Plum 360 Infusion System, List Numbers: 300100405… recallZ-2430-2023 · ICU Medical Inc · initiated Jul 13, 2023

Class II Open, Classified All recalls posted in August 2023

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from ICU Medical Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to a manufacturing defect of a supplier provided component, there is a potential that the audible signal for an alarm may not sound under certain conditions.

FDA root cause category: Nonconforming Material/Component

What the company is doing

ICU Medical issued an URGENT: MEDICAL DEVICE CORRECTION notice on 07/13/2023 via traceable mailing. The notice explained the issue, potential risk, and requested the following: Actions for Clinical Users: If your Plum 360 pump does not power on or indicates an Audio Alarm failure, please remove the pump from use and send the pump to Biomedical Engineering. Actions for Biomedical Engineering: If you have Plum 360 pumps with Audio Alarm failures, please contact ICU Medical. 1. Locate all affected pumps in your possession and ensure all users or potential users of these devices are immediately made aware of this notification and required actions. 2. Complete and return the attached Response Form to [email protected] within ten days of receipt to acknowledge your understanding of this notification. 3. DISTRIBUTORS: If you have distributed affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to [email protected]. A similar notice was distributed specifically to consignees of instruments that were serviced which lists the specific serial numbers. The issue, potential risk, and actions are the same as the general notice.

Recall details

Recall number
Z-2430-2023
Recall class
Class II
Status
Open, Classified
Initiated by the company
Jul 13, 2023
Posted by the FDA
Aug 18, 2023
Recalling firm
ICU Medical Inc
Quantity
93135 devices
Product code
FRN
FDA submissions
K141789, K161469
Distribution
Worldwide

Product description

Plum 360 Infusion System, List Numbers: 300100405, 300100409, 300100413, 300100414, 300100415, 300100418, 300100451, 300100452, 300100453, 300100481, 300100483, 300100613, 300101013, 300101113, 300101114, 300101313, 300101379, 300102213, 300102713, 300102913, 300103413, 300103414, 300104213, 300105413, 300105713, 300106313, 300106513, 300106581, 300107813, 300107814, 300108513, 300108813, 300108814, 300109913; infusion pump

Code information (models, lots, serial numbers)

UDI/DI M335300101, All Plum 360 pumps manufactured between July 2020 and December 2021

Other devices under the same FDA submission

35 devices cleared under K141789, K161469 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.