520 2112001521 2112001522 2112001523 2112001524 2112001525 2112001526 2112001527 2112001528 2112001529 2112001530 2112001531 2112001532 2112001533 2112001534 2112001535 2112001536 2202000473 2202000474 2202000475 2202000476 2202000477 2202000478 2202000479 2202000480 2202000481 2202000482 2202000483 2202000484 2202000485 2202000487 2202000488 2202000489 2202000490 2202000491 2202000492 2202000797 2202000798 2202000799 2202000800 2202000801 2202000802 2202000803 2202000804 2202000805 2202000806 2202000807 2202000808 2202000809 2202000810 2202000811 2202000812 2202000813 2202000814 2202000815 2202000816 2202000817 2202000819 2202000820 2202000821 2202000822 2202000823 2202000824 2202000825 2202000826 2202001299 2202001300 2202001301 2202001302 2202001303 2202001304 2202001305 2202001306 2202001307 2202001308 2202001309 2202001311 2202001312 2202001313 2202001314 2202001315 2202001316 2202001317 2202001318 2202001518 2202001519 2202001520 2202001521 2202001522 2202001523 2202001524 2202001526 2202001535 2202001539 2202001541 2202001544 2202001545 2202001546 2202001547 2202001548 2202001549 2202001552 2202001554 2202001555 2202001556 2202001557 2202001558 2202001560 2202001561 2202001562 2202001563 2202001564 2202001565 2202001566 2202001568 2202001569 2202001570 2202001572 2202001573 2202001574 2202001575 2202001576 2202001577 2202001578 2202001579 2202001580 2202001582 2202001584 2202001585 2202001587 2202001590 2202001591 2202001592 2202001593 2202001594 2202001597 2202001795 2202001796 2202001797 2203001606 2203001607 2203001608 2203001609 2203001611 2203001612 2203001613 2203001614 2203001615 2203001619 2203001620 2203001719 2203001720 2203001721 2203001723 2203001742 2203001743 2203001744 2203001745 2203001746 2203001747 2203001748 2203001750 2203001924 2203001925 2203001926 2203001927 2203001928 2203001929 2203001930 2203001931 2203001932 2203001934 2203001935 2203001936 2203001937 2203001938 2203001939 2203001940 2203001941 2203001942 2203001943 2203001944 2203001945 2203001946 2203001947 2203001948 2203001949 2203001950 2203001951 2203001952 2203001953 2203001954 2203001955 2203001956 2203001957 2203002386 2203002387 2203002388 2203002389 2203002390 2203002391 2203002392 2203002393 2203002394 2203002395 2203002406 2203002408 2203002409 2203002410 2203002411 2203002412 2203002413 2203002414 2203002415 2203002416 2203002417 2203002418 2203002650 2203002651 2203002652 2203002653 2203002654 2203002655 2203002656 2203002657 2203002658 2203002659 2203002660 2203002663 2203002664 2203002665 2203002666 2203002667 2203002668 2203002669 2204000074 2204000075 2204000076 2204000077 2204000078 2204000080 2204000081 2204000082 2204000084 2204000085 2204000086 2204000087 2204000088 2204000089 2204000092 2204000093 2204000094 2204000095 2204000097 2204000098 2204000099 2204000101 2204000103 2204000104 2204000105 2204000108 2204000109 2204000111 2204000112 2204000114 2204000115 2204000116 2204000117 2204000118 2204000269 2204000270 2204000271 2204000272 2204000273 2204000274 2204000275 2204000276 2204000277 2204000279 2204000280 2204000282 2204000283 2204000284 2204000285 2204000286 2204000287 2204000288 2204000289 2204000290 2204000291 2204000292 2204000293 2204000294 2204000295 2204000296 2204000298 2204000299 2204000300 2204000301 2204000302 2204000303 2204000304 2204000305 2204000306 2204000308 2204000309 2204000310 2204000311 2204000327 2204000328 2204000329 2204000330 2204000331 2204000332 2204000339 2204000340 2204000341 2204000343 2204000344 2204000345 2204000346 2204000348 2204000350 2204000351 2204000352 2204000353 2204000354 2204000355 2204000356 2204000357 2204000358 2204000359 2204000360 2204000361 2204000362 2204000363 2204000364 2204000365 2204000366 2204000367 2204000368 2204000369 2204000370 2204000371 2204000372 2204000373 2204000374 2204000375 2204000376 2204000377 2204000379 2204000381 2204000382 2204000383 2204000394 2204000395 2204000396 2204000397 2204000399 2204000400 2204000401 2204000402 2204000403 2204000404 2204000405 2204000406 2204000407 2204000408 2204000409 2204000411 2204000412 2204000414 2204000415 2204000416 2204000418 2204000419 2204000420 2204000422 2204000423 2204000424 2204000425 2204000426 2204000427 2204000428 2204000429 2204000430 2204000431 2204000432 2204000433 2204000434 2204000435 2204000437 2204000438 2204000439 2204000440 2204000441 2204000442 2204000443 2204000444 2204000445 2204000446 2204000447 2204000449 2204000450 2204000452 2204000453 2204000632 2204000633 2204000634 2204000635 2204000636 2204000637 2204000638 2204000639 2204000640 2204000641 2204000642 2204000643 2204000644 2204000645 2204000646 2204000647 2204000648 2204000649 2204000650 2204000651 2204000652 2204000653 2204000654 2204000655 2204000656 2204000657 2204000658 2204000659 2204000660 2204000661 2204000662 2204000663 2204000664 2204000665 2204000666 2204000667 2204000669 2204000670 2204000671 2204000672 2204000673 2204000674 2204000675 2204000676 2204000677 2204000678 2204000679 2204000680 2204000681 2204000682 2204000683 2204000684 2204000685 2204000686 2204000687 2204000688 2204000689 2204000690 2204000691 2204000692 2204000693 2204000694 2205000094 2205000095 2205000096 2205000097 2205000098 2205000099 2205000100 2205000101 2205000102 2205000103 2205000105 2205000106 2205000107 2205000108 2205000109 2205000110 2205000111 2205000112 2205000113 2205000114 2205000115 2205000116 2205000117 2205000118 2205000119 2205000120 2205000121 2205000122 2205000123 2205000124 2205000125 2205000126 2205000127 2205000128 2205000129 2205000130 2205000131 2205000132 2205000133 2205000135 2205000136 2205000137 2205000138 2205000139 2205000140 2205000141 2205000142 2205000143 2205000144 2205000145 2205000146 2205000147 2205000148 2205000149 2205000150 2205000151 2205000152 2205000153 2205000154 2205000155 2205000156 2205000157 2205000158 2205000159 2205000160 2205000161 2205000162 2205000163 2205000164 2205000165 2205000166 2205000167 2205000168 2205000169 2205000170 2205000171 2205000172 2205000173 2208001658 2208001659 2208001660 2208001661 2208001662 2208001663 2208001664 2208001667 2208001668 2208001671 2208002061 2208002065 2208002066 2208002068 2208002069 2208002070 2208002071 2208002072 2208002581 2208002582 2208002583 2208002584 2208002585 2208002586 2208002587 2208002588 2208002589 2208002590 2208002591 2208002592 2208002593 2208002594 2208002595 2208002596 2208002597 2208002598 2208002599 2209000766 2209000767 2209000768 2209000769 2209000770 2209000771 2209000772 2209000773 2209000774 2209000775 2209000776 2209000777 2209000778 2209000779 2209000780 2209000781 2209000782 2209000783 2209000784 2209000785 2209001486 2209001487 2209001488 2209001489 2209001490 2209001491 2209001492 2209001493 2209001494 2209001495 2209001496 2209001497 2209001498 2209001499 2209001500 2209001501 2209001502 2209001503 2209001504 2209001505 2209001506 2209001507 2209001508 2209001510 2209001512 2209001513 2209001514 2209001515 2209001516 2209001517 2209001518 2209001519 2209001520 2209001522 2209001523 2209001524 2209001526 2209001527 2209001528 2209001529 2209001531 2209002223 2209002224 2209002225 2209002226 2209002227 2209002228 2209002229 2209002230 2209002231 2209002232 2209002233 2209002234 2209002235 2209002236 2209002237 2209002238 2209002239 2209002240 2209002241 2209002242 2209002243 2209002244 2209002245 2209002246 2209002247 2209002248 2209002249 2209002250 2209002251 2209002252 2209002253 2209002254 2209002255 2209002256 2209002257 2209002258 2209002259 2209002260 2209002261 2209002262 2209002263 2209002264 2209002265 2209002266 2209002267 2209002268 2209002269 2209002270 2209002436 2209002437 2209002438 2209002439 2209002440 2209002441 2209002443 2209002444 2209002445 2209002446 2209002448 2209002449 2209002450 2209002451 2209002452 2209002453 2209002454 2209002455 2209002456 2209002457 2209002458 2209002459 2209002460 2209002461 2209002462 2209002463 2209002464 2209002465 2209002466 2209002467 2209002468 2209002469 2209002470 2209002471 2209002472 2209002473 2209002474 2209002475 2209002476 2209002477 2209002478 2209002479 2209002480 2209002481 2209002482 2209002483 2210000056 2210000057 2210000058 2210000059 2210000060 2210000061 2210000062 2210000063 2210000064 2210000065 2210000066 2210000067 2210000068 2210000069 2210000070 2210000071 2210000072 2210000073 2210000075 2210000076 2210000077 2210000078 2210000079 2210000080 2210000081 2210000082 2210000083 2210000084 2210000085 2210000086 2210000087 2210000088 2210000089 2210000090 2210000091 2210000092 2210000093 2210000094 2210000095 2210000288 2210000289 2210000290 2210000291 2210000292 2210000293 2210000294 2210000295 2210000296 2210000297 2210000298 2210000299 2210000300 2210000301 2210000302 2210000303 2210000304 2210000305 2210000306 2210000307 2210000308 2210000309 2210000310 2210000311 2210000312 2210000313 2210000314 2210000315 2210000316 2210000317 2210000318 2210000319 2210000320 2210000321 2210000322 2210000323 2210000324 2210000325 2210000326 2210000327 2210000328 2210000329 2210000330 2210000331 2210000332 2210000333 2210000334 2210000335 2210000336 2210000337 2210000338 2210000339 2210000340 2210000341 2210000342 2210000343 2210000344 2210000345 2210000346 2210000347 2210000509 2210000510 2210000511 2210000512 2210000513 2210000514 2210000515 2210000516 2210000517 2210000518 2210000519 2210000520 2210000521 2210000522 2210000523 2210000524 2210000525 2210000526 2210000527 2210000528 2210000529 2210000530 2210000531 2210000532 2210000533 2210000534 2210000535 2210000536 2210000537 2210000538 2210000539 2210000540 2210000541 2210000542 2210000543 2210000544 2210000545 2210000546 2210000547 2210000548 2210001413 2210001414 2210001415 2210001416 2210001417 2210001418 2210001419 2210001420 2210001433 2210001434 2210001435 2210001436 2210001437 2210001438 2210001439 2210001440 2210001441 2210001442 2210001443 2210001444 2210001445 2210001446 2210001447 2210001448 2210001449 2210001450 2210001451 2210001452 2210001453 2210001454 2210001455 2210001456 2210001457 2210001458 2210001459 2210001460 2210001461 2210001462 2210001464 2210001528 2210001529 2210001530 2210001531 2210001532 2210001533 2210001534 2210001535 2210001536 2210001537 2210001538 2210001539 2210001540 2210001541 2210001542 2210001543 2210001544 2210001545 2210001546 2210001564 2210001565 2210001566 2210001567 2210001568 2210001569 2210001570 2210001571 2210001572 2210001573 2210001574 2210001575 2210001576 2210001577 2210001578 2210001579 2210001580 2210001581 2210001582 2210001583 2210001584 2210001585 2210001586 2210001587 2210001588 2210001589 2210001590 2210001591 2210001592 2210001593 2210001594 2210001595 2210001596 2210001597 2210001598 2210001599 2210001600 2210001601 2210001602 2210001603 2210001604 2210001605 2210001606 2210001607 2210001608 2210001609 2210001610 2210001611 2210001612 2210001613 2210001614 2210001615 2210001616 2210001617 2210001618 2210001620 2210001621 2210001622 2210001623 2210002438 2210002439 2210002440 2210002441 2210002442 2210002443 2210002444 2210002445 2210002446 2210002447 2210002448 2210002449 2210002450 2210002451 2210002452 2210002453 2210002454 2210002455 2210002456 2210002457 2210002951 2210002952 2210002953 2210002954 2210002955 2210002956 2210002957 2210002958 2210002959 2210002960 2210002961 2210002962 2210002963 2210002964 2210002965 2210002966 2210002967 2210002968 2210002969 2210002970 2210002971 2210002972 2210002973 2210002974 2210002975 2210002976 2210002977 2210002978 2210002979 2210002980 2210002981 2210002982 2210002983 2210002984 2210002985 2210003064 2210003065 2210003066 2210003067 2210003068 2210003069 2210003070 2210003071 2210003072 2210003073 2210003074 2210003075 2210003076 2210003117 2210003118 2210003119 2210003120 2210003121 2210003122 2210003123 2210003124 2210003125 2210003126 2210003127 2210003128 2210003129 2210003130 2210003131 2210003132 2210003133 2210003134 2210003135 2210003136 2210003137 2210003138 2210003139 2210003140 2210003141 2210003142 2210003143 2210003144 2210003145 2210003146 2210003147 2210003148 2211000340 2211000341 2211000342 2211000343 2211000344 2211000345 2211000346 2211000347 2211000348 2211000349 2211000350 2211000351 2211000352 2211000353 2211000354 2211000355 2211000356 2211000357 2211000358 2211000359 2211000361 2211000362 2211000363 2211000364 2211000365 2211000366 2211000367 2211000368 2211000369 2211000371 2211000372 2211000374 2211000375 2211000376 2211000377 2211000378 2211000379 2211000380 2211000381 2211000382 2211000383 2211000384 2211000385 2211000386 2211000387 2211000388 2211000389 2211000390 2211000391 2211000392 2211000393 2211000997 2211000998 2211000999 2211001000 2211001001 2211001002 2211001003 2211001004 2211001125 2211001126 2211001267 2211001268 2211001269 2211001270 2211001271 2211001272 2211001273 2211001274 2211001471 2211001472 2211001473 2211001474 2211001519 2211001531 2211001532 2211001533 2211001534 2211001535 2211001536 2211001537 2211001538 2211001539 2211001540 2211001569 2211001570 2211001571 2211001572 2211001573 2211001574 2211001575 2211001576 2211001577 2211001578 2211001579 2211001580 2211001581 2211001582 2211001583 2211001584 2211001585 2211001586 2211001587 2211001588 2211001589 2211001590 2211001591 2211001594 2211001596 2211001597 2211001598 2211001599 2211001600 2211001601 2211001610 2211001611 2211001612 2211001613 2211001614 2211001645 2211001646 2211001647 2211001648 2211001649 2211001650 2211001651 2211001652 2211001653 2211001654 2211003256 2211003270 2211003274 2211003275 2211003312 2211003316 2211003363 2211003365 2211003367 2211003369 2211003371 2211003373 2211003375 2211003377 2211003379 2211003380 2211003574 2211003575 2211003576 2211003577 2211003578 2211003579 2211003580 2211003581 2211003582 2211003583 2211003584 2211003585 2211003586 2211003587 2211003588 2211003589 2211003590 2211003591 2211003592 2211003593 2211003594 2211003595 2211003596 2211003597 2211003598 2211003599 2211003600 2211003601 2211003602 2211003603 2211003604 2211003605 2211003606 2211003608 2211003609 2211003610 2211003611 2211003612 2211003613 2211003614 2211003615 2211003616 2211003617 2211003618 2211003619 2211003620 2211003621 2211003622 2211003623 2211003624 2211003625 2211003626 2211003627 2211003628 2211003629 2211003630 2211003631 2211003632 2211003633 2211003634 2211003635 2211003636 2211003637 2211003638 2211003639 2211003640 2211003641 2211003642 2211003643 2211003644 2211003645 2211003646 2211003647 2211003648 2211003649 2211003650 2211003651 2211003652 2211003653 2211003655 2211003656 2211003657 2211003658 2211003659 2211003660 2211003661 2211003662 2211003663 2211003664 2211003665 2211003666 2211003667 2211003668 2211003669 2211003670 2211003671 2211003672 2211003673 2211003674 2211003675 2211003676 2211003677 2211003678 2211003679 2211003680 2211003681 2211003682 2211003683 2211003684 2211003685 2211003686 2211003687 2211003688 2211003689 2211003690 2211003691 2211003692 2211003693 2211003694 2211003695 2211003696 2211003697 2211003698 2211003699 2211003700 2211003701 2211003702 2211003703 2211003705 2211003706 2211003707 2211003708 2211003709 2211003710 2211003711 2211003713 2211003714 2211003715 2211003716 2211003717 2211003718 2211003720 2211003721 2211003722 2211003723 2211003725 2211003726 2211003727 2211003728 2211003729 2211003730 2211003731 2211003732 2211003733 2211003734 2211003735 2211003736 2211003737 2211003738 2211003739 2211003740 2211003741 2211003742 2211003743 2211003745 2211003746 2211003747 2211003748 2211003749 2211003750 2211003752 2211003753 2211003754 2211003755 2211003756 2211003757 2211003758 2211003759 2211003760 2211003761 2211003762 2211003763 2211003764 2211003765 2211003766 2211003767 2211003768 2211003769 2211003771 2211003772 2211003773 2211003775 2211003776 2211003777 2211003779 2211003780 2211003781 2211003782 2211003783 2211003784 2211003786 2211003787 2211003788 2211003789 2211003790 2211003791 2211003792 2211003793 2211003794 2211003795 2211003796 2211003797 2211003798 2211003799 2211003800 2211003801 2211003802 2211003803 2211003804 2211003805 2211003806 2211003807 2211003808 2211003809 2211003810 2211003811 2211003812 2211003813 2211003814 2211003815 2211003816 2211003817 2211003818 2211003819 2211003822 2211003823 2211003824 2211003825 2211003826 2211003827 2211003828 2211003829 2211003830 2211003831 2211003832 2211003833 2211003834 2211003835 2211003836 2211003837 2211003838 2211003839 2211003840 2211003841 2211003842 2211003843 2211003844 2211003845 2211003846 2211003847 2211003848 2211003849 2211003850 2211003851 2211003852 2211003853 2211003854 2211003855 2211003856 2211003857 2211003858 2211003859 2211003860 2211003861 2211003862 2211003863 2211003864 2211003865 2211003866 2211003867 2211003868 2211003869 2211003870 2211003871 2211003872 2211003873 2211003874 2211003875 2211003876 2211003877 2211003878 2211003879 2211003880 2211003881 2211003882 2211003884 2211003885 2211003886 2211003887 2211003888 2211003889 2211003890 2211003891 2211003892 2211003893 2211003894 2211003895 2211003896 2211003897 2211003898 2211003899 2211003900 2211003901 2211003902 2211003903 2211003904 2211003905 2211003906 2211003907 2211003908 2211003909 2211003910 2211003911 2211003912 2211003913 2211003914 2211003915 2211003916 2211003918 2211003919 2211003920 2211003921 2211003922 2211003923 2211003924 2211003925 2211003926 2211003927 2211003928 2211003929 2211003931 2211003932 2211003933 2211003934 2211003935 2211003936 2211003937 2211003938 2211003939 2211003940 2211003941 2211003942 2211003943 2211003944 2211003945 2211003946 2211003947 2211003948 2211003949 2211003950 2211003951 2211003952 2211003953 2211003954 2211003955 2211003956 2211003957 2211003958 2211003959 2211003960 2211003961 2211003962 2211003963 2211003965 2211003966 2211003967 2211003968 2211003969 2211003970 2211003971 2211003972 2211003973 2211003974 2211003975 2211003976 2211003977 2211003978 2211003979 2211003980 2211003981 2211003982 2211003983 2211003984 2211003985 2211003986 2211003987 2211003988 2211003989 2211003990 2211003991 2211003992 2211003993 2211003994 2211003995 2211003996 2211003997 2211003998 2211003999 2211004000 2211004001 2211004002 2211004003 2211004004 2211004005 2211004006 2211004007 2211004008 2211004009 2211004010 2211004011 2211004012 2211004013 2211004014 2211004015 2211004016 2211004017 2211004018 2211004019 2211004020 2211004021 2211004022 2211004023 2211004024 2211004026 2211004027 2211004028 2211004029 2211004030 2211004031 2211004032 2211004033 2211004034 2211004035 2211004036 2211004037 2211004038 2211004039 2211004496 2212002484 2212002485 2212002486 2212002487 2212002488 2212002489 2212002490 2212002491 2212002492 2212002493 2212002494 2212002495 2212002497 2212002498 2212002499 2212002500 2212002502 2212002503 2212002505 2212002506 2212002507 2212002508 2212002509 2212002510 2212002511 2212002512 2212002513 2212002514 2212002515 2212002516 2212002518 2212002519 2212002520 2212002521 2212002522 2212002523 2212002524 2212002525 2212002526 2212002527 2212002528 2212002529 2212002530 2212002531 2212002532 2212002533 2212002534 2212002536 2212002537 2212002538 2212002539 2212002540 2212002541 2212002542 2212002543 2212002544 2212002545 2212002546 2212002547 2212002548 2212002549 2212002550 2212002551 2212002552 2212002553 2212002554 2212002556 2212002557 2212002558 2212003303 2212003304 2212003305 2212003306 2212003307 2212003308 2212003309 2212003310 2212003311 2212003312 2212003313 2212003314 2212003317 2212003318 2212003319 2212003320 2212003321 2212003322 2212003323 2212003324 2212003325 2212003327 2212003328 2212003329 2212003330 2212003331 2212003332 2212003333 2212003334 2212003335 2212003336 2212003337 2212003338 2212003339 2212003340 2212003341 2212003342 2212003343 2212003344 2212003346 2212003347 2212003348 2212003349 2212003350 2212003351 2212003352 2212003353 2212003354 2212003355 2212003356 2212003357 2212003358 2212003359 2212003360 2212003361 2212003362 2212003363 2212003364 2212003365 2212003366 2212003367 2212003368 2212003369 2212003370 2212003371 2212003372 2212003373 2212003374 2212003375 2212003376 2212003377 2212003381 2212003382 2212003383 2212003384 2212003385 2212003386 2212003387 2212003388 2212003389 2212003390 2212003391 2212003393 2212003395 2212003396 2212003399 2212003400 2212003402 2212003403 2212003404 2212003405 2212003407 2212003408 2212003409 2212003410 2212003411 2212003412 2212003413 2212003414 2212003416 2212003417 2212003418 2212003419 2212003420 2212003421 2212003422 2212003425 2212003426 2212003427 2212003428 2212003429 2212003430 2212003431 2212003432 2212003433 2212003434 2212003435 2212003436 2212003437 2212003438 2212003439 2212003440 2212003441 2212003443 2212003444 2212003446 2212003447 2212003448 2212003449 2212003450 2212003451 2212003452 2212003453 2212003454 2212003494 2212003495 2212003496 2212003497 2212003498 2212003499 2212003500 2212003502 2212003505 2212003506 2212003507 2212003508 2212003509 2212003512 2212003513 2212003514 2212003515 2212003517 2212003520 2212003521 2212003522 2212003523 2212003524 2212003525 2212003527 2212003528 2212003529 2212003530 2212003531 2212003532 2212003533 2212003534 2212003538 2212003539 2212003542 2212003543 2212003544 2212003545 2212003546 2212003547 2212003549 2212003550 2212003554 2212003555 2212003556 2212003557 2212003558 2212003559 2212003561 2212003562 2212003563 2212003566 2212003567 2212003569 2212003570 2212003571 2212003573 2212003574 2212003576 2212003577 2212003581 2212003594 2212003596 2212003597 2212003603
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- Power-PRO 2, High Configuration 650705550001Stryker Corporation · STRETCHER, WHEELED2 recalls2 deaths99 reports
- Power-PRO 2 650700000000Stryker Corporation · STRETCHER, WHEELED2 recalls1 death42 reports
- Power-PRO 2, Mid Configuration 650705550002Stryker Corporation · STRETCHER, WHEELED1 recall17 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page